| | Class 2 Device Recall Exact Couch with Exact Couch Top |  |
| Date Initiated by Firm | August 11, 2008 |
| Date Posted | October 20, 2008 |
| Recall Status1 |
Terminated 3 on November 27, 2010 |
| Recall Number | Z-0117-2009 |
| Recall Event ID |
49399 |
| 510(K)Number | K033339 K904364 |
| Product Classification |
Linear Accelerator - Product Code IYE
|
| Product | Exact Couch with Exact Couch Top, a component of the C-Series Clinacs, Acuity and Ximatron treatment systems. Model numbers H14, H18, H27, H72, H77. A powered radiation therapy patient support assembly is an electrically powered adjustable couch intended to support a patient during radiation therapy. |
| Code Information |
Serial numbers: H140001 H140044 H140086 H180113 H180439 H180505 H180544 H180586 H180631 H140002 H140045 H140087 H180114 H180446 H180506 H180545 H180587 H180632 H140003 H140046 H140088 H180115 H180448 H180507 H180546 H180588 H180633 H140004 H140047 H140089 H180130 H180449 H180508 H180547 H180590 H180634 H140005 H140048 H140090 H180132 H180452 H180509 H180548 H180591 H180635 H140006 H140049 H140091 H180136 H180468 H180510 H180549 H180592 H180636 H140007 H140050 H140092 H180138 H180469 H180511 H180550 H180593 H180637 H140008 H140051 H140093 H180141 H180470 H180512 H180551 H180594 H180638 H140009 H140052 H140094 H180147 H180471 H180513 H180552 H180595 H180639 H140010 H140053 H140095 H180157 H180472 H180514 H180553 H180596 H180640 H140011 H140054 H140096 H180167 H180473 H180515 H180554 H180597 H180641 H140012 H140055 H140097 H180170 H180474 H180517 H180555 H180598 H180642 H140014 H140056 H140098 H180172 H180475 H180518 H180556 H180599 H180643 H140015 H140057 H140099 H180173 H180476 H180519 H180557 H180600 H180644 H140016 H140058 H140100 H180183 H180477 H180520 H180558 H180601 H180645 H140017 H140059 H140101 H180186 H180478 H180521 H180559 H180602 H180646 H140018 H140060 H140102 H180197 H180479 H180522 H180560 H180603 H180647 H140019 H140061 H140103 H180204 H180480 H180522 H180561 H180604 H180648 H140020 H140062 H140105 H180207 H180482 H180523 H180561 H180605 H180649 H140021 H140063 H140106 H180208 H180483 H180524 H180562 H180606 H180650 H140022 H140064 H140107 H180210 H180484 H180525 H180563 H180607 H180651 H140023 H140065 H140108 H180222 H180485 H180526 H180564 H180610 H180652 H140024 H140066 H140109 H180231 H180486 H180527 H180565 H180611 H180653 H140025 H140067 H140110 H180242 H180487 H180528 H180566 H180612 H180654 H140026 H140068 H140111 H180248 H180488 H180529 H180568 H180613 H180655 H140027 H140069 H140112 H180256 H180489 H180529 H180569 H180614 H180656 H140028 H140070 H140113 H180260 H180490 H180530 H180570 H180615 H180657 H140029 H140071 H140114 H180283 H180491 H180531 H180571 H180616 H180658 H140030 H140072 H140115 H180285 H180491 H180532 H180572 H180617 H180659 H140031 H140073 H180007 H180292 H180492 H180533 H180573 H180618 H180660 H140032 H140074 H180018 H180298 H180493 H180534 H180574 H180619 H180661 H140033 H140075 H180029 H180300 H180494 H180535 H180575 H180620 H180662 H140034 H140076 H180045 H180309 H180494 H180536 H180576 H180621 H180663 H140035 H140077 H180053 H180313 H180495 H180537 H180577 H180622 H180664 H140036 H140078 H180058 H180404 H180496 H180538 H180578 H180623 H180665 H140037 H140079 H180064 H180409 H180498 H180539 H180579 H180624 H180667 H140038 H140080 H180066 H180412 H180499 H180539 H180580 H180625 H180668 H140039 