• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall Welch Allyn AED10

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall Welch Allyn AED10see related information
Date Initiated by FirmOctober 26, 2007
Date PostedNovember 06, 2007
Recall Status1 Terminated 3 on November 09, 2011
Recall NumberZ-0129-2008
Recall Event ID 45476
510(K)NumberK021168 
Product Classification Automated External Defibrillator - Product Code MKJ
ProductWelch Allyn AED 10, Automated External Defibrillator, a portable 12V internal battery powered defibrillator; Model AED10; Part Numbers: 970302E,970308E,970309E, 970310E and 970311E;
Code Information Serial numbers of units affected: NA036960, NA036970, NA037000, NA037004, NA037005, NA037009, NA037010, NA037012, NA037014, NA037015, NA037020, NA037022, NA037025, NA037027, NA037029, NA037030, NA037030, NA037031, NA037032, NA037034, NA037035, NA037036, NA037044, NA037045, NA037046, NA037048, NA037050, NA037051, NA037052, NA037053, NA037056, NA037057, NA037058, NA037088, NA037090, NA037091, NA037092, NA037093, NA037094, NA037095, NA037097, NA037098, NA037108, NA037110, NA037111, NA037112, NA037114, NA037115, NA037116, NA037118, NA037119, NA037120, NA037121, NA037122, NA037123, NA037124, NA037125, NA037126, NA037127, NA037128, NA037129, NA037132, NA037134, NA037139, NA037140, NA037143, NA037145, NA037146, NA037147, NA037148, NA037149, NA037150, NA037154, NA037155, NA037157, NA037158, NA037159, NA037160, NA037161, NA037162, NA037163, NA037164, NA037166, NA037167, NA037168, NA037169, NA037170, NA037171, NA037172, NA037173, NA037174, NA037175, NA037177, NA037179, NA037180, NA037182, NA037183, NA037184, NA037185, NA037186, NA037187, NA037188, NA037189, NA037190, NA037193, NA037194, NA037203, NA037206, NA037207, NA037209, NA037210, NA037213, NA037217, NA037219, NA037220, NA037221, NA037223, NA037225, NA037228, NA037234, NA037256, NA037354, NA037499, NA037511, NA037512, NA037513, NA037515, NA037516, NA037517, NA037518, NA037519, NA037523, NA037524, NA037525, NA037532, NA037533, NA037534, NA037535, NA037536, NA037538, NA037539, NA037540, NA037541, NA037542, NA037543, NA037544, NA037545, NA037546, NA037547, NA037548, NA037549, NA037550, NA037551, NA037552, NA037553, NA037554, NA037555, NA037556, NA037557, NA037558, NA037559, NA037560, NA037561, NA037562, NA037563, NA037564, NA037565, NA037566, NA037567, NA037568, NA037569, NA037570, NA037571, NA037572, NA037573, NA037574, NA037575, NA037576, NA037577, NA037578, NA037579, NA037580, NA037581, NA037582, NA037583, NA037584, NA037585, NA037586, NA037587, NA037588, NA037589, NA037590, NA037591, NA037592, NA037593, NA037594, NA037595, NA037596, NA037597, NA037598, NA037599, NA037600, NA037601, NA037602, NA037603, NA037604, NA037605, NA037606, NA037607, NA037608, NA037609, NA037610, NA037614, NA037615, NA037616, NA037617, NA037619, NA037620, NA037621, NA037622, NA037624, NA037625, NA037626, NA037627, NA037628, NA037629, NA037630, NA037631, NA037632, NA037633, NA037634, NA037635, NA037636, NA037637, NA037638, NA037639, NA037640, NA037641, NA037643, NA037644, NA037645, NA037646, NA037647, NA037648, NA037649, NA037650, NA037651, NA037652, NA037653, NA037654, NA037655, NA037656, NA037658, NA037659, NA037660, NA037661