 Class 2 Device Recall LIFEPAK Express AEDsee related information
Date Initiated by FirmJuly 24, 2008
Date PostedOctober 30, 2008
Recall Status1 Terminated 3 on August 08, 2013
Recall NumberZ-0150-2009
Recall Event ID 49056
510(K)NumberK011144 
Product Classification Automated External Defibrillator - Product Code MKJ
ProductLIFEPAK Express Automated External Defibrillator (AED). Labeling on the device states LIFEPAK CR Plus/LIFEPAK EXPRESS Defibrillators with ADAPTIV Biphasic Technology; Medtronic Emergency Response Systems, Inc. 11811 Willows Road NE Redmond, WA. The battery operated, small portable device is primarily deployed in Public Access Defibrillator (PAD) environments where the expected use for any one device is once every 20 months. Product is capable of 30 full discharges. LIFEPAK Express AEDs are intended for use on victims/patients in cardiac arrest. It includes a patented shock advisory system that notifies the user (i.e. minimally trained responder) if it detects a cardiac rhythm that requires a shock. The individual must be unresponsive (unconscious), not breathing normally, and showing no signs of circulation (for example, no pulse, no coughing, or no movement). With Infant/Child Reduced Energy Defibrillation Electrodes, the LIFEPAK Express AEDs may be used on children who are up to 8 years old or who weigh up to 25kg (55 lb). The device is designed specifically for infrequent use and for use by people whose only training is in CPR and in using AEDs.
Code Information PART NUMBER: 3202177-000, 3202177-001, 3202177-002, 3202177-006, 3202177-024, 3202177-026, 3202177-046, 3202177-066, 3202177-080, 3202177-086, and 3202177-106; SERIAL NUMBER: 35057999, 35558399, 35057996, 35076760, 35086872, 35086873, 35086874, 35086875, 35086876, 35086877, 35086878, 35086879, 35086880, 35086881, 35087002, 35087003, 35087004, 35112123, 35120681, 35120682, 35120683, 35174476, 35174477, 35174478, 35174479, 35174487, 35189989, 35190004, 35190371, 35190372, 35237901, 35363559, 35388553, 35415805, 35457181, 35606013, 35627387, 35667618, 35702342, 35843530, 36084979, 36106000, 35086839, 35142569, 35142570, 35142571, 35142572, 35142573, 35142574, 35266844, 35266845, 35266878, 35266879, 35266881, 35266883, 35266884, 35266886, 35266887, 35266888, 35266889, 35266890, 35266891, 35266892, 35266893, 35266899, 35266900, 35314291, 35416139, 35450316, 35450324, 35058947, 35058948, 35058949, 35058950, 35058951, 35058952, 35058953, 35058954, 35058956, 35058958, 35058961, 35058968, 35058969, 35058970, 35058971, 35058972, 35058981, 35058982, 35058986, 35058988, 35058993, 35058994, 35058999, 35059002, 35060160, 35060308, 35060309, 35060310, 35060311, 35060312, 35060313, 35060314, 35060315, 35060316, 35060317, 35060589, 35060721, 35060961, 35061061, 35061139, 35061205, 35061206, 35061207, 35064900, 35064901, 35064902, 35064903, 35064904, 35064905, 35064907, 35064908, 35064909, 35064910, 35064911, 35064912, 35064913, 35064914, 35064915, 35064916, 35064917, 35064918, 35064919, 35064920, 35064921, 35064922, 35064923, 35064924, 35064925, 35064926, 35064927, 35064928, 35064929, 35064930, 35064931, 35064932, 35064933, 35064934, 35064935, 35064936, 35064937, 35064938, 35064939, 35064940, 35064941, 35064942, 35064943, 35064944, 