| | Class 2 Device Recall Novation XLE |  |
| Date Initiated by Firm | December 31, 2024 |
| Date Posted | February 12, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1166-2025 |
| Recall Event ID |
96102 |
| 510(K)Number | K173583 |
| Product Classification |
Hip, semi-constrained, cemented, metal/ceramic/polymer + additive, porous uncemented - Product Code OQI
|
| Product | Novation XLE
Item Number / Product Description:
140-22-70 NV EHXL NTL LNR G00 22MM
140-28-50 NV EHXLNTRL LNR G0 28MM
140-28-51 NV EHXLNTRL LNR G1 28MM
140-28-52 NV EHXL NTRL LNR G228MM
140-32-51 NV EHXL NTRL LNR G132MM
140-32-52 NV EHXL NTRL LNR G232MM
140-32-53 NV EHXL NTRL LNR G332MM
140-32-54 NV EHXLNTRL LNR G4 32MM
140-32-55 NV EHXLNTRL LNR G5 32MM
140-36-52 NV EHXLNTRL LNR G2 36MM
140-36-53 NV EHXL NTRL LNR G336MM
140-36-54 NV EHXLNTRL LNR G4 36MM
140-36-55 NV EHXLNTRL LNR G5 36MM
140-40-53 NV EHXLNTRL LNR G3 40MM
140-40-54 NV EHXLNTRL LNR G4 40MM
140-40-55 NV EHXLNTRL LNR G5 40MM
142-22-68 NV EHXL ALPLNR G00 22MM
142-22-70 NV EHXL LIPLNR G00 22MM
142-28-50 NV EHXL LIP LNR G0 28MM
142-28-51 NV EHXL LIP LNR G1 28MM
142-28-52 NV EHXL LIP LNR G2 28MM
142-28-60 NV EHXL ALIPLNR G0 28MM
142-28-61 NV EHXL ALIP LNR G1 28MM
142-28-62 NV EHXL ALIP LNR G2 28MM
142-32-51 NV EHXL LIP LNR G1 32MM
142-32-52 NV EHXL LIP LNR G2 32MM
142-32-53 NV EHXL LIP LNR G3 32MM
142-32-54 NV EHXL LIP LNR G4 32MM
142-32-55 NV EHXL LIP LNR G5 32MM
142-32-61 NV EHXL ALIP LNR G1 32MM
142-32-62 NV EHXL ALIP LNR G2 32MM
142-32-63 NV EHXL ALIP LNR G3 32MM
142-32-64 NV EHXL ALIP LNR G4 32MM
142-32-65 NV EHXL ALIP LNR G5 32MM
142-36-52 NV EHXL LIP LNR G2 36MM
142-36-53 NV EHXL LIP LNR G3 36MM
142-36-54 NV EHXL LIP LNR G4 36MM
142-36-55 NV EHXL LIP LNR G5 36MM
142-36-62 NV EHXL ALIP LNR G2 36MM
142-36-63 NV EHXL ALIP LNR G3 36MM
142-36-64 NV EHXL ALIP LNR G4 36MM
142-36-65 NV EHXL ALIP LNR G5 36MM
142-40-53 NV EHXL LIP LNR G3 40MM
142-40-54 NV EHXL LIP LNR G4 40MM
142-40-55 NV EHXL LIP LNR G5 40MM
142-40-63 NV EHXL ALIP LNR G3 40MM
142-40-64 NV EHXL ALIP LNR G4 40MM
142-40-65 NV EHXL ALIP LNR G5 40MM
146-22-70 NV EHXL LATLNR G00 22MM
146-28-51 NV EHXL LAT LNR G1 28MM
146-32-51 NV EHXL LAT LNR G1 32MM
146-32-52 NV EHXL LAT LNR G2 32MM
146-32-53 NV EHXL LAT LNR G3 32MM
146-32-54 NV EHXL LAT LNR G4 32MM
146-32-55 NV EHXL LAT LNR G5 32MM
146-36-52 NV EHXL LAT LNR G2 36MM
146-36-53 NV EHXL LAT LNR G3 36MM
146-36-54 NV EHXL LAT LNR G4 36MM
146-36-55 NV EHXL LAT LNR G5 36MM
146-40-53 NV EHXL LAT LNR G3 40MM
146-40-54 NV EHXL LAT LNR G4 40MM
146-40-55 NV EHXL LAT LNR G5 40MM
148-32-51 NV EHXL 10 LNR G1 32MM
148-36-52 NV EHXL 10 LNR G2 36MM
148-36-53 NV EHXL 10 LNR G3 36MM
148-36-54 NV EHXL 10 LNR G4 36MM
148-36-55 NV EHXL 10 LNR G5 36MM
148-40-53 NV EHXL 10 LNR G3 40MM
148-40-54 NV EHXL 10 LNR G4 40MM
148-40-55 NV EHXL 10 LNR G5 40MM
All Exactech Hip Systems are indicated for use in skeletally mature individuals undergoing primary surgery for hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment. |
| Code Information |
Item Number / Product Description/ UDI-DI code:
140-22-70 NV EHXL NTL LNR G00 22MM 10885862537225
