| Date Initiated by Firm | June 22, 2006 |
| Date Posted | August 05, 2006 |
| Recall Status1 |
Terminated 3 on April 07, 2008 |
| Recall Number | Z-1364-06 |
| Recall Event ID |
35022 |
| 510(K)Number | K973662 |
| Product Classification |
Blood Glucose Meter - Product Code LFR
|
| Product | ACCU-CHEK Complete blood glucose monitor, Ref. no. 516; Catalog nos. 03266770001 and 03871983001 with meter model no. 200 and Catalog no. 03871983001 with meter model number 250. |
| Code Information |
All units. |
| FEI Number |
1823260
|
Recalling Firm/ Manufacturer |
Roche Diagnostics Corp. 9115 Hague Rd Indianapolis IN 46250-0416
|
| For Additional Information Contact | 800-440-3638 |
Manufacturer Reason for Recall | The meter gives an error message that can "actually" mean either a problem with the strip or a blood glucose too low to measure, but the meter error message only reports that there is a bad strip. |
FDA Determined Cause 2 | Other |
| Action | A recall notification letter dated April 24, 2006 was sent to each customer alerting them to the fact that this error code has a dual meaning. New software will be installed in all meters when it becomes available. |
| Quantity in Commerce | 55,189 meters (for all Recalls Z-1354-1371-06) |
| Distribution | Worldwide, USA, Germany, Guam, North Mariana Islands, Panama and Virgin Islands. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = LFR
|