Date Initiated by FirmJanuary 06, 2025
Date PostedJanuary 28, 2025
Recall Status1 Open3, Classified
Recall NumberZ-1038-2025
Recall Event ID 96090
510(K)NumberK960279 
Product Classification Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer - Product Code JWH
ProductNexGen LCCK Legacy Articular Surface with Locking Screw - Striped Purple/C,D - Height 14 mm, Item Number 00-5994-022-14
Code Information UDI-DI: 00889024635647; Lot Numbers: 66602503 66520665 66881918 66949906
Recalling Firm/
Manufacturer
Zimmer, Inc.
1800 W Center St
Warsaw IN 46580-2304
For Additional Information Contact411 Technical Services
574-371-3071
Manufacturer Reason
for Recall
It was identified internally that two commingle events occurred where the metal support post within the polyethylene articular surface was assembled incorrectly for three units within the scope.
FDA Determined
Cause 2
Process change control
ActionAn URGENT MEDICAL DEVICE RECALL - REMOVAL notification letter dated 1/6/25 was sent to customers. Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the recall notice for hospitals and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form shall be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to orporateQuality.PostMarket@zimmerbiomet.com. Distributor Responsibilities: 1. Review this notification and ensure that affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Immediately return all affected product from your distributorship and from affected hospitals within your territory. a. Complete Attachment 1 Inventory Return Certification Form for each return and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form must be returned even if you do not have affected products available to return in your territory. b. For International Returns, request an IRA by emailing zimmerbiometintlirareques
Quantity in Commerce155 units
DistributionWorldwide - US Nationwide distribution in the states of CA, FL, GA, KS, MO, NE, NJ, PA, TN and the countries of CHINA, INDIA, KOREA, MALAYSIA, NETHERLANDS, SINGAPORE, TAIWAN.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = JWH
-
-