| | Class 1 Device Recall VALITUDE CRTPs |  |
| Date Initiated by Firm | August 20, 2025 |
| Date Posted | October 10, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0101-2026 |
| Recall Event ID |
97467 |
| PMA Number | P030005 |
| Product Classification |
Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P) - Product Code NKE
|
| Product | Model Number U125, VALITUDE CRT-P EL |
| Code Information |
GTIN 00802526559389, Serial numbers: 100095, 100182, 100194, 100221, 100236, 100239, 100242, 100243, 100247, 100254, 100258, 100259, 100263, 100270, 100285, 100287, 100292, 100298, 100301, 100302, 100307, 100310, 100315, 100323, 100331, 100333, 100339, 100342, 100343, 100348, 100349, 100350, 100357, 100358, 100359, 100360, 100362, 100366, 100367, 100371, 100372, 100375, 100378, 100386, 100390, 100392, 100394, 100396, 100397, 100401, 100402, 100403, 100406, 100409, 100412, 100413, 100418, 100419, 100420, 100423, 100429, 100430, 100432, 100435, 100438, 100448, 100452, 100453, 100455, 100456, 100458, 100462, 100466, 100467, 100472, 100473, 100475, 100476, 100480, 100481, 100482, 100485, 100486, 100489, 100492, 100493, 100495, 100496, 100497, 100498, 100499, 100500, 100502, 100505, 100510, 100514, 100518, 100526, 100529, 100532, 100533, 100536, 100537, 100539, 100540, 100541, 100544, 100546, 100548, 100549, 100552, 100560, 100566, 100570, 100572, 100573, 100575, 100576, 100578, 100582, 100583, 100585, 100591, 100593, 100598, 100603, 100605, 100612, 100614, 100618, 100625, 100626, 100632, 100635, 100636, 100640, 100642, 100643, 100645, 100646, 100647, 100648, 100650, 100653, 100657, 100658, 100661, 100666, 100671, 100672, 100674, 100676, 100679, 100683, 100684, 100687, 100690, 100692, 100693, 100697, 100699, 100700, 100703, 100707, 100711, 100713, 100721, 100722, 100723, 100725, 100726, 100730, 100731, 100735, 100740, 100743, 100746, 100751, 100754, 100755, 100756, 100757, 100758, 100759, 100760, 100761, 100763, 100767, 100770, 100772, 100774, 100775, 100776, 100777, 100778, 100779, 100780, 100781, 100784, 100786, 100788, 100789, 100790, 100792, 100799, 100802, 100803, 100805, 100813, 100814, 100823, 100830, 100833, 100846, 100864, 100866, 100870, 100871, 100873, 100886, 100891, 100893, 100910, 700191, 700207, 700250, 700252, 700272, 700306, 700507, 700582, 700630, 700713, 700726, 700735, 700762, 700796, 700801, 700802, 700820, 700832, 700834, 700837, 700839, 700852, 700853, 700854, 700856, 700860, 700863, 700867, 700985, 701002, 701003, 701015, 701023, 701033, 701060, 701104, 701168, 701218, 701231, 701232, 701249, 701250, 701252, 701256, 701266, 701270, 701271, 701273, 701278, 701281, 701288, 701291, 701296, 701301, 701302, 701312, 701323, 701325, 701332, 701341, 701351, 701355, 701357, 701414, 701416, 701432, 701451, 701485, 701520, 701539, 701560, 701562, 701563, 701564, 701566, 701569, 701570, 701571, 701573, 701574, 701587, 701593, 701599, 701600, 701601, 701603, 701606, 701609, 701612, 701613, 701616, 701617, 701618, 701623, 701624, 701625, 701638, 701640, 701641, 701642, 701646, 701648, 701651, 701653, 701656, 701657, 701659, 701662, 701665, 701667, 701669, 701671, 701675, 701676, 701678, 701681, 701682, 701683, 701684, 701691, 701692, 701693, 701696, 701697, 701699, 701702, 701703, 701704, 701705, 701707, 701709, 701718, 701720, 701722, 701724, 701725, 701726, 701733, 701739, 701742, 701745, 701746, 701751, 701752, 701755, 701757, 701761, 701763, 701764, 701766, 701767, 701770, 701771, 701772, 701774, 701780, 701781, 701789, 701790, 701791, 701795, 701798, 701816, 701818, 701823, 701828, 701829, 701841, 701848, 701858, 701869, 701873, 701879, 701887, 701889, 701895, 701896, 701910, 701916, 701921, 701934, 701954, 701966, 701970, 701981, 701983, 701989, 701991, 701992, 701999, 702002, 702011, 702012, 702020, 702024, 702029, 702030, 702033, 702040, 702045, 702046, 702047, 702052, 702056, 702059, 702060, 702065, 702066, 702080, 702087, 702091, 702102, 702105, 702116, 702137, 702139, 702156, 702157, 702162, 702220, 702221, 702231, 702235, 702237, 702242, 702245, 702256, 702260, 702263, 702264, 702275, 702288, 702353, 702357, 702360, 702361, 702364, 702367, 702384, 702395, 702402, 702404, 702413, 702414, 702427, 702431, 702433, 702436, 702438, 702439, 702441, 702445, 702455, 702458, 702466, 702467, 702469, 702470, 702475, 702476, 702478, 702482, 702494, 702498, 702505, 702506, 702511, 702512, 702515, 702516, 702517, 702542, 702553, 702569, 702587, 702592, 702596, 702616, 702619, 702620, 702627, 702628, 702630, 702638, 702644, 702649, 702652, 702659, 702665, 702686, 702687, 702691, 702692, 702702, 702707, 702710, 702724, 702732, 702737, 702743, 702744, 702746, 702748, 702751, 702758, 702768, 702774, 702780, 702803, 702804, 702808, 702813, 702814, 702839, 702851, 702863, 702880, 702881, 702889, 702896, 702922, 702929, 702943, 702953, 702957, 702968, 702976, 702977, 702985, 703000, 703012, 703014, 703020, 703047, 703067, 703096, 703101, 703102, 703103, 703117, 703118, 703120, 703121, 703128, 703131, 703135, 703141, 703142, 703144, 703146, 703147, 703150, 703152, 703154, 703156, 703157, 703163, 703167, 703176, 703189, 703193, 703195, 703200, 703204, 703211, 703214, 703215, 703222, 703223, 703225, 703227, 703229, 703234, 703236, 703240, 703248, 703249, 703250, 703252, 703253, 703254, 703256, 703257, 703268, 703281, 703284, 703290, 703293, 703297, 703301, 703304, 703305, 703326, 703338, 703342, 703344, 703351, 703356, 703364, 703366, 703369, 703371, 703374, 703375, 703376, 703379, 703383, 703386, 703405, 703410, 703414, 703417, 703420, 703428, 703449, 703455, 703457, 703459, 703460, 703478, 703479, 703487, 703488, 703505, 703523, 703524, 703532, 703544, 703553, 703562, 703598, 703604, 703613, 703639, 703649, 703653, 703656, 703661, 703671, 703675, 703677, 703681, 703687, 703702, 703709, 703714, 703715, 703720, 703724, 703727, 703733, 703734, 703744, 703746, 703760, 703763, 703764, 703766, 703767, 703768, 703769, 703771, 703775, 703780, 703785, 703796, 703804, 703809, 703825, 703835, 703841, 703844, 703848, 703855, 703878, 703884, 703892, 703896, 703901, 703914, 703918, 703922, 703924, 703937, 703942, 703943, 703944, 703946, 703950, 703952, 703956, 703959, 703967, 703978, 703988, 703991, 703993, 703996, 704000, 704001, 704002, 704005, 704007, 704010, 704014, 704020, 704024, 704026, 704028, 704035, 704041, 704042, 704045, 704050, 704052, 704054, 704055, 704058, 704059, 704069, 704072, 704074, 704077, 704080, 704085, 704089, 704095, 704096, 704101, 704110, 704112, 704116, 704117, 704118, 704120, 704121, 704122, 704124, 704134, 704138, 704153, 704158, 704160, 704162, 704163, 704164, 704166, 704167, 704169, 704171, 704183, 704185, 704186, 704206, 704220, 704222, 704226, 704246, 704248, 704253, 704255, 704257, 704260, 704262, 704278, 704280, 704287, 704289, 704290, 704291, 704293, 704299, 704300, 704307, 704309, 704311, 704314, 704315, 704321, 704340, 704342, 704351, 704354, 704402, 704438, 704441, 704449, 704456, 704459, 704460, 704461, 704464, 704465, 704468, 704469, 704471, 704476, 704482, 704483, 704484, 704490, 704491, 704492, 704498, 704499, 704503, 704504, 704505, 704507, 704509, 704514, 704520, 704521, 704522, 704525, 704527, 704528, 704530, 704537, 704539, 704542, 704545, 704547, 704551, 704557, 704560, 704563, 704564, 704571, 704588, 704591, 704598, 704602, 704604, 704605, 704608, 704615, 704616, 704626, 704655, 704656, 704659, 704662, 704663, 704665, 704668, 704671, 704674, 704678, 704683, 704688, 704690, 704692, 704698, 704701, 704703, 704704, 704710, 704711, 704712, 704713, 704719, 704722, 704723, 704726, 704728, 704729, 704730, 704731, 704734, 704739, 704740, 704742, 704747, 704748, 704773, 704777, 704793, 704801, 704803, 704820, 704826, 704828, 704830, 704834, 704838, 704847, 704851, 704860, 704863, 704868, 704869, 704883, 704895, 704896, 704902, 704906, 704911, 704923, 704925, 704928, 704930, 704931, 704932, 704935, 704938, 704939, 704940, 704942, 704943, 704952, 704953, 704957, 704965, 704966, 704970, 704973, 704976, 704980, 704981, 704983, 704985, 704986, 704987, 704988, 704994, 704997, 705001, 705002, 705004, 705005, 705008, 705011, 705012, 705013, 705016, 705018, 705020, 705021, 705027, 705041, 705044, 705045, 705046, 705047, 705048, 705050, 705051, 705052, 705056, 705059, 705062, 705066, 705071, 705075, 705078, 705084, 705097, 705107, 705108, 705116, 705117, 705121, 705132, 705136, 705148, 705190, 705193, 705197, 705221, 705226, 705239, 705248, 705270, 705273, 705277, 705291, 705292, 705301, 705320, 705328, 705336, 705337, 705342, 705343, 705347, 705350, 705352, 705353, 705361, 705362, 