| | Class 2 Device Recall IntelliVue Patient Monitor MX450 |  |
| Date Initiated by Firm | October 31, 2025 |
| Date Posted | December 02, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0863-2026 |
| Recall Event ID |
97917 |
| 510(K)Number | K130849 |
| Product Classification |
Detector and alarm, arrhythmia - Product Code DSI
|
| Product | IntelliVue Patient Monitor MX450. Product Number: 866062. |
| Code Information |
Product Number: 866062; UDI-DI: 00884838038769; All Serial No. |
| FEI Number |
3016618143
|
Recalling Firm/ Manufacturer |
Philips North America Llc
222 Jacobs St
Cambridge
MA
02141-2289
|
| For Additional Information Contact | Philips Customer Service 800-722-9377 |
Manufacturer Reason for Recall | Potential issue where the IntelliVue monitors did not alarm. |
FDA Determined Cause 2 | Use error |
| Action | An URGENT Medical Device Correction notice, dated 10/31/25, was mailed to customers and distributors.
Distributors were notified of this correction and instructed to forward the provided recall notification to all consignees. Additionally, distributors are asked to include their contact information for consignees in the provided response form and perform a good faith effort for each customer to obtain a completed response form.
Consignees are instructed to notify all users of this recall notification and forward the notice to all who need to be aware within their organization or where affected devices were potentially transferred. Consignees are to re-evaluate and reconfigure device configuration per the provided instructions. Once recall actions have been completed, consignees are to complete the provided response form either via QR code, by email to Recall.Response@philips.com, or by fax to 877-499-7223.
Philips is planning to develop a software update which will enforce the additional confirmation step when the user enables infinite Alarms Off for IntelliVue MP5 models only.
Consignees with any questions can contact Philips at 800-722-9377. |
| Quantity in Commerce | 1,913,441 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Angola, Argentina, Armenia, Aruba, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia, Botswana, Brazil, Brunei, Bulgaria, Burundi, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Cote D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, El Salvador, Equat Guinea, Estonia, Ethiopia, Faroe Islands, Finland, French Polynesia, France, French Guiana, Gabon, Georgia, Germany, Ghana, Gibraltar, Greece, Greenland, Guadeloupe, Guam, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea Rep, Kuwait, Lao, Latvia, Lebanon, Lesotho, Liberia, Libya, Lithuania, Luxembourg, Macao, Macedonia, Madagascar, Malaysia, Maldives, Mali, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Monaco, Mongolia, Montenegro, Morocco, Mozambique, Myanmar, Namibia, Nepal, Netherlands, Netherlands Ant, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, , R¿union, Romania, Russia, Saint Vincent, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, St. Pierre Mqln, Suriname, Sweden, Switzerland, Syrian Arab Rep, Taiwan, Tajikistan, Thailand, Trinidad Tobago, Tunisia, T¿rkiye, Turkmenistan, Uganda, Ukraine, United Kingdom, United States, Uruguay, Utd Arab Emir, Uzbekistan, Venezuela, Viet Nam, Virgin Isl (Br), Virgin Isl (US), White Russia, Yemen, Zambia, Zimbabwe. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = DSI
|
|
|
|