 Class 1 Device Recall ALTRUA Pacemakerssee related information
Date Initiated by FirmAugust 20, 2025
Date PostedOctober 10, 2025
Recall Status1 Open3, Classified
Recall NumberZ-0099-2026
Recall Event ID 97467
Product Classification Implantable pulse generator, pacemaker (non-CRT) - Product Code LWP
ProductModel Number S702, ALTRUA 2 DR SL Pacemaker
Code Information GTIN 00802526559358, Serial numbers: 100002, 100003, 100004, 100005, 100006, 100007, 100008, 100009, 100010, 100011, 100012, 100013, 100014, 100015, 100016, 100017, 100018, 100020, 100021, 100025, 100026, 100028, 100029, 100030, 100031, 100032, 100033, 100034, 100035, 100036, 100037, 100038, 100040, 100042, 100045, 100046, 100050, 100051, 100052, 100053, 100054, 100055, 100056, 100057, 100058, 100059, 100060, 100061, 100062, 100063, 100064, 100065, 100066, 100067, 100068, 100069, 100070, 100071, 100072, 100073, 100074, 100075, 100076, 100077, 100078, 100079, 100080, 100081, 100082, 100083, 100084, 100085, 100086, 100087, 100088, 100089, 100090, 100092, 100093, 100094, 100095, 100096, 100097, 100098, 100099, 100100, 100101, 100102, 100103, 100104, 100105, 100106, 100107, 100108, 100109, 100110, 100112, 100113, 100115, 100116, 100117, 100118, 100119, 100120, 100121, 100122, 100123, 100125, 100126, 100128, 100130, 100132, 100133, 100134, 100135, 100136, 100137, 100138, 100139, 100140, 100141, 100142, 100143, 100144, 100145, 100146, 100147, 100148, 100149, 100150, 100151, 100152, 100153, 100154, 100155, 100158, 100159, 100160, 100161, 100162, 100163, 100164, 100165, 100166, 100167, 100168, 100170, 100171, 100172, 100173, 100174, 100175, 100176, 100177, 100179, 100180, 100181, 100182, 100183, 100184, 100185, 100186, 100187, 100188, 100189, 100190, 100191, 100193, 100194, 100195, 100197, 100198, 100199, 100200, 100201, 100202, 100203, 100204, 100205, 100208, 100209, 100211, 100212, 100213, 100214, 100215, 100216, 100217, 100218, 100219, 100220, 100221, 100222, 100223, 100224, 100225, 100226, 100227, 100228, 100230, 100231, 100232, 100234, 100235, 100236, 100237, 100238, 100239, 100240, 100241, 100242, 100243, 100244, 100245, 100246, 100247, 100248, 100249, 100250, 100251, 100252, 100253, 100254, 100255, 100256, 100257, 100258, 100259, 100260, 100261, 100262, 100263, 100265, 100266, 100267, 100268, 100269, 100270, 100271, 100272, 100273, 100274, 100275, 100276, 100277, 100278, 100279, 100280, 100281, 100282, 100283, 100284, 100285, 100286, 100287, 100289, 100290, 100291, 100292, 100293, 100294, 100297, 100298, 100299, 100300, 100301, 100303, 100305, 100306, 100307, 100308, 100310, 100311, 100312, 100313, 100314, 100315, 100316, 100318, 100319, 100320, 100321, 100322, 100323, 100324, 100325, 100326, 100329, 100330, 100331, 100332, 100333, 100334, 100335, 100336, 100338, 100340, 100342, 100343, 100344, 100345, 100346, 100347, 100348, 100349, 100350, 100351, 100352, 100353, 100354, 100355, 100357, 100358, 100363, 100365, 100366, 100367, 100368, 100370, 100373, 100374, 100375, 100376, 100377, 100378, 100379, 100381, 100383, 100387, 100388, 100389, 100391, 100392, 100393, 100394, 100396, 100400, 100402, 100403, 100406, 100407, 100409, 100411, 100412, 100413, 100414, 100415, 100416, 100417, 100418, 100420, 100423, 100424, 100425, 100427, 100430, 100431, 100432, 100434, 100435, 100436, 100437, 100438, 100439, 100440, 100441, 100442, 100444, 100445, 100446, 100447, 100448, 100449, 100450, 100452, 100453, 100454, 100456, 100457, 100458, 100459, 100460, 100461, 100462, 100463, 100464, 100465, 100466, 100467, 100468, 100469, 100470, 100471, 100472, 100473, 100474, 100475, 100476, 100477, 100478, 100479, 100480, 100481, 100482, 100484, 100485, 100486, 100487, 100488, 100489, 100490, 100491, 100493, 100495, 100496, 100497, 100498, 100499, 100500, 100501, 100503, 100505, 100506, 100507, 100511, 100513, 100515, 100517, 100518, 100519, 100520, 100521, 100523, 100524, 100525, 100526, 100527, 100528, 100529, 100530, 100531, 100532, 100533, 100534, 100535, 100538, 100539, 100540, 100541, 100543, 100544, 100545, 100546, 100548, 100550, 100551, 100552, 100553, 100554, 100555, 100556, 100557, 100558, 100559, 100560, 100561, 100562, 100563, 100564, 100565, 100566, 100568, 100569, 100574, 100576, 100577, 100578, 100581, 100582, 100583, 100585, 100586, 100588, 100591, 100592, 100593, 100594, 100595, 100596, 100597, 100598, 100599, 100600, 100601, 100602, 100603, 100604, 100605, 100606, 100607, 100608, 100609, 