| Date Initiated by Firm | October 12, 2006 |
| Date Posted | October 27, 2006 |
| Recall Status1 |
Terminated 3 on August 30, 2012 |
| Recall Number | Z-0095-2007 |
| Recall Event ID |
36515 |
| 510(K)Number | K943327 |
| Product Classification |
Laparoscopic Instrument - Product Code GCJ
|
| Product | ConMed Laparoscopic Instrument: Babcock Disposable Grasper, 5mm x 33cm length, REF/Product Code 60-6047-001. --- The product is distributed sterile in heat-sealed tray and labeled with an expiration date that pertains only to the sterility. |
| Code Information |
REF/Product Code 60-6047-001, all lot codes manufactured between 08/25/2003 and 03/24/2005. [Lot codes on unopened boxes and packaging contain a lot code in the following form: for example 0409271 - 04 represents the year 2004, 09 represents the month of the year, 27 represents the day of the month, and the last digit 1 represents the manufacturing shift code. Lot codes are located on the device in this form: for example 40927 - 4 represents the year 2004, 09 represents the month of the year, and 27 represents the day of the month.] |
Recalling Firm/ Manufacturer |
Conmed Corporation 525 French Road Utica NY 13502-5945
|
| For Additional Information Contact | Ms. Patricia Cotter 315-624-3533 |
Manufacturer Reason for Recall | The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | On 10/12/06, ConMed sent "URGENT DEVICE RECALL" packages (letter dated 10/11/06) to the domestic accounts via UPS Ground, and via UPS Express for the International accounts. Devices are to be removed from use and returned. |
| Quantity in Commerce | 287 units |
| Distribution | Worldwide |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|