| | Class 2 Device Recall Medline Convenience kits |  |
| Date Initiated by Firm | March 03, 2025 |
| Date Posted | April 17, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1597-2025 |
| Recall Event ID |
96510 |
| Product Classification |
Ear, nose, and throat surgical tray - Product Code OGR
|
| Product | Medline Convenience kits used for various procedures:
1) T A PACK, Model Number: DYNJ67791A;
2) NASAL PACK, Model Number: DYNJ66514;
3) ENT PACK, Model Number: DYNJ67775A;
4) T/A ENDO PACK, Model Number: DYNJ83158;
5) T & A PACK, Model Number: DYNJ0634878I;
6) PPM-ICD PACK-LF, Model Number: DYNJ15949D;
7) TONSIL PACK, Model Number: DYNJ44525C;
8) ENT TRAY, Model Number: DYNJ44391K;
9) ENT PACK, Model Number: DYNJ49300D;
10) NASAL PACK, Model Number: DYNJ61555B;
11) HEAD AND NECK, Model Number: DYNJ57287A;
12) SINUS ENDOSCOPY-LF, Model Number: DYNJ36720F;
13) ENT PACK, Model Number: DYNJ54035B;
14) ENT PACK, Model Number: DYNJ36516;
15) FESS PACK, Model Number: DYNJ59502B;
16) T AND A PACK, Model Number: DYNJ80790;
17) PK,ENT, Model Number: DYNJ57583;
18) ENT/LAP PACK, Model Number: DYNJ84876;
19) TONSIL PACK, Model Number: DYNJ50109B;
20) TONSIL, Model Number: DYNJ908452C;
21) CSS SINUS PACK, Model Number: DYNJ17005I;
22) ENT, Model Number: DYNJ910514;
23) ADULT ENT - OAKBEND, Model Number: DYNJ905762;
24) EENT SPLIT PACK-LF, Model Number: LLSCAENT01;
25) ENT PACK, Model Number: DYNJ31453D;
26) ENT PACK, Model Number: DYNJ59447A;
27) TONSIL PACK, Model Number: DYNJ26441R;
28) ORAL ENT PACK, Model Number: DYNJ58408A;
29) ENT PACK, Model Number: DYNJ51085;
30) DR LANO NASAL CHRISTUS, Model Number: DYNJ61196A;
31) ENT PACK, Model Number: DYNJ66178D;
32) BWNBORO EENT PACK-LF, Model Number: DYNJ37551A;
33) HEAD AND NECK PACK, Model Number: DYNJ68872F;
34) EENT PACK, Model Number: DYNJ50301C;
35) ENT PACK, Model Number: DYNJ40498B;
36) SHN-C-THYROIDECTOMY-LF, Model Number: DYNJ46869A;
37) MINOR CYSTO PACK, Model Number: DYNJ58727D;
38) TONSILS AND ADENOIDS, Model Number: DYNJ45171C;
39) T&A PACK, Model Number: DYNJ56543J;
40) ENT PACK, Model Number: DYNJ54035B;
41) TONSIL PACK, Model Number: DYNJ50109B;
42) ENT PACK, Model Number: DYNJ40139D;
43) NASAL/SEPTO PACK, Model Number: DYNJ32099A;
44) EENT SPLIT PACK-LF, Model Number: LLSCAENT01;
45) ENT PACK, Model Number: DYNJ53573A;
46) CHS ENT PACK II, Model Number: DYNJCH908;
47) SIMPLE NECK DISSECTION, Model Number: DYNJ47481F;
48) TONSIL PACK, Model Number: DYNJ81479A;
49) E.N.T. TRAY, Model Number: DYNJ22357K;
50) NASAL PACK, Model Number: DYNJ39317B;
51) HEAD AND NECK PACK, Model Number: DYNJ54321D;
52) T A PACK, Model Number: DYNJ67791A;
53) ENT PACK, Model Number: DYNJ42451B;
54) HEAD AND NECK PACK-FOX, Model Number: DYNJ53348C;
55) HEAD AND NECK PACK, Model Number: DYNJ58596B;
56) SINUS PACK, Model Number: DYNJ56262;
57) ENT PACK, Model Number: DYNJ51085;
58) TONSIL T AND A PACK SHD, Model Number: DYNJ68417C;
59) ENT PACK, Model Number: DYNJ48595;
60) T AND A PACK-LF, Model Number: DYNJ52611;
61) HEAD AND NECK PACK, Model Number: DYNJ81897B;
62) ENSEMBLE ORL THYROIDE-LF, Model Number: DYNJ43390A;
63) EENT PACK-LF, Model Number: DYNJ34728D;