100610, 100611, 100613, 100614, 100615, 100616, 100617, 100618, 100620, 100621, 100622, 100623, 100624, 100625, 100626, 100627, 100628, 100630, 100631, 100632, 100634, 100635, 100636, 100637, 100638, 100639, 100640, 100641, 100642, 100643, 100644, 100645, 100646, 100647, 100648, 100649, 100650, 100651, 100652, 100653, 100654, 100655, 100656, 100657, 100658, 100659, 100660, 100661, 100662, 100663, 100664, 100666, 100667, 100668, 100669, 100670, 100671, 100672, 100673, 100674, 100675, 100676, 100677, 100678, 100679, 100682, 100683, 100684, 100685, 100686, 100689, 100690, 100691, 100692, 100693, 100694, 100695, 100696, 100698, 100699, 100700, 100701, 100702, 100703, 100704, 100705, 100707, 100708, 100709, 100710, 100711, 100712, 100713, 100715, 100717, 100720, 100721, 100722, 100723, 100724, 100725, 100726, 100727, 100728, 100729, 100730, 100731, 100733, 100735, 100736, 100737, 100738, 100739, 100740, 100741, 100742, 100743, 100744, 100745, 100746, 100747, 100748, 100749, 100750, 100752, 100753, 100754, 100755, 100756, 100757, 100759, 100760, 100762, 100763, 100764, 100765, 100766, 100767, 100768, 100769, 100770, 100771, 100772, 100773, 100774, 100775, 100776, 100777, 100778, 100779, 100780, 100781, 100782, 100783, 100787, 100789, 100790, 100791, 100793, 100794, 100795, 100796, 100797, 100799, 100800, 100801, 100802, 100804, 100805, 100806, 100807, 100808, 100809, 100810, 100811, 100812, 100813, 100814, 100815, 100816, 100817, 100819, 100820, 100821, 100822, 100823, 100825, 100826, 100828, 100829, 100830, 100831, 100833, 100836, 100837, 100838, 100839, 100841, 100843, 100845, 100846, 100847, 100851, 100852, 100853, 100854, 100855, 100856, 100870, 100874, 100879, 100881, 100882, 100884, 100885, 100887, 100889, 100892, 100894, 100895, 100897, 100898, 100899, 100903, 100907, 100909, 100912, 100913, 100914, 100915, 100916, 100919, 100920, 100921, 100922, 100923, 100929, 100932, 100937, 100938, 100942, 100943, 100946, 100947, 100949, 100951, 100957, 100958, 100960, 100961, 700001, 700002, 700004, 700005, 700006, 700007, 700009, 700011, 700014, 700025, 700026, 700028, 700029, 700030, 700033, 700039, 700042, 700046, 700050, 700056, 700058, 700060, 700063, 700065, 700067, 700068, 700069, 700071, 700073, 700074, 700075, 700076, 700078, 700079, 700080, 700081, 700082, 700083, 700084, 700086, 700087, 700088, 700089, 700091, 700095, 700097, 700099, 700100, 700101, 700102, 700103, 700104, 700105, 700107, 700108, 700109, 700110, 700111, 700113, 700114, 700115, 700117, 700118, 700122, 700124, 700127, 700128, 700129, 700130, 700134, 700136, 700138, 700140, 700141, 700142, 700145, 700146, 700147, 700151, 700152, 700153, 700154, 700156, 700158, 700159, 700160, 700161, 700162, 700164, 700165, 700166, 700167, 700168, 700169, 700170, 700171, 700172, 700173, 700174, 700175, 700176, 700177, 700178, 700179, 700180, 700181, 700182, 700183, 700184, 700185, 700187, 700188, 700190, 700192, 700193, 700195, 700196, 700197, 700198, 700199, 700200, 700201, 700202, 700203, 700204, 700208, 700209, 700210, 700212, 700213, 700214, 700215, 700216, 700218, 700219, 700220, 700221, 700222, 700225, 700227, 700228, 700231, 700232, 700233, 700234, 700235, 700238, 700239, 700241, 700242, 700244, 700245, 700248, 700250, 700251, 700252, 700253, 700254, 700255, 700256, 700258, 700260, 700262, 700265, 700267, 700268, 700269, 700270, 700273, 700274, 700276, 700277, 700278, 700279, 700280, 700281, 700282, 700283, 700284, 700286, 700287, 700289, 700290, 700292, 700293, 700294, 700295, 700298, 700299, 700301, 700303, 700305, 700306, 700307, 700308, 700309, 700310, 700311, 700313, 700314, 700315, 700316, 700317, 700318, 700319, 700320, 700321, 700324, 700325, 700327, 700328, 700330, 700331, 700332, 700334, 700335, 700336, 700338, 700339, 700340, 700341, 700342, 700343, 700345, 700346, 700347, 700348, 700349, 700350, 700351, 700352, 700354, 700355, 700356, 700358, 700359, 700360, 700361, 700363, 700364, 700366, 700367, 700368, 700370, 700371, 700372, 700375, 700376, 700377, 700379, 700380, 700381, 700382, 700384, 700387, 700388, 700389, 700390, 700391, 700392, 700394, 700395, 700396, 700398, 700399, 700400, 700401, 700402, 700404, 700405, 700406, 700407, 700408, 700409, 700410, 700411, 700412, 700413, 700414, 700416, 700417, 700418, 700419, 700420, 700422, 700425, 700426, 700427, 700428, 700430, 700431, 700432, 700433, 700434, 700436, 700437, 700438, 700439, 700440, 700441, 700442, 700443, 