705363, 705370, 705375, 705378, 705387, 705390, 705393, 705401, 705408, 705411, 705413, 705416, 705420, 705423, 705426, 705431, 705433, 705434, 705436, 705441, 705442, 705445, 705446, 705447, 705452, 705453, 705455, 705459, 705460, 705461, 705467, 705469, 705475, 705480, 705481, 705482, 705492, 705495, 705502, 705510, 705511, 705521, 705533, 705535, 705554, 705570, 705575, 705576, 705580, 705581, 705582, 705589, 705594, 705595, 705596, 705599, 705605, 705606, 705613, 705616, 705622, 705628, 705631, 705639, 705642, 705646, 705647, 705662, 705663, 705666, 705685, 705688, 705692, 705703, 705712, 705715, 705722, 705724, 705725, 705727, 705734, 705737, 705743, 705746, 705753, 705755, 705759, 705762, 705764, 705768, 705783, 705799, 705803, 705808, 705820, 705821, 705829, 705831, 705832, 705833, 705842, 705845, 705848, 705849, 705852, 705855, 705859, 705860, 705868, 705884, 705886, 705888, 705889, 705890, 705893, 705894, 705896, 705898, 705899, 705902, 705904, 705905, 705920, 705928, 705936, 705942, 705943, 705949, 705950, 705953, 705954, 705958, 705962, 705966, 705972, 705980, 705982, 705983, 705987, 705988, 705989, 705990, 705994, 705996, 705999, 706016, 706022, 706027, 706042, 706045, 706048, 706052, 706054, 706055, 706056, 706059, 706067, 706075, 706076, 706083, 706088, 706096, 706103, 706104, 706115, 706117, 706119, 706120, 706131, 706147, 706174, 706189, 706193, 706197, 706198, 706201, 706208, 706211, 706212, 706214, 706217, 706232, 706243, 706248, 706251, 706255, 706260, 706261, 706266, 706267, 706283, 706288, 706303, 706310, 706314, 706320, 706334, 706341, 706343, 706344, 706354, 706355, 706362, 706363, 706367, 706372, 706375, 706377, 706380, 706381, 706383, 706384, 706388, 706391, 706403, 706405, 706407, 706410, 706411, 706413, 706414, 706415, 706416, 706418, 706419, 706421, 706422, 706424, 706425, 706429, 706434, 706438, 706442, 706444, 706445, 706453, 706454, 706457, 706461, 706463, 706466, 706471, 706472, 706479, 706484, 706485, 706490, 706499, 706500, 706502, 706505, 706508, 706525, 706527, 706529, 706535, 706536, 706541, 706543, 706544, 706548, 706551, 706553, 706555, 706557, 706562, 706564, 706574, 706578, 706585, 706589, 706590, 706597, 706598, 706605, 706624, 706640, 706642, 706645, 706647, 706655, 706660, 706665, 706667, 706671, 706673, 706676, 706677, 706678, 706679, 706682, 706683, 706685, 706686, 706694, 706696, 706697, 706698, 706700, 706703, 706708, 706712, 706714, 706715, 706716, 706721, 706733, 706735, 706740, 706744, 706750, 706753, 706754, 706755, 706763, 706765, 706769, 706771, 706775, 706783, 706784, 706789, 706800, 706802, 706804, 706805, 706806, 706814, 706822, 706824, 706825, 706826, 706827, 706828, 706830, 706831, 706832, 706833, 706834, 706836, 706842, 706844, 706845, 706847, 706849, 706854, 706855, 706856, 706857, 706860, 706861, 706863, 706866, 706867, 706869, 706870, 706873, 706874, 706880, 706883, 706886, 706890, 706893, 706897, 706903, 706907, 706915, 706933, 706936, 706977, 706983, 707007, 707049, 707104, 707105, 707107, 707108, 707118, 707122, 707136, 707138, 707141, 707143, 707152, 707160, 707163, 707168, 707172, 707175, 707176, 707183, 707228, 707230, 707238, 707241, 707246, 707247, 707250, 707257, 707260, 707265, 707266, 707269, 707277, 707281, 707283, 707285, 707286, 707289, 707290, 707291, 707292, 707295, 707296, 707297, 707299, 707300, 707308, 707309, 707313, 707321, 707322, 707323, 707325, 707341, 707351, 707352, 707353, 707355, 707357, 707358, 707359, 707360, 707361, 707363, 707364, 707369, 707375, 707376, 707377, 707379, 707387, 707389, 707402, 707409, 707418, 707428, 707435, 707449, 707452, 707456, 707475, 707481, 707485, 707486, 707488, 707491, 707493, 707497, 707498, 707500, 707504, 707513, 707517, 707519, 707520, 707526, 707530, 707531, 707535, 707537, 707541, 707555, 707571, 707616, 707624, 707628, 707638, 707642, 707643, 707653, 707656, 707662, 707665, 707667, 707668, 707675, 707676, 707685, 707687, 707690, 707691, 707692, 707693, 707694, 707696, 707699, 707700, 707702, 707703, 707705, 707707, 707711, 707712, 707714, 707717, 707724, 707730, 707737, 707741, 707746, 707750, 707754, 707759, 707762, 707765, 707770, 707772, 707776, 707781, 707793, 707799, 707806, 707811, 707815, 707816, 707817, 707819, 707822, 707825, 707830, 707832, 707833, 707835, 707837, 707841, 707842, 707845, 707846, 707856, 707858, 707863, 707864, 707867, 707869, 707870, 707875, 707876, 707877, 707879, 707888, 707890, 707897, 707898, 707906, 707907, 707912, 707913, 707916, 707919, 707924, 707925, 707934, 707943, 707949, 707954, 707959, 707969, 707979, 707983, 707984, 707986, 708027, 708029, 708030, 708032, 708047, 708049, 708050, 708060, 708061, 708064, 708075, 708077, 708079, 708082, 708084, 708086, 708097, 708099, 708100, 708105, 708111, 708113, 708114, 708125, 708131, 708134, 708173, 708178, 708180, 708182, 708185, 708189, 708198, 708204, 708207, 708209, 708225, 708226, 708237, 708240, 708245, 708252, 708253, 708267, 708280, 708285, 708287, 708289, 708290, 708296, 708300, 708313, 708315, 708337, 708342, 708343, 708351, 708360, 708361, 708377, 708378, 708379, 708385, 708386, 708387, 708391, 708393, 708394, 708395, 708396, 708399, 708417, 708419, 708421, 708423, 708424, 708425, 708426, 708428, 708433, 708435, 708436, 708438, 708439, 708440, 708442, 708446, 708449, 708450, 708451, 708453, 708456, 708460, 708462, 708463, 708466, 708467, 708468, 708470, 708471, 708473, 708474, 708476, 708477, 708479, 708480, 708481, 708484, 708485, 708490, 708491, 708493, 708495, 708496, 708498, 708499, 708500, 708501, 708502, 708503, 708504, 708506, 708507, 708512, 708513, 708515, 708518, 708520, 708524, 708526, 708528, 708530, 708532, 708537, 708539, 708541, 708543, 708544, 708551, 708552, 708553, 708556, 708558, 708563, 708583, 708584, 708586, 708594, 708603, 708605, 708606, 708611, 708614, 708616, 708619, 708620, 708626, 708651, 708698, 708701, 708730, 708735, 708739, 708744, 708751, 708753, 708754, 708758, 708759, 708762, 708771, 708780, 708783, 708791, 708798, 708799, 708801, 708807, 708838, 708842, 708855, 708856, 708867, 708868, 708873, 708875, 708919, 708923, 708928, 708932, 708933, 708938, 708950, 708952, 708962, 708967, 708987, 708991, 708996, 709007, 709008, 709011, 709015, 709016, 709034, 709039, 709040, 709043, 709046, 709051, 709053, 709054, 709058, 709059, 709060, 709062, 709063, 709065, 709066, 709071, 709073, 709074, 709076, 709078, 709081, 709082, 709083, 709086, 709090, 709092, 709099, 709101, 709103, 709107, 709111, 709114, 709122, 709123, 709127, 709129, 709130, 709134, 709146, 709148, 709149, 709167, 709175, 709177, 709178, 709180, 709181, 709186, 709190, 709200, 709201, 709205, 709216, 709225, 709233, 709237, 709242, 709244, 709251, 709261, 709262, 709270, 709272, 709274, 709278, 709280, 709282, 709288, 709290, 709291, 709293, 709294, 709295, 709296, 709299, 709301, 709314, 709316, 709317, 709318, 709324, 709327, 709336, 709340, 709343, 709347, 709348, 709352, 709353, 709354, 709356, 709361, 709374, 709379, 709381, 709384, 709399, 709403, 709405, 709408, 709414, 709415, 709427, 709434, 709454, 709457, 709465, 709468, 709495, 709496, 709499, 709500, 709506, 709511, 709524, 709525, 709527, 709534, 709546, 709547, 709550, 709551, 709552, 709567, 709577, 709580, 709583, 709587, 709588, 709593, 709598, 709606, 709611, 709619, 709621, 709625, 709627, 709632, 709634, 709638, 709644, 709646, 709654, 709656, 709663, 709664, 709667, 709671, 709674, 709676, 709682, 709684, 709694, 709697, 709698, 709701, 709705, 709708, 709712, 709719, 709732, 709738, 709739, 709740, 709743, 709747, 709748, 709749, 709752, 709755, 709758, 709761, 709762, 709767, 709769, 709775, 709776, 709781, 709808, 709818, 709846, 709847, 709859, 709866, 709869, 709872, 709873, 709876, 709877, 709879, 709885, 709887, 709908, 709927, 709929, 709932, 709934, 710010, 710021, 710022, 710025, 710028, 710034, 710035, 710036, 710037, 710038, 710042, 710047, 710049, 710050, 710051, 710053, 710057, 710061, 710063, 710064, 710066, 710069, 710071, 710072, 710073, 710074, 710077, 710084, 710085, 710090, 710092, 710097, 710108, 710118, 710119, 710120, 710121, 710122, 710124, 710126, 710127, 710129, 710131, 710132, 710135, 710136, 710137, 710138, 710141, 710142, 710144, 710145, 710146, 710151, 710157, 710161, 710166, 710168, 710170, 710177, 710178, 710188, 710199, 710200, 710201, 710211, 710212, 710213, 710214, 710228, 710229, 710237, 710238, 710242, 710245, 710248, 710253, 710254, 710258, 710264, 710268, 710270, 710274, 710279, 710296, 710300, 710302, 710303, 710608, 710611, 710627, 710628, 710629, 710631, 710637, 710642, 710645, 