64) ENT, Model Number: DYNJ910514;
65) TONSIL ADENOID CKHA-LF, Model Number: DYNJ41771C;
66) ENT PACK, Model Number: DYNJ68679B;
67) PACK,T & A ASC, Model Number: DYNJ906914;
68) EENT SPLIT PACK-LF, Model Number: LLSCAENT01;
69) ENT PACK, Model Number: DYNJ83603;
70) ENT PACK, Model Number: DYNJ32247;
71) PPM-ICD PACK-LF, Model Number: DYNJ15949D;
72) ENT PACK, Model Number: DYNJ31453D;
73) SINOSCOPY PACK-LF, Model Number: DYNJ25600L;
74) T & A PACK, Model Number: DYNJ38978C;
75) CSC-N T&A, Model Number: DYNJ50870D;
76) ENT I-LF, Model Number: DYNJS0702;
77) ENT KIT, Model Number: DYNJ902836F;
78) ENT PACK, Model Number: DYNJ66039B;
79) TONSIL CDS, Model Number: DYNJ907844D;
80) PACK,T & A ASC, Model Number: DYNJ906914 |
| Code Information |
1) DYNJ67791A, UDI-DI: 10193489941715(each), 40193489941716(case), Lot Number: 24IBE758;
2) DYNJ66514, UDI-DI: 10193489359282(each), 40193489359283(case), Lot Number: 24IBE817;
3) DYNJ67775A, UDI-DI: 10193489996258(each), 40193489996259(case), Lot Number: 24IBG346;
4) DYNJ83158, UDI-DI: 10195327236885(each), 40195327236886(case), Lot Number: 24IBG449;
5) DYNJ0634878I, UDI-DI: 10195327694500(each), 40195327694501(case), Lot Number: 24IBG548;
6) DYNJ15949D, UDI-DI: 10195327573805(each), 40195327573806(case), Lot Number: 24IBG570;
7) DYNJ44525C, UDI-DI: 10193489232042(each), 40193489232043(case), Lot Number: 24IBG608;
8) DYNJ44391K, UDI-DI: 10195327457648(each), 40195327457649(case), Lot Number: 24IBG848;
9) DYNJ49300D, UDI-DI: 10193489297287(each), 40193489297288(case), Lot Number: 24IBH734;
10) DYNJ61555B, UDI-DI: 10195327608101(each), 40195327608102(case), Lot Number: 24IBH769;
11) DYNJ57287A, UDI-DI: 10193489854152(each), 40193489854153(case), Lot Number: 24IBH959;
12) DYNJ36720F, UDI-DI: 10198459079306(each), 40198459079307(case), Lot Number: 24IBQ081;
13) DYNJ54035B, UDI-DI: 10195327680541(each), 40195327680542(case), Lot Number: 24IBQ933;
14) DYNJ36516, UDI-DI: 10884389544112(each), 40884389544113(case), Lot Number: 24IBR105;
15) DYNJ59502B, UDI-DI: 10193489372939(each), 40193489372930(case), Lot Number: 24IBR121;
16) DYNJ80790, UDI-DI: 10195327079482(each), 40195327079483(case), Lot Number: 24IBR337;
17) DYNJ57583, UDI-DI: 10889942755332(each), 40889942755333(case), Lot Number: 24IBR873;
18) DYNJ84876, UDI-DI: 10195327431778(each), 40195327431779(case), Lot Number: 24IBR895;
19) DYNJ50109B, UDI-DI: 10198459115776(each), 40198459115777(case), Lot Number: 24IBS352;
20) DYNJ908452C, UDI-DI: 10195327610937(each), 40195327610938(case), Lot Number: 24JBD799;
21) DYNJ17005I, UDI-DI: 10193489241860(each), 40193489241861(case), Lot Number: 24JBF130;
22) DYNJ910514, UDI-DI: 10198459021411(each), 40198459021412(case), Lot Number: 24JBK654;
23) DYNJ905762, UDI-DI: 10193489507966(each), 40193489507967(case), Lot Number: 24JBK682;
24) LLSCAENT01, UDI-DI: 10080196120802(each), 40080196120803(case), Lot Number: 24JBK897;
25) DYNJ31453D, UDI-DI: 10195327286224(each), 40195327286225(case), Lot Number: 24JBM367;
26) DYNJ59447A, UDI-DI: 10193489707663(each), 40193489707664(case), Lot Number: 24JBM926;
27) DYNJ26441R, UDI-DI: 10198459085888(each), 40198459085889(case), Lot Number: 24JBN335;