700445, 700447, 700448, 700449, 700450, 700453, 700454, 700457, 700458, 700459, 700460, 700461, 700462, 700463, 700464, 700465, 700467, 700468, 700469, 700472, 700474, 700475, 700476, 700477, 700478, 700479, 700480, 700481, 700486, 700487, 700489, 700490, 700491, 700492, 700493, 700494, 700495, 700497, 700498, 700500, 700501, 700502, 700504, 700505, 700506, 700507, 700508, 700509, 700510, 700511, 700512, 700513, 700514, 700515, 700516, 700517, 700518, 700519, 700520, 700521, 700522, 700523, 700524, 700525, 700526, 700527, 700528, 700532, 700533, 700534, 700535, 700536, 700537, 700539, 700540, 700543, 700544, 700547, 700549, 700550, 700552, 700553, 700554, 700555, 700556, 700557, 700558, 700559, 700560, 700561, 700562, 700563, 700564, 700565, 700566, 700567, 700570, 700571, 700572, 700573, 700574, 700577, 700578, 700579, 700580, 700581, 700582, 700583, 700584, 700585, 700587, 700588, 700589, 700590, 700591, 700592, 700593, 700594, 700595, 700596, 700597, 700598, 700599, 700600, 700602, 700603, 700604, 700606, 700607, 700608, 700609, 700610, 700611, 700612, 700613, 700614, 700616, 700617, 700618, 700619, 700620, 700625, 700626, 700628, 700629, 700630, 700632, 700633, 700634, 700635, 700636, 700637, 700639, 700640, 700641, 700643, 700644, 700645, 700646, 700647, 700649, 700652, 700653, 700654, 700655, 700656, 700658, 700660, 700661, 700664, 700665, 700666, 700668, 700669, 700671, 700672, 700674, 700675, 700676, 700677, 700678, 700679, 700680, 700682, 700683, 700684, 700685, 700686, 700687, 700689, 700690, 700692, 700693, 700694, 700695, 700697, 700698, 700699, 700700, 700701, 700702, 700703, 700704, 700705, 700706, 700707, 700708, 700709, 700710, 700711, 700712, 700713, 700714, 700715, 700716, 700717, 700718, 700719, 700720, 700721, 700722, 700723, 700725, 700727, 700728, 700731, 700733, 700734, 700735, 700736, 700737, 700738, 700739, 700740, 700741, 700742, 700743, 700744, 700746, 700747, 700748, 700749, 700750, 700752, 700754, 700755, 700758, 700759, 700760, 700761, 700762, 700763, 700764, 700765, 700767, 700768, 700769, 700770, 700771, 700772, 700773, 700774, 700775, 700776, 700777, 700778, 700779, 700780, 700782, 700783, 700784, 700785, 700787, 700788, 700789, 700790, 700792, 700793, 700794, 700795, 700796, 700798, 700800, 700801, 700802, 700805, 700807, 700808, 700810, 700815, 700816, 700819, 700820, 700821, 700824, 700826, 700827, 700828, 700829, 700830, 700831, 700832, 700833, 700834, 700835, 700836, 700837, 700839, 700840, 700841, 700842, 700843, 700844, 700845, 700846, 700847, 700848, 700849, 700850, 700851, 700852, 700853, 700855, 700856, 700857, 700858, 700859, 700860, 700861, 700862, 700863, 700864, 700865, 700867, 700868, 700869, 700870, 700871, 700872, 700873, 700874, 700875, 700876, 700877, 700878, 700879, 700880, 700881, 700882, 700883, 700886, 700887, 700888, 700889, 700890, 700891, 700892, 700893, 700894, 700895, 700896, 700897, 700898, 700899, 700900, 700901, 700903, 700904, 700905, 700908, 700909, 700911, 700913, 700914, 700915, 700916, 700917, 700918, 700919, 700920, 700921, 700922, 700923, 700924, 700925, 700926, 700927, 700928, 700929, 700930, 700931, 700932, 700934, 700935, 700936, 700937, 700938, 700939, 700940, 700942, 700943, 700944, 700945, 700946, 700947, 700948, 700949, 700950, 700951, 700952, 700953, 700954, 700955, 700956, 700957, 700958, 700959, 700960, 700961, 700962, 700963, 700965, 700966, 700967, 700968, 700969, 700970, 700971, 700972, 700973, 700974, 700975, 700976, 700977, 700978, 700979, 700980, 700981, 700982, 700983, 700985, 700986, 700987, 700989, 700990, 700991, 700992, 700993, 700994, 700995, 700996, 700997, 700998, 700999, 701000, 701001, 701002, 701003, 701004, 701005, 701006, 701012, 701013, 701014, 701015, 701016, 701018, 701019, 701020, 701022, 701023, 701024, 701025, 701026, 701027, 701028, 701029, 701030, 701032, 701033, 701034, 701035, 701036, 701037, 701038, 701039, 701040, 701041, 701042, 701043, 701044, 701045, 701046, 701047, 701048, 701049, 701050, 701051, 701052, 701053, 701054, 701055, 701056, 701057, 701058, 701059, 701060, 701061, 701062, 701064, 701065, 701067, 701068, 701069, 701070, 701071, 701072, 701073, 701074, 701076, 701079, 701080, 701081, 701082, 701083, 701084, 701085, 701086, 701087, 701088, 701089, 701090, 701091, 701092, 701093, 701096, 701100, 701101, 701104, 701105, 701107, 701109, 701112, 701113, 701114, 701116, 701118, 701119, 701120, 701122, 701123, 701124, 701125, 