140-28-50 NV EHXLNTRL LNR G0 28MM 10885862537232
140-28-51 NV EHXLNTRL LNR G1 28MM 10885862537249
140-28-52 NV EHXL NTRL LNR G228MM 10885862537256
140-32-51 NV EHXL NTRL LNR G132MM 10885862537263
Serial Numbers:
6594943
6970850
6595376
6595380
6858220
6858230
6973123
6594924
6970840
6595383
6595385
6595381
6970835
6970838
6970842
6595922
6510276
6973139
140-32-52 NV EHXL NTRL LNR G232MM 10885862537270
140-32-53 NV EHXL NTRL LNR G332MM 10885862537287
140-32-54 NV EHXLNTRL LNR G4 32MM 10885862537294
140-32-55 NV EHXLNTRL LNR G5 32MM 10885862537300
140-36-52 NV EHXLNTRL LNR G2 36MM 10885862537317
Serial Numbers:
6558356
6422337
6631675
6504196
6631677
6631695
6631712
6844304
6631689
6631699
6631725
6558360
6844285
6844298
6844281
6348961
6504177
6504195
6504202
6504204
6631686
6844321
140-36-53 NV EHXL NTRL LNR G336MM 10885862537324
Serial Numbers:
6600076
6600101
6600054
6600057
6600093
140-36-54 NV EHXLNTRL LNR G4 36MM 10885862537331
140-36-55 NV EHXLNTRL LNR G5 36MM 10885862537348
140-40-53 NV EHXLNTRL LNR G3 40MM 10885862537355
Serial Numbers:
6510314
6510327
140-40-54 NV EHXLNTRL LNR G4 40MM 10885862537362
140-40-55 NV EHXLNTRL LNR G5 40MM 10885862537379
142-22-68 NV EHXL ALPLNR G00 22MM 10885862537546
142-22-70 NV EHXL LIPLNR G00 22MM 10885862537386
142-28-50 NV EHXL LIP LNR G0 28MM 10885862537393
142-28-51 NV EHXL LIP LNR G1 28MM 10885862537409
142-28-52 NV EHXL LIP LNR G2 28MM 10885862537416
142-28-60 NV EHXL ALIPLNR G0 28MM 10885862537553
142-28-61 NV EHXL ALIP LNR G1 28MM 10885862537560
142-28-62 NV EHXL ALIP LNR G2 28MM 10885862537577
142-32-51 NV EHXL LIP LNR G1 32MM 10885862537423
142-32-52 NV EHXL LIP LNR G2 32MM 10885862537430
142-32-53 NV EHXL LIP LNR G3 32MM 10885862537447
142-32-54 NV EHXL LIP LNR G4 32MM 10885862537454
142-32-55 NV EHXL LIP LNR G5 32MM 10885862537461
142-32-61 NV EHXL ALIP LNR G1 32MM 10885862537584
Serial Numbers:
7040076
6842549
6554448
6596005
6554464
6554458
6866177
6596019
6554457
6595993
6503270
142-32-62 NV EHXL ALIP LNR G2 32MM 10885862537591
142-32-63 NV EHXL ALIP LNR G3 32MM 10885862537607
142-32-64 NV EHXL ALIP LNR G4 32MM 10885862537614
142-32-65 NV EHXL ALIP LNR G5 32MM 10885862537621
142-36-52 NV EHXL LIP LNR G2 36MM 10885862537478
142-36-53 NV EHXL LIP LNR G3 36MM 10885862537485
142-36-54 NV EHXL LIP LNR G4 36MM 10885862537492
142-36-55 NV EHXL LIP LNR G5 36MM 10885862537508
142-36-62 NV EHXL ALIP LNR G2 36MM 10885862537638
Serial Numbers:
6831997
6679288
6831980
7032226
6679275
6831983
6335167
6832001
6555554
7032248
6679286
6679321
6831992
6927281
142-36-63 NV EHXL ALIP LNR G3 36MM 10885862537645
142-36-64 NV EHXL ALIP LNR G4 36MM 10885862537652
142-36-65 NV EHXL ALIP LNR G5 36MM 10885862537669
142-40-53 NV EHXL LIP LNR G3 40MM 10885862537515
142-40-54 NV EHXL LIP LNR G4 40MM 10885862537522
142-40-55 NV EHXL LIP LNR G5 40MM 10885862537539
142-40-63 NV EHXL ALIP LNR G3 40MM 10885862537676
Serial Numbers:
6452595
142-40-64 NV EHXL ALIP LNR G4 40MM 10885862537683
142-40-65 NV EHXL ALIP LNR G5 40MM 10885862537690
146-22-70 NV EHXL LATLNR G00 22MM 10885862537751
146-28-51 NV EHXL LAT LNR G1 28MM 10885862537768
146-32-51 NV EHXL LAT LNR G1 32MM 10885862537775
