| | Class 2 Device Recall Foundation Hip System Bipolar Modular |  |
| Date Initiated by Firm | August 02, 2007 |
| Date Posted | November 07, 2007 |
| Recall Status1 |
Terminated 3 on January 09, 2008 |
| Recall Number | Z-0089-2008 |
| Recall Event ID |
44788 |
| 510(K)Number | K953510 |
| Product Classification |
hip prosthesis - Product Code KWY
|
| Product | Foundation Hip System Bipolar Modular, Size 47mm OD; Cat. #412-01-047; Sterile; assembly required by physician prior to implant; manufactured by Encore Medical, L.P., Austin, TX 78758. |
| Code Information |
Lot #: 602731, 677071, 677081, 677091, 678001, 678011, 678731, 678741, 678751, 678761, 693891, 693901, 695101, 697801, 699141, 699151, 732841, 208522, 208532, 208542, 208552, 733961, 734521, 735721, 736521, 739121, 770121, 777431, 780391, 783951, 788491, 796481, 797041, 797341, 814321, 814331, 830931, 830941, 835561, 836181, 842381, 844601, 848701, 853551, 855191, 855201, 857761, 857771, 875591, 875601, 878401, 878411, 878421, 878431, 885391, 892341, 907151, 907161, 907171, 907181, 907191, 907201, 925821, 933801, 933811, 933821, 933831, 949501, 949511, 964361, 964371, 982731 and 421511A. |
| FEI Number |
1000116912
|
Recalling Firm/ Manufacturer |
Encore Medical, Lp
9800 Metric Blvd
Austin
TX
78758
|
| For Additional Information Contact | 512-834-6235 |
Manufacturer Reason for Recall | Device mislabeled; hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date. |
FDA Determined Cause 2 | Packaging change control |
| Action | Firm notified consignees of recall via phone, e-mail and overnight letters beginning on 08/02/07. Consignees asked to return devices for a properly labeled replacement. |
| Quantity in Commerce | 478 units. |
| Distribution | Worldwide, including USA, Greece, Japan, and Saudi Arabia. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = KWY
|
|
|
|