H140081 H180073 H180413 H180500 H180540 H180581 H180626 H180669 H140040 H140082 H180082 H180425 H180501 H180541 H180582 H180627 H180670 H140041 H140083 H180087 H180426 H180502 H180542 H180583 H180628 H180671 H140042 H140084 H180102 H180427 H180503 H180543 H180584 H180629 H180672 H140043 H140085 H180104 H180436 H180504 H180543 H180585 H180630 H180673 H180674 H180716 H180758 H180799 H180841 H270243 H270439 H270667 H270808 H180675 H180717 H180759 H180800 H180842 H270244 H270449 H270668 H270817 H180676 H180718 H180760 H180801 H180843 H270251 H270461 H270671 H270822 H180677 H180719 H180761 H180802 H180844 H270260 H270463 H270672 H270826 H180678 H180720 H180762 H180803 H180845 H270264 H270477 H270675 H270827 H180679 H180721 H180763 H180804 H180846 H270269 H270485 H270679 H270829 H180680 H180722 H180764 H180805 H180847 H270280 H270489 H270685 H270830 H180681 H180723 H180765 H180806 H230004 H270281 H270491 H270687 H270833 H180682 H180724 H180766 H180807 H230097 H270296 H270496 H270689 H270836 H180683 H180725 H180767 H180808 H230143 H270299 H270499 H270690 H270837 H180684 H180726 H180768 H180809 H230205 H270302 H270501 H270692 H270840 H180685 H180727 H180769 H180810 H230250 H270311 H270503 H270695 H270842 H180686 H180728 H180770 H180811 H260902 H270316 H270504 H270703 H270843 H180687 H180729 H180771 H180812 H260905 H270321 H270505 H270710 H270844 H180688 H180730 H180772 H180813 H260913 H270322 H270507 H270717 H270846 H180689 H180731 H180773 H180814 H260924 H270323 H270513 H270718 H270848 H180690 H180732 H180774 H180815 H260928 H270327 H270514 H270719 H270852 H180691 H180733 H180775 H180816 H260947 H270338 H270521 H270721 H270855 H180692 H180734 H180776 H180817 H270004 H270345 H270522 H270726 H270857 H180693 H180735 H180777 H180818 H270006 H270347 H270524 H270727 H270861 H180694 H180736 H180778 H180819 H270009 H270349 H270532 H270728 H270862 H180695 H180737 H180779 H180820 H270010 H270359 H270535 H270735 H270863 H180696 H180738 H180780 H180821 H270056 H270362 H270537 H270737 H270868 H180697 H180739 H180781 H180822 H270058 H270369 H270541 H270739 H270875 H180698 H180740 H180782 H180823 H270060 H270370 H270548 H270740 H270880 H180699 H180741 H180783 H180824 H270073 H270378 H270549 H270746 H270882 H180700 H180742 H180784 H180825 H270084 H270379 H270600 H270749 H270883 H180701 H180743 H180785 H180826 H270092 H270386 H270602 H270757 H270884 H180702 H180744 H180786 H180827 H270094 H270391 H270603 H270761 H270888 H180703 H180745 H180787 H180828 H270096 H270392 H270604 H270762 H270890 H180704 H180746 H180788 H180829 H270106 H270393 H270614 H270767 H270894 H180705 H180747 H180789 H180830 H270108 H270396 H270616 H270775 H271000 H180706 H180748 H180790 H180831 H270126 H270399 H270635 H270780 H271002 H180707 H180749 H180791 H180832 H270136 H270401 H270639 H270781 H271005 H180708 H180750 H180792 H180833 H270138 H270407 H270640 H270784 H271006 H180709 H180751 H180793 H180834 H270201 H270408 H270644 H270785 H271007 H180710 H180752 H180794 H180835 H270206 H270416 H270646 H270787 H271011 H180711 H180753 H180795 H180836 H270214 H270421 H270647 H270788 H271016 H180712 H180754 H180795 H180837 H270219 H270430 H270649 H270797 H271017 H180713 H180755 H180796 H180838 H270227 H270431 H270652 H270801 H271018 H180714 H180756 H180797 H180839 H270228 H270437 H270655 H270803 H271021 H180715 H180757 H180798 H180840 H270241 H270438 H270656 H270804 H271023 