28) DYNJ58408A, UDI-DI: 10889942969029(each), 40889942969020(case), Lot Number: 24JBN611;
29) DYNJ51085, UDI-DI: 10889942117284(each), 40889942117285(case), Lot Number: 24JBN781;
30) DYNJ61196A, UDI-DI: 10195327604240(each), 40195327604241(case), Lot Number: 24JBN797;
31) DYNJ66178D, UDI-DI: 10198459100307(each), 40198459100308(case), Lot Number: 24JBN811;
32) DYNJ37551A, UDI-DI: 10888277585720(each), 40888277585721(case), Lot Number: 24JBT022;
33) DYNJ68872F, UDI-DI: 10198459079269(each), 40198459079260(case), Lot Number: 24JBT065;
34) DYNJ50301C, UDI-DI: 10198459052330(each), 40198459052331(case), Lot Number: 24JBT456;
35) DYNJ40498B, UDI-DI: 10193489806779(each), 40193489806770(case), Lot Number: 24JBV199;
36) DYNJ46869A, UDI-DI: 10198459069949(each), 40198459069940(case), Lot Number: 24JBV309;
37) DYNJ58727D, UDI-DI: 10198459048425(each), 40198459048426(case), Lot Number: 24JBV608;
38) DYNJ45171C, UDI-DI: 10889942485260(each), 40889942485261(case), Lot Number: 24JBV814;
39) DYNJ56543J, UDI-DI: 10195327453602(each), 40195327453603(case), Lot Number: 24JBW129;
40) DYNJ54035B, UDI-DI: 10195327680541(each), 40195327680542(case), Lot Number: 24JBW183;
41) DYNJ50109B, UDI-DI: 10198459115776(each), 40198459115777(case), Lot Number: 24JBW623;
42) DYNJ40139D, UDI-DI: 10195327663506(each), 40195327663507(case), Lot Number: 24JBX313;
43) DYNJ32099A, UDI-DI: 10195327610890(each), 40195327610891(case), Lot Number: 24JBY276;
44) LLSCAENT01, UDI-DI: 10080196120802(each), 40080196120803(case), Lot Number: 24JBY278;
45) DYNJ53573A, UDI-DI: 10193489201635(each), 40193489201636(case), Lot Number: 24JBY404;
46) DYNJCH908, UDI-DI: 10193489393347(each), 40193489393348(case), Lot Number: 24KBA907;
47) DYNJ47481F, UDI-DI: 10195327067335(each), 40195327067336(case), Lot Number: 24KBB421;
48) DYNJ81479A, UDI-DI: 10195327378370(each), 40195327378371(case), Lot Number: 24KBB455;
49) DYNJ22357K, UDI-DI: 10193489926736(each), 40193489926737(case), Lot Number: 24KBB793;
50) DYNJ39317B, UDI-DI: 10889942775729(each), 40889942775720(case), Lot Number: 24KBC590;
51) DYNJ54321D, UDI-DI: 10195327679323(each), 40195327679324(case), Lot Number: 24KBC663;
52) DYNJ67791A, UDI-DI: 10193489941715(each), 40193489941716(case), Lot Number: 24KBC754;
53) DYNJ42451B, UDI-DI: 10193489740318(each), 40193489740319(case), Lot Number: 24KBC910;
54) DYNJ53348C, UDI-DI: 10195327453565(each), 40195327453566(case), Lot Number: 24KBD155;
55) DYNJ58596B, UDI-DI: 10195327635947(each), 40195327635948(case), Lot Number: 24KBD243;
56) DYNJ56262, UDI-DI: 10889942624331(each), 40889942624332(case), Lot Number: 24KBD661;
57) DYNJ51085, UDI-DI: 10889942117284(each), 40889942117285(case), Lot Number: 24KBE334;
58) DYNJ68417C, UDI-DI: 10195327341633(each), 40195327341634(case), Lot Number: 24KBE396;
59) DYNJ48595, UDI-DI: 10888277915183(each), 40888277915184(case), Lot Number: 24KBE590;
60) DYNJ52611, UDI-DI: 10889942300853(each), 40889942300854(case), Lot Number: 24KBF355;
61) DYNJ81897B, UDI-DI: 10198459155710(each), 40198459155711(case), Lot Number: 24KBF577;
