Date Initiated by FirmJune 10, 2004
Date PostedJuly 27, 2004
Recall Status1 Terminated 3 on November 01, 2004
Recall NumberZ-1210-04
Recall Event ID 29300
PMA NumberP020018 
Product Classification System, Endovascular Graft, Aortic Aneurysm Treatment - Product Code MIH
ProductCook brand Zenith AAA Iliac leg graft with the H&L-B One-Shot Introducer System, preloaded iliac leg graft - distal cuff diameter 24 mm - usable length 71mm - delivery system 16 FR; Catalog no. TFLE-24-71.
Code Information Lots F1573048, F1571074, F1561902, F1569266, F1565685, F1569268, F1565686, F1565687, F1565689, F1573057, F1573058, F1573059, F1573060, F1573061, F1573049, F1573050, F1565688, F1571072, F1571073, F1571075, F1561419, F1573052, F1575274, F1575275, F1575365, F1575366, F1575367 and F1575368. 
Recalling Firm/
Manufacturer
Cook, Inc.
750 Daniels Way
Bloomington IN 47404-9120
For Additional Information ContactCustomer Relations
800-457-4500
Manufacturer Reason
for Recall
Introducer tubing may contain a vein of unextruded material and carbon black or a strand of tubing material may be displaced from the introducer during use.
FDA Determined
Cause 2
Other
ActionThe firm issued recall letters to customers on 6/10/04. The recall was extended and URGENT EXTENSION OF PRODUCT RECALL notices were sent out on June 23, 2004.
Quantity in Commerce28
DistributionNationwide, Canada, Denmark and Japan.
Total Product Life CycleTPLC Device Report