, NA037662, NA037663, NA037664, NA037665, NA037666, NA037667, NA037668, NA037669, NA037670, NA037671, NA037672, NA037673, NA037674, NA037675, NA037676, NA037677, NA037678, NA037679, NA037680, NA037681, NA037682, NA037683, NA037684, NA037685, NA037686, NA037688, NA037691, NA037696, NA037699, NA037700, NA037702, NA037704, NA037705, NA037706, NA037707, NA037708, NA037712, NA037713, NA037716, NA037717, NA037719, NA037720, NA037723, NA037724, NA037725, NA037729, NA037730, NA037731, NA037732, NA037738, NA037742, NA037743, NA037745, NA037746, NA037747, NA037749, NA037750, NA037751, NA037752, NA037757, NA037761, NA037762, NA037765, NA037766, NA037768, NA037769, NA037770, NA037773, NA037775, NA037779, NA037780, NA037781, NA037782, NA037784, NA037785, NA037786, NA037787, NA037788, NA037789, NA037790, NA037794, NA037795, NA037796, NA037797, NA037798, NA037800, NA037801, NA037802, NA037803, NA037804, NA037805, NA037806, NA037808, NA037809, NA037810, NA037811, NA037814, NA037816, NA037817, NA037818, NA037819, NA037821, NA037822, NA037823, NA037824, NA037825, NA037826, NA037828, NA037829, NA037831, NA037832, NA037834, NA037835, NA037836, NA037837, NA037838, NA037839, NA037840, NA037841, NA037842, NA037843, NA037844, NA037845, NA037846, NA037847, NA037848, NA037849, NA037850, NA037851, NA037852, NA037853, NA037854, NA037855, NA037856, NA037857, NA037858, NA037859, NA037860, NA037861, NA037862, NA037863, NA037864, NA037866, NA037867, NA037869, NA037870, NA037871, NA037872, NA037873, NA037874, NA037875, NA037876, NA037877, NA037878, NA037879, NA037880, NA037881, NA037882, NA037883, NA037884, NA037885, NA037886, NA037887, NA037888, NA037889, NA037890, NA037891, NA037892, NA037893, NA037894, NA037895, NA037896, NA037897, NA037898, NA037899, NA037900, NA037901, NA037902, NA037903, NA037904, NA037905, NA037906, NA037907, NA037908, NA037909, NA037911, NA037912, NA037913, NA037915, NA037918, NA037919, NA037920, NA037922, NA037923, NA037924, NA037926, NA037927, NA037928, NA037929, NA037930, NA037931, NA037932, NA037933, NA037934, NA037935, NA037936, NA037937, NA037939, NA037940, NA037942, NA037943, NA037944, NA037946, NA037947, NA037948, NA037950, NA037952, NA037954, NA037959, NA037960, NA037963, NA037969, NA037971, NA037972, NA037973, NA037975, NA037977, NA037978, NA037979, NA037980, NA038000, NA038004, NA038005, NA038012, NA038027, NA038028, NA038029, NA038030, NA038031, NA038032, NA038033, NA038035, NA038036, NA038037, NA038038, NA038039, NA038040, NA038041, NA038042, NA038043, NA038044, NA038045, NA038046, NA038047, NA038048, NA038049, NA038050, NA038051, NA038052, NA038053, NA038054, NA038056, NA038057, NA038058, NA038059, NA038060, NA038061, NA038062, NA038063, NA038064, NA038065, NA038066, NA038067, NA038068, NA038069, NA038070, NA038072, NA038073, NA038074, NA038075, NA038076, NA038077, NA038078, NA038080, NA038081, NA038082, NA038083, NA038084, NA038085, NA038086, NA038087, NA038088, NA038089, NA038090, NA038091, NA038092, NA038093, NA038094, NA038095, NA038096, NA038097, NA038098, NA038099, NA038100, NA038101, NA038102, NA038103, NA038104, NA038105, NA038106, NA038107, NA038108, NA038109, NA038110, NA038111, NA038112, NA038113, NA038114, NA038116, NA038117, NA038118, NA038119, NA038120, NA038121, NA038122, NA038123, NA038124, NA038125, NA038126, NA038127, NA038128, NA038129, NA038130, NA038131, NA038132, NA038133, NA038134, NA038135, NA038136, NA038137, NA038138, NA038139, NA038140, NA038141, NA038142, NA038143, NA038144, NA038145, NA038146, NA038147, NA038148, NA038149, NA038150, NA038151, NA038152, NA038153, NA038154, NA038155, NA038156, NA038157, NA038158, NA038159, NA038160, NA038161, NA038162, NA038163, NA038164, NA038165, NA038166, NA038167, NA038168, NA038169, NA038170, NA038171, NA038172, NA038174, NA038175, NA038176, NA038177, NA038178, NA038179, NA038180, NA038181, NA038182, NA038183, NA038184, NA038185, NA038186, NA038187, NA038188, NA038189, NA038190, NA038191, NA038193, NA038194, NA038195, NA038196, NA038197, NA038198, NA038199, NA038200, NA038201, NA038202, NA038203, NA038204, NA038205, NA038206, NA038207, NA038208, NA038209, NA038210, NA038211, NA038213, NA038214, NA038215, NA038216, NA038217, NA038218, NA038219, NA038220, NA038221, NA038222, NA038223, NA038224, NA038225, NA038226, NA038227, NA038228, NA038229, NA038230, NA038231, NA038232, NA038233, NA038234, NA038235, NA038236, NA038237, NA038238, NA038239, NA038240, NA038241, NA038242, NA038243, NA038244, NA038245, NA038246, NA038247, NA038249, NA038250, NA038251, NA038252, NA038253, NA038254, NA038255, NA038256, NA038257, NA038258, NA038259, NA038260, NA038261, NA038262, NA038263, NA038264, NA038265, NA038266, NA038267, NA038268, NA038269, NA038270, NA038271, NA038272, NA038273, NA038274, NA038275, NA038276, NA038277, NA038278, NA038279, NA038280, NA038281, NA038282, NA038283, NA038284, NA038285, NA038286, NA038287, NA038288, NA038289, NA038290, NA038291, NA038292, NA038293, NA038294, NA038295, NA038296, NA038297, NA038298, NA038299, NA038300, NA038301, NA038302, NA038303, NA038304, NA038305, NA038306, NA038307, NA038308, NA038309, NA038310, NA038311, NA038312, NA038313, NA038314, NA038315, NA038316, NA038317, NA038318, NA038319, NA038320, NA038321, NA038322, NA038323, NA038324, NA038325, NA038326, NA038327, NA038328, NA038329, NA038330, NA038331, NA038332, NA038333, NA038334, NA038335, NA038336, NA038337, NA038338, NA038339, NA038340, NA038341, NA038342, NA038343, NA038344, NA038345, NA038346, NA038347, NA038348, NA038349, NA038350, NA038351, NA038352, NA038353, NA038354, NA038355, NA038356, NA038357, NA038358, NA038359, NA038360, NA038361, NA038362, NA038363, NA038364, NA038365, NA038366, NA038367, NA038368, NA038369, NA038370, NA038371, NA038372, NA038373, NA038374, NA038375, NA038376, NA038377, NA038378, NA038379, NA038380, NA038381, NA038382, NA038383, NA038384, NA038385, NA038388, NA038389, NA038390, NA038392, NA038393, NA038396, NA038397, NA038398, NA038399, NA038400, NA038403, NA038404, NA038406, NA038407, NA038409, NA038410, NA038412, NA038413, NA038414, NA038415, NA038416, NA038417, NA038420, NA038421, NA038422, NA038423, NA038425, NA038427, NA038428, NA038429, NA038430, NA038432, NA038434, NA038435, NA038436, NA038437, NA038438, NA038439, NA038440, NA038441, NA038442, NA038443, NA038447, NA038448, NA038449, NA038450, NA038452, NA038453, NA038454, NA038455, NA038456, NA038457, NA038458, NA038459, NA038465, NA038467, NA038468, NA038471, NA038472, NA038473, NA038475, NA038476, NA038477, NA038478, NA038479, NA038483, NA038484, NA038485, NA038486, NA038489, NA038490, NA038492, NA038493, NA038494, NA038496, NA038498, NA038499, NA038500, NA038501, NA038502, NA038503, NA038504, NA038505, NA038506, NA038507, NA038508, NA038509, NA038510, NA038511, NA038512, NA038513, NA038514, NA038515, NA038516, NA038517, NA038518, NA038519, NA038520, NA038521, NA038522, NA038523, NA038524, NA038526, NA038527, NA038528, NA038529, NA038531, NA038532, NA038533, NA038536, NA038538, NA038539, NA038541, NA038542, NA038543, NA038544, NA038545, NA038547, NA038549, NA038550, NA038551, NA038552, NA038554, NA038555, NA038556, NA038557, NA038558, NA038559, NA038560, NA038561, NA038562, NA038564, NA038565, NA038566, NA038567, NA038568, NA038569, NA038570, NA038571, NA038572, NA038573, NA038574, NA038575, NA038576, NA038577, NA038578, NA038579, NA038580, NA038581, NA038582, NA038583, NA038585, NA038586, NA038587, NA038589, NA038590, NA038591, NA038592, NA038593, NA038594, NA038595, NA038596, NA038597, NA038598, NA038599, NA038600, NA038601, NA038602, NA038603, NA038604, NA038605, NA038607, NA038608, NA038609, NA038610, NA038612, NA038613, NA038614, NA038616, NA038618, NA038619, NA038624, NA038625, NA038626, NA038627, NA038628, NA038629, NA038630, NA038632, NA038634, NA038635, NA038636, NA038637, NA038638, NA038639, NA038640, NA038641, NA038642, NA038643, NA038644, NA038645, NA038646, NA038648, NA038649, NA038650, NA038651, NA038652, NA038653, NA038654, NA038658, NA038659, NA038662, NA038665, NA038666, NA038667, NA038668, NA038669, NA038671, NA038672, NA038673, NA038674, NA038675, NA038676, NA038677, NA038678, NA038679, NA038680, NA038683, NA038684, NA038685, NA038686, NA038687, NA038688, NA038690, NA038691, NA038692, NA038693, NA038694, NA038696, NA038697, NA038699, NA038700, NA038702, NA038704, NA038705, NA038706, NA038707, NA038709, NA038711, NA038712, NA038713, NA038714, NA038715, NA038716, NA038717, NA038718, NA038719, NA038720, NA038721, NA038722, NA038723, NA038725, NA038726, NA038727, NA038728, NA038729, NA038730, NA038731, NA038732, NA038733, NA038734, NA038735, NA038736, NA038737, NA038738, NA038739, NA038740, NA038741, NA038742, NA038743, NA038744, NA038745, NA038746, NA038747, NA038748, NA038749, NA038750, NA038751, NA038752, NA038753, NA038754, NA038755, NA038756, NA038757, NA038758, NA038759, NA038760, NA038761, NA038762, NA038763, NA038764, NA038765, NA038766, NA038767, NA038768, NA038769, NA038770, NA038771, NA038772, NA038773, NA038774, NA038775, NA038776, NA038777, NA038778, NA038779, NA038780, NA038781, NA038782, NA038783, NA038784, NA038785, NA038786, NA038787, NA038788, NA038789, NA038791, NA038792, NA038793, NA038794, NA038795, NA038796, NA038797, NA038798, NA038799, NA038800, NA038801, NA038802, NA038803, NA038804, NA038805, NA038806, NA038807, NA038808, NA038809, NA038810, NA038811, NA038814, NA038815, NA038816, NA038817, NA038818, NA038819, NA038820, NA038821, NA038822, NA038823, NA038824, NA038825, NA038826, NA038827, NA038828, NA038829, NA038831, NA038832, NA038833, NA038834, NA038835, NA038836, NA038837, NA038838, NA038839, NA038840, NA038841, NA038842, NA038843, NA038845, NA038846, NA038847, NA038848, NA038849, NA038850, NA038851, NA038852, NA038853, NA