| | Class 1 Device Recall Ivenix Infusion System (IIS), Large Volume Pump LVP0004 |  |
| Date Initiated by Firm | December 06, 2024 |
| Date Posted | January 08, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0763-2025 |
| Recall Event ID |
95891 |
| 510(K)Number | K183311 |
| Product Classification |
Pump, infusion - Product Code FRN
|
| Product | Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004 |
| Code Information |
Model Number: LVP-0004, UDI Number: 00811505030320.
Serial Numbers: 9233532106, 9233421830, 9233421828, 9233532158, 9233522076, 9233522041, 9233421833, 9233421829, 9233421808, 9233421654, 9233421626, 9233421621, 9233321340, 2215401517, 2215401545, 9233421789, 9233421809, 9233532167, 9233532100, 9233522070, 9233421825, 9233421824, 9233421794, 9233421783, 9233421781, 9233421771, 9233421768, 9233421756, 9233421736, 9233421713, 9233421702, 9233421617, 9233421611, 9233391513, 9233391491, 9233532098, 9233421822, 9233421797, 9233421678, 9233421666, 9233421601, 9233532114, 9233421792, 9233421766, 9233421687, 2215401372, 9233522040, 9233522037, 9233421764, 9233421752, 9233421665, 9232490602, 9233522062, 9233421817, 9233421832, 9233522094, 9233421805, 9233421740, 9233532161, 9233532111, 9233522061, 9233421836, 2212200949, 2212200956, 2212200959, 2212200960, 2212200964, 2212200965, 2212200983, 2212200999, 2212201000, 2212201001, 2212201006, 2212201008, 2212201009, 2212201011, 2212201017, 2212201020, 2212201021, 2212201035, 2212201037, 2212201038, 2212201043, 2215401572, 2219200223, 2219200257, 2219200268, 2219200273, 2219200304, 2219200317, 2219200381, 2223703322, 2223703362, 2223703372, 2223703378, 2223703388, 2223703421, 2223703437, 2223703461, 2223703484, 2223703490, 2223703494, 2223703508, 2223703523, 2223703534, 2223703541, 2223703547, 2318600285, 2318600382, 2318600387, 2321600019, 2321600020, 2321600022, 2321600023, 2321600026, 2321600027, 2321600039, 2321600046, 2321600048, 2321600049, 2321600051, 2321600052, 2321600053, 2321600054, 2321600057, 2321600058, 2321600059, 2321600061, 2321600062, 2321600063, 2321600065, 2321600066, 2321600067, 2321600069, 2321600070, 2321600072, 2321600077, 2321600079, 2321600081, 2321600082, 2321600083, 2321600086, 2321600087, 2321600088, 2321600089, 2321600090, 2321600091, 2321600093, 2321600094, 2321600095, 2321600096, 2321600170, 2323000031, 2323000032, 2323000114, 2323000115, 2323000117, 2323000119, 2323000164, 2323000167, 2323000168, 2323000169, 2323000170, 2324100704, 2332600105, 2333300378, 2333300539, 2333300544, 2333300545, 2333300547, 2333300548, 2333300549, 2333300550, 2333300552, 2333300553, 2333300554, 2333300555, 2333300557, 2333300558, 2333300559, 2333300560, 2333300561, 2333300588, 2333300591, 2333300592, 2333300593, 2333300594, 2333300595, 2333300596, 2333300597, 2334800040, 2334800043, 2334800044, 2334800045, 2334800046, 2334800047, 2334800049, 2334800050, 2334800052, 2334800054, 2334800056, 2334800188, 2335600401, 2335600402, 2335600409, 2335600410, 2335600412, 2335600413, 2335600424, 2335600427, 2335600430, 2335600431, 2335600432, 2335600433, 2335600435, 2335600437, 2335600438, 2335600439, 2335600440, 2335600442, 2335600444, 2400300002, 2212200950, 2212200951, 2212200968, 2212201054, 2219200183, 2321600021, 2321600042, 2323000165, 2323000166, 2333300532, 2333300589, 2333300598, 2335600411, 2335600417, 2212200955, 2212200997, 2219200240, 2321600179, 2215401536, 