| | Class 2 Device Recall Oncor Impression Medical Linear Accelerator |  |
| Date Initiated by Firm | January 11, 2008 |
| Date Posted | April 02, 2008 |
| Recall Status1 |
Terminated 3 on May 06, 2011 |
| Recall Number | Z-1398-2008 |
| Recall Event ID |
46940 |
| 510(K)Number | K993425 |
| Product Classification |
Medical Linear Accelerator - Product Code IYE
|
| Product | Oncor Impression Medical Linear Accelerator, Material Number : 5857920, with control console software versions 6.5/7.2, 8.0, 9.0 and 9.1, Manufactured by Siemens Medical Solutions USA, Inc, Concord, CA |
| Code Information |
Serial numbers; 4078 70-4143 70-4257 3878 70-4344 70-4207 70-4235 4021 70-4296 70-4165 70-4167 3729 5022 5035 5074 3811 3845 5108 70-4238 70-4386 3894 4048 70-4276 70-4277 70-4281 70-4083 3974 3962 3973 70-4152 and 3883 |
| FEI Number |
2937457
|
Recalling Firm/ Manufacturer |
Siemens Medical Solutions USA, Inc
4040 Nelson Ave
Concord
CA
94520-1200
|
| For Additional Information Contact | 925-246-8407 |
Manufacturer Reason for Recall | Linear accelerator gantry may rotate in an unexpected direction and may cause patient injury on impact. |
FDA Determined Cause 2 | Software design |
| Action | This is a mandatory software upgrade for consignees. The firm's representatives have contacted consignees via Urgent Device Notification letters dated February 14, 2008, notifying them of the new update and making arrangements to install it. |
| Quantity in Commerce | 31 units |
| Distribution | Worldwide-USA including states of AL, AK, AZ, AR, CA, CO, DE, DC, FL, GA, IL, IN, KS, KY, LA, MD, ME, MI, MN, MS, MO, MN, NE, NV, NH, NJ, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, VT, VA, WA, WV, WI, and WY and countries of Angola, Australia, Austria,Belgium,Brazil,Canada,Chile,Colombia,Costa Rica, Croatia,Cyprus,The Czech Republic,Denmark,DPR of Korea,El Salvador,France,Germany,Greece,Hungary,India,Iran,Ireland,Italy,Japan,Jordan,Malaysia,Mexico, The Netherlands,New Zealand,Norway,P.R. China,Pakistan,Peru,Philippines,Poland,Portugal,Republic of Korea,Romania,Russian Fed.Saudi Arabia,Singapore,Slovakia,South Africa,Spain,Sweden,Taiwan,Thailand,Trinidad,Tobago,Turkey, U.A.E.,Ukraine, The United Kingdom, and Venezuela |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|