62) DYNJ43390A, UDI-DI: 10889942670321(each), 40889942670322(case), Lot Number: 24KBG019;
63) DYNJ34728D, UDI-DI: 10198459061851(each), 40198459061852(case), Lot Number: 24KBG035;
64) DYNJ910514, UDI-DI: 10198459021411(each), 40198459021412(case), Lot Number: 24KBG072;
65) DYNJ41771C, UDI-DI: 10889942599189(each), 40889942599180(case), Lot Number: 24KBG178;
66) DYNJ68679B, UDI-DI: 10195327129095(each), 40195327129096(case), Lot Number: 24KBG186;
67) DYNJ906914, UDI-DI: 10193489280173(each), 40193489280174(case), Lot Number: 24KBG371;
68) LLSCAENT01, UDI-DI: 10080196120802(each), 40080196120803(case), Lot Number: 24KBG399;
69) DYNJ83603, UDI-DI: 10195327284886(each), 40195327284887(case), Lot Number: 24KBG953;
70) DYNJ32247, UDI-DI: 10884389203873(each), 40884389203874(case), Lot Number: 24KBG965;
71) DYNJ15949D, UDI-DI: 10195327573805(each), 40195327573806(case), Lot Number: 24KBG993;
72) DYNJ31453D, UDI-DI: 10195327286224(each), 40195327286225(case), Lot Number: 24KBG996;
73) DYNJ25600L, UDI-DI: 10889942686247(each), 40889942686248(case), Lot Number: 24KBH226;
74) DYNJ38978C, UDI-DI: 10195327583590(each), 40195327583591(case), Lot Number: 24KBH242;
75) DYNJ50870D, UDI-DI: 10195327165666(each), 40195327165667(case), Lot Number: 24KBH280;
76) DYNJS0702, UDI-DI: 10080196798728(each), 40080196798729(case), Lot Number: 24KBI176;
77) DYNJ902836F, UDI-DI: 10198459002816(each), 40198459002817(case), Lot Number: 24KBI638;
78) DYNJ66039B, UDI-DI: 10195327151997(each), 40195327151998(case), Lot Number: 24KBL783;
79) DYNJ907844D, UDI-DI: 10195327387334(each), 40195327387335(case), Lot Number: 24KBP945;
80) DYNJ906914, UDI-DI: 10193489280173(each), 40193489280174(case), Lot Number: 24LBA080 |
| FEI Number |
1417592
|
Recalling Firm/ Manufacturer |
MEDLINE INDUSTRIES, LP - Northfield
3 Lakes Dr
Northfield
IL
60093-2753
|
| For Additional Information Contact | Haley Barclay 866-359-1704 |
Manufacturer Reason for Recall | There is a potential for open/weak seals on purchased breather pouches used for packaging of select sterile convenience kits. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | An IMMEDIATE ACTION REQUIRED MEDLINE DEVICE RECALL notification letter dated 3/3/25 was sent to customers.
REQUIRED ACTION:
1.Immediately check your stock for the affected item number and the affected lot numbers which can be located within the recall portal.
2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form.
3.Please destroy any affected inventory. Medline will issue credit for the product based upon the quantity and lot numbers identified on the submitted response form
The login for completing the response form is: Website link: https://recalls.medline.com Recall Reference #: R-25-050-FG Recall Code: RECALL CODE
4.If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please request that customers document and destroy any affected product, and include your customer s quantities on your response form.
If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com. |
| Quantity in Commerce | 9216 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of Canada, Mexico, Panama, Virgin Islands. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|