2226400082, 9233522074, 9233421777, 9233421763, 9233421758, 9233421670, 9233391553, 9233391489, 9233311310, 9233201214, 9233201198, 9233201193, 9233201158, 9232960976, 9232490688, 9232330412, 9233421746, 9233421733, 9233421730, 9233421728, 9233421718, 9233421705, 9233421698, 9233421697, 9233421695, 9233421692, 9233421681, 9233421679, 9233421677, 9233421674, 9233421661, 9233421659, 9233421658, 9233421653, 9233421652, 9233421646, 9233421635, 9233421634, 9233421633, 9233421631, 9233421618, 9233421615, 9233421612, 9233421610, 9233421602, 9233421593, 9233421592, 9233421590, 9233421589, 9233421588, 9233421587, 9233391584, 9233391582, 9233391580, 9233391579, 9233391577, 9233391576, 9233391575, 9233391574, 9233391573, 9233391572, 9233391571, 9233391570, 9233391567, 9233391565, 9233391564, 9233391561, 9233391560, 9233391559, 9233391557, 9233391556, 9233391555, 9233391554, 9233391552, 9233391551, 9233391550, 9233391549, 9233391548, 9233391547, 9233391546, 9233391542, 9233391541, 9233391539, 9233391538, 9233391536, 9233391535, 9233391531, 9233391530, 9233391529, 9233391524, 9233391522, 9233391517, 9233391514, 9233391511, 9233391510, 9233391509, 9233391508, 9233391507, 9233391505, 9233391504, 9233391502, 9233391500, 9233391498, 9233391497, 9233391496, 9233391495, 9233391494, 9233391493, 9233391492, 9233391488, 9233391487, 9233391486, 9233391485, 9233391484, 9233391481, 9233391480, 9233381479, 9233381478, 9233381477, 9233381475, 9233381473, 9233381472, 9233381471, 9233381470, 9233381469, 9233381468, 9233381467, 9233381466, 9233381465, 9233381464, 9233381462, 9233381461, 9233381460, 9233381457, 9233381456, 9233381455, 9233381454, 9233381453, 9233381451, 9233381450, 9233381447, 9233381446, 9233381445, 9233351443, 9233351440, 9233351439, 9233351437, 9233351436, 9233351435, 9233351434, 9233351433, 9233351431, 9233351430, 9233351429, 9233351428, 9233351427, 9233351426, 9233351425, 9233351424, 9233351423, 9233351422, 9233351421, 9233351420, 9233341419, 9233341416, 9233341415, 9233341414, 9233341413, 9233341412, 9233341411, 9233341410, 9233341409, 9233331132, 9233331131, 9233331130, 9233331129, 9233331128, 9233331126, 9233331124, 9233331123, 9233331122, 9233331121, 9233331120, 9233331119, 9233331118, 9233331116, 9233331115, 9233331114, 9233331113, 9233331112, 9233331111, 9233331110, 9233331109, 9233331108, 9233331107, 9233331106, 9233331105, 9233331104, 9233331103, 9233331102, 9233331101, 9233331100, 9233331099, 9233331098, 9233331097, 9233331096, 9233331095, 9233331093, 9233331091, 9233331090, 9233331089, 9233331088, 9233331087, 9233331084, 9233331082, 9233331078, 9233331077, 9233331076, 9233331075, 9233331073, 9233331070, 9233331067, 9233331066, 9233331062, 9233331060, 9233331056, 9233331055, 9233331052, 9233331051, 9233331050, 9233331044, 9233331041, 9233331040, 9233331039, 9233331036, 9233331034, 9233321407, 9233321406, 9233321405, 9233321404, 9233321402, 9233321401, 9233321400, 9233321399, 9233321398, 9233321397, 9233321396, 9233321395, 9233321394, 9233321392, 9233321391, 9233321390, 9233321389, 9233321388, 9233321387, 9233321386, 9233321384, 9233321383, 9233321381, 9233321379, 9233321378, 9233321377, 9233321376, 9233321375, 9233321373, 9233321372, 9233321371, 9233321370, 9233321369, 9233321368, 9233321365, 9233321364, 9233321363, 9233321362, 9233321361, 9233321360, 9233321359, 9233321357, 9233321355, 9233321353, 9233321351, 9233321350, 