710647, 710649, 710651, 710652, 710655, 710667, 710675, 710676, 710678, 710679, 710680, 710681, 710682, 710687, 710699, 710701, 710710, 710716, 710719, 710723, 710724, 710728, 710729, 710735, 710738, 710743, 710744, 710752, 710753, 710756, 710757, 710759, 710760, 710765, 710766, 710769, 710772, 710773, 710776, 710777, 710778, 710779, 710782, 710784, 710789, 710800, 710801, 710807, 710809, 710810, 710811, 710818, 710825, 710827, 710828, 710830, 710831, 710838, 710839, 710842, 710846, 710852, 710855, 710862, 710866, 710869, 710875, 710878, 710882, 710891, 710892, 710901, 710906, 710910, 710912, 710914, 710915, 710920, 710924, 710932, 710933, 710955, 710958, 710960, 710962, 710964, 710965, 710968, 710969, 710970, 710973, 710976, 710977, 710978, 710981, 710986, 710987, 710988, 710990, 710991, 711013, 711015, 711018, 711021, 711023, 711024, 711025, 711031, 711032, 711034, 711036, 711038, 711039, 711040, 711041, 711042, 711045, 711046, 711047, 711050, 711051, 711052, 711055, 711056, 711059, 711061, 711063, 711065, 711069, 711073, 711075, 711078, 711079, 711080, 711082, 711083, 711087, 711093, 711099, 711117, 711120, 711129, 711141, 711143, 711145, 711147, 711153, 711154, 711162, 711168, 711170, 711182, 711183, 711184, 711185, 711194, 711199, 711202, 711209, 711211, 711212, 711238, 711243, 711248, 711252, 711279, 711282, 711283, 711288, 711289, 711290, 711291, 711295, 711296, 711299, 711300, 711302, 711306, 711312, 711316, 711318, 711332, 711333, 711334, 711338, 711343, 711348, 711356, 711358, 711359, 711363, 711366, 711382, 711393, 711398, 711403, 711404, 711405, 711406, 711408, 711409, 711410, 711411, 711412, 711413, 711424, 711426, 711432, 711435, 711443, 711459, 711463, 711466, 711470, 711473, 711484, 711500, 711501, 711506, 711510, 711512, 711515, 711518, 711528, 711536, 711539, 711540, 711543, 711547, 711550, 711560, 711583, 711601, 711605, 711624, 711630, 711631, 711639, 711640, 711642, 711644, 711657, 711662, 711667, 711668, 711669, 711675, 711678, 711679, 711685, 711695, 711696, 711723, 711724, 711732, 711736, 711740, 711744, 711745, 711750, 711757, 711762, 711764, 711773, 711775, 711791, 711808, 711818, 711831, 711832, 711833, 711834, 711836, 711839, 711844, 711846, 711847, 711854, 711857, 711868, 711871, 711892, 711893, 711902, 711904, 711909, 711911, 711912, 711914, 711918, 711922, 711923, 711926, 711942, 711949, 711952, 711967, 711969, 711989, 712000, 712007, 712011, 712017, 712020, 712021, 712026, 712038, 712054, 712058, 712059, 712071, 712074, 712075, 712079, 712091, 712092, 712106, 712110, 712119, 712126, 712129, 712130, 712134, 712137, 712139, 712149, 712153, 712154, 712155, 712156, 712157, 712158, 712160, 712161, 712162, 712166, 712167, 712172, 712174, 712189, 712208, 712210, 712216, 712218, 712224, 712225, 712232, 712238, 712242, 712243, 712255, 712257, 712258, 712260, 712268, 712293, 712298, 712301, 712331, 712334, 712435, 712442, 712445, 712453, 712455, 712458, 712460, 712461, 712462, 712463, 712464, 712469, 712473, 712474, 712477, 712479, 712481, 712482, 712483, 712488, 712491, 712492, 712493, 712495, 712498, 712499, 712502, 712503, 712506, 712507, 712508, 712517, 712523, 712531, 712532, 712534, 712535, 712536, 712537, 712538, 712540, 712542, 712545, 712546, 712547, 712553, 712555, 712557, 712558, 712567, 712571, 712577, 712612, 712613, 712625, 712626, 712631, 712633, 712639, 712645, 712654, 712657, 712686, 712687, 712688, 712709, 712710, 712731, 712737, 712741, 712751, 712754, 712756, 712764, 712773, 712776, 712788, 712789, 712792, 712793, 712796, 712797, 712802, 712807, 712830, 712838, 712842, 712853, 712860, 712869, 712871, 712875, 712879, 712881, 712885, 712887, 712891, 712892, 712894, 712895, 712911, 712919, 712921, 712922, 712925, 712926, 712930, 712931, 712933, 712939, 712940, 712941, 712942, 712945, 712948, 712954, 712955, 712956, 712959, 712967, 712972, 712976, 712979, 712983, 712984, 712990, 712994, 713006, 713026, 713030, 713032, 713033, 713035, 713036, 713041, 713043, 713045, 713049, 713050, 713053, 713058, 713060, 713063, 713069, 713070, 713074, 713075, 713076, 713078, 713081, 713084, 713086, 713088, 713092, 713095, 713097, 713103, 713105, 713110, 713111, 713115, 713116, 713120, 713127, 713131, 713133, 713134, 713160, 713169, 713343, 713344, 713346, 713352, 713358, 713360, 713364, 713393, 713607, 713669, 713675, 713691, 713699, 713701, 713705, 713706, 713718, 713720, 713728, 713729, 713732, 713743, 713744, 713747, 713748, 713752, 713753, 713760, 713766, 713767, 713768, 713770, 713776, 713779, 713780, 713785, 713789, 713791, 713792, 713793, 713794, 713796, 713800, 713801, 713804, 713807, 713809, 713812, 713813, 713818, 713826, 713829, 713840, 713841, 713846, 713849, 713851, 713852, 713855, 713856, 713860, 713862, 713865, 713866, 713867, 713868, 713869, 713870, 713871, 713872, 713873, 713875, 713878, 713879, 713880, 713884, 713887, 713888, 713890, 713893, 713895, 713898, 713900, 713901, 713902, 713904, 713905, 713908, 713912, 713915, 713919, 713920, 713921, 713922, 713923, 713924, 713925, 713935, 713936, 713937, 713939, 713940, 713941, 713946, 713947, 713949, 713953, 713955, 713963, 713964, 713965, 713970, 713973, 713974, 713979, 713980, 713983, 713984, 713986, 713987, 713989, 713990, 713991, 713994, 713996, 713997, 714001, 714002, 714003, 714006, 714007, 714008, 714011, 714012, 714016, 714017, 714019, 714022, 714030, 714037, 714043, 714044, 714059, 714061, 714064, 714065, 714066, 714067, 714073, 714076, 714078, 714081, 714091, 714094, 714096, 714098, 714103, 714115, 714116, 714120, 714122, 714130, 714148, 714149, 714152, 714154, 714155, 714157, 714158, 714162, 714165, 714169, 714179, 714180, 714182, 714183, 714184, 714185, 714188, 714189, 714193, 714200, 714201, 714204, 714205, 714206, 714207, 714208, 714210, 714213, 714215, 714222, 714224, 714226, 714227, 714228, 714230, 714232, 714234, 714235, 714237, 714240, 714241, 714245, 714246, 714254, 714256, 714261, 714262, 714263, 714266, 714267, 714269, 714273, 714275, 714281, 714282, 714283, 714284, 714287, 714289, 714290, 714291, 714294, 714295, 714296, 714297, 714305, 714322, 714333, 714334, 714336, 714348, 714350, 714353, 714357, 714365, 714368, 714370, 714371, 714374, 714375, 714380, 714393, 714394, 714405, 714407, 714409, 714412, 714416, 714418, 714419, 714421, 714424, 714425, 714426, 714435, 714438, 714440, 714441, 714443, 714445, 714446, 714449, 714452, 714453, 714454, 714455, 714456, 714459, 714462, 714467, 714469, 714470, 714472, 714473, 714476, 714477, 714479, 714481, 714483, 714484, 714490, 714491, 714492, 714493, 714495, 714496, 714501, 714502, 714504, 714508, 714509, 714512, 714517, 714519, 714524, 714526, 714537, 714538, 714541, 714542, 714551, 714552, 714554, 714564, 714575, 714581, 714583, 714585, 714588, 714593, 714599, 714602, 714604, 714607, 714612, 714613, 714615, 714626, 714632, 714638, 714640, 714641, 714642, 714645, 714647, 714650, 714651, 714653, 714654, 714655, 714656, 714657, 714658, 714671, 714674, 714681, 714682, 714683, 714684, 714686, 714691, 714702, 714706, 714707, 714708, 714709, 714710, 714712, 714713, 714714, 714715, 714717, 714718, 714729, 714738, 714742, 714750, 714751, 714752, 714753, 714754, 714758, 714764, 714787, 714830, 714834, 714835, 714850, 714857, 714858, 714861, 714862, 714866, 714867, 714868, 714869, 714871, 714873, 714874, 714875, 714876, 714878, 714879, 714880, 714882, 714886, 714887, 714888, 714918, 714920, 714923, 714933, 714936, 714959, 714960, 714962, 714963, 714964, 714965, 714967, 714982, 714983, 714985, 714987, 715019, 715021, 715023, 715024, 715025, 715026, 715027, 715028, 715059, 715061, 715063, 715099, 715103, 715104, 715106, 715107, 715139, 715141, 715142, 715146, 715147, 715148, 715191, 715192, 715194, 715195, 715197, 715213, 715216, 715229, 715230, 715231, 715233, 715235, 715236, 715237, 715238, 715251, 715256, 715312, 715315, 715318, 715409, 715410, 715412, 715413, 715414, 715415, 715418, 715423, 715426, 715427, 715428, 715430, 715433, 715434, 715435, 715436, 715441, 715446, 715448, 715451, 715453, 715455, 715458, 715465, 715467, 715469, 715470, 715473, 715474, 715475, 715476, 715477, 715478, 715479, 715480, 715482, 715483, 715484, 715487, 715488, 715570, 715571, 715573, 715575, 715576, 715580, 715581, 715582, 715583, 715584, 715585, 715586, 715589, 715592, 715593, 715594, 715596, 715598, 715604, 715610, 715611, 715613, 715614, 715616, 715620, 715623, 715628, 715652, 715655, 715656, 715701, 715707, 715781, 715782, 715785, 715812, 715870, 715871, 715875, 715876, 715909, 715916, 715917, 715930, 715931, 715932, 715934, 715935, 715936, 715937, 715938, 715941, 