038854, NA038855, NA038856, NA038857, NA038858, NA038859, NA038860, NA038861, NA038862, NA038863, NA038864, NA038865, NA038867, NA038868, NA038869, NA038870, NA038871, NA038872, NA038873, NA038874, NA038875, NA038876, NA038877, NA038878, NA038879, NA038880, NA038881, NA038882, NA038883, NA038884, NA038885, NA038886, NA038887, NA038888, NA038889, NA038890, NA038892, NA038894, NA038896, NA038897, NA038898, NA038899, NA038900, NA038901, NA038902, NA038903, NA038904, NA038905, NA038906, NA038907, NA038908, NA038909, NA038911, NA038912, NA038913, NA038914, NA038915, NA038916, NA038917, NA038918, NA038919, NA038920, NA038921, NA038922, NA038923, NA038924, NA038925, NA038926, NA038927, NA038928, NA038929, NA038930, NA038931, NA038932, NA038933, NA038934, NA038935, NA038936, NA038937, NA038938, NA038939, NA038940, NA038942, NA038943, NA038944, NA038945, NA038946, NA038947, NA038951, NA038952, NA038953, NA038956, NA038957, NA038958, NA038959, NA038960, NA038961, NA038963, NA038964, NA038965, NA038966, NA038969, NA038970, NA038971, NA038973, NA038975, NA038979, NA038980, NA038981, NA038982, NA038983, NA038985, NA038986, NA038988, NA038989, NA038991, NA038993, NA038994, NA038995, NA038996, NA038997, NA039004, NA039005, NA039006, NA039007, NA039008, NA039009, NA039010, NA039011, NA039012, NA039014, NA039015, NA039016, NA039017, NA039018, NA039019, NA039020, NA039021, NA039022, NA039025, NA039026, NA039027, NA039028, NA039029, NA039030, NA039031, NA039032, NA039033, NA039034, NA039035, NA039036, NA039037, NA039038, NA039039, NA039040, NA039041, NA039042, NA039047, NA039048, NA039049, NA039050, NA039051, NA039052, NA039053, NA039054, NA039055, NA039056, NA039057, NA039058, NA039059, NA039060, NA039061, NA039062, NA039063, NA039064, NA039065, NA039066, NA039067, NA039068, NA039069, NA039070, NA039071, NA039072, NA039073, NA039074, NA039075, NA039076, NA039077, NA039078, NA039079, NA039080, NA039081, NA039083, NA039084, NA039085, NA039086, NA039087, NA039088, NA039089, NA039090, NA039091, NA039092, NA039093, NA039094, NA039095, NA039096, NA039097, NA039098, NA039099, NA039100, NA039101, NA039102, NA039103, NA039104, NA039105, NA039106, NA039107, NA039108, NA039109, NA039110, NA039111, NA039112, NA039113, NA039114, NA039115, NA039116, NA039117, NA039118, NA039119, NA039120, NA039121, NA039122, NA039123, NA039124, NA039125, NA039126, NA039127, NA039128, NA039129, NA039130, NA039131, NA039132, NA039133, NA039134, NA039135, NA039136, NA039137, NA039138, NA039139, NA039140, NA039141, NA039142, NA039143, NA039144, NA039145, NA039146, NA039147, NA039148, NA039149, NA039150, NA039151, NA039152, NA039153, NA039154, NA039155, NA039156, NA039157, NA039158, NA039159, NA039160, NA039161, NA039162, NA039163, NA039164, NA039165, NA039166, NA039167, NA039168, NA039169, NA039170, NA039171, NA039172, NA039173, NA039174, NA039175, NA039176, NA039177, NA039178, NA039179, NA039180, NA039181, NA039182, NA039183, NA039184, NA039185, NA039186, NA039187, NA039188, NA039189, NA039190, NA039191, NA039192, NA039193, NA039194, NA039195, NA039196, NA039197, NA039198, NA039200, NA039201, NA039202, NA039203, NA039204, NA039205, NA039206, NA039207, NA039208, NA039209, NA039210, NA039211, NA039212, NA039213, NA039214, NA039