9233321349, 9233321347, 9233321346, 9233321345, 9233321344, 9233321343, 9233321341, 9233311337, 9233311336, 9233311335, 9233311334, 9233311332, 9233311331, 9233311328, 9233311327, 9233311326, 9233311323, 9233311322, 9233311320, 9233311317, 9233311316, 9233311314, 9233311312, 9233311311, 9233311309, 9233311308, 9233311306, 9233311304, 9233311303, 9233311302, 9233311299, 9233311298, 9233311297, 9233311296, 9233311295, 9233311294, 9233311292, 9233311290, 9233311288, 9233311287, 9233311286, 9233311285, 9233311284, 9233311283, 9233311282, 9233311280, 9233311279, 9233311278, 9233311277, 9233311275, 9233311274, 9233311273, 9233311272, 9233311271, 9233311270, 9233311268, 9233311267, 9233311266, 9233311264, 9233311263, 9233311262, 9233311261, 9233311260, 9233311259, 9233311258, 9233311257, 9233311256, 9233311255, 9233311254, 9233311252, 9233311251, 9233311250, 9233311249, 9233311246, 9233311244, 9233311242, 9233311241, 9233311240, 9233311238, 9233311237, 9233311236, 9233311235, 9233311234, 9233201232, 9233201231, 9233201228, 9233201227, 9233201226, 9233201225, 9233201224, 9233201223, 9233201222, 9233201221, 9233201220, 9233201219, 9233201218, 9233201217, 9233201216, 9233201215, 9233201213, 9233201212, 9233201211, 9233201210, 9233201209, 9233201208, 9233201207, 9233201206, 9233201205, 9233201204, 9233201203, 9233201202, 9233201201, 9233201199, 9233201196, 9233201194, 9233201192, 9233201191, 9233201190, 9233201189, 9233201188, 9233201187, 9233201186, 9233201185, 9233201184, 9233201183, 9233201181, 9233201178, 9233201177, 9233201176, 9233201174, 9233201173, 9233201172, 9233201171, 9233201169, 9233201168, 9233201166, 9233201163, 9233201162, 9233201159, 9233201156, 9233201155, 9233201152, 9233201151, 9233201150, 9233201149, 9233201147, 9233201146, 9233201145, 9233201144, 9233201143, 9233201142, 9233201141, 9233201139, 9233201138, 9233201137, 9233201135, 9233201133, 9232961008, 9232961007, 9232961002, 9232961001, 9232961000, 9232960998, 9232960997, 9232960991, 9232960990, 9232960989, 9232960987, 9232960986, 9232960984, 9232960982, 9232960980, 9232960979, 9232960975, 9232960974, 9232960973, 9232960969, 9232960967, 9232960965, 9232960964, 9232960957, 9232960953, 9232960946, 9232960943, 9232960941, 9232960938, 9232960935, 9232750931, 9232750927, 9232750925, 9232750911, 9232750902, 9232750900, 9232750899, 9232750898, 9232750886, 9232750884, 9232750873, 9232750865, 9232750833, 9232750832, 9232750830, 9232750829, 9232750824, 9232750814, 9232750807, 9232750806, 9232720800, 9232720794, 9232720790, 9232720789, 9232720786, 9232720784, 9232720783, 9232720780, 9232720777, 9232720776, 9232720773, 9232720769, 9232720764, 9232720759, 9232720757, 9232720747, 9232720736, 9232490730, 9232490724, 9232490697, 9232490685, 9232490667, 9232490640, 9232490631, 9232490618, 9232490615, 9232490614, 9232490612, 9232490582, 9232490575, 9232420507, 9232420505, 9232330437, 9232330426, 9232300400, 9232300386, 9232190324, 9232190290, 9232190257, 9232090192, 9231910075, 9231910039, 9233421708, 9233421682, 9233421640, 9233421632, 9233421620, 9233391516, 9233391506, 9233391483, 9233351441, 9233321385, 9233321338, 9233311330, 9233311329, 9233311318, 9233311315, 9233311313, 9233311307, 9233311305, 9233311281, 9233311276, 9233311253, 9233311247, 9233311233, 9233201165, 9232961012, 9232961011, 9232961005, 9232961004, 9232960951, 9232750919, 9232750888, 9232750882, 9232750869, 9232750862, 