H271024 H271108 H271151 H271197 H271240 H271282 H271324 H271365 H271409 H271027 H271109 H271152 H271198 H271241 H271283 H271325 H271367 H271410 H271029 H271110 H271153 H271199 H271242 H271284 H271326 H271368 H271411 H271030 H271111 H271154 H271200 H271243 H271285 H271327 H271369 H271412 H271033 H271112 H271155 H271201 H271244 H271286 H271328 H271370 H271413 H271041 H271113 H271157 H271202 H271245 H271287 H271329 H271371 H271414 H271042 H271114 H271158 H271203 H271246 H271288 H271330 H271372 H271415 H271044 H271115 H271159 H271204 H271247 H271289 H271331 H271373 H271416 H271045 H271116 H271160 H271205 H271248 H271290 H271332 H271374 H271417 H271050 H271117 H271161 H271206 H271249 H271291 H271333 H271375 H271418 H271056 H271118 H271162 H271207 H271250 H271292 H271334 H271376 H271419 H271057 H271119 H271163 H271208 H271251 H271293 H271335 H271377 H271420 H271061 H271120 H271164 H271209 H271252 H271294 H271336 H271378 H271421 H271061 H271121 H271165 H271210 H271253 H271295 H271337 H271379 H271422 H271065 H271122 H271166 H271212 H271254 H271296 H271338 H271380 H271423 H271066 H271123 H271167 H271213 H271255 H271297 H271339 H271381 H271424 H271070 H271124 H271168 H271214 H271256 H271298 H271339 H271382 H271425 H271071 H271125 H271169 H271215 H271257 H271299 H271340 H271383 H271426 H271072 H271126 H271170 H271216 H271258 H271300 H271341 H271384 H271427 H271074 H271127 H271171 H271217 H271259 H271301 H271342 H271385 H271428 H271075 H271128 H271172 H271218 H271260 H271302 H271343 H271386 H271429 H271080 H271129 H271173 H271219 H271261 H271303 H271344 H271387 H271430 H271081 H271130 H271174 H271220 H271262 H271304 H271345 H271388 H271431 H271082 H271131 H271175 H271222 H271263 H271305 H271346 H271389 H271432 H271083 H271132 H271176 H271223 H271264 H271306 H271347 H271390 H271433 H271084 H271133 H271177 H271224 H271265 H271307 H271348 H271391 H271434 H271085 H271135 H271178 H271225 H271266 H271308 H271349 H271392 H271435 H271093 H271136 H271179 H271226 H271267 H271309 H271350 H271393 H271436 H271094 H271137 H271182 H271227 H271268 H271310 H271351 H271394 H271437 H271095 H271138 H271183 H271228 H271269 H271311 H271352 H271395 H271438 H271096 H271139 H271184 H271229 H271270 H271312 H271353 H271396 H271439 H271097 H271140 H271185 H271230 H271271 H271313 H271354 H271397 H271440 H271098 H271141 H271186 H271231 H271272 H271314 H271355 H271398 H271441 H271099 H271142 H271188 H271232 H271273 H271315 H271356 H271399 H271442 H271100 H271143 H271189 H271233 H271274 H271316 H271357 H271400 H271443 H271101 H271144 H271190 H271234 H271275 H271317 H271358 H271401 H271444 H271102 H271145 H271191 H271234 H271276 H271318 H271359 H271402 H271445 H271103 H271146 H271192 H271235 H271277 H271319 H271360 H271403 H271446 H271104 H271147 H271193 H271236 H271278 H271320 H271361 H271404 H271447 H271105 H271148 H271194 H271237 H271279 H271321 H271362 H271406 H271448 H271106 H271149 H271195 H271238 H271280 H271322 H271363 H271407 H271449 H271107 H271150 H271196 H271239 H271281 H271323 H271364 H271408 H271450 H271451 H271494 H271535 H271577 H271621 H271664 H271707 H271749 H271791 H271452 H271494 H271536 H271578 H271622 H271665 H271708 H271750 H271792 H271453 H271495 H271537 H271579 H271623 H271666 H271709 H271751 H271793 H271454 H271496 H271538 H271580 H271624 H271667 H271710 H271752 H271794 H271455 H271497 H271539 H271581 H271625 H271668 H271711 H271753 H271795 