35064945, 35064946, 35064947, 35064948, 35064949, 35064950, 35064951, 35064952, 35064953, 35064954, 35064955, 35064956, 35064957, 35064958, 35064959, 35064960, 35064961, 35064962, 35064963, 35064964, 35064965, 35064966, 35064967, 35064968, 35064969, 35064970, 35064971, 35064972, 35064973, 35064974, 35064975, 35065305, 35065306, 35065307, 35065308, 35065309, 35065310, 35065311, 35065312, 35065313, 35065314, 35065315, 35065316, 35065317, 35065318, 35065319, 35065320, 35065321, 35065325, 35065327, 35065336, 35065337, 35065338, 35065379, 35065381, 35065384, 35065385, 35065392, 35065393, 35065394, 35065395, 35065396, 35065397, 35065400, 35065401, 35065403, 35065408, 35065411, 35066342, 35066343, 35066464, 35074843, 35074861, 35074863, 35074866, 35074867, 35074868, 35074869, 35074870, 35074871, 35074872, 35075602, 35075604, 35075605, 35075606, 35075607, 35075608, 35075609, 35075610, 35075611, 35075612, 35075613, 35075615, 35075616, 35075617, 35075618, 35075619, 35075620, 35075621, 35075622, 35075624, 35075626, 35075627, 35075628, 35075629, 35075630, 35075631, 35075632, 35075633, 35075634, 35075637, 35075638, 35075639, 35075640, 35075641, 35075642, 35075643, 35075644, 35075645, 35075646, 35075647, 35075648, 35075649, 35075650, 35075651, 35075652, 35075653, 35075654, 35075655, 35075662, 35075663, 35075664, 35075665, 35075666, 35075667, 35075668, 35075669, 35075670, 35075671, 35075672, 35075673, 35075674, 35075675, 35075676, 35075677, 35075678, 35075679, 35075680, 35075681, 35075682, 35075683, 35075684, 35075685, 35076506, 35076756, 35076757, 35076766, 35076767, 35076768, 35076769, 35076820, 35076821, 35076822, 35076823, 35076824, 35076825, 35076826, 35076827, 35076829, 35076830, 35076831, 35076832, 35076833, 35076834, 35076835, 35076836, 35076837, 35076838, 35076839, 35076840, 35076841, 35076842, 35076843, 35076844, 35076845, 35076846, 35076847, 35076848, 35076849, 35076850, 35076851, 35076852, 35076853, 35076854, 35076855, 35076856, 35076857, 35076858, 35076859, 35076860, 35076861, 35076862, 35076863, 35076864, 35076865, 35076866, 35076867, 35076868, 35076869, 35076870, 35077750, 35079092, 35079093, 35079094, 35079095, 35079096, 35079097, 35079098, 35079099, 35079100, 35079101, 35079102, 35079103, 35079104, 35079105, 35079106, 35079108, 35081319, 35081320, 35081321, 35081322, 35081323, 35081332, 35081333, 35081334, 35081335, 35081336, 35081337, 35081338, 35081339, 35081340, 35081342, 35081344, 35081345, 35081346, 35081347, 35081348, 35081349, 35081350, 35081351, 35081352, 35081354, 35081356, 35081357, 35081358, 35081360, 35081442, 35081443, 35081444, 35081445, 35081446, 35081447, 35081448, 35081449, 35081450, 35081451, 35081452, 35081453, 35081454, 35081455, 35081456, 35081457, 35081458, 35081459, 35081460, 35081461, 35081462, 35081463, 35081464, 35081465, 35081466, 35081467, 35081468, 35081469, 35081470, 35081471, 35081472, 35081473, 35081474, 35081475, 35081476, 35081477, 35081478, 35081479, 35081480, 35081481, 35081482, 35081483, 35081484, 35081485, 35081486, 35081487, 35081488, 35081489, 35081490, 35081491, 35081492, 35081493, 35081494, 35081495, 35081496, 35081497, 35081498, 35081499, 35081500, 35081501, 35081502, 35081503, 35081504, 35081506, 35081507, 35081508, 35081509, 35081510, 35081511, 35081512, 