9232750847, 9232750835, 9232750815, 9232720778, 9232490682, 9232490649, 9232490599, 9232490592, 9232490570, 9232330439, 9232300381, 9232750928, 9232490576, 9232490565, 9232330499, 9232330497, 9232330494, 9232330492, 9232330447, 9232330444, 9232330420, 9232300407, 9232300404, 9232300401, 9232300391, 9232190369, 9232190297, 9233421748, 9233421747, 9233421738, 9233421727, 9233421724, 9233421723, 9233421721, 9233421720, 9233421719, 9233421714, 9233421707, 9233421706, 9233421701, 9233421700, 9233421699, 9233421696, 9233421683, 9233421680, 9233421673, 9233421668, 9233421667, 9233421663, 9233421656, 9233421655, 9233421649, 9233421648, 9233421644, 9233421643, 9233421642, 9233421641, 9233421639, 9233421637, 9233421636, 9233421629, 9233421624, 9233421623, 9233421619, 9233421609, 9233421608, 9233421605, 9233421604, 9233421603, 9233421600, 9233421599, 9233421598, 9233421596, 9233421595, 9233421594, 9233421585, 9233391563, 9233391558, 9233391534, 9233391532, 9233391525, 9233391521, 9233391520, 9233391512, 9233391503, 9233391499, 9233391482, 9233331127, 9233331094, 9233331092, 9233331086, 9233331085, 9233331083, 9233331080, 9233331074, 9233331072, 9233331069, 9233331065, 9233331064, 9233331061, 9233331059, 9233331058, 9233331054, 9233331053, 9233331049, 9233331047, 9233331046, 9233331043, 9233331042, 9233331038, 9233331037, 9233331035, 9233331033, 9233201195, 9233201170, 9233201161, 9233201154, 9233201153, 9233201148, 9233201140, 9232961010, 9232961009, 9232961006, 9232961003, 9232960999, 9232960995, 9232960994, 9232960993, 9232960992, 9232960988, 9232960985, 9232960983, 9232960981, 9232960978, 9232960977, 9232960970, 9232960968, 9232960966, 9232960963, 9232960961, 9232960960, 9232960959, 9232960956, 9232960955, 9232960952, 9232960950, 9232960948, 9232960945, 9232960944, 9232960940, 9232960939, 9232960937, 9232960936, 9232960934, 9232960933, 9232750932, 9232750930, 9232750929, 9232750926, 9232750924, 9232750923, 9232750922, 9232750921, 9232750920, 9232750918, 9232750917, 9232750916, 9232750915, 9232750914, 9232750913, 9232750912, 9232750909, 9232750908, 9232750907, 9232750906, 9232750905, 9232750896, 9232750893, 9232750892, 9232750875, 9232750872, 9232750871, 9232750863, 9232750860, 9232750859, 9232750856, 9232750854, 9232750853, 9232750850, 9232750849, 9232750844, 9232750842, 9232750841, 9232750839, 9232750838, 9232750825, 9232750820, 9232750812, 9232720750, 9232720737, 9232490731, 9232490703, 9232490694, 9232490686, 9232490684, 9232490678, 9232490676, 9232490675, 9232490674, 9232490673, 9232490661, 9232490656, 9232490652, 9232490650, 9232490646, 9232490643, 9232490641, 9232490639, 9232490638, 9232490635, 9232490628, 9232490627, 9232490626, 9232490625, 9232490624, 9232490623, 9232490621, 9232490619, 9232490617, 9232490601, 9232490596, 9232490595, 9232490591, 9232490587, 9232490584, 9232490583, 9232490580, 9232490579, 9232490574, 9232490573, 9232490571, 9232490569, 9232490568, 9232490567, 9232490566, 9232490564, 9232490563, 9232490562, 9232490561, 9232490560, 9232490558, 9232490557, 9232490556, 9232490555, 9232490554, 9232490553, 9232490552, 9232490551, 9232490550, 9232490549, 9232490548, 9232490547, 9232490546, 9232490544, 9232490543, 9232490542, 9232490541, 9232490540, 9232490539, 9232490538, 9232490537, 9232490535, 9232490534, 9232490533, 9232420531, 9232420529, 9232420527, 9232420526, 9232420525, 9232420524, 9232420523, 9232420521, 