| Date Initiated by Firm | September 30, 2025 |
| Date Posted | November 14, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0528-2026 |
| Recall Event ID |
97846 |
| Product Classification |
Obstetrical kit - Product Code OKV
|
| Product | MEDLINE convenience kits labeled as: GP-LEEP PACK-LF, REF PHS396969004F |
| Code Information |
UDI/DI 10195327454227 (each) 40195327454228 (case), Lot Numbers: 23HDA742
|
Recalling Firm/ Manufacturer |
Medline Industries, LP 3 Lakes Dr Northfield IL 60093-2753
|
| For Additional Information Contact | Haley Barclay 800-369-1704 |
Manufacturer Reason for Recall | Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | Medline issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 09/30/2025 and 10/6/2025 via email and mail. The notice explained the issue with the component, potential risk, and requested the following actions:
1. Immediately check stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product.
2. Please use the link and the information found in the notice to complete the response form. Please list the quantity of affected product in inventory on the form. Even if no affected product is in inventory, please complete and submit the form.
3. Upon receipt of the submitted response form, the customer account will receive over-labels to place on affected inventory, with instructions for the customer staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable.
4. For distributors or those who have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product.
If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com. |
| Quantity in Commerce | 18 units |
| Distribution | US Nationwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|