715942, 715943, 715944, 715945, 715946, 715947, 715948, 715951, 715952, 715953, 715954, 715955, 715956, 715957, 715958, 715959, 715960, 715962, 715963, 715964, 715965, 715966, 715967, 715968, 715969, 715971, 715972, 715977, 715980, 715981, 715985, 715987, 715990, 715991, 715992, 715993, 715994, 715995, 715996, 715997, 716005, 716007, 716012, 716015, 716017, 716021, 716022, 716023, 716025, 716038, 716100, 716191, 716195, 716200, 716201, 716204, 716205, 716209, 716210, 716232, 716237, 716238, 716243, 716265, 716266, 716268, 716272, 716275, 716284, 716288, 716290, 716295, 716300, 716301, 716309, 716310, 716311, 716312, 716315, 716319, 716321, 716325, 716328, 716329, 716337, 716341, 716344, 716345, 716346, 716347, 716353, 716359, 716362, 716364, 716368, 716369, 716371, 716375, 716377, 716378, 716380, 716383, 716384, 716387, 716393, 716394, 716404, 716408, 716431, 716434, 716458, 716461, 716466, 716467, 716473, 716475, 716480, 716488, 716489, 716495, 716499, 716500, 716502, 716503, 716505, 716506, 716513, 716514, 716517, 716518, 716520, 716522, 716523, 716525, 716528, 716532, 716533, 716534, 716549, 716554, 716557, 716558, 716561, 716562, 716563, 716565, 716568, 716573, 716576, 716577, 716578, 716580, 716581, 716585, 716587, 716588, 716589, 716590, 716594, 716603, 716608, 716612, 716613, 716614, 716621, 716624, 716628, 716641, 716643, 716656, 716671, 716687, 716703, 716705, 716708, 716711, 716721, 716722, 716723, 716725, 716728, 716731, 716734, 716737, 716741, 716743, 716747, 716750, 716755, 716757, 716758, 716759, 716762, 716763, 716764, 716766, 716768, 716776, 716777, 716778, 716780, 716782, 716796, 716799, 716803, 716804, 716805, 716806, 716807, 716810, 716811, 716815, 716817, 716822, 716825, 716827, 716829, 716839, 716842, 716843, 716844, 716845, 716846, 716847, 716848, 716849, 716850, 716851, 716856, 716859, 716860, 716861, 716862, 716863, 716864, 716869, 716878, 716880, 716881, 716883, 716884, 716885, 716886, 716887, 716888, 716890, 716891, 716892, 716893, 716895, 716897, 716899, 716900, 716903, 716923, 716924, 716925, 716928, 717095, 717097, 717245, 717267, 717287, 717291, 717296, 717307, 717318, 717326, 717335, 717355, 717360, 717363, 717372, 717394, 717395, 717396, 717404, 717409, 717410, 717414, 717418, 717420, 717426, 717428, 717429, 717430, 717448, 717449, 717451, 717452, 717456, 717457, 717458, 717459, 717461, 717463, 717464, 717465, 717466, 717467, 717473, 717476, 717480, 717483, 717485, 717487, 717488, 717493, 717498, 717500, 717502, 717505, 717506, 717508, 717535, 717536, 717541, 717559, 717567, 717586, 717644, 717659, 717660, 717679, 717680, 717683, 717685, 717687, 717698, 717699, 717700, 717702, 717710, 717712, 717715, 717719, 717724, 717726, 717730, 717736, 717742, 717746, 717749, 717760, 717766, 717768, 717769, 717770, 717776, 717777, 717780, 717782, 717783, 717784, 717785, 717786, 717790, 717793, 717794, 717796, 717797, 717798, 717801, 717802, 717803, 717804, 717805, 717807, 717810, 717812, 717813, 717815, 717816, 717817, 717822, 717826, 717827, 717828, 717842, 717963, 717966, 717968, 717979, 717981, 717986, 717993, 717994, 717995, 717997, 717998, 718004, 718008, 718009, 718011, 718013, 718014, 718016, 718019, 718023, 718025, 718033, 718035, 718041, 718043, 718044, 718058, 718066, 718071, 718072, 718074, 718077, 718078, 718081, 718085, 718087, 718088, 718094, 718096, 718097, 718099, 718100, 718101, 718104, 718106, 718110, 718116, 718118, 718120, 718122, 718129, 718131, 718132, 718136, 718138, 718140, 718147, 718151, 718156, 718159, 718161, 718163, 718164, 718166, 718172, 718175, 718177, 718178, 718180, 718181, 718182, 718184, 718185, 718186, 718188, 718191, 718193, 718197, 718200, 718202, 718208, 718210, 718211, 718218, 718228, 718263, 718276, 718288, 718399, 718401, 718408, 718439, 718441, 718445, 718447, 718448, 718449, 718453, 718454, 718455, 718456, 718457, 718482, 718506, 718525, 718529, 718531, 718534, 718535, 718536, 718540, 718546, 718550, 718551, 718554, 718556, 718561, 718573, 718577, 718578, 718589, 718590, 718591, 718594, 718597, 718608, 718716, 718721, 718737, 718747, 718759, 718761, 718762, 718763, 718764, 718765, 718767, 718768, 718771, 718773, 718774, 718775, 718777, 718778, 718779, 718780, 718781, 718783, 718784, 718785, 718787, 718788, 718793, 718797, 718802, 718804, 718807, 718808, 718811, 718812, 718814, 718816, 718823, 718824, 718833, 718835, 718837, 718843, 718846, 718848, 718854, 718856, 718857, 718859, 718861, 718872, 718875, 718878, 718881, 718885, 718893, 718919, 718929, 718934, 718935, 718938, 718942, 718943, 718946, 718947, 718950, 718958, 718959, 718969, 718972, 718973, 718977, 718979, 718983, 718988, 718995, 718996, 718997, 719020, 719023, 719028, 719029, 719038, 719041, 719047, 719056, 719062, 719076, 719078, 719108, 719110, 719116, 719150, 719152, 719154, 719157, 719158, 719159, 719160, 719164, 719166, 719181, 719182, 719185, 719191, 719199, 719200, 719201, 719207, 719208, 719209, 719210, 719212, 719213, 719215, 719216, 719217, 719218, 719219, 719221, 719222, 719225, 719226, 719227, 719229, 719231, 719233, 719234, 719235, 719236, 719237, 719238, 719240, 719243, 719251, 719252, 719253, 719254, 719255, 719256, 719257, 719261, 719262, 719264, 719265, 719266, 719269, 719270, 719275, 719276, 719277, 719280, 719282, 719283, 719284, 719289, 719291, 719293, 719294, 719295, 719297, 719303, 719304, 719305, 719306, 719308, 719309, 719310, 719311, 719312, 719313, 719314, 719315, 719317, 719320, 719321, 719323, 719324, 719328, 719329, 719331, 719332, 719335, 719337, 719341, 719343, 719346, 719354, 719355, 719357, 719358, 719360, 719361, 719372, 719376, 719378, 719379, 719383, 719384, 719387, 719389, 719395, 719399, 719400, 719401, 719403, 719404, 719406, 719407, 719409, 719410, 719411, 719415, 719417, 719420, 719421, 719422, 719425, 719435, 719437, 719439, 719442, 719443, 719444, 719445, 719447, 719449, 719455, 719457, 719458, 719459, 719461, 719463, 719466, 719467, 719482, 719512, 719513, 719562, 719563, 719647, 719671, 719700, 719703, 719711, 719713, 719719, 719720, 719721, 719723, 719724, 719726, 719732, 719733, 719734, 719735, 719736, 719738, 719739, 719740, 719743, 719746, 719756, 719757, 719759, 719760, 719761, 719770, 719782, 719785, 719791, 719799, 719807, 719822, 719825, 719826, 719831, 719843, 719846, 719848, 719853, 719855, 719857, 719858, 719859, 719867, 719870, 719872, 719874, 719883, 719884, 719886, 719890, 719891, 719892, 719896, 719901, 719906, 719907, 719908, 719909, 719911, 719913, 719915, 719916, 719917, 719919, 719923, 719950, 719953, 719954, 719956, 719957, 719960, 719963, 719965, 719968, 719979, 719986, 719990, 720009, 720015, 720024, 720033, 720051, 720052, 720058, 720060, 720061, 720062, 720063, 720064, 720066, 720078, 720084, 720088, 720094, 720096, 720097, 720103, 720105, 720108, 720109, 720110, 720111, 720114, 720117, 720121, 720130, 720145, 720151, 720154, 720159, 720167, 720172, 720179, 720180, 720181, 720182, 720187, 720188, 720191, 720193, 720198, 720204, 720206, 720209, 720210, 720211, 720212, 720213, 720214, 720215, 720216, 720220, 720221, 720228, 720229, 720234, 720235, 720236, 720237, 720239, 720245, 720247, 720250, 720252, 720261, 720262, 720268, 720274, 720276, 720284, 720292, 720294, 720298, 720300, 720301, 720302, 720304, 720306, 720325, 720328, 720329, 720333, 720339, 720342, 720350, 720358, 720367, 720370, 720372, 720374, 720382, 720402, 720434, 720441, 720443, 720446, 720453, 720456, 720458, 720459, 720461, 720469, 720472, 720474, 720492, 720495, 720501, 720502, 720508, 720509, 720514, 720516, 720519, 720521, 720527, 720551, 720557, 720561, 720563, 720567, 720568, 720569, 720570, 720576, 720578, 720580, 720587, 720589, 720593, 720595, 720597, 720598, 720981, 720982, 720985, 720995, 721002, 721014, 721083, 721088, 721089, 721090, 721093, 721095, 721097, 721099, 721100, 721102, 721103, 721104, 721107, 721108, 721109, 721110, 721114, 721115, 721119, 721121, 721122, 721123, 721131, 721132, 721133, 721134, 721137, 721140, 721289, 721347, 721356, 721357, 721363, 721380, 721421, 721436, 721437, 721492, 721513, 721521, 721523, 721528, 721540, 721543, 721544, 721546, 721549, 721557, 721558, 721571, 721587, 721589, 721593, 721599, 721701, 721707, 721713, 721716, 721721, 721722, 721723, 721724, 721726, 721728, 721731, 721732, 721738, 721739, 721752, 721754, 721757, 721760, 721762, 721763, 721766, 721771, 721787, 721789, 721795, 721799, 721802, 721804, 721821, 721902, 721922, 721932, 721937, 721938, 721955, 721956, 721958, 721959, 721984, 721990, 722053, 722091, 722092, 722099, 722103, 722104, 722105, 722107, 722139, 722140, 722148, 