• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall BD 5 ml Syringe Luer Lok Tip

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall BD 5 ml Syringe Luer Lok Tipsee related information
Date Initiated by FirmJanuary 07, 2008
Date PostedApril 16, 2008
Recall Status1 Terminated 3 on May 19, 2008
Recall NumberZ-0914-2008
Recall Event ID 46347
Product Classification Piston syringe - Product Code FMF
ProductBD 5 ml Syringe Luer Lok Tip Bulk Sterile Convenience Pak. Latex Free . Sterile. Reorder Number: 309703, BD, Franklin Lakes, NJ 07417.
Code Information Reorder # 309703: Lot numbers: 7123237, 7156175, 7156177, 7212946, 7212948, 7239565, 7239566, 7271878.
FEI Number 2243072
Recalling Firm/
Manufacturer
Becton Dickinson & Company
1 Becton Dr
Franklin Lakes NJ 07417-1815
For Additional Information ContactGreg Morgan
201-847-4344
Manufacturer Reason
for Recall
Defective Seal: The convenience trays may have open seals, which can adversely impact tray sterility.
FDA Determined
Cause 2
Packaging
ActionBD notified distributors by an Urgent Product Recall Letter sent via UPS on January 7, 2008. In addition, the firm sent an Urgent Product Recall Letter on January 11, 2008 to customers identified by the distributors. The letter instructed the users to remove the product from its inventory and to return affected products to the firm using an enclosed customer packing list. For additional information, contact 1-888-237-2762.
Quantity in Commerce1,754,100
DistributionNationwide, including the states of CA, NY, MO, MA, IL, OH, TN, TX, MI, LA, MO, ID, GA, NE, SC and FL.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-