35081513, 35081514, 35081516, 35081517, 35081518, 35081519, 35081520, 35081521, 35081522, 35081523, 35081524, 35081525, 35081526, 35081527, 35081528, 35081529, 35081530, 35081532, 35081533, 35081534, 35081535, 35081536, 35081537, 35081538, 35081539, 35081540, 35081541, 35081542, 35081543, 35081544, 35081545, 35081546, 35081547, 35081548, 35081549, 35081550, 35081551, 35081552, 35081553, 35081554, 35081555, 35081557, 35081558, 35081559, 35081560, 35081561, 35081562, 35081563, 35081564, 35081565, 35081566, 35081567, 35081568, 35081569, 35081570, 35081571, 35081572, 35081573, 35081574, 35081575, 35081576, 35081577, 35081578, 35081579, 35081580, 35081581, 35081582, 35081583, 35081584, 35081585, 35081586, 35081587, 35081588, 35081589, 35081590, 35081591, 35081592, 35081593, 35081594, 35081595, 35081596, 35081597, 35081598, 35081599, 35081601, 35081602, 35081603, 35081604, 35081605, 35081606, 35081607, 35081608, 35081609, 35084506, 35084507, 35084699, 35084700, 35084701, 35084702, 35084703, 35084704, 35084705, 35084861, 35084862, 35084895, 35085212, 35085217, 35085222, 35085313, 35086860, 35086987, 35086988, 35086989, 35119386, 35119387, 35119388, 35119389, 35119390, 35119391, 35119392, 35119393, 35119394, 35119395, 35119396, 35119397, 35119398, 35119399, 35119400, 35119401, 35119402, 35119403, 35119404, 35119405, 35127147, 35127148, 35135990, 35135991, 35136661, 35136662, 35136675, 35136682, 35136689, 35136690, 35142494, 35142495, 35142496, 35142497, 35142498, 35142499, 35142500, 35142686, 35142696, 35142697, 35142698, 35142702, 35142703, 35142704, 35142705, 35142706, 35142709, 35142711, 35142725, 35142727, 35142728, 35142729, 35142730, 35142731, 35142732, 35142733, 35142734, 35142741, 35142744, 35142751, 35142752, 35142753, 35142754, 35142755, 35142762, 35142764, 35142765, 35142767, 35164275, 35165280, 35165459, 35165460, 35165462, 35165463, 35165465, 35165466, 35165467, 35165470, 35165473, 35165474, 35165475, 35165476, 35165478, 35165479, 35165481, 35165482, 35165483, 35165484, 35165485, 35165486, 35165881, 35165882, 35165883, 35165885, 35165889, 35165890, 35165894, 35165895, 35166530, 35166555, 35166561, 35166662, 35166667, 35166668, 35166669, 35173005, 35180729, 35180730, 35180731, 35180732, 35180733, 35180734, 35180735, 35180736, 35180737, 35180738, 35180739, 35180740, 35180741, 35180742, 35180743, 35180744, 35180745, 35180746, 35180747, 35180748, 35180749, 35180750, 35180752, 35180753, 35180755, 35180756, 35180757, 35180758, 35180759, 35180760, 35180761, 35180762, 35180763, 35180764, 35180765, 35180766, 35180767, 35180768, 35180769, 35180770, 35182559, 35188809, 35188848, 35188851, 35188853, 35188855, 35188858, 35188860, 35188863, 35188865, 35188868, 35188870, 35188872, 35188874, 35188876, 35188878, 35188881, 35188884, 35188891, 35188895, 35189847, 35189982, 35189994, 35190002, 35190012, 35190021, 35190045, 35190057, 35190065, 35190071, 35190072, 35190076, 35190077, 35190078, 35200862, 35200863, 35200866, 35200867, 35200869, 35200871, 35200873, 35200875, 35200877, 35200880, 35200882, 35200884, 35200887, 35200889, 35200891, 35200894, 35200895, 35200916, 35200917, 35200918, 35200919, 35200920, 35200921, 35200922, 35200923, 35200924, 35200925, 35200926, 35200927, 35200928, 35200929, 35200930, 