722152, 722155, 722156, 722157, 722158, 722160, 722161, 722164, 722168, 722169, 722170, 722171, 722174, 722175, 722176, 722177, 722182, 722184, 722187, 722188, 722190, 722207, 722261, 722263, 722269, 722415, 722421, 722425, 722428, 722430, 722492, 722497, 722501, 722506, 722507, 722510, 722513, 722514, 722516, 722518, 722519, 722528, 722530, 722537, 722554, 722564, 722577, 722579, 722582, 722586, 722587, 722595, 722600, 722602, 722604, 722605, 722607, 722609, 722610, 722611, 722612, 722614, 722618, 722620, 722653, 722655, 722662, 722680, 722686, 722689, 722694, 722696, 722703, 722705, 722707, 722714, 722715, 722721, 722741, 722762, 722787, 722804, 722849, 722850, 722852, 722885, 722921, 722931, 722991, 723067, 723080, 723081, 723134, 723156, 723158, 723166, 723170, 723173, 723175, 723181, 723189, 723194, 723198, 723201, 723218, 723233, 723266, 723289, 723295, 723311, 723328, 723337, 723359, 723361, 723364, 723368, 723384, 723385, 723388, 723437, 723443, 723448, 723450, 723454, 723473, 723477, 723488, 723505, 723514, 723517, 723534, 723540, 723543, 723545, 723547, 723554, 723555, 723558, 723559, 723562, 723565, 723571, 723572, 723575, 723581, 723582, 723583, 723585, 723588, 723589, 723591, 723611, 723653, 723703, 723708, 723709, 723710, 723711, 723712, 723713, 723715, 723716, 723722, 723727, 723731, 723734, 723736, 723737, 723738, 723740, 723744, 723761, 723765, 723766, 723770, 723771, 723779, 723780, 723839, 723846, 723849, 723905, 723911, 723914, 723916, 723932, 723933, 723936, 723940, 723948, 723952, 723953, 723967, 723973, 723974, 723978, 723979, 723982, 723986, 723997, 724004, 724005, 724017, 724019, 724020, 724024, 724025, 724026, 724035, 724044, 724046, 724047, 724153, 724158, 724160, 724162, 724168, 724183, 724193, 724199, 724203, 724206, 724208, 724222, 724273, 724282, 724283, 724284, 724285, 724288, 724316, 724330, 724333, 724334, 724336, 724340, 724355, 724357, 724360, 724384, 724403, 724424, 724426, 724428, 724439, 724453, 724501, 724515, 724523, 724526, 724527, 724531, 724533, 724534, 724541, 724550, 724555, 724590, 724610, 724630, 724635, 724658, 724663, 724669, 724694, 724696, 724698, 724701, 724706, 724718, 724722, 724725, 724728, 724731, 724734, 724737, 724742, 724745, 724751, 724753, 724754, 724757, 724758, 724765, 724768, 724769, 724776, 724780, 724782, 724784, 724791, 724795, 724798, 724813, 724822, 724828, 724840, 724880, 724940, 724941, 724944, 724945, 724946, 724947, 724949, 724950, 724951, 724959, 724961, 724971, 724972, 724973, 724978, 724982, 724987, 725008, 725027, 725030, 725031, 725036, 725045, 725051, 725061, 725066, 725068, 725071, 725075, 725077, 725079, 725083, 725086, 725103, 725111, 725118, 725119, 725129, 725144, 725145, 725148, 725163, 725170, 725172, 725189, 725205, 725229, 725305, 725309, 725332, 725333, 725335, 725344, 725345, 725350, 725356, 725360, 725362, 725380, 725383, 725389, 725390, 725394, 725395, 725396, 725400, 725402, 725403, 725405, 725406, 725410, 725411, 725412, 725414, 725424, 725427, 725429, 725430, 725431, 725432, 725434, 725435, 725436, 725438, 725439, 725446, 725447, 725512, 725517, 725533, 725552, 725559, 725570, 725572, 725587, 725591, 725594, 725598, 725603, 725613, 725617, 725624, 725630, 725632, 725637, 725640, 725650, 725652, 725656, 725662, 725663, 725668, 725738, 725742, 725756, 725761, 725768, 725810, 725812, 725814, 725815, 725817, 725818, 725828, 725831, 725834, 725838, 725842, 725843, 725845, 725847, 725872, 725874, 725966, 725975, 725979, 725988, 725992, 725993, 725998, 725999, 726000, 726001, 726003, 726008, 726015, 726017, 726023, 726024, 726026, 726029, 726034, 726035, 726038, 726039, 726040, 726055, 726067, 726069, 726075, 726091, 726104, 726107, 726123, 726134, 726137, 726151, 726154, 726156, 726161, 726164, 726166, 726174, 726179, 726261, 726277, 726293, 726294, 726317, 726328, 726334, 726349, 726353, 726355, 726387, 726419, 726424, 704295, 711286, 711460, 717496, 718918, 719827, 726022.
GTIN 00802526559389, Serial numbers: 721134.
GTIN 00802526559396, Serial numbers: 100002, 100003, 100004, 100005, 100006, 100007, 100008, 100009, 100010, 100011, 100013, 100015, 100020, 100024, 100026, 100028, 100029, 100033, 100050, 100054, 100057, 100058, 100062, 100063, 100066, 100067, 100068, 100069, 100071, 100074, 100076, 100079, 100081, 100084, 100085, 100086, 100091, 100102, 100106, 100110, 100133, 100136, 100146, 100148, 100149, 100150, 100157, 100159, 100163, 100164, 100165, 100171, 100175, 100188, 100192, 100199, 100200, 100201, 100202, 100203, 100205, 100208, 100209, 100210, 100211, 100213, 100214, 100215, 100216, 100217, 100218, 100219, 100220, 100222, 100224, 100226, 100228, 100231, 100234, 100235, 100237, 100244, 100245, 100248, 100249, 100250, 100251, 100253, 100256, 100265, 100267, 100268, 100272, 100274, 100275, 100277, 100279, 100283, 100284, 100290, 100308, 100317, 100322, 100324, 100326, 100334, 100336, 100338, 100341, 100353, 100355, 100363, 100391, 100416, 100451, 100507, 100509, 100511, 100512, 100520, 100521, 100556, 100606, 100662, 100665, 100668, 100709, 100739, 100741, 100796, 100797, 100815, 100816, 100820, 100831, 100836, 100838, 100839, 100848, 100855, 100857, 100859, 100862, 100863, 100867, 100872, 100882, 100885, 100898, 100899, 100903, 100911, 100914, 100918, 100920, 100927, 100936, 100937, 700001, 700004, 700005, 700008, 700010, 700011, 700012, 700014, 700019, 700022, 700026, 700028, 700029, 700031, 700034, 700036, 700037, 700038, 700041, 700049, 700051, 700054, 700055, 700057, 700058, 700061, 700062, 700065, 700068, 700069, 700070, 700076, 700079, 700086, 700087, 700091, 700093, 700095, 700097, 700098, 700101, 700111, 700112, 700113, 700114, 700119, 700123, 700127, 700129, 700131, 700133, 700135, 700137, 700139, 700144, 700154, 700158, 700161, 700163, 700164, 700166, 700168, 700169, 700171, 700173, 700174, 700175, 700177, 700178, 700182, 700183, 700184, 700185, 700187, 700194, 700195, 700196, 700199, 700219, 700228, 700231, 700233, 700236, 700237, 700240, 700241, 700242, 700244, 700247, 700248, 700251, 700254, 700255, 700257, 700258, 700260, 700263, 700265, 700266, 700267, 700269, 700275, 700276, 700282, 700283, 700287, 700289, 700290, 700312, 700404, 700436, 700589, 700747, 701193, 701197, 701208, 701243, 701258, 701276, 701342, 701352, 701572, 701589, 701594, 701597, 701621, 701627, 701630, 701632, 701673, 701686, 701711, 701748, 701769, 701784, 701799, 701804, 701808, 701817, 701847, 701850, 701851, 701854, 701857, 701876, 701880, 701881, 701883, 701897, 701902, 701906, 701912, 701919, 701922, 701927, 701932, 701951, 701956, 701963, 701965, 701967, 701969, 701974, 701980, 701985, 701990, 702003, 702006, 702009, 702010, 702013, 702014, 702015, 702016, 702017, 702022, 702027, 702032, 702036, 702037, 702039, 702042, 702055, 702058, 702067, 702068, 702069, 702071, 702081, 702089, 702090, 702101, 702110, 702111, 702113, 702117, 702120, 702131, 702134, 702140, 702152, 702195, 702219, 702223, 702224, 702228, 702238, 702284, 702293, 702347, 702351, 702375, 702452, 702490, 702492, 702497, 702519, 702550, 702551, 702554, 702556, 702571, 702572, 702575, 702577, 702585, 702598, 702599, 702605, 702608, 702611, 702613, 702614, 702633, 702642, 702670, 702683, 702684, 702697, 702761, 702767, 702785, 702787, 702788, 702798, 702821, 702827, 702830, 702833, 702856, 702866, 702872, 702897, 702931, 702942, 702947, 702960, 702964, 702966, 702970, 702974, 702978, 702984, 702987, 703001, 703005, 703009, 703010, 703025, 703027, 703029, 703034, 703054, 703097, 703099, 703109, 703113, 703140, 703168, 703172, 703175, 703183, 703199, 703208, 703209, 703213, 703219, 703220, 703233, 703241, 703243, 703260, 703261, 703274, 703352, 703357, 703362, 703367, 703380, 703381, 703385, 703392, 703399, 703400, 703404, 703406, 703409, 703411, 703423, 703425, 703435, 703441, 703443, 703447, 703451, 703453, 703463, 703470, 703474, 703497, 703521, 703531, 703546, 703549, 703595, 703684, 703728, 703736, 703742, 703748, 703753, 703782, 703787, 703791, 703792, 703793, 703797, 703799, 703805, 703807, 703811, 703818, 703819, 703820, 703821, 703828, 703829, 703833, 703836, 703857, 703870, 703877, 703887, 703889, 703898, 703905, 703906, 703915, 703919, 703920, 703934, 703940, 703947, 703948, 703962, 703963, 704003, 704009, 704017, 704018, 704023, 704027, 704030, 704032, 704039, 704060, 704088, 704090, 704107, 704114, 704115, 704119, 704125, 704126, 704135, 704140, 704142, 704143, 704147, 704148, 704151, 704154, 704156, 704157, 704168, 704234, 704236, 704239, 704269, 704298, 704326, 704332, 704339, 704345, 704355, 704365, 704437, 704442, 704443, 704448, 