| | Class 2 Device Recall Baxa ExactaMix 600 Pharmacy compounding system |  |
| Date Initiated by Firm | February 11, 2008 |
| Date Posted | August 19, 2008 |
| Recall Status1 |
Terminated 3 on March 16, 2012 |
| Recall Number | Z-1371-2008 |
| Recall Event ID |
46771 |
| Product Classification |
Automated pharmacy compounding system - Product Code NEP
|
| Product | Exacta-Mix 600 Pharmacy Compounding System is used in the pharmacy to compound multiple source ingredients into on final solution based on user-specified sequences and volumes. |
| Code Information |
Model 071B-RS232, all serial numbers |
| FEI Number |
1419106
|
Recalling Firm/ Manufacturer |
Baxa Corporation 14445 Grasslands Dr Englewood CO 80112-7062
|
| For Additional Information Contact | 303-617-2242 |
Manufacturer Reason for Recall | Flush solution from TPN compounder may be added to patient's TPN bag. |
FDA Determined Cause 2 | Software design |
| Action | Consignees were notified by letter (Exacta-Mix Compounder Software Safety Alert) on 02/11/2008 advising them of the necessity to install a waste bag prior to a machine flush. |
| Quantity in Commerce | 221 units |
| Distribution | Nationwide, including military facilities in CA, FL, HI, MS, NC, TX, VA and VA facilities in NY, OR, SC, UT. Foreign distribution to Australia, Canada, China and New Zealand. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|