H271456 H271498 H271540 H271582 H271626 H271669 H271712 H271754 H271796 H271457 H271499 H271541 H271583 H271627 H271670 H271713 H271755 H271797 H271458 H271500 H271542 H271584 H271628 H271671 H271714 H271756 H271798 H271459 H271501 H271543 H271585 H271629 H271672 H271715 H271757 H271799 H271460 H271502 H271544 H271586 H271630 H271673 H271716 H271758 H271800 H271461 H271503 H271545 H271587 H271631 H271674 H271717 H271759 H271801 H271462 H271504 H271546 H271588 H271632 H271675 H271718 H271760 H271802 H271463 H271505 H271547 H271589 H271634 H271676 H271719 H271761 H271803 H271464 H271506 H271548 H271590 H271635 H271677 H271720 H271762 H271804 H271465 H271507 H271549 H271591 H271636 H271678 H271721 H271763 H271805 H271466 H271508 H271550 H271592 H271637 H271679 H271722 H271764 H271806 H271467 H271509 H271551 H271593 H271638 H271680 H271723 H271765 H271807 H271468 H271510 H271552 H271594 H271639 H271681 H271724 H271766 H271808 H271469 H271511 H271553 H271595 H271640 H271682 H271725 H271767 H271809 H271470 H271512 H271554 H271596 H271641 H271683 H271726 H271768 H271810 H271471 H271513 H271555 H271597 H271642 H271684 H271727 H271769 H271811 H271472 H271514 H271556 H271598 H271643 H271685 H271728 H271770 H271812 H271473 H271515 H271557 H271599 H271644 H271686 H271729 H271771 H271813 H271474 H271516 H271558 H271600 H271645 H271687 H271730 H271772 H271814 H271475 H271517 H271559 H271601 H271646 H271688 H271731 H271773 H271815 H271476 H271518 H271560 H271602 H271647 H271689 H271732 H271774 H271816 H271477 H271519 H271561 H271603 H271648 H271690 H271733 H271775 H271817 H271478 H271520 H271562 H271604 H271649 H271691 H271734 H271776 H271818 H271479 H271521 H271563 H271605 H271650 H271692 H271735 H271777 H271819 H271480 H271522 H271564 H271606 H271651 H271693 H271736 H271778 H271820 H271481 H271523 H271565 H271607 H271652 H271694 H271737 H271779 H271821 H271482 H271524 H271566 H271608 H271653 H271695 H271738 H271780 H271822 H271483 H271525 H271567 H271610 H271654 H271697 H271739 H271781 H271823 H271484 H271526 H271568 H271611 H271655 H271698 H271740 H271782 H271824 H271485 H271527 H271569 H271612 H271656 H271699 H271741 H271783 H271825 H271486 H271528 H271570 H271613 H271657 H271700 H271742 H271784 H271826 H271487 H271529 H271571 H271614 H271658 H271701 H271743 H271785 H271827 H271488 H271530 H271572 H271615 H271659 H271702 H271744 H271786 H271828 H271490 H271531 H271573 H271616 H271660 H271703 H271745 H271787 H271829 H271491 H271532 H271574 H271618 H271661 H271704 H271746 H271788 H271830 H271492 H271533 H271575 H271619 H271662 H271705 H271747 H271789 H271831 H271493 H271534 H271576 H271620 H271663 H271706 H271748 H271790 H271832 H271833 H271875 H271917 H271959 H272001 H272044 H272086 H272128 H272170 H271834 H271876 H271918 H271960 H272002 H272045 H272087 H272129 H272171 H271835 H271877 H271919 H271961 H272003 H272046 H272088 H272130 H272172 H271836 H271878 H271920 H271962 H272004 H272047 H272089 H272131 H272173 H271837 H271879 H271921 H271963 H272005 H272048 H272090 H272132 H272174 H271838 H271880 H271922 H271964 H272006 H272049 H272091 H272133 H272175 H271839 H271881 H271923 H271965 H272007 H272050 H272092 H272134 H272176 H271840 H271882 H271924 H271966 H272008 H272051 H272093 H272135 H272177 H271841 H271883 H271925 H271967 H272009 H272052 H272094 H272136 H272178 H271842 H271884 H271926 H271968 H272010 H272053 H272095 H272137 H272179 