| Date Initiated by Firm | November 13, 2007 |
| Date Posted | June 11, 2008 |
| Recall Status1 |
Terminated 3 on May 23, 2011 |
| Recall Number | Z-1158-2008 |
| Recall Event ID |
45875 |
| 510(K)Number | K062213 |
| Product Classification |
EBV IgM in-vitro diagnostic test kit - Product Code LSE
|
| Product | BIOPLEX 2200 EBV IgM KIT, CALIBRATORS, AND CONTROLS: 100 tests - Catalog No. 665-1350 - manufactured by Bio-Rad Laboratories in Redmond, WA. The product is a multiplex flow immunoassay intended for the qualitative detection of two (2) analytes; Epstein-Barr Virus Viral Capsid Antigen (EBV VCA) IgM antibodies and Heterophile antibodies in human serum. |
| Code Information |
Lot Numbers: 943712 Exp. 7/30/07 and 943713 Exp. 4/30/08 |
| FEI Number |
3008272466
|
Recalling Firm/ Manufacturer |
Bio-Rad Laboratories 6565 185th Ave NE Redmond WA 98052-5039
|
| For Additional Information Contact | Scott J. Dennis 425-498-1709 |
Manufacturer Reason for Recall | False negative results due to reagent packs exhibiting low signal. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | On 11/13/2007, the firm faxed Voluntary Field Correction Letters (which are dated 11/9/2007) to its consignees and also contacted them via telephone. The Voluntary Field Correction Letters: a) informed them of the potential for false negative results; b) recommended that they perform QC testing daily with each reagent pack; c) provided a new insert for the Instructions for Use; and d) provided a Response Sheet for each consignee to complete and return to the firm. |
| Quantity in Commerce | 996 kits |
| Distribution | Worldwide Distribution -- USA, Canada, France, United Kingdom, Italy, Germany, South Africa, and Switzerland. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = LSE
|