9232420520, 9232420519, 9232420518, 9232420517, 9232420516, 9232420515, 9232420514, 9232420513, 9232420512, 9232420511, 9232420508, 9232420506, 9232420504, 9232420502, 9232340501, 9232330500, 9232330498, 9232330496, 9232330495, 9232330491, 9232330490, 9232330489, 9232330488, 9232330487, 9232330486, 9232330484, 9232330454, 9232330453, 9232330452, 9232330451, 9232330450, 9232330449, 9232330446, 9232330445, 9232330442, 9232330441, 9232330438, 9232330436, 9232330435, 9232330434, 9232330432, 9232330430, 9232330429, 9232330428, 9232330427, 9232330424, 9232330423, 9232330422, 9232330421, 9232330419, 9232330418, 9232330416, 9232330415, 9232330413, 9232330411, 9232330410, 9232330409, 9232330408, 9232300406, 9232300405, 9232300402, 9232300399, 9232300398, 9232300396, 9232300395, 9232300394, 9232300393, 9232300392, 9232300390, 9232300389, 9232300388, 9232300387, 9232300379, 9232300378, 9232220377, 9232210376, 9232190371, 9232190368, 9232190367, 9232190365, 9232190363, 9232190360, 9232190359, 9232190356, 9232190355, 9232190354, 9232190349, 9232190345, 9232190341, 9232190339, 9232190336, 9232190334, 9232190329, 9232190328, 9232190322, 9232190317, 9232190316, 9232190315, 9232190311, 9232190310, 9232190308, 9232190307, 9232190305, 9232190303, 9232190302, 9232190301, 9232190299, 9232190298, 9232190295, 9232190294, 9232190293, 9232190291, 9232190289, 9232190288, 9232190287, 9232190285, 9232190284, 9232190283, 9232190282, 9232190281, 9232190280, 9232190279, 9232190278, 9232190276, 9232190274, 9232190273, 9232190272, 9232190270, 9232190267, 9232190265, 9232190264, 9232190261, 9232190260, 9232190258, 9232190255, 9232190254, 9232190253, 9232190252, 9232190251, 9232190250, 9232190248, 9232190247, 9232190244, 9232190243, 9232190241, 9232190239, 9232190233, 9232190232, 9232190231, 9232190228, 9232090227, 9232090226, 9232090224, 9232090221, 9232090218, 9232090216, 9232090211, 9232090199, 9232090191, 9232090180, 9232090178, 9232090119, 9232090118, 9232090115, 9232090112, 9232090108, 9232090101, 9232090098, 9232090088, 9231910065, 9231910059, 9231910052, 9231910048, 9231910042, 9231910033, 9233421689, 9232750904, 9232750843, 9232750810, 9232720796, 9232490654, 9232490651, 9232420530, 9232330483, 9232300403, 9232300382, 9232190353, 9232190259, 9232720771, 9233421762, 9233311333, 9232750866, 9232490726, 9232490665, 9233321367, 9233321352, 9233311321, 9233311293, 9233201230, 9233201157, 9232750910, 9232750903, 9232750901, 9232750897, 9232750895, 9232750894, 9232750891, 9232750890, 9232750889, 9232750887, 9232750885, 9232750881, 9232750880, 9232750879, 9232750878, 9232750877, 9232750876, 9232750874, 9232750870, 9232750868, 9232750867, 9232750864, 9232750861, 9232750858, 9232750857, 9232750852, 9232750851, 9232750848, 9232750846, 9232750837, 9232750836, 9232750834, 9232750831, 9232750828, 9232750827, 9232750826, 9232750823, 9232750822, 9232750821, 9232750818, 9232750817, 9232750816, 9232750813, 9232750811, 9232750809, 9232750808, 9232750805, 9232750804, 9232750803, 9232750802, 9232750801, 9232720799, 9232720797, 9232720795, 9232720793, 9232720792, 9232720788, 9232720787, 9232720785, 9232720782, 9232720781, 9232720779, 9232720775, 9232720772, 9232720770, 9232720768, 9232720767, 9232720766, 9232720765, 9232720763, 9232720762, 9232720761, 9232720760, 9232720758, 9232720755, 9232720752, 9232720751, 9232720749, 9232720748, 9232720746, 9232720745, 9232720744, 9232720743, 