701134, 701135, 701137, 701139, 701140, 701141, 701142, 701143, 701145, 701147, 701148, 701149, 701151, 701152, 701153, 701157, 701159, 701160, 701164, 701168, 701169, 701170, 701172, 701174, 701176, 701177, 701178, 701179, 701180, 701181, 701183, 701184, 701185, 701186, 701189, 701191, 701194, 701196, 701197, 701203, 701204, 701205, 701212, 701217, 701218, 701219, 701222, 701236, 701238, 701239, 701242, 701243, 701245, 701246, 701247, 701251, 701253, 701254, 701255, 701257, 701258, 701259, 701262, 701263, 701265, 701266, 701267, 701268, 701275, 701280, 701281, 701285, 701288, 701291, 701293, 701295, 701296, 701301, 701303, 701304, 701305, 701307, 701308, 701309, 701310, 701312, 701313, 701315, 701316, 701320, 701323, 701324, 701325, 701327, 701328, 701329, 701330, 701333, 701334, 701335, 701336, 701338, 701340, 701341, 701343, 701347, 701348, 701349, 701351, 701354, 701357, 701362, 701363, 701364, 701365, 701366, 701367, 701371, 701373, 701374, 701375, 701376, 701377, 701381, 701382, 701391, 701392, 701400, 701405, 701406, 701407, 701408, 701409, 701410, 701411, 701412, 701414, 701415, 701416, 701417, 701418, 701419, 701420, 701421, 701422, 701424, 701425, 701426, 701427, 701428, 701429, 701430, 701431, 701432, 701435, 701438, 701440, 701443, 701444, 701445, 701451, 701452, 701454, 701455, 701456, 701457, 701458, 701459, 701460, 701461, 701462, 701465, 701466, 701467, 701472, 701474, 701475, 701477, 701478, 701480, 701486, 701487, 701488, 701491, 701494, 701496, 701498, 701500, 701501, 701509, 701510, 701512, 701513, 701516, 701520, 701523, 701532, 701536, 701537, 701538, 701540, 701542, 701545, 701546, 701547, 701553, 701556, 701558, 701559, 701561, 701562, 701563, 701564, 701565, 701566, 701567, 701568, 701569, 701570, 701571, 701572, 701574, 701576, 701580, 701583, 701585, 701586, 701604, 701609, 701627, 701659, 701660, 701662, 701664, 701669, 701670, 701672, 701674, 701676, 701677, 701678, 701692, 701693, 701698, 701702, 701716, 701718, 701721, 701723, 701724, 701727, 701729, 701734, 701735, 701736, 701737, 701738, 701742, 701744, 701746, 701749, 701750, 701751, 701752, 701753, 701754, 701755, 701756, 701757, 701759, 701760, 701762, 701763, 701766, 701773, 701774, 701777, 701778, 701779, 701780, 701782, 701783, 701784, 701785, 701786, 701787, 701788, 701790, 701791, 701792, 701793, 701795, 701796, 701797, 701798, 701799, 701800, 701801, 701802, 701803, 701805, 701806, 701807, 701808, 701809, 701811, 701812, 701813, 701815, 701817, 701819, 701820, 701822, 701823, 701824, 701827, 701828, 701829, 701830, 701831, 701832, 701833, 701836, 701837, 701838, 701839, 701840, 701842, 701843, 701844, 701845, 701847, 701849, 701850, 701852, 701853, 701855, 701856, 701858, 701859, 701864, 701866, 701867, 701868, 701870, 701871, 701872, 701873, 701874, 701875, 701876, 701878, 701879, 701880, 701882, 701899, 701900, 701903, 701908, 701910, 701911, 701916, 701921, 701928, 701929, 701930, 701931, 701939, 701940, 701943, 701944, 701946, 701949, 701950, 701952, 701953, 701955, 701958, 701966, 701967, 701970, 701971, 701972, 701981, 701984, 701985, 701986, 701989, 701990, 701991, 701992, 701993, 701995, 701997, 701998, 701999, 702003, 702004, 702008, 702009, 702015, 702016, 702018, 702019, 702021, 702023, 702026, 702027, 702031, 702032, 702033, 702034, 702035, 702036, 702037, 702038, 702039, 702040, 702041, 702042, 702043, 702046, 702047, 702048, 702049, 702050, 702051, 702052, 702053, 702056, 702058, 702059, 702060, 702061, 702076, 702078, 702081, 702091, 702096, 702098, 702101, 702106, 702112, 702132, 702138, 702142, 702143, 702153, 702165, 702179, 702181, 702182, 702187, 702189, 702192, 702193, 702208, 702216, 702217, 702222, 702227, 702244, 702245, 702246, 702247, 702248, 702249, 702250, 702251, 702252, 702253, 702254, 702257, 702259, 702260, 702261, 702262, 702264, 702265, 702267, 702269, 702271, 702274, 702276, 702277, 702280, 702281, 702283, 702284, 702285, 702287, 702289, 702291, 702293, 702294, 702295, 702296, 702297, 702318, 702328, 702333, 702335, 702338, 702346, 702352, 702358, 702364, 702373, 702379, 702380, 702382, 702391, 702393, 702398, 702399, 702400, 702439, 702440, 702441, 702453, 702479, 702490, 702491, 702501, 702513, 702517, 702527, 702533, 702538, 702541, 702542, 702543, 702546, 702548, 702550, 702557, 702566, 702571, 702580, 702587, 702590, 702591, 702592, 702594, 702597, 702599, 702603, 702613, 702616, 702623, 702626, 702631, 