704467, 704470, 704475, 704480, 704481, 704485, 704489, 704494, 704500, 704508, 704519, 704526, 704536, 704538, 704541, 704543, 704548, 704549, 704550, 704561, 704566, 704572, 704573, 704574, 704575, 704576, 704580, 704585, 704587, 704651, 704653, 704673, 704685, 704686, 704695, 704705, 704706, 704717, 704720, 704732, 704743, 704745, 704754, 704756, 704759, 704764, 704768, 704769, 704770, 704774, 704778, 704781, 704782, 704788, 704790, 704792, 704805, 704816, 704824, 704840, 704842, 704844, 704845, 704846, 704848, 704849, 704861, 704867, 704889, 704900, 704904, 704921, 704933, 704944, 704946, 704948, 704954, 704960, 704964, 704969, 704995, 704998, 705006, 705022, 705028, 705058, 705060, 705068, 705073, 705088, 705089, 705092, 705093, 705100, 705109, 705118, 705124, 705126, 705130, 705134, 705137, 705155, 705156, 705159, 705175, 705180, 705191, 705212, 705230, 705274, 705302, 705330, 705357, 705364, 705366, 705377, 705380, 705385, 705388, 705389, 705391, 705395, 705396, 705397, 705398, 705406, 705407, 705414, 705435, 705439, 705448, 705471, 705490, 705500, 705501, 705523, 705537, 705543, 705547, 705550, 705551, 705553, 705558, 705560, 705567, 705569, 705572, 705574, 705583, 705604, 705608, 705610, 705612, 705615, 705617, 705621, 705626, 705636, 705638, 705640, 705641, 705649, 705654, 705658, 705659, 705661, 705676, 705681, 705684, 705691, 705694, 705702, 705707, 705708, 705709, 705716, 705719, 705723, 705728, 705730, 705736, 705738, 705740, 705741, 705744, 705749, 705754, 705773, 705781, 705784, 705789, 705800, 705806, 705811, 705816, 705840, 705847, 705851, 705861, 705876, 705880, 705885, 705907, 705921, 705923, 705924, 705925, 705926, 705929, 705930, 705933, 705937, 705938, 705940, 705944, 705947, 705956, 705957, 705959, 705968, 705985, 705986, 705991, 705998, 706001, 706002, 706019, 706020, 706023, 706093, 706100, 706107, 706112, 706116, 706128, 706129, 706220, 706221, 706229, 706253, 706281, 706301, 706304, 706321, 706328, 706332, 706337, 706349, 706397, 706408, 706428, 706458, 706460, 706469, 706482, 706498, 706503, 706522, 706538, 706545, 706567, 706570, 706577, 706591, 706593, 706595, 706600, 706603, 706608, 706610, 706619, 706628, 706629, 706634, 706723, 706745, 706846, 706943, 706960, 706962, 706964, 706971, 706979, 706981, 706994, 707002, 707003, 707004, 707006, 707008, 707016, 707021, 707022, 707036, 707047, 707054, 707062, 707069, 707071, 707073, 707077, 707079, 707086, 707091, 707092, 707093, 707095, 707097, 707099, 707100, 707102, 707113, 707114, 707115, 707116, 707117, 707119, 707120, 707123, 707126, 707133, 707139, 707140, 707144, 707145, 707146, 707147, 707156, 707162, 707167, 707169, 707188, 707206, 707220, 707229, 707232, 707236, 707240, 707301, 707306, 707314, 707315, 707318, 707319, 707331, 707332, 707343, 707350, 707368, 707370, 707371, 707372, 707410, 707412, 707413, 707421, 707422, 707442, 707443, 707446, 707450, 707455, 707458, 707460, 707462, 707468, 707471, 707472, 707478, 707484, 707496, 707527, 707529, 707556, 707557, 707573, 707575, 707601, 707614, 707627, 707631, 707633, 707635, 707637, 707639, 707640, 707641, 707646, 707648, 707649, 707650, 707681, 707683, 707684, 707704, 707710, 707738, 707756, 707757, 707777, 707784, 707792, 707798, 707803, 707805, 707812, 707839, 707865, 707899, 707909, 707910, 707918, 707920, 707926, 707928, 707957, 707965, 707971, 707985, 707989, 707990, 707992, 707996, 708002, 708026, 708038, 708039, 708046, 708048, 708078, 708087, 708095, 708098, 708117, 708118, 708122, 708124, 708128, 708129, 708135, 708140, 708144, 708150, 708154, 708163, 708169, 708231, 708234, 708283, 708306, 708327, 708334, 708339, 708350, 708380, 708402, 708410, 708422, 708437, 708441, 708443, 708445, 708483, 708487, 708489, 708497, 708517, 708525, 708538, 708591, 708604, 708607, 708625, 708630, 708632, 708640, 708647, 708650, 708660, 708687, 708690, 708695, 708702, 708710, 708711, 708715, 708716, 708717, 708727, 708728, 708733, 708736, 708738, 708745, 708746, 708749, 708755, 708757, 708761, 708765, 708766, 708768, 708772, 708773, 708774, 708775, 708777, 708778, 708779, 708781, 708782, 708786, 708787, 708790, 708793, 708794, 708795, 708817, 708821, 708830, 708832, 708833, 708834, 708836, 708839, 708852, 708865, 708879, 708884, 708911, 709113, 709118, 709174, 709247, 709256, 709307, 709319, 709332, 709334, 709346, 709358, 709366, 709368, 709369, 709370, 709375, 709376, 709401, 709410, 709420, 709421, 709440, 709441, 709444, 709451, 709459, 709461, 709463, 709469, 709473, 709481, 709485, 709486, 709487, 709488, 709489, 709491, 709498, 709502, 709504, 709505, 709512, 709526, 709528, 709532, 709539, 709543, 709545, 709548, 709556, 709564, 709597, 709600, 709602, 709617, 709623, 709628, 709635, 709640, 709641, 709645, 709665, 709668, 709677, 709690, 709696, 709706, 709790, 709826, 709834, 709842, 709849, 709865, 709867, 709884, 709892, 709898, 709903, 709909, 709921, 709922, 709924, 709928, 709936, 709939, 709961, 709967, 709970, 710000, 710040, 710081, 710093, 710107, 710181, 710218, 710219, 710224, 710232, 710236, 710244, 710249, 710255, 710269, 710280, 710292, 710294, 710304, 710312, 710315, 710321, 710325, 710327, 710336, 710339, 710344, 710346, 710360, 710364, 710365, 710368, 710370, 710384, 710389, 710402, 710407, 710411, 710421, 710423, 710428, 710431, 710432, 710433, 710445, 710448, 710455, 710478, 710480, 710484, 710486, 710488, 710491, 710495, 710496, 710507, 710520, 710525, 710530, 710534, 710535, 710551, 710574, 710577, 710580, 710731, 710742, 710746, 710750, 710751, 710774, 710775, 710786, 710826, 710854, 710928, 710929, 711319, 711322, 711353, 711354, 711400, 711417, 711418, 711559, 711561, 711562, 711570, 711571, 711573, 711581, 711582, 711584, 711586, 711587, 711590, 711595, 711598, 711610, 711620, 711627, 711653, 711670, 711674, 711680, 711687, 711697, 711711, 711800, 711840, 711851, 711856, 711862, 711966, 711968, 711971, 711972, 711974, 711978, 711992, 711993, 711996, 712004, 712005, 712006, 712008, 712019, 712040, 712044, 712049, 712070, 712072, 712089, 712094, 712111, 712164, 712168, 712175, 712177, 712178, 712181, 712183, 712190, 712272, 712277, 712278, 712286, 712290, 712294, 712297, 712308, 712311, 712321, 712324, 712330, 712332, 712333, 712335, 712337, 712340, 712344, 712345, 712364, 712371, 712375, 712520, 712541, 712575, 712588, 712592, 712614, 712616, 712617, 712622, 712624, 712630, 712634, 712636, 712650, 712666, 712671, 712678, 712683, 712696, 712854, 712859, 712867, 712876, 712897, 712900, 713004, 713031, 713038, 713055, 713108, 713117, 713126, 713143, 713146, 713147, 713156, 713163, 713164, 713165, 713168, 713172, 713174, 713177, 713180, 713181, 713200, 713207, 713210, 713211, 713320, 713363, 713378, 713400, 713583, 713593, 713598, 713612, 713629, 713645, 713667, 713716, 713781, 713798, 713802, 713810, 713817, 713837, 713843, 713848, 713913, 713914, 713942, 713956, 713957, 713961, 713969, 713975, 713976, 713993, 713999, 714015, 714020, 714027, 714054, 714057, 714536, 714594, 714621, 714672, 714719, 714720, 714728, 714748, 714761, 714765, 714767, 714769, 714776, 714778, 714784, 714789, 714791, 714813, 714817, 714824, 714826, 714842, 714891, 714895, 714896, 714990, 715008, 715032, 715115, 715121, 715127, 715132, 715138, 715151, 715153, 715158, 715161, 715162, 715168, 715261, 715264, 715344, 715347, 715357, 715369, 715437, 715495, 715497, 715525, 715657, 715700, 715705, 715712, 715718, 715735, 715749, 715751, 715756, 715763, 715764, 715766, 715767, 715788, 715790, 715791, 715823, 715832, 715848, 715857, 715873, 715883, 715886, 715898, 715906, 715914, 715920, 715921, 715927, 715928, 716046, 716050, 716052, 716058, 716059, 716060, 716061, 716062, 716064, 716066, 716067, 716068, 716069, 716070, 716071, 716073, 716075, 716076, 716079, 716080, 716081, 716082, 716084, 716087, 716089, 716090, 716094, 716124, 716127, 716129, 716130, 716136, 716138, 716623, 716629, 716630, 716642, 716649, 716660, 716662, 716664, 716669, 716688, 716696, 716697, 716698, 716727, 716772, 716781, 716788, 716790, 716794, 716808, 716830, 716831, 716832, 716833, 716834, 716836, 716838, 716915, 716932, 716937, 716960, 716963, 716981, 716983, 716984, 716988, 716990, 716999, 717002, 717003, 717008, 717025, 717026, 717027, 717028, 717034, 717035, 717039, 717049, 717050, 717052, 717054, 717058, 717063, 717068, 717087, 717295, 717400, 717412, 717415, 717416, 717417, 717419, 717425, 717484, 717509, 717523, 717618, 717621, 717647, 717654, 717667, 717675, 717690, 717823, 717825, 717831, 717832, 717833, 717837, 717838, 717843, 717844, 717848, 717849, 717850, 717851, 717853, 717866, 717868, 