9232720742, 9232720741, 9232720740, 9232720738, 9232720735, 9232720734, 9232720733, 9232490732, 9232490729, 9232490728, 9232490723, 9232490722, 9232490721, 9232490720, 9232490719, 9232490718, 9232490717, 9232490716, 9232490715, 9232490714, 9232490713, 9232490712, 9232490711, 9232490710, 9232490709, 9232490707, 9232490706, 9232490704, 9232490702, 9232490701, 9232490700, 9232490699, 9232490698, 9232490695, 9232490693, 9232490691, 9232490690, 9232490689, 9232490687, 9232490683, 9232490681, 9232490680, 9232490679, 9232490677, 9232490670, 9232490669, 9232490668, 9232490666, 9232490664, 9232490663, 9232490662, 9232490660, 9232490657, 9232490655, 9232490653, 9232490648, 9232490647, 9232490645, 9232490644, 9232490642, 9232490637, 9232490636, 9232490632, 9232490630, 9232490629, 9232490622, 9232490620, 9232490613, 9232490611, 9232490610, 9232490609, 9232490608, 9232490607, 9232490606, 9232490605, 9232490604, 9232490603, 9232490600, 9232490598, 9232490597, 9232490593, 9232490589, 9232490588, 9232490586, 9232490585, 9232490581, 9232490577, 9232490545, 9232490532, 9232330493, 9232330440, 9232330414, 9232300397, 9232300385, 9232300383, 9232300380, 9232190304, 9232190269, 9232190263, 9232090186, 9231910067, 9232750855, 9232490708, 9232490594, 9233532105, 9233522073, 9233522058, 9233421815, 9233421796, 9233421795, 9233421784, 9233421757, 9233421745, 9233421734, 9233421732, 9233421725, 9233421676, 9233421671, 9233421591, 9233391518, 9233522071, 9233421811, 9233421803, 9233421802, 9233421799, 9233421775, 9233421774, 9233421773, 9233421772, 9233421769, 9233421760, 9233421759, 9233421754, 9233421751, 9233421750, 9233421742, 9233421737, 9233421729, 9233421709, 9233421704, 9233421690, 9233421657, 9233421651, 9233421638, 9233421625, 9233391515, 9233201182, 9232750840, 9232190266. |
Recalling Firm/ Manufacturer |
Fresenius Kabi USA, LLC 50 High St Ste 50 North Andover MA 01845-2620
|
| For Additional Information Contact | 978-775-8050 |
Manufacturer Reason for Recall | Potential pneumatic valve failure that may cause the LVP to stop an active infusion and/or could prevent the pump from being used. |
FDA Determined Cause 2 | Process control |
| Action | Fresenius Kabi notified consignees on 12/06/2024 via "URGENT - Medical Device Recall" letter. Consignees were instructed to do the following:
1. Review the facilities list of affected Serial Numbers and post the list in a public place.
2. If an LVP exhibits a Pump Problem during use, remove the LVP from circulation, and notify Fresenius Kabi.
3. Inform potential users of the product within your organization of this notification.
4. Notify customers if product was further distributed.
5. Complete and return the Customer Reply Form included in the letter to acknowledge receipt.
Customers will be contacted by a Fresenius Kabi service representative to create a plan for repairing any affected LVPs at your facility at no cost. Fresenius Kabi will also repair LVPs that may encounter pneumatic valve failures at no additional charge.
For customer support and questions, call 1-855-354-6387 normal hours are from Monday thru Friday 8:30am-5pm Easter Time, after hours supported by voicemail messaging or email: Ivenix_support@fresnius-kabi.com. |
| Quantity in Commerce | 1,520 units |
| Distribution | US distribution to states of: ID, MN, NV, TX, VA, WI. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = FRN
|
|
|
|