| Class 2 Device Recall Clinac |  |
Date Initiated by Firm | August 31, 2006 |
Date Posted | November 16, 2006 |
Recall Status1 |
Terminated 3 on December 04, 2007 |
Recall Number | Z-0194-2007 |
Recall Event ID |
36555 |
510(K)Number | K904264 K913119 |
Product Classification |
linear accelerator - Product Code IYE
|
Product | Clinac High Energy Medical Linear Accelerator, Medical Charged-particle radiation therapy system; Model #s: 2100(C,C/D EX) and 2300 C, C/D EX) |
Code Information |
Serial numbers H272325, H291060, H271879, H273040, H272187, H272434 |
Recalling Firm/ Manufacturer |
Varian Medical Systems Inc 911 Hansen Way Palo Alto CA 94304
|
For Additional Information Contact | Dale Reynolds 650-424-6640 |
Manufacturer Reason for Recall | The chain holding the device gantry in position may break. |
FDA Determined Cause 2 | Process control |
Action | On 8/31/06, the firm issued letters via Federal Express to all its direct consignees, informing them of the affected product and providing instructions on the recall. |
Quantity in Commerce | 254 units |
Distribution | Worldwide, including USA, England, Taiwan, Switzerland, Sweden, South Korea, Portugal, Poland, Norway, New Zealand, Netherlands, Malaysia, Japan, Italy, Germany, France, Denmark, Czech Republic, Colombia, China, Canada, Belgium, Australia and Venezuela. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = IYE 510(K)s with Product Code = IYE
|
|
|
|