717889, 717890, 717892, 717898, 717902, 717903, 717906, 717907, 717908, 717912, 717913, 717916, 717918, 717920, 717923, 717924, 717925, 717928, 717982, 717984, 717987, 717989, 717991, 717992, 717996, 717999, 718000, 718003, 718005, 718007, 718010, 718015, 718018, 718021, 718022, 718026, 718027, 718059, 718270, 718271, 718272, 718274, 718275, 718292, 718294, 718320, 718357, 718359, 718362, 718363, 718365, 718381, 718383, 718386, 718394, 718395, 718653, 718655, 718656, 718658, 718661, 718680, 718693, 718712, 718726, 718744, 718760, 718766, 718772, 718786, 718794, 718817, 718818, 718819, 718829, 718830, 718834, 718838, 718839, 718840, 718852, 718858, 718862, 718866, 718867, 718868, 718880, 718882, 718883, 718887, 718888, 719307, 719542, 719584, 719585, 719586, 719588, 719592, 719593, 719596, 719604, 719605, 719608, 719620, 719623, 719636, 719651, 719653, 719712, 719722, 719727, 719745, 719748, 719750, 719754, 719776, 719780, 719932, 719935, 719938, 719940, 719975, 720023, 720092, 720098, 720125, 720134, 720135, 720160, 720230, 720251, 720288, 720327, 720338, 720340, 720341, 720353, 720354, 720356, 720360, 720361, 720363, 720365, 720376, 720387, 720400, 720412, 720414, 720415, 720419, 720422, 720423, 720435, 720540, 720602, 720607, 720636, 720637, 720639, 720721, 720723, 720724, 720726, 720732, 720734, 720740, 720742, 720743, 720746, 720750, 720753, 720755, 720756, 720769, 720781, 720893, 720912, 720914, 720915, 720916, 720923, 720924, 720925, 720927, 720934, 721101, 721111, 721117, 721118, 721129, 721130, 721136, 721138, 721141, 721192, 721200, 721211, 721273, 721274, 721276, 721278, 721295, 721317, 721318, 721319, 721323, 721326, 721339, 721359, 721370, 721372, 721377, 721395, 721400, 721460, 721465, 721555, 721556, 721560, 721564, 721566, 721572, 721573, 721574, 721576, 721577, 721578, 721690, 721751, 721775, 721801, 721806, 721808, 721809, 721811, 721817, 721833, 721837, 721843, 721846, 721850, 721921, 721923, 721934, 721972, 721975, 721977, 721979, 721980, 721982, 721983, 721987, 721997, 722019, 722047, 722049, 722050, 722178, 722211, 722223, 722241, 722248, 722251, 722283, 722290, 722291, 722298, 722300, 722302, 722305, 722306, 722307, 722308, 722309, 722324, 722328, 722332, 722337, 722342, 722453, 722454, 722456, 722459, 722467, 722484, 722485, 722489, 722565, 722571, 722580, 722591, 722616, 722687, 722708, 722716, 722782, 722784, 722785, 722820, 722832, 722833, 722834, 722841, 722843, 722845, 722853, 722854, 722861, 722863, 722870, 722871, 722886, 722890, 722891, 722892, 722897, 722900, 722906, 722908, 722912, 722918, 722919, 722920, 722928, 722934, 722999, 723001, 723002, 723003, 723004, 723005, 723007, 723008, 723009, 723012, 723015, 723148, 723153, 723366, 723367, 723369, 723371, 723375, 723377, 723378, 723379, 723390, 723392, 723400, 723405, 723406, 723408, 723409, 723414, 723419, 723424, 723426, 723429, 723430, 723451, 723469, 723598, 723614, 723621, 723633, 723695, 723718, 723929, 723972, 723976, 723977, 723980, 723984, 723985, 723987, 723988, 723991, 723992, 723994, 723995, 723998, 723999, 724001, 724003, 724006, 724008, 724010, 724012, 724031, 724036, 724038, 724040, 724041, 724048, 724051, 724052, 724053, 724054, 724061, 724245, 724247, 724249, 724250, 724272, 724291, 724342, 724366, 724411, 724474, 724481, 724494, 724495, 724512, 724525, 724560, 724565, 724568, 724575, 724588, 724597, 724599, 724604, 724879, 724899, 724902, 724903, 724920, 724960, 724964, 724965, 724968, 724969, 724977, 724985, 725012, 725050, 725058, 725067, 725069, 725080, 725084, 725121, 725142, 725156, 725159, 725162, 725168, 725193, 725195, 725209, 725218, 725223, 725227, 725244, 725247, 725249, 725261, 725265, 725312, 725327, 725330, 725342, 725346, 725348, 725361, 725578, 725580, 725584, 725586, 725596, 725614, 725623, 725626, 725636, 725657, 725664, 725672, 725718, 725719, 725720, 725721, 725736, 725746, 725751, 725753, 725869, 725875, 725953, 725962, 725963, 725969, 725972, 725973, 726004, 726009, 726014, 726036, 726054, 726063, 726065, 726074, 726096, 726103, 726116, 726117, 726153, 726191, 726192, 726199, 726214, 726216, 726221, 726222, 726223, 726226, 726227, 726231, 726235, 726237, 726242, 726244, 726247, 726248, 726250, 726255, 726256, 726297, 726301, 726310, 726332, 726339.
GTIN 00802526577109, Serial numbers: 100417, 100426, 100450, 100621, 700009, 700020, 700053, 700090, 700107, 700136, 700147, 700162, 700167, 700176, 700192, 700204, 700223, 700224, 700229, 700230, 700238, 700259, 700402, 700700, 700771, 700786, 700806, 700838, 701388, 701394, 701442, 701933, 701943, 702266, 702270, 702681, 702879, 702993, 703008, 703165, 703171, 703184, 703355, 703492, 703506, 703683, 703806, 703813, 703853, 703893, 704043, 704048, 704302, 704439, 704510, 704512, 704934, 704955, 705054, 705122, 705128, 705173, 705174, 705187, 705241, 705313, 705334, 705345, 705399, 705415, 705538, 705629, 705867, 706010, 706026, 706121, 706125, 706262, 706263, 706323, 706333, 706353, 706420, 706475, 706514, 706539, 706565, 706596, 706609, 706616, 706702, 706706, 706717, 706722, 706724, 706725, 706726, 706738, 707171, 707253, 707254, 707272, 707275, 707284, 707317, 707447, 707508, 707518, 707536, 707636, 707677, 707682, 707686, 707688, 707701, 707742, 707761, 707771, 707775, 707801, 707862, 707891, 707931, 707950, 707960, 707967, 708040, 708145, 708444, 708447, 708568, 708646, 708649, 708653, 708688, 708707, 708729, 708748, 708760, 708800, 709112, 709155, 709156, 709162, 709185, 709189, 709240, 709259, 709267, 709271, 709273, 709305, 709321, 709326, 709382, 709392, 709412, 709464, 709779, 709799, 709839, 709845, 709858, 709864, 709871, 709874, 709896, 709904, 709906, 709913, 709931, 709937, 710029, 710059, 710186, 710307, 710313, 710316, 710320, 710328, 710330, 710332, 710351, 710352, 710363, 710372, 710373, 710376, 710377, 710403, 710536, 710570, 710881, 710916, 710943, 710948, 710995, 710999, 711009, 711127, 711188, 711236, 711247, 711287, 711293, 711301, 711313, 711317, 711323, 711324, 711325, 711328, 711331, 711335, 711357, 711360, 711361, 711394, 711451, 711592, 711794, 711866, 711872, 711960, 712001, 712124, 712343, 712448, 712514, 712527, 712548, 712551, 712560, 712596, 712673, 712690, 712695, 712707, 712708, 712714, 712717, 712746, 712760, 712847, 712865, 712884, 712917, 712928, 712951, 712974, 712982, 712998, 713044, 713048, 713056, 713057, 713107, 713190, 713197, 713204, 713213, 713248, 713285, 713608, 713694, 713712, 713714, 713726, 713734, 713745, 713758, 713777, 713778, 713797, 713811, 713821, 713909, 713910, 714164, 714302, 714304, 714317, 714321, 714338, 714345, 714346, 714347, 714349, 714383, 714385, 714521, 714523, 714543, 714595, 714620, 714705, 714819, 714957, 714993, 714995, 714997, 715049, 715227, 715242, 715265, 715268, 715272, 715280, 715281, 715288, 715289, 715291, 715295, 715334, 715374, 715385, 715393, 715490, 715552, 715564, 715568, 715716, 715844, 716913, 716930, 716938, 717168, 717186, 717231, 717233, 717236, 717242, 717269, 717273, 717275, 717283, 717288, 717289, 717300, 717302, 717304, 717306, 717308, 717309, 717310, 717313, 717316, 717334, 717340, 717347, 717374, 717378, 717481, 717491, 717600, 717604, 717608, 717629, 718017, 718036, 718038, 718047, 718050, 718055, 718060, 718065, 718067, 718068, 718277, 718293, 718321, 718329, 718332, 718333, 718350, 718356, 718384, 718387, 719507, 719565, 719657, 719696, 719702, 719731, 719749, 719861, 719866, 719994, 719995, 720158, 720273, 720520, 720548, 720621, 720624, 720626, 720628, 720629, 720631, 720632, 720641, 720642, 720646, 720647, 720720, 720885, 720902, 720909, 720911, 720918, 720919, 720922, 721144, 721190, 721280, 721292, 721308, 721322, 721324, 721327, 721328, 721337, 721474, 721604, 721622, 721624, 721629, 721645, 721646, 721649, 721851, 721874, 721911, 721913, 721917, 721920, 721950, 721957, 721969, 722106, 722294, 722301, 722315, 722316, 722317, 722323, 722327, 722333, 722429, 722432, 722438, 722455, 722457, 722493, 722640, 722646, 722672, 722681, 722682, 722683, 722692, 722698, 722723, 722732, 722733, 722734, 722735, 722737, 722740, 723192, 723376, 723395, 723401, 723404, 723413, 723442, 723459, 723480, 723863, 723890, 723975, 724058, 724187, 724244, 724410, 724491, 724554, 724561, 724629, 724671, 724726, 724756, 724759, 724761, 724763, 724764, 724767, 724785, 724799, 724801, 724806, 724810, 725016, 725096, 725128, 725181, 725251, 725262, 725322, 725351, 725354, 725408, 725409, 725426, 725441, 725450, 725472, 725488, 725522, 725625, 725628, 725641, 726032, 726160, 726196, 726200, 726240, 726249, 726253, 726265, 726307, 726380, 726408, 726418.