146-32-52 NV EHXL LAT LNR G2 32MM 10885862537782
146-32-53 NV EHXL LAT LNR G3 32MM 10885862537799
146-32-54 NV EHXL LAT LNR G4 32MM 10885862537805
146-32-55 NV EHXL LAT LNR G5 32MM 10885862537812
146-36-52 NV EHXL LAT LNR G2 36MM 10885862537829
146-36-53 NV EHXL LAT LNR G3 36MM 10885862537836
146-36-54 NV EHXL LAT LNR G4 36MM 10885862537843
146-36-55 NV EHXL LAT LNR G5 36MM 10885862537850
146-40-53 NV EHXL LAT LNR G3 40MM 10885862537867
Serial Numbers:
6602255
6602261
6602265
6602268
6602269
6602270
6602272
6602256
6602263
6602267
6602277
6602257
6602258
6602273
146-40-54 NV EHXL LAT LNR G4 40MM 10885862537874
146-40-55 NV EHXL LAT LNR G5 40MM 10885862537881
148-32-51 NV EHXL 10¿ LNR G1 32MM 10885862537898
148-36-52 NV EHXL 10¿ LNR G2 36MM 10885862537904
148-36-53 NV EHXL 10¿ LNR G3 36MM 10885862537911
148-36-54 NV EHXL 10¿ LNR G4 36MM 10885862537928
148-36-55 NV EHXL 10¿ LNR G5 36MM 10885862537935
148-40-53 NV EHXL 10¿ LNR G3 40MM 10885862537942
148-40-54 NV EHXL 10¿ LNR G4 40MM 10885862537959
148-40-55 NV EHXL 10¿ LNR G5 40MM 10885862537966
|
| FEI Number |
1038671
|
Recalling Firm/ Manufacturer |
Exactech, Inc. 2320 Nw 66th Ct Gainesville FL 32653-1630
|
| For Additional Information Contact | Matthew Collins 352-377-1140 |
Manufacturer Reason for Recall | Due to the lack of the Ethylene Vinyl Alcohol (EVOH) barrier layer. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | On 12/31/2024, the firm sent an "URGENT VOLUNTARY MEDICAL DEVICE RECALL" letter via FedEx to customers initially informing them that the firm was recalling specific hip polyethylene liner lots that were inadvertently packaged without the ethylene vinyl alcohol (EVOH) layer in the innermost bag.
An updated customer notification letter was communicated to affected customer via email on 01/17/2025.
Customers are instructed to:
-Immediate Discontinuation: Immediately discontinue use of all affected products
-Quarantine and Return: Segregate and quarantine affected products and arrange for their return to Exactech. The Recall Confirmation Form section of this communication provides product return instructions.
Clinical Guidance:
1. Patient Monitoring: Surgeons should regularly monitor patients with affected devices for potential risks such as device wear, failure, component cracking or fracture, new or worsening pain, bone loss and/or swelling, as per the instructions for use.
2. Diagnostic Considerations: Diagnostic imaging, such as X-rays, may be considered if device failure is suspected.
3. Revision Considerations: Revision of well-functioning devices is not recommended for patients.
4. Resources: Healthcare providers and patients with questions can access educational resources on our website exac.com/recall-information and use the device serial look-up tool.
Public Contact - 352-377-1140 or email matt.collins@exac.com |
| Quantity in Commerce | 87 |
| Distribution | U.S.: AR, CA, CO, FL, GA, HI, IL, IN, KS, KY, LA, MD, MN, NV, NY, OH, OK, PR,SC, TN, TX and VA
O.U.S.: Argentina, Colombia, Germany, Guatemala, Japan and Spain |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = OQI
|
|
|
|