| | Class 2 Device Recall LinkSymphoKnee |  |
| Date Initiated by Firm | April 03, 2025 |
| Date Posted | April 30, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1702-2025 |
| Recall Event ID |
96643 |
| 510(K)Number | K202924 |
| Product Classification |
Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer - Product Code JWH
|
| Product | Tibial Augment; Item Numbers: (1) 880-331/11 (Sz. 1-2 - cemented x-sm), (2) 880-331/12 (Sz. 1-2 - cemented sm), (3) 880-331/13 (Sz. 1-2 - cemented md), (4) 880-331/21 (Sz. 1-2 - cemented lg), (5) 880-331/22 (Sz. 1-2 - cemented x-lg), (6) 880-331/23 (Sz. 1-2 - cemented xx-lg), (7) 880-333/11 (Sz. 3-4 - cemented x-sm), (8) 880-333/12 (Sz. 3-4 - cemented sm), (9) 880-333/13 (Sz. 3-4 - cemented md), (10) 880-333/21 (Sz. 3-4 - cemented lg), (11) 880-333/22 (Sz. 3-4 - cemented x-lg), (12) 880-333/23 (Sz. 3-4 - cemented xx-lg), (13) 880-335/11 (Sz. 5-6 - cemented x-sm), (14) 880-335/12 (Sz. 5-6 - cemented sm), (15) 880-335/13 (Sz. 5-6 - cemented md), (16) 880-335/21 (Sz. 5-6 - cemented lg), (17) 880-335/22 (Sz. 5-6 - cemented lg), (18) 880-335/23 (Sz. 5-6 - cemented xx-lg), (19) 880-337/11 (Sz. 7-8 - cemented x-sm), (20) 880-337/12 (Sz. 5-6 - cemented sm), (21) 880-337/13 (Sz. 7-8 - cemented md), (22) 880-337/21 (Sz. 7-8 - cemented lg), (23) 880-337/22 (Sz. 7-8 - cemented x-lg), (24) 880-337/23 (Sz. 7-8 - cemented xx-lg), (25) 880-339/11 (Sz. 9-10 - cemented x-sm), (26) 880-339/12 (Sz. 9-10 - cemented sm), (27) 880-339/13 (Sz. 9-10 - cemented md), (28) 880-339/21 (Sz. 9-10 - cemented lg), (29) 880-339/22 (Sz. 9-10 - cemented x-lg), (30) 880-339/23 (Sz. 9-10 - cemented xx-lg); |
| Code Information |
Item Number/UDI-DI: (1) 880-331/11 (04026575258192), (2) 880-331/12 (04026575258208), (3) 880-331/13 (04026575258215), (4) 880-331/21 (04026575258222), (5) 880-331/22 (04026575258246), (6) 880-331/23 (04026575258253), (7) 880-333/11 (04026575258260), (8) 880-333/12 (04026575258277), (9) 880-333/13 (04026575258284), (10) 880-333/21 (04026575258291), (11) 880-333/22 (04026575258307), (12) 880-333/23 (04026575258314), (13) 880-335/11 (04026575258321), (14) 880-335/12 (04026575258338), (15) 880-335/13 (04026575258345), (16) 880-335/21 (04026575258352), (17) 880-335/22 (04026575258369), (18) 880-335/23 (04026575258376), (19) 880-337/11 (04026575258383), (20) 880-337/12 (04026575258390), (21) 880-337/13 (04026575258406), (22) 880-337/21 (04026575258413), (23) 880-337/22 (04026575258420), (24) 880-337/23 (04026575258437), (25) 880-339/11 (04026575258475), (26) 880-339/12 (04026575258505), (27) 880-339/13 (04026575258529), (28) 880-339/21 (04026575258536), (29) 880-339/22 (04026575258543), (30) 880-339/23 (04026575258550); Applies to all lots of the LINK SymphoKnee Femoral Augments listed, manufactured up until 03/01/2025. |
| FEI Number |
3003386935
|
Recalling Firm/ Manufacturer |
Waldemar Link GmbH & Co. KG (Mfg Site) Oststr. 4-10 Norderstedt Germany
|
| For Additional Information Contact | Jennifer Lubrecht 862-477-0370 |
Manufacturer Reason for Recall | The device was delivered with a preassembled Femoral Augment screw that was missing its thread. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | On April 3, 2025 URGENT: MEDICAL DEVICE CORRECTION letters were emailed to customers.
Actions to be taken by the Customer/User:
Please take note of the following instructions for replacing any screws found missing the thread. No product return is required.
Instructions for disassembling and replacing a defective screw:
Disassembly
1. Use the 2.5 mm Torque Wrench article REF (15-2545) to unscrew the screw from the Augment while pushing the tip of the screw towards the Augment.
2. Exchange the screw.
Open an alternative Augment and use its screw. Alternative sizes of Augments are always available in the operating room during the surgery.
For all Femoral Augment sizes and types the same long screw type is preassembled and for all Tibial Augments the same short screw type is preassembled, see table 1.
Assembly/Replacement
3. Use the 2.5 mm Torque Wrench article REF (15-2545) to screw in the replacement screw into the Augment.
4. The screw head must be countersunk in the Augment and the screw tip must protrude. Follow the LinkSymphoKnee surgical technique for all other steps.
Please respond using the Distributor Reply form to indicate your understanding of this Device Correction and of the need to apply the instructions provided above to replace any affected screws.
Type of Action by the Company:
LINK will provide sterilized replacement screws (for both Femoral Augments and Tibial Augments) for each affected distributor as a back-up for upcoming surgeries.
Other Information:
Opening an additional Augment to exchange the screw will not incur any costs to you. Should you have any questions about acquiring backup/replacement screws for forthcoming surgeries (once available), please contact Customer Service at LinkBio (order@linkbio.com).
Please contact LinkBio at 973-625-1333 or complaints@linkbio.com for any additional questions. |
| Quantity in Commerce | 1771 units |
| Distribution | US Nationwide distribution in the states of AL, AZ, CA, CO, FL, GA, IN, IL, KS, LA, MD, MO, NV, OH, TX, VA, WI. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|