35200931, 35200932, 35200933, 35200934, 35200935, 35213744, 35213746, 35213749, 35213751, 35219822, 35219823, 35219824, 35219825, 35219826, 35219827, 35219828, 35219829, 35219830, 35219831, 35219832, 35219833, 35219834, 35219835, 35219836, 35219837, 35219838, 35219839, 35219840, 35219841, 35219842, 35219843, 35219844, 35219846, 35219847, 35219848, 35219849, 35219850, 35219851, 35219852, 35219853, 35219854, 35219855, 35219856, 35219857, 35219858, 35219859, 35219860, 35219861, 35219862, 35219863, 35219864, 35219865, 35219866, 35219867, 35219868, 35219869, 35219870, 35219871, 35219872, 35219873, 35219874, 35219875, 35219876, 35219877, 35219878, 35219879, 35219880, 35219881, 35219882, 35219883, 35219884, 35219885, 35219886, 35219887, 35219888, 35219889, 35219890, 35219891, 35219892, 35219893, 35219894, 35219895, 35219896, 35219897, 35219898, 35219899, 35219900, 35219901, 35219902, 35219903, 35219904, 35219905, 35219906, 35219907, 35219908, 35219909, 35219910, 35219911, 35219912, 35219913, 35219914, 35219915, 35219916, 35219917, 35219918, 35219919, 35219920, 35219921, 35219922, 35219923, 35219925, 35219926, 35219927, 35219928, 35219929, 35219930, 35219931, 35219932, 35219933, 35219934, 35219935, 35219936, 35219937, 35219938, 35219939, 35219940, 35219941, 35219942, 35219943, 35219944, 35219945, 35219946, 35219947, 35219948, 35219949, 35219950, 35219951, 35219952, 35219953, 35219954, 35219955, 35219956, 35219957, 35219958, 35219959, 35219960, 35219961, 35219962, 35219963, 35219964, 35219965, 35219966, 35219967, 35219968, 35219969, 35219970, 35219971, 35219972, 35219973, 35219974, 35219975, 35219976, 35219977, 35219978, 35219979, 35219980, 35219981, 35219982, 35219983, 35219984, 35219985, 35219986, 35219987, 35219988, 35219989, 35219990, 35219991, 35219992, 35219993, 35219994, 35219995, 35219996, 35219997, 35219998, 35219999, 35220000, 35220001, 35220002, 35220003, 35220004, 35220005, 35220006, 35220007, 35220009, 35220010, 35220011, 35220012, 35220013, 35220014, 35220015, 35220016, 35220017, 35220018, 35220019, 35220020, 35220021, 35220022, 35220023, 35220024, 35220025, 35220026, 35220027, 35220028, 35220029, 35220030, 35220031, 35220032, 35220033, 35220034, 35220035, 35220036, 35220037, 35220038, 35220039, 35220040, 35220041, 35220042, 35220043, 35220044, 35220045, 35220046, 35220047, 35220048, 35220049, 35220050, 35220051, 35220052, 35220053, 35220643, 35220644, 35220645, 35220646, 35220647, 35220649, 35220651, 35220655, 35220656, 35220657, 35220663, 35220664, 35220666, 35220668, 35220708, 35232342, 35232346, 35232352, 35232358, 35232362, 35232364, 35237919, 35237920, 35237921, 35237922, 35237923, 35237924, 35237926, 35237927, 35237928, 35237929, 35237930, 35243430, 35450318, 35450322, 35527546, 35573219, 35582098, 35589905, 35606041, 35627778, 35882828, 35882831, 35889950, 35906079, 35927909, 35927913, 35927915, 35927918, 35988794, 35988795, 35988797, 35988798, 35988799, 35988800, 35988801, 35988802, 35988803, 35988804, 35988805, 35988806, 35988807, 35988808, 35988809, 35988810, 35988811, 35988812, 35988813, 35988814, 35988815, 35988816, 35988817, 35988818, 35988819, 35988820, 35988821, 35988822, 35988823, 35988824, 35988825, 35988826, 35988827, 35988828, 35988829, 35988830, 35988831, 35988837, 35992978, 