702636, 702640, 702642, 702647, 702651, 702654, 702655, 702656, 702658, 702659, 702662, 702665, 702668, 702671, 702675, 702680, 702688, 702693, 702698, 702702, 702703, 702704, 702707, 702711, 702716, 702718, 702719, 702720, 702728, 702732, 702737, 702738, 702743, 702744, 702747, 702752, 702755, 702760, 702765, 702771, 702772, 702773, 702777, 702780, 702788, 702789, 702790, 702793, 702794, 702800, 702802, 702803, 702805, 702806, 702807, 702808, 702810, 702811, 702813, 702816, 702817, 702823, 702832, 702833, 702836, 702837, 702838, 702847, 702848, 702851, 702854, 702880, 702890, 702892, 702893, 702894, 702896, 702897, 702898, 702901, 702902, 702903, 702914, 702920, 702921, 702933, 702935, 702952, 702954, 702974, 702977, 702978, 702981, 702986, 702988, 702992, 703001, 703004, 703009, 703010, 703018, 703021, 703025, 703030, 703034, 703035, 703036, 703037, 703038, 703039, 703043, 703044, 703046, 703047, 703048, 703049, 703050, 703051, 703052, 703053, 703054, 703056, 703060, 703063, 703065, 703069, 703080, 703081, 703083, 703084, 703085, 703086, 703087, 703090, 703098, 703100, 703104, 703108, 703109, 703110, 703111, 703119, 703122, 703124, 703125, 703126, 703162, 703172, 703174, 703180, 703183, 703186, 703189, 703202, 703211, 703215, 703222, 703223, 703225, 703230, 703231, 703235, 703255, 703262, 703270, 703274, 703281, 703283, 703308, 703309, 703312, 703404, 703405, 703426, 703488, 703495, 703500, 703511, 703512, 703513, 703515, 703516, 703520, 703529, 703531, 703533, 703534, 703535, 703536, 703537, 703543, 703544, 703547, 703550, 703554, 703560, 703563, 703565, 703568, 703570, 703572, 703575, 703576, 703577, 703579, 703580, 703583, 703584, 703586, 703590, 703592, 703593, 703594, 703596, 703597, 703598, 703599, 703601, 703605, 703606, 703607, 703608, 703610, 703611, 703613, 703615, 703618, 703619, 703620, 703621, 703624, 703631, 703634, 703636, 703785, 703795, 703807, 703810, 703814, 703816, 703821, 703834, 703835, 703839, 703844, 703845, 703846, 703848, 703849, 703850, 703861, 703906, 703908, 703909, 703910, 703917, 703942, 703943, 703957, 703986, 703990, 704001, 704012, 704019, 704027, 704029, 704035, 704037, 704038, 704039, 704040, 704044, 704048, 704052, 704053, 704054, 704055, 704056, 704058, 704059, 704061, 704063, 704064, 704069, 704070, 704073, 704077, 704083, 704084, 704088, 704089, 704101, 704107, 704114, 704116, 704118, 704121, 704122, 704127, 704130, 704131, 704134, 704135, 704136, 704140, 704142, 704144, 704148, 704150, 704151, 704153, 704154, 704167, 704172, 704175, 704177, 704181, 704183, 704184, 704187, 704188, 704196, 704202, 704203, 704206, 704211, 704229, 704248, 704251, 704288, 704338, 704505, 704515, 704567, 704577, 704578, 704613, 704623, 704627, 704635, 704637, 704639, 704640, 704641, 704642, 704644, 704645, 704647, 704648, 704650, 704659, 704662, 704663, 704668, 704669, 704670, 704671, 704672, 704682, 704683, 704685, 704688, 704689, 704692, 704693, 704698, 704699, 704700, 704711, 704713, 704717, 704718, 704719, 704739, 704745, 704746, 704753, 704772, 704788, 704792, 704813, 704824, 704827, 704828, 704832, 704836, 704838, 704839, 704840, 704845, 704847, 704851, 704852, 704853, 704856, 704861, 704866, 704868, 704871, 704876, 704904, 704905, 704910, 705079, 705122, 705154, 705184, 705262, 705264, 705267, 705274, 705277, 705279, 705287, 705289, 705293, 705294, 705304, 705309, 705310, 705316, 705317, 705332, 705349, 705361, 705364, 705365, 705368, 705373, 705377, 705378, 705387, 705388, 705390, 705393, 705395, 705396, 705401, 705462, 705466, 705471, 705674, 705682, 705689, 705690, 705703, 705710, 705731, 705732, 705735, 705742, 705746, 705747, 705778, 705819, 705824, 705854, 705947, 705973, 705990, 705999, 706058, 706063, 706068, 706110, 706187, 706193, 706196, 706197, 706201, 706206, 706208, 706229, 706231, 706292, 706295, 706337, 706353, 706355, 706359, 706390, 706401, 706438, 706439, 706440, 706445, 706450, 706452, 706456, 706457, 706470, 706472, 706483, 706504, 706505, 706511, 706520, 706523, 706526, 706538, 706539, 706541, 706552, 706553, 706554, 706555, 706559, 706560, 706592, 706593, 706595, 706596, 706597, 706598, 706599, 706600, 706602, 706603, 706605, 706606, 706609, 706610, 706619, 706623, 706634, 706637, 706640, 706722, 706727, 706734, 706739, 706741, 706743, 706759, 706760, 706762, 706765, 706766, 706767, 706770, 706772, 706773, 706776, 706777, 706779, 706780, 706781, 706829, 706830, 706849, 706850, 706851, 707324, 