U.S. Department of Health and Human Services

Class 2 Device Recall CXP Analysis Software

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall CXP Analysis Software see related information
Date Initiated by Firm August 29, 2006
Date Posted February 08, 2007
Recall Status1 Terminated 3 on April 19, 2012
Recall Number Z-0420-2007
Recall Event ID 37083
510(K)Number K030828  
Product Classification cell analyzer software - Product Code GKZ
Product CXP Analysis Software, Versions 2.0 & 2.1: Part #722396 CXP Analysis Software Kit V2.1, Single User; Part #722395 CXP Analysis Software Kit V2.1, Single User Upgrade; Part #722397 CXP Analysis Software Kit V2.1, 3 User Pack; Part #722398 CXP Analysis Software Kit V2.1, 5 User Pack; Part #722399 CXP Analysis Software Kit V2. 1, 15 User Pack; Part #
722593 CXP Analysis Software Kit V2.1, 5 Network User Pack
Part #722594 CXP Analysis Software Kit V2.1, 10 Network User Pack; Part #722595 CXP Analysis Software Kit V2.1, 15 Network User Pack; Part #629640 CXP Analysis Software Kit V2.0, Single User; Part #629639 CXP Analysis Software Kit V2.0, Single User Upgrade; Part #629641 CXP Analysis Software Kit V2.0, 3 User Pack; Part #629642 CXP Analysis Software Kit V2.0, 5 User Pack; Part #629643 CXP Analysis Software Kit V2.0, 15 User Pack; Part #629644 CXP Analysis Software Kit V2.0, 5 Network User Pack; Part #629645 CXP Analysis Software Kit V2.0, 10 Network User Pack; Part #629646 CXP Analysis Software Kit V2.0, 15 Network User Pack. The Cytomics FC 500 is a system for the qualitative and quantitative measurement of biological and physical properties of cells and other particles.
Code Information Part Numbers: 722396, 722395, 722397, 722398, 722399,  722593, 722594, 722595, 629640, 629639, 629641, 629642,  629643, 629644, 629645, 629646.  Software, Versions 2.0 & 2.1
Recalling Firm/
Manufacturer
Beckman Coulter Inc
200 S Kraemer Blvd
Brea CA 92822-6208
For Additional Information Contact Nora Zerounian
714-961-3634
Manufacturer Reason
for Recall
Two workflow scenarios associated with renaming regions may produce incorrect results.
FDA Determined
Cause 2
Other
Action A Product Corrective Action (PCA) letter was sent on the week of August 29, 2006 to all accounts that have an FC500 or have ordered CXP or MXP software to inform them that a) incorrect statistics are generated when a polygonal region with vertical re-entrant vertices is renamed using the Region Properties dialog box. b)when a region is copied from one plot to another, and then the new region is renamed, the statistics are not updated with the new region name. (Letter was sent by US mail). Letter includes immediate user preventive action to remedy both issues. A response form is included for consignees to return to recalling firm.
Quantity in Commerce 200 units nationwide
Distribution Nationwide & Canada
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = GKZ and Original Applicant = BECKMAN COULTER, INC.
-
-