35992991, 35992996, 35992997, 35992998, 35993000, 35993011, 35993028, 35993031, 35993033, 36085007, 36103006, 36103012, 36103013, 36103015, 36103016, 36103017, 36103022, 36103028, 36103029, 36103030, 36103033, 36121401, 36121402, 36121403, 36138709, 36151002, 36151004, 36151006, 36151007, 36151008, 36151009, 36151010, 36151011, 36151012, 36151013, 36151014, 36151015, 36151016, 36151017, 36151018, 36151019, 36154390, 36154391, 36154392, 36154393, 36154394, 36154395, 36154396, 36154397, 36154398, 36154399, 36154400, 36154401, 36154402, 36154403, 36154404, 36154405, 36154406, 36154407, 36154408, 36154410, 36154411, 36154412, 36154413, 36154414, 36154415, 36154416, 36154417, 36154418, 36154419, 36154420, 36154421, 36154422, 36154467, 36154468, 36154469, 36154470, 36154471, 36154472, 36154473, 36154474, 36154475, 36154476, 36154477, 36154478, 36154479, 36154481, 36154482, 36154483, 36154484, 36154490, 36358206, 36358208, 36358209, 35232338, 35140669, 35140672, 35189992, 35270549, 35270550, 35270551, 35288235, 35346019, 35362977, 35362978, 35362979, 35377871, 35408792, 35424159, 35424160, 35424161, 35424162, 35424163, 35424164, 35424165, 35424166, 35424167, 35424168, 35452838, 35452839, 35452840, 35452841, 35452842, 35452844, 35452845, 35452846, 35452860, 35469945, 35494575, 35566908, 35566911, 35601550, 35601551, 35602423, 35616846, 35616849, 35616850, 35616856, 35616857, 35616858, 35616862, 35711122, 35711123, 35711134, 35711141, 35711144, 35711145, 35711148, 35737235, 35737237, 35737239, 35892100, 35892107, 35892111, 35892112, 35892113, 35893568, 35893581, 35893582, 35893583, 35893584, 35893585, 35893586, 35893587, 35893588, 35893589, 35893590, 35957829, 35957832, 35957884, 35957885, 35957886, 35957892, 35957893, 35957894, 35957895, 35957897, 35957924, 35957926, 35957968, 35957969, 35972399, 35983054, 35983055, 35983056, 35983059, 35983060, 35983061, 35983062, 35983063, 35983064, 35983065, 35983066, 35983067, 35983068, 35983069, 35983070, 35983071, 35983072, 35983073, 35983074, 35983075, 35987947, 36121408, 36121410, 36121414, 36121428, 36234171, 36235067, 36307120, 36307121, 36307122, 36307130, 36328427, 36328429, 36328430, 36328431, 36328434, 36328435, 36263073, 35506587, 35506593, 35506594, 35506595, 35506596, 35506597, 35506598, 35506599, 35506600, 35506601, 35506602, 35506603, 35506604, 35506605, 35506606, 35506607, 35506608, 35506609, 35506610, 35506611, 35506612, 35506613, 35506614, 35506615, 35506616, 35506617, 35506618, 35506619, 35506620, 35506621, 35647495, 35647496, 35647497, 35647498, 35647499, 35647500, 35647501, 35647502, 35647503, 35647504, 35719649, 35719650, 35719651, 35719652, 35719653, 35719844, 35719845, 35719846, 35719847, 35719848, 35719849, 35756648, 35756649, 35756650, 35756651, 35894094, 35894095, 35894096, 35894097, 35894098, 35894099, 35894100, 35894101, 35894102, 35894103, 35894104, 35894105, 35894106, 35894107, 35957977, 35957983, 35957984, 35957985, 35957986, 36203676, 36203677, 36203678, 36203679, 36203680, 36203681, 36203682, 36203684, 36203685, 36203686, 36203687, 36203688, 36203689, 36203690, 36203691, 36203698, 36203699, 36203700, 36203701, 36203702, 36235080, 36235081, 36235082, 36235083, 36235084, 36235085, 36235087, 36235088, 36235089, 36235090, 36257929, 36257932, 