707356, 707357, 707359, 707361, 707369, 707906, 707908, 707909, 707927, 707948, 707959, 707978, 707979, 707981, 708003, 708026, 708030, 708035, 708039, 708055, 708056, 708058, 708059, 708211, 708236, 708675, 708695, 708710, 708913, 708916. GTIN 00802526576508, Serial Numbers: 100385, 100419, 100483, 100502, 100508, 100537, 100542, 100547, 100584, 100629, 100714, 100716, 100734, 100761, 100798, 700003, 700008, 700010, 700012, 700015, 700016, 700017, 700019, 700021, 700022, 700023, 700024, 700031, 700032, 700034, 700035, 700036, 700041, 700043, 700045, 700047, 700048, 700049, 700051, 700052, 700054, 700055, 700057, 700061, 700062, 700064, 700066, 700070, 700072, 700085, 700090, 700093, 700094, 700098, 700106, 700112, 700119, 700120, 700121, 700125, 700126, 700131, 700132, 700133, 700139, 700143, 700144, 700148, 700150, 700155, 700157, 700163, 700186, 700189, 700191, 700217, 700357, 700365, 700369, 700385, 700551, 700605, 700623, 700651, 700659, 700663, 700756, 701302, 701339, 701345, 701369, 701442, 701601, 701626, 701637, 701715, 701758, 701776, 701781, 701816, 701846, 701861, 701884, 701885, 701886, 701890, 701892, 701893, 701897, 701912, 701926, 701951, 701959, 701962, 702022, 702411, 702437, 702529, 702758, 702822, 702828, 702830, 702913, 703096, 703101, 703103, 703105, 703248, 703263, 703269, 703276, 703288, 703295, 703303, 703311, 703325, 703329, 703337, 703344, 703350, 703353, 703357, 703361, 703369, 703372, 703376, 703380, 703383, 703402, 703415, 703416, 703425, 703429, 703430, 703433, 703447, 703456, 703484, 703517, 703540, 703553, 703588, 704104, 704178, 704180, 704186, 704192, 704198, 704210, 704215, 704216, 704218, 704221, 704227, 704245, 704247, 704448, 704490, 704493, 704496, 704499, 704507, 704513, 704514, 704517, 704518, 704519, 704521, 704535, 704549, 704557, 704588, 704603, 704606, 706920, 706930, 706932, 706935, 706936, 706937, 706938, 706940, 706949, 707006, 707012, 707014, 707015, 708020, 708027, 708033, 708037, 708038, 708052, 709683, 709695, 709713, 709723, 709733, 709734, 709736, 709737, 709741, 709852, 709977, 710385, 710412, 710422, 710423, 710425, 710426, 710428, 710430, 710435, 710439, 710445, 710456, 710458, 710460, 710480, 710494, 710498, 710502, 710509, 710513, 710531, 710544, 710614, 710616, 710632, 710640, 710645, 710728, 710783, 710949, 711048, 711058, 711082, 711112, 711149, 711462, 711463, 711464, 711466, 711468, 711470, 711474, 711476, 711478, 711479, 711481. GTIN 00802526577000, Serial Numbers: 100827, 100834, 100840, 100868, 100880, 100888, 100890, 100896, 100900, 100904, 100905, 100908, 100925, 100926, 100927, 100928, 100930, 100931, 100933, 100935, 100936, 100939, 100940, 100944, 700207, 700211, 700229, 700243, 700249, 700257, 700259, 700263, 700264, 700266, 700271, 700272, 700275, 700291, 700296, 700297, 700302, 700304, 700312, 700322, 700326, 700329, 700333, 700378, 700435, 700503, 700586, 700745, 701099, 701127, 701290, 701380, 701384, 701390, 701413, 701423, 701503, 701504, 701506, 701679, 701680, 701682, 701683, 701684, 701685, 701686, 701687, 701688, 701689, 701690, 701691, 701695, 701696, 701964, 701965, 701968, 701974, 702064, 702065, 702066, 702069, 702071, 702072, 702073, 702074, 702075, 702079, 702082, 702086, 702087, 702089, 702090, 702093, 702095, 702097, 702099, 702100, 702102, 702103, 702104, 702105, 702107, 702108, 702109, 702110, 702111, 702128, 702131, 702133, 702134, 702135, 702136, 702139, 702140, 702144, 702145, 702148, 702150, 702152, 702161, 702174, 702175, 702176, 702177, 702180, 702185, 702186, 702235, 702270, 702288, 702302, 702305, 702306, 702307, 702310, 702353, 702354, 702355, 702363, 702367, 702368, 702370, 702371, 702375, 702401, 702402, 702404, 702421, 702426, 702448, 702449, 702450, 702451, 702456, 702460, 702512, 702514, 702516, 702518, 702519, 702520, 702521, 702522, 702524, 702528, 702537, 702539, 702572, 702573, 702574, 702575, 702577, 702578, 702581, 702582, 702583, 702584, 702585, 702588, 702589, 702657, 702660, 702705, 702735, 702783, 702784, 702785, 702787, 702799, 702825, 702888, 702889, 702945, 702946, 702949, 702950, 702951, 702961, 702962, 702963, 702966, 702967, 702968, 702975, 702982, 702985, 702996, 702999, 703000, 703033, 703045, 703064, 703066, 703067, 703093, 703097, 703099, 703107, 703113, 703114, 703115, 703116, 703117, 703118, 703121, 703141, 703170, 703173, 703175, 703179, 703181, 703342, 703343, 703354, 703355, 703356, 703360, 703392, 703393, 