Class 2 Device Recall CXP Analysis Software |
 |
Date Initiated by Firm |
August 29, 2006 |
Date Posted |
February 08, 2007 |
Recall Status1 |
Terminated 3 on April 19, 2012 |
Recall Number |
Z-0420-2007 |
Recall Event ID |
37083 |
510(K)Number |
K030828
|
Product Classification |
cell analyzer software - Product Code GKZ
|
Product |
CXP Analysis Software, Versions 2.0 & 2.1: Part #722396 CXP Analysis Software Kit V2.1, Single User; Part #722395 CXP Analysis Software Kit V2.1, Single User Upgrade; Part #722397 CXP Analysis Software Kit V2.1, 3 User Pack; Part #722398 CXP Analysis Software Kit V2.1, 5 User Pack; Part #722399 CXP Analysis Software Kit V2. 1, 15 User Pack; Part # 722593 CXP Analysis Software Kit V2.1, 5 Network User Pack Part #722594 CXP Analysis Software Kit V2.1, 10 Network User Pack; Part #722595 CXP Analysis Software Kit V2.1, 15 Network User Pack; Part #629640 CXP Analysis Software Kit V2.0, Single User; Part #629639 CXP Analysis Software Kit V2.0, Single User Upgrade; Part #629641 CXP Analysis Software Kit V2.0, 3 User Pack; Part #629642 CXP Analysis Software Kit V2.0, 5 User Pack; Part #629643 CXP Analysis Software Kit V2.0, 15 User Pack; Part #629644 CXP Analysis Software Kit V2.0, 5 Network User Pack; Part #629645 CXP Analysis Software Kit V2.0, 10 Network User Pack; Part #629646 CXP Analysis Software Kit V2.0, 15 Network User Pack. The Cytomics FC 500 is a system for the qualitative and quantitative measurement of biological and physical properties of cells and other particles. |
Code Information |
Part Numbers: 722396, 722395, 722397, 722398, 722399, 722593, 722594, 722595, 629640, 629639, 629641, 629642, 629643, 629644, 629645, 629646. Software, Versions 2.0 & 2.1 |
Recalling Firm/ Manufacturer |
Beckman Coulter Inc 200 S Kraemer Blvd Brea CA 92822-6208
|
For Additional Information Contact |
Nora Zerounian 714-961-3634
|
Manufacturer Reason for Recall |
Two workflow scenarios associated with renaming regions may produce incorrect results.
|
FDA Determined Cause 2 |
Other |
Action |
A Product Corrective Action (PCA) letter was sent on the week of August 29, 2006 to all accounts that have an FC500 or have ordered CXP or MXP software to inform them that a) incorrect statistics are generated when a polygonal region with vertical re-entrant vertices is renamed using the Region Properties dialog box. b)when a region is copied from one plot to another, and then the new region is renamed, the statistics are not updated with the new region name. (Letter was sent by US mail). Letter includes immediate user preventive action to remedy both issues. A response form is included for consignees to return to recalling firm. |
Quantity in Commerce |
200 units nationwide |
Distribution |
Nationwide & Canada |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = GKZ and Original Applicant = BECKMAN COULTER, INC.
|
|
|
|