GTIN 00802526578793, Serial Numbers: 700074, 700172, 700179, 700190, 700193, 700208, 700253, 700300, 701639, 701995, 701997, 702049, 702178, 702191, 702201, 702208, 702225, 702276, 702277, 702280, 702281, 702299, 702300, 702305, 702308, 702310, 702311, 702333, 702335, 702336, 702338, 702339, 702345, 702362, 702363, 702385, 702394, 702420, 702422, 702426, 702440, 702447, 702451, 702460, 702479, 702480, 702487, 702491, 702493, 702496, 702499, 702500, 702503, 702504, 702532, 702533, 702535, 702538, 702563, 702583, 702589, 702625, 702647, 702675, 702677, 702739, 702750, 702766, 702801, 702802, 702954, 702999, 703022, 703030, 703231, 703239, 703384, 703468, 703533, 703680, 703706, 703718, 703798, 703800, 703801, 703802, 703803, 703824, 703832, 703834, 703837, 703839, 703843, 703856, 703859, 704104, 704746, 704772, 704785, 704806, 704821, 704909, 704990, 705268, 705443, 705494, 705540, 705552, 705559, 705563, 705564, 705571, 705573, 705645, 705650, 705653, 705655, 705696, 705699, 705721, 705769, 705793, 705825, 705862, 705873, 705878, 706006, 706021, 706028, 706029, 706259, 706286, 706291, 706293, 706294, 706295, 706299, 706324, 706327, 706331, 706335, 706339, 706352, 706361, 706373, 706376, 706749, 706801, 706868, 706922, 706928, 706931, 706938, 706950, 706961, 706974, 706978, 706985, 706986, 706992, 707011, 707018, 707027, 707101, 707181, 707182, 707258, 707279, 707288, 707345, 707565, 707582, 707603, 707612, 707672, 707697, 707723, 707788, 707794, 707809, 707820, 707853, 707859, 707893, 707921, 707956, 708142, 708167, 708170, 708175, 708177, 708186, 708199, 708213, 708229, 708235, 708242, 708275, 708278, 708284, 708286, 708292, 708294, 708301, 708307, 708308, 708354, 708382, 708408, 708908, 708918, 708921, 708924, 708937, 708939, 708942, 708945, 708946, 708949, 708955, 708957, 708959, 708960, 708969, 708971, 708972, 708974, 708990, 708994, 708999, 709002, 709005, 709024, 709030, 709032, 709292, 709438, 709452, 709490, 709514, 709520, 709557, 709569, 709574, 709582, 709584, 709594, 709604, 709616, 709631, 709651, 709660, 709670, 709686, 709692, 709699, 709709, 709714, 709770, 709832, 709833, 709841, 709863, 709888, 709900, 709905, 709911, 709923, 709930, 709933, 709940, 709947, 709949, 709951, 709953, 709971, 709973, 709975, 709988, 710019, 710032, 710067, 710109, 710112, 710114, 710116, 710123, 710152, 710301, 710477, 710500, 710501, 710518, 710522, 710532, 710542, 710569, 710581, 710582, 710584, 710596, 710607, 710621, 710623, 710669, 710706, 710890, 710949, 710951, 710952, 710957, 711094, 711138, 711139, 711140, 711146, 711160, 711164, 711179, 711190, 711206, 711239, 711246, 711281, 711305, 711307, 711314, 711347, 711371, 711373, 711374, 711383, 711401, 711445, 711448, 711455, 711464, 711480, 711482, 711483, 711489, 711492, 711514, 711522, 711523, 711525, 711527, 711532, 711534, 711551, 711556, 711558, 711565, 711566, 711572, 711585, 711602, 711606, 711615, 711622, 711625, 711628, 711691, 711729, 711730, 711738, 711755, 711994, 712476, 712522, 712530, 712584, 712585, 712589, 712593, 712600, 712604, 712606, 712607, 712608, 712621, 712629, 712658, 712663, 712672, 712679, 712692, 712700, 712701, 712703, 712705, 713113, 713139, 713145, 713148, 713152, 713154, 713175, 713187, 713189, 713192, 713194, 713195, 713198, 713202, 713208, 713217, 713219, 713227, 713234, 713235, 713239, 713241, 713245, 713260, 713271, 713280, 713288, 713321, 713325, 713330, 713336, 713388, 713596, 713621, 713635, 713660, 713684, 713722, 713724, 713725, 713741, 713750, 713756, 713759, 713784, 713795, 713815, 713899, 713927, 713928, 714308, 714344, 714359, 714387, 714400, 714408, 714442, 714447, 714465, 714503, 714506, 714507, 714520, 714540, 714550, 714568, 714569, 714570, 714574, 714577, 714584, 714591, 714592, 714596, 714600, 714606, 714619, 714627, 714660, 714676, 714677, 714685, 714688, 714700, 714701, 714704, 714722, 714744, 714745, 714771, 714772, 714773, 714779, 714782, 714783, 714786, 714790, 714792, 714793, 714796, 714797, 714799, 714802, 714803, 714808, 714809, 714811, 714814, 714816, 714818, 714822, 714827, 714828, 714840, 714841, 714845, 714846, 714847, 714890, 716801, 717047, 717082, 717109, 717149, 717192, 717195, 717197, 717206, 717214, 717224, 717229, 717230, 717234, 717237, 717240, 717243, 717246, 717248, 717250, 717253, 717254, 717255, 717256, 717258, 717270, 717278, 717315, 717325, 717331, 717339, 717341, 717343, 717346, 717377, 717471, 717482, 717486, 717489, 717490, 717492, 717494, 717497, 717537, 717695, 717728, 717744, 717820, 717821, 717830, 717834, 717839, 717857, 717911, 718133, 718192, 718300, 718319, 718353, 718366, 718367, 718368, 718372, 718376, 718380, 718409, 718411, 718414, 718423, 718424, 718432, 718446, 718485, 718486, 718494, 718502, 718514, 718524, 718528, 718539, 718549, 718559, 718560, 718567, 718638, 718650, 718651, 718654, 718657, 718739, 718746, 718748, 719645, 719688, 719689, 719690, 719691, 719694, 720131, 720146, 720147, 720265, 720267, 720297, 720312, 720319, 720357, 720393, 720404, 720405, 720408, 720409, 720413, 720416, 720428, 720429, 720447, 720481, 720534, 720537, 720543, 720545, 720546, 720547, 720549, 720550, 720553, 720555, 720559, 720560, 720577, 721005, 721034, 721043, 721044, 721049, 721050, 721057, 721060, 721135, 721145, 721149, 721150, 721151, 721158, 721182, 721183, 721186, 721189, 721194, 721196, 721212, 721214, 721216, 721220, 721238, 721241, 721243, 721244, 721246, 721247, 721248, 721249, 721250, 721253, 721255, 721265, 721270, 721753, 721774, 721805, 721847, 721852, 721853, 721854, 721855, 721857, 721858, 721859, 721862, 721863, 721864, 721866, 721867, 721868, 721869, 721870, 721871, 721872, 721873, 721875, 721876, 721877, 721878, 721879, 721880, 721891, 721892, 721893, 721894, 721895, 721896, 721897, 721898, 721899, 721900, 722025, 722071, 722073, 722083, 722289, 722338, 722531, 722551, 722555, 722558, 722560, 722578, 722651, 722704, 722706, 722727, 722770, 722783, 722788, 722799, 722816, 722819, 722821, 722830, 722835, 722836, 722837, 722846, 722847, 722851, 722856, 722857, 722866, 722868, 722873, 722876, 722879, 722881, 722894, 722895, 722896, 722898, 722904, 722909, 722914, 722916, 722925, 722926, 722930, 722937, 722950, 722968, 722971, 722973, 722976, 722982, 722987, 723019, 723022, 723047, 723053, 723074, 723075, 723078, 723079, 723087, 723152, 723203, 723223, 723229, 723235, 723241, 723262, 723268, 723271, 723276, 723282, 723292, 723298, 723303, 723307, 723312, 723316, 723318, 723320, 723326, 723327, 723346, 723353, 723461, 724185, 724271, 724279, 724280, 724281, 724286, 724295, 724301, 724320, 724321, 724329, 724331, 724332, 724347, 724363, 724367, 724377, 724390, 724394, 724395, 724420, 724427, 724437, 724440, 724452, 724455, 724468, 724569, 724570, 724582, 724583, 724693, 724702, 724704, 724711, 724715, 724716, 724723, 724724, 724735, 724760, 724762, 724770, 724771, 724774, 724775, 724779, 724781, 724783, 724788, 724802, 724804, 724805, 724808, 724809, 724811, 724833, 724835, 724841, 724855, 724858, 724861, 724862, 724864, 724871, 725104, 725155, 725185, 725194, 725220, 725221, 725228, 725233, 725239, 725242, 725254, 725260, 725263, 725267, 725562, 725582, 725585, 725590, 725602, 725609, 725618, 725627, 725635, 725646, 725667, 725671, 725765, 725911, 726110, 726119, 726190, 726193, 726197, 726207, 726217, 726219, 726220, 726234, 726238, 726258, 726298, 726439, 726469, 726503. |
Recalling Firm/ Manufacturer |
Boston Scientific Corporation 4100 Hamline Ave N Saint Paul MN 55112-5700
|
| For Additional Information Contact | United States Technical Services 800-227-3422 |
Manufacturer Reason for Recall | Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps). |
FDA Determined Cause 2 | Other |
| Action | Boston Scientific issued an "Urgent Medical Device Correction" notice to its customers on 08/18/2025 via letter. The notice provides an update to the December 2024 customer communication regarding the potential for the ACCOLADE family of pacemakers to initiate Safety Mode due to a high battery impedance state and contains recommendations timely upgrade of pacemaker software to prevent the occurrence of Safety Mode in an ambulatory setting. On 09/05/2025 an update to the 08/18/2025 notice (via letter), superseding the prior notice to address two unintended behaviors associated with the Model 3869 v2.04 software (designed to prevent potential initiation of Safety Mode due to a high battery impedance condition). The behaviors are:
1. Incomplete disablement of wandless automatic telemetry after detecting a high battery impedance condition. If an ACCOLADE family device with an elevated battery impedance induced Code-1003 remains in service, the software s new telemetry activated battery test is designed to continue assessing battery impedance and disable all wandless telemetry operation before high impedance is achieved and device resets can initiate Safety Mode.
2. Potential for the software s new daily battery test to misinterpret measurements in the presence of a magnet, initiating a false positive response that results in disablement of wandless telemetry sessions. For patients monitored via LATITUDE, this will trigger an alert message indicating an erroneous explant indicator (e.g., explant indicator was reached January 1, 2000) if the device had not previously reached replacement indication, stating that LATITUDE is no longer available, and prompting to contact Technical Services. Patients were notified regarding battery performance. Advisory is published at www.BostonScientific.com/advisory.
If you have any questions, contact US Technical Services at 1.800.CARDIAC (227.3422) or email to: tech.services@bsci.com. |
| Quantity in Commerce | 3620 units |
| Distribution | Worldwide |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| PMA Database | PMAs with Product Code = NKE
|
|
|
|