36257933, 36257934, 36257935, 36257937, 36257938, 36257939, 36257940, 36257941, 36257943, 36257945, 36257946, 36257947, 36257949, 36257951, 36257952, 36257953, 36257954, 36257955, 36257956, 36257957, 36257959, 36257960, 36257961, 36257962, 36257963, 36257965, 36257966, 36257967, 36257968, 36257969, 36257973, 36257975, 36257976, 35401334, 36263074, 36263075, 36263076, 36263077, 35247955, 35247957, 35247959, 35247963, 35247965, 35247967, 35247969, 35295014, 35295018, 35295020, 35295022, 35295024, 35295026, 35295028, 35295031, 35295033, 35295035, 35295803, 35295807, 35295812, 35295818, 35295822, 35295826, 35295832, 35295837, 35295844, 35295850, 35295855, 35295858, 35295863, 35295866, 35295870, 35295875, 35295879, 35295882, 35295883, 35295886, 35295887, 35295888, 35295889, 35295890, 35295891, 35295892, 35295893, 35295894, 35295895, 35295896, 35295897, 35295898, 35295899, 35295900, 35295902, 35295903, 35295904, 35295905, 35295906, 35295907, 35295908, 35295916, 35295917, 35295918, 35295926, 35295927, 35295928, 35295929, 35295931, 35295932, 35295933, 35295939, 35295940, 35295941, 35295950, 35295951, 35295957, 35295958, 35295959, 35295964, 35402024, 35402025, 35402026, 35402027, 35402028, 35402029, 35402030, 35402031, 35402032, 35402033, 35402034, 35402035, 35402036, 35402037, 35402038, 35402039, 35402040, 35402041, 35402042, 35402043, 35402044, 35402045, 35402046, 35402047, 35402048, 35414627, 35414628, 35414629, 35414630, 35414631, 35414632, 35414633, 35414634, 35414635, 35414636, 35414637, 35414638, 35414639, 35414640, 35414641, 35414642, 35414643, 35414644, 35414645, 35414646, 35414647, 35414648, 35414649, 35414650, 35414651, 35414652, 35414653, 35414654, 35414655, 35414656, 35414657, 35414658, 35414659, 35414660, 35414661, 35414662, 35414663, 35414664, 35414665, 35414666, 35414667, 35414668, 35414669, 35414670, 35414671, 35414672, 35414673, 35414674, 35414675, 35414676, 35414677, 35414678, 35414679, 35414680, 35414681, 35414682, 35414683, 35414684, 35414685, 35414686, 35414687, 35414688, 35414689, 35414690, 35414691, 35414692, 35414693, 35414694, 35414695, 35414696, 35414697, 35414698, 35415274, 35415275, 35415276, 35415277, 35415278, 35415279, 35415280, 35415281, 35415282, 35415283, 35415284, 35415285, 35415286, 35415287, 35415288, 35415289, 35415290, 35415291, 35415292, 35415293, 35415294, 35415295, 35415296, 35415297, 35415298, 35415300, 35415301, 35415302, 35415303, 35415304, 35415305, 35415306, 35415307, 35415308, 35415309, 35415984, 35415985, 35415986, 35415987, 35415988, 35415989, 35415990, 35415991, 35415992, 35415993, 35415994, 35415995, 35415996, 35415997, 35415998, 35415999, 35416000, 35416001, 35601469, 35601470, 35616859, 35616860, 35730506, 35730507, 35730508, 35730509, 35730510, 35730511, 35730512, 35730513, 35730514, 35730515, 35730516, 35730517, 35730518, 35730519, 35730520, 35730521, 35730522, 35730523, 35730524, 35730525, 35730526, 35730527, 35730528, 35730529, 35730530, 35730531, 35730532, 35730533, 35730534, 35730535, 35730536, 35730537, 35730538, 35730539, 35730540, 35730541, 35730542, 35730543, 35730544, 35730545, 35730546, 35730547, 35730548, 35730549, 35730550, 35730551, 35730552, 35730553, 35730554, 35730555, 35730556, 35730557, 35730558, 35730559, 35730560, 35730561, 