215, NA039216, NA039217, NA039218, NA039219, NA039220, NA039221, NA039222, NA039223, NA039224, NA039225, NA039226, NA039227, NA039228, NA039229, NA039230, NA039231, NA039232, NA039233, NA039234, NA039235, NA039236, NA039237, NA039238, NA039239, NA039240, NA039241, NA039242, NA039243, NA039244, NA039245, NA039246, NA039247, NA039248, NA039249, NA039250, NA039251, NA039252, NA039253, NA039254, NA039255, NA039256, NA039257, NA039258, NA039259, NA039260, NA039261, NA039262, NA039263, NA039264, NA039265, NA039266, NA039267, NA039268, NA039269, NA039270, NA039271, NA039272, NA039273, NA039274, NA039275, NA039276, NA039277, NA039278, NA039279, NA039280, NA039281, NA039282, NA039283, NA039284, NA039285, NA039286, NA039287, NA039288, NA039289, NA039290, NA039291, NA039292, NA039293, NA039294, NA039295, NA039296, NA039297, NA039298, NA039299, NA039300, NA039301, NA039302, NA039303, NA039304, NA039305, NA039306, NA039307, NA039308, NA039309, NA039310, NA039311, NA039312, NA039313, NA039314, NA039315, NA039316, NA039317, NA039318, NA039319, NA039320, NA039321, NA039322, NA039324, NA039327, NA039328, NA039329, NA039330, NA039331, NA039332, NA039334, NA039335, NA039336, NA039337, NA039340, NA039341, NA039342, NA039343, NA039344, NA039345, NA039346, NA039347, NA039349, NA039350, NA039351, NA039352, NA039353, NA039354, NA039355, NA039356, NA039357, NA039358, NA039359, NA039360, NA039361, NA039362, NA039363, NA039364, NA039366, NA039367, NA039368, NA039369, NA039370, NA039371, NA039372, NA039373, NA039374, NA039375, NA039376, NA039377, NA039378, NA039379, NA039380, NA039381, NA039382, NA039383, NA039385, NA039386, NA039387, NA039388, NA039389, NA039390, NA039391, NA039392, NA039393, NA039394, NA039395, NA039396, NA039397, NA039398, NA039399, NA039400, NA039401, NA039402, NA039403, NA039404, NA039405, NA039406, NA039407, NA039408, NA039409, NA039410, NA039411, NA039412, NA039413, NA039414, NA039415, NA039416, NA039417, NA039418, NA039419, NA039420, NA039421, NA039422, NA039423, NA039424, NA039425, NA039426, NA039427, NA039428, NA039429, NA039430, NA039431, NA039432, NA039433, NA039434, NA039435, NA039436, NA039437, NA039438, NA039439, NA039440, NA039441, NA039442, NA039443, NA039444, NA039445, NA039446, NA039447, NA039448, NA039449, NA039450, NA039451, NA039452, NA039453, NA039454, NA039455, NA039456, NA039457, NA039458, NA039459, NA039460, NA039461, NA039462, NA039463, NA039464, NA039465, NA039466, NA039467, NA039468, NA039469, NA039470, NA039471, NA039472, NA039473, NA039474, NA039475, NA039476, NA039477, NA039478, NA039479, NA039480, NA039481, NA039482, NA039483, NA039484, NA039485, NA039486, NA039487, NA039488, NA039489, NA039490, NA039491, NA039492, NA039493, NA039494, NA039495, NA039497, NA039498, NA039499, NA039500, NA039501, NA039502, NA039503, NA039504, NA039505, NA039506, NA039507, NA039508, NA039510, NA039511, NA039512, NA039514, NA039515, NA039516, NA039517, NA039518, NA039519, NA039520, NA039521, NA039522, NA039523, NA039524, NA039525, NA039526, NA039527, NA039528, NA039529, NA039530, NA039531, NA039532, NA039533, NA039534, NA039535, NA039536, NA039537, NA039538, NA039539, NA039540, NA039541, NA039542, NA039543, NA039544, NA039545, NA039546, NA039547, NA039548, NA039549, NA039550