703394, 703395, 703396, 703397, 703398, 703399, 703400, 703403, 703407, 703408, 703409, 703437, 703464, 703472, 703473, 703474, 703475, 703476, 703477, 703478, 703479, 703480, 703481, 703482, 703485, 703486, 703487, 703489, 703490, 703491, 703493, 703496, 703562, 703566, 703567, 703569, 703571. GTIN 00802526578106, Serial Numbers: 100573, 100680, 100849, 100850, 100858, 100860, 100865, 100875, 100911, 700323, 700444, 700446, 700484, 700496, 700546, 700624, 700627, 700670, 700673, 700696, 700726, 700781, 700791, 700799, 700823, 700906, 701021, 701031, 701063, 701075, 701077, 701078, 701102, 701108, 701111, 701115, 701117, 701121, 701126, 701128, 701129, 701130, 701131, 701132, 701136, 701138, 701144, 701146, 701154, 701155, 701156, 701161, 701163, 701165, 701166, 701167, 701173, 701175, 701182, 701193, 701198, 701199, 701200, 701201, 701202, 701206, 701207, 701208, 701209, 701210, 701211, 701213, 701214, 701215, 701216, 701221, 701223, 701226, 701227, 701228, 701229, 701230, 701231, 701232, 701233, 701234, 701235, 701237, 701240, 701241, 701244, 701248, 701249, 701252, 701256, 701260, 701264, 701269, 701270, 701271, 701272, 701273, 701274, 701278, 701279, 701283, 701284, 701286, 701287, 701289, 701292, 701294, 701297, 701298, 701299, 701300, 701306, 701311, 701314, 701317, 701318, 701319, 701321, 701322, 701326, 701331, 701332, 701337, 701342, 701344, 701346, 701350, 701352, 701353, 701355, 701356, 701358, 701360, 701361, 701368, 701370, 701372, 701378, 701379, 701383, 701385, 701387, 701388, 701389, 701393, 701394, 701395, 701396, 701397, 701398, 701399, 701402, 701403, 701404, 701446, 701468, 701471, 701522, 701591, 701638, 701646, 701652, 701667, 701675, 701681, 701694, 701697, 701699, 701701, 701704, 701705, 701707, 701708, 701709, 701710, 701713, 701717, 701719, 701720, 701722, 701775, 701789, 701794, 701804, 701814, 701818, 701821, 701825, 701826, 701834, 701835, 701848, 701851, 701854, 701857, 701860, 701869, 701881, 701883, 701894, 701895, 701898, 701901, 701902, 701905, 701906, 701919, 701922, 701923, 701927, 701932, 701933, 701934, 701935, 701936, 701938, 701941, 701945, 701947, 701954, 701956, 701957, 701960, 701961, 701973, 701983, 701987, 701988, 701994, 702000, 702006, 702007, 702017, 702024, 702028, 702045, 702054, 702067, 702068, 702070, 702083, 702084, 702085, 702088, 702092, 702113, 702115, 702122, 702125, 702130, 702137, 702141, 702147, 702151, 702155, 702158, 702171, 702172, 702224, 702351, 702357, 702365, 702366, 702369, 702372, 702374, 702377, 702378, 702383, 702384, 702385, 702386, 702389, 702394, 702395, 702406, 702409, 702412, 702413, 702414, 702417, 702419, 702427, 702428, 702429, 702432, 702433, 702435, 702442, 702447, 702454, 702455, 702457, 702469, 702473, 702477, 702482, 702486, 702489, 702499, 702505, 702510, 702511, 702515, 702523, 702531, 702532, 702535, 702544, 702555, 702560, 702567, 702576, 702579, 702586, 702604, 702608, 702612, 702615, 702620, 702621, 702622, 702629, 702634, 702637, 702638, 702643, 702644, 702645, 702663, 702664, 702666, 702672, 702676, 702678, 702679, 702681, 702691, 702694, 702695, 702696, 702699, 702701, 702710, 702712, 702714, 702715, 702717, 702721, 702731, 702733, 702734, 702739, 702740, 702741, 702748, 702749, 702750, 702751, 702753, 702757, 702764, 702770, 702774, 702775, 702776, 702778, 702779, 702781, 702782, 702792, 702795, 702796, 702797, 702798, 702801, 702804, 702812, 702814, 702824, 702826, 702829, 702831, 702834, 702835, 702839, 702840, 702841, 702842, 702843, 702844, 702845, 702846, 702850, 702852, 702855, 702859, 702861, 702862, 702863, 702864, 702865, 702866, 702867, 702868, 702869, 702874, 702876, 702878, 702881, 702882, 702883, 702885, 702886, 702891, 702895, 702899, 702900, 702904, 702905, 702906, 702908, 702909, 702911, 702912, 702916, 702917, 702919, 702925, 702957, 702958, 702959, 702960, 702964, 702965, 702969, 702970, 702971, 702972, 702973, 702979, 702983, 702984, 702991, 702993, 702994, 702995, 702997, 703002, 703005, 703006, 703007, 703008, 703014, 703015, 703022, 703027, 703031, 703041, 703058, 703062, 703068, 703072, 703075, 703078, 703079, 703082, 703088, 703091, 703094, 703095, 703102, 703106, 703112, 703120, 703127, 703136, 703148, 703151, 703154, 703157, 703164, 703166, 703169, 703176, 703177, 703178, 703182, 703187, 703194, 703196, 703198, 703200, 703201, 703203, 703205, 703213, 703216, 703218, 703220, 703224, 703227, 703228, 703243, 703244, 703245, 