• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Continuous ventilators

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Continuous ventilatorssee related information
Date Initiated by FirmOctober 15, 2007
Date PostedApril 24, 2008
Recall Status1 Terminated 3 on September 01, 2009
Recall NumberZ-1085-2008
Recall Event ID 45545
510(K)NumberK842876 K982454 
Product Classification ventilation accessories - Product Code CBK
ProductContinuous ventilator accessories under the following brand names: a) RP-Seal, Spring-5 pk model number F35465; b) RP-Seal, Greece, Ballscrew model number F35036; c) RP-Tub PVC 1/8 in x 50 ft model number 1008841; d) RP-Vision Filter Cover model number 1003444; e) RP-Vision RS-232 HRN ABV 106K model number 1004699; and f) RP-Wrap, Cord model number F35195. Filter, mask, and/or accessories for sleep therapy and ventilation devices, Respironics, Inc., Murrysville, PA 15668.
Code Information Lot Numbers 071010, 071011, and 071012.
FEI Number 2518422
Recalling Firm/
Manufacturer
Respironics, Inc.
1001 Murry Ridge Ln
Murrysville PA 15668-8550
For Additional Information ContactKatherine dePadua
724-387-7770
Manufacturer Reason
for Recall
Foreign material: Glass fragments may present in the plastic bag material used to ship components.
FDA Determined
Cause 2
Environmental control
ActionThe recalling firm issued a Recall Notification dated 10/18/07 to its customers. Replacement products were to be sent.
Quantity in Commerce20950 units
DistributionThe products were shipped nationwide to medical facilities.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = CBK
510(K)s with Product Code = CBK
-
-