, NA039551, NA039552, NA039553, NA039554, NA039555, NA039556, NA039558, NA039560, NA039561, NA039562, NA039563, NA039564, NA039565, NA039566, NA039567, NA039568, NA039569, NA039570, NA039571, NA039572, NA039573, NA039574, NA039575, NA039577, NA039578, NA039579, NA039581, NA039582, NA039584, NA039585, NA039586, NA039588, NA039589, NA039591, NA039592, NA039593, NA039595, NA039598, NA039599, NA039600, NA039601, NA039603, NA039604, NA039605, NA039607, NA039608, NA039609, NA039612, NA039616, NA039618, NA039619, NA039620, NA039621, NA039623, NA039624, NA039629, NA039630, NA039632, NA039633, NA039639, NA039644, NA039645, NA039650, NA039655, NA039664, NA039674, NA039692, NA039701, NA039702, NA039703, NA039711, NA039715, NA039719, NA039720, NA039722, NA039723, NA039724, NA039726, NA039728, NA039729, NA039730, NA039732, NA039733, NA039734, NA039735, NA039736, NA039737, NA039741, NA039742, NA039743, NA039744, NA039747, NA039748, NA039750, NA039751, NA039753, NA039754, NA039755, NA039756, NA039757, NA039758, NA039759, NA039760, NA039761, NA039762, NA039763, NA039764, NA039766, NA039767, NA039771, NA039772, NA039773, NA039774, NA039776, NA039778, NA039780, NA039781, NA039782, NA039784, NA039785, NA039786, NA039787, NA039790, NA039793, NA039795, NA039796, NA039797, NA039798, NA039799, NA039800, NA039802, NA039803, NA039804, NA039805, NA039806, NA039808, NA039809, NA039810, NA039814, NA039828, NA039839, NA039842, NA039843, NA039844, NA039846, NA039847, NA039848, NA039849, NA039850, NA039851, NA039852, NA039853, NA039854, NA039855, NA039856, NA039857, NA039858, NA039859, NA039860, NA039865, NA039866, NA039868, NA039871, NA039876, NA039877, NA039878
Recalling Firm/
Manufacturer
Welch Allyn Protocol, Inc
8500 Sw Creekside Pl
Beaverton OR 97008
For Additional Information ContactRobert W. Berry
503-526-8500
Manufacturer Reason
for Recall
Failure to Deliver Shock; a defective capacitor may cause the delay or non-delivery of the defibrillating shock which may result in failure to resuscitate the patient
FDA Determined
Cause 2
Nonconforming Material/Component
ActionThe firm issued a press release on 10/30/07, and has set up a call center for customers, and plans to send three (3) certified mailings to notify affected customers, alert them of the issue, and provide instructions on how to return their device. The first letter dated 10/26/07 was sent via certified mail to customers on 10/26/07. The letter informs customers of the firm recalling the product to correct faulty capacitors on the circuit board, and which could result in a malfunction that prevents or delays delivery of a defibrillating shock. The letter indicates the faulty AED10 devices were manufactured between 3/29/07 and 8/9/07. The letter also indicates the serial numbers affected, and where the serial number is located - upper left hand corner of the back panel of the device. In addition the letter includes a response form, the firm's contact information, and indicates that the firm will exchange the recalled defibrillator with a replacement and new five (5) year warranty.
Quantity in Commerce1,794 distributed - 1176 in US and 618 internationally
DistributionWorldwide.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = MKJ
-
-