703249, 703250, 703252, 703253, 703254, 703259, 703265, 703271, 703272, 703280, 703285, 703289, 703293, 703294, 703296, 703306, 703307, 703314, 703315, 703318, 703322, 703327, 703331, 703753, 703919, 703949, 703973, 703974, 703976, 703977, 703978, 703993, 703994, 703995, 703996, 703997, 703998, 703999, 704000, 704002, 704003, 704004, 704005, 704006, 704007, 704009, 704010, 704014, 704015, 704016, 704017, 704018, 704020, 704021, 704022, 704023, 704025, 704026, 704028, 704030, 704031, 704033, 704036, 704042, 704043, 704045, 704046, 704047, 704049, 704050, 704051, 704057, 704080, 704082, 704085, 704086, 704102, 704103, 704105, 704109, 704111, 704112, 704119, 704401, 704404, 704441, 704450, 704460, 704462, 704464, 704465, 704468, 704484, 704492, 704494, 704501, 704512, 704523, 704524, 704526, 704534, 704544, 704551, 704576, 704587, 704591, 704592, 704593, 704595, 704596, 704597, 704598, 704601, 704602, 704605, 704607, 704610, 704611, 705032, 705042, 705043, 705047, 705052, 705058, 705064, 705066, 705071, 705074, 705075, 705078, 705082, 705083, 705085, 705087, 705088, 705089, 705091, 705092, 705094, 705095, 705096, 705097, 705098, 705099, 705101, 705125, 705126, 705127, 705130, 705142, 705146, 705151, 705152, 705157, 705163, 705164, 705166, 705167, 705169, 705170, 705171, 705176, 705198, 705209, 705210, 705212, 705224, 705229, 705253, 705254, 705319, 705328, 705335, 705336, 705341, 705382, 705403, 705404, 705407, 705408, 705413, 705415, 705416, 705417, 705418, 705420, 705421, 705441, 705444, 705445, 705447, 705449, 705541, 705590, 705612, 705653, 705664, 705669, 705672, 705673, 705676, 705680, 705688, 705691, 705698, 705699, 705700, 705701, 705702, 705711, 705724, 705727, 705781, 705800, 705843, 705861, 705862, 705884, 705888, 705889, 705925, 705933, 705974, 705983, 705994, 706002, 706034, 706046, 706047, 706054, 706065, 706090, 706096, 706118, 706133, 706135, 706138, 706142, 706144, 706149, 706172, 706683, 706684, 706686. GTIN 00802526593208, Serial numbers: 701730, 701731, 701732, 701877, 701915, 702129, 702178, 702214.
FEI Number 2124215
Recalling Firm/
Manufacturer
Boston Scientific Corporation
4100 Hamline Ave N
Saint Paul MN 55112-5700
For Additional Information ContactUnited States Technical Services
800-227-3422
Manufacturer Reason
for Recall
Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
FDA Determined
Cause 2
Other
ActionBoston Scientific issued an "Urgent Medical Device Correction" notice to its customers on 08/18/2025 via letter. The notice provides an update to the December 2024 customer communication regarding the potential for the ACCOLADE family of pacemakers to initiate Safety Mode due to a high battery impedance state and contains recommendations timely upgrade of pacemaker software to prevent the occurrence of Safety Mode in an ambulatory setting. On 09/05/2025 an update to the 08/18/2025 notice (via letter), superseding the prior notice to address two unintended behaviors associated with the Model 3869 v2.04 software (designed to prevent potential initiation of Safety Mode due to a high battery impedance condition). The behaviors are: 1. Incomplete disablement of wandless automatic telemetry after detecting a high battery impedance condition. If an ACCOLADE family device with an elevated battery impedance induced Code-1003 remains in service, the software s new telemetry activated battery test is designed to continue assessing battery impedance and disable all wandless telemetry operation before high impedance is achieved and device resets can initiate Safety Mode. 2. Potential for the software s new daily battery test to misinterpret measurements in the presence of a magnet, initiating a false positive response that results in disablement of wandless telemetry sessions. For patients monitored via LATITUDE, this will trigger an alert message indicating an erroneous explant indicator (e.g., explant indicator was reached January 1, 2000) if the device had not previously reached replacement indication, stating that LATITUDE is no longer available, and prompting to contact Technical Services. Patients were notified regarding battery performance. Advisory is published at www.BostonScientific.com/advisory. If you have any questions, contact US Technical Services at 1.800.CARDIAC (227.3422) or email to: tech.services@bsci.com.
Quantity in Commerce4055 units
DistributionWorldwide
Total Product Life CycleTPLC Device Report