35730562, 35730563, 35730564, 35730565, 35730566, 35730567, 35730568, 35730569, 35730570, 35730571, 35730572, 35730573, 35730574, 35730575, 35730576, 35730577, 35730578, 35730579, 35730580, 35730581, 35730582, 35730583, 35730584, 35730585, 35730586, 35730587, 35730588, 35730589, 35730590, 35730591, 35730592, 35730593, 35730594, 35730595, 35730596, 35730597, 35730598, 35730599, 35730600, 35730601, 35730602, 35730603, 35730604, 35730605, 35730606, 35730607, 35730608, 35730609, 35730610, 35730611, 35730612, 35730613, 35730614, 35730615, 35730616, 35730617, 35730618, 35730619, 35730620, 35730621, 35730622, 35730623, 35730624, 35730625, 35730626, 35730627, 35730628, 35730629, 35730630, 35730631, 35730632, 35730633, 35730635, 35730637, 35730641, 35730642, 35730643, 35730644, 35730645, 35730646, 35730647, 35730648, 35730649, 35730650, 35730651, 35730652, 35730653, 35730654, 35730655, 35730656, 35730657, 35730658, 35730659, 35730660, 35737251, 35737252, 35737253, 35737254, 35737255, 35737256, 35737257, 35737258, 35737259, 35737260, 35892105, 35892356, 35993705, 35993706, 35993707, 35993708, 35993709, 35993710, 35993711, 35993712, 35993713, 35993714, 36210042, 36210043, 36210044, 36210045, 36210047, 36210048, 36210049, 36210050, 36210051, 36210052, 36210053, 36210054, 36210055, 36210056, 36210057, 36210058, 36210059, 36210060, 36210061, 36210062, 36210063, 36210064, 36210065, 36210066, 36210067, 36210068, 36210069, 36210070, 36210071, 36210072, 36210073, 36210074, 36210075, 36210077, 36210078, 36210079, 36210080, 36210081, 36210082, 36210083, 36210084, 36210085, 36210086, 36210087, 36210088, 36210089, 36210090, 36210091, 36210092, 36210093, 36210094, 36210095, 36210096, 36210097, 36210098, 36210099, 36210100, 36210102, 36210103, 36210104, 36210105, 36210106, 36210107, 36210108, 36210109, 36210110, 36210111, 36210112, 36210113, 36210114, 36210115, 36210116, 36210117, 36210118, 36210119, 36210120, 36210121, 36210122, 36210123, 36210124, 36210125, 36210126, 36210127, 36210128, 36210129, 36210130, 36210131, 36210132, 36210133, 36210135, 36210136, 36210137, 36210138, 36210139, 36210140, 36210141, 36210142, 36210143, 36210144, 36210145, 36223650, 36223651, 36223652, 36223654, 36223655, 36223656, 36223657, 36223658, 36223659, 36245049, 36245050, 36245051, 36245052, 36245053, 36245054, 36245055, 36245057, 36245058, 36245059, 36245061, 36245062, 36245063, 36245064, 36245065, 36278623, 36278624, 36278625, 36278626, 36278627, 36278628, 36278629, 36278630, 36278631, 36278632, 36278633, 36278634, 36278635, 36278636, 36278637, 36278638, 36278639, 36278640, 36278641, 36278642, 36278643, 36278644, 36278645, 36278646, 36278647, 36278648, 36278649, 36278650, 36278651, 36278652, 36278653, 36278654, 36278655, 36278656, 36278657, 36278658, 36278659, 36278660, 36278661, 36278662, 36278663, 36278664, 36278665, 36278666, 36278668, 36278673, 36278676, 36278677, 36278678, 36278679, 36278680, 36278681, 36278682, 36278683, 36278684, 36278685, 36278686, 36278687, 36278688, 36278689, 36278690, 36278691, 36278692, 36278693, 36278694, 36278695, 36278696, 36278697, 36278698, 36278699, 36278700, 36278701, 36278702, 36278703, 36278704, 36278705, 36278707, 36278708, 36278709, 36278710, 36278711, 36278713, 36278714, 36278715, 36278716, 36278717, 36278718, 36278719, 36278720, 36278721, 36278722, 36278723, 36278724, 36278725, 36278726, 36278727, 36278728, 36278729, 36278730, 36278731, 36278732, 36278733, 36278734, 36278735, 36278736, 36278737, 36278738, 36278739, 36278740, 36278741, 36278742, 36278743, 36278744, 36278745, 36278746, 36278747, 36278748, 36278749, 36278750, 36278751, 36278752, 36278753, 36278754, 36278755, 36278756, 36278757, 36278758, 36278759, 36278760, 36278761, 36278762, 36278763, 36278764, 36278766, 36278767, 36278768, 36278769, 36278770, 36278771, 36278772, 36278773, 36278774, 36278775, 36278777, 36278778, 36278779, 36278780, 36278781, 36278782, 36278783, 36278785, 36278786, 36278787, 36278789, 36278790, 36278791, 36278792, 36278793, 36278794, 36278796, 36278797, 36278798, 36278799, 36278800, 36278802, 36278803, 36278804, 36278805, 36278806, 36278807, 36278808, 36278809, 36278810, 36278811, 36278812, 36278813, 36278814, 36278815, 36278816, 36278817, 36278818, 36278819, 36278820, 36278821, 36278822, 36278823, 36278824, 36278825, 36278826, 36278827, 36278828, 36278829, 36278830, 36278831, 36278832, 36278833, 36278834, 36278835, 36278836, 36278837, 36328444, 36328445, 36328446, 36328447, and 36328448.
FEI Number 3015876
Recalling Firm/
Manufacturer
Physio Control, Inc.
11811 Willows Rd Ne
Redmond WA 98052-2003
For Additional Information Contact
425-867-4000 Ext. 4644
Manufacturer Reason
for Recall
The device may not power on although it indicates it is ready for use and would not be able to provide defibrillation therapy.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionOn 07/24/08, the firm sent an "URGENT MEDICAL DEVICE RECALL" letter with a confirmation sheet via certified return receipt to customers. The letter notified the customer of the recall and the hazard associated with the recall and provided instructions to customers: 1) Immediately verify the AED powers on and voice prompts begin. After verification, press and hold the On/Off button for approximately two seconds to turn the AED off. 2) Keep the AED in use and perform monthly inspections that include the additional AED power on test. Perform a monthly inspection which is consistent with the device labeling. 3) If "OK" is visible on the Readiness Display and AED powers on, it is ready for use. 4) If, at any time, the AED does not power on or if any other indicator displays (i.e. battery or CHARGE-PAK symbol, attention symbol or wrench symbol) immediately call Physio-Control Technical Support (1-800-442-1142). The letter advises customers a Physio-Control representative will contact them within 60 days to arrange a replacement. A confirmation sheet was sent with the letter; to be returned to Physio-Control, Inc. For specific questions or additional information, contact Physio-Control, Inc. at 1-800-442-1142 weekdays between 6:00 a.m. and 4:00 p.m. (PST) or at www.physio-control.com.
Quantity in Commerce2,230 Units
DistributionWorldwide including USA (and Puerto Rico) and countries of Argentina, Australia, Bahamas, Barbados, Brazil, Canada , Chile, Colombia, Dominica, Germany, Hong Kong, Hungary, Jamaica, Japan, Mexico, Netherlands, New Zealand, Singapore, Uruguay, Venezuela, and Virgin Islands.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = MKJ
-
-