H271843 H271885 H271927 H271969 H272011 H272054 H272096 H272138 H272180 H271844 H271886 H271928 H271970 H272012 H272055 H272097 H272139 H272181 H271845 H271887 H271929 H271971 H272013 H272056 H272098 H272140 H272182 H271846 H271888 H271930 H271972 H272014 H272057 H272099 H272141 H272183 H271847 H271889 H271931 H271973 H272015 H272058 H272100 H272142 H272184 H271848 H271890 H271932 H271974 H272016 H272059 H272101 H272143 H272185 H271849 H271891 H271933 H271975 H272017 H272060 H272102 H272144 H272186 H271850 H271892 H271934 H271976 H272018 H272061 H272103 H272145 H272187 H271851 H271893 H271935 H271977 H272019 H272062 H272104 H272146 H272188 H271852 H271894 H271936 H271978 H272020 H272063 H272105 H272147 H272189 H271853 H271895 H271937 H271979 H272021 H272064 H272106 H272148 H272190 H271854 H271896 H271938 H271980 H272022 H272065 H272107 H272149 H272191 H271855 H271897 H271939 H271981 H272023 H272066 H272108 H272150 H272192 H271856 H271898 H271940 H271982 H272024 H272067 H272109 H272151 H272193 H271857 H271899 H271941 H271983 H272025 H272068 H272110 H272152 H272194 H271858 H271900 H271942 H271984 H272026 H272069 H272111 H272153 H272195 H271859 H271901 H271943 H271985 H272027 H272070 H272112 H272154 H272196 H271860 H271902 H271944 H271986 H272028 H272071 H272113 H272155 H272197 H271861 H271903 H271945 H271987 H272029 H272072 H272114 H272156 H272198 H271862 H271904 H271946 H271988 H272030 H272073 H272115 H272157 H272199 H271863 H271905 H271947 H271989 H272031 H272074 H272116 H272158 H272200 H271864 H271906 H271948 H271990 H272032 H272075 H272117 H272159 H272201 H271865 H271907 H271949 H271991 H272033 H272076 H272118 H272160 H272202 H271866 H271908 H271950 H271992 H272034 H272077 H272119 H272161 H272203 H271867 H271909 H271951 H271993 H272035 H272078 H272120 H272162 H272204 H271868 H271910 H271952 H271994 H272036 H272079 H272121 H272163 H272205 H271869 H271911 H271953 H271995 H272037 H272080 H272122 H272164 H272206 H271870 H271912 H271954 H271996 H272038 H272081 H272123 H272165 H272207 H271871 H271913 H271955 H271997 H272039 H272082 H272124 H272166 H272208 H271872 H271914 H271956 H271998 H272040 H272083 H272125 H272167 H272209 H271873 H271915 H271957 H271999 H272041 H272084 H272126 H272168 H272210 H271874 H271916 H271958 H272000 H272043 H272085 H272127 H272169 H272211 H272212 H272254 H272296 H272338 H272380 H272422 H272464 H272810 H272854 H272213 H272255 H272297 H272339 H272381 H272423 H272465 H272811 H272855 H272214 H272256 H272298 H272340 H272382 H272424 H272466 H272812 H272856 H272215 H272257 H272299 H272341 H272383 H272425 H272467 H272813 H272857 H272216 H272258 H272300 H272342 H272384 H272426 H272468 H272814 H272858 H272217 H272259 H272301 H272343 H272385 H272427 H272469 H272815 H272859 H272218 H272260 H272302 H272344 H272386 H272428 H272470 H272816 H272860 H272219 H272261 H272303 H272345 H272387 H272429 H272471 H272817 H272861 H272220 H272262 H272304 H272346 H272388 H272430 H272472 H272819 H272862 H272221 H272263 H272305 H272347 H272389 H272431 H272473 H272820 H272863 H272222 H272264 H272306 H272348 H272390 H272432 H272475 H272821 H272864 H272223 H272265 H272307 H272349 H272391 H272433 H272476 H272822 H272865 H272224 H272266 H272308 H272350 H272392 H272434 H272477 H272823 H272866 H272225 H272267 H272309 H272351 H272393 H272435 H272478 H272824 H272867 H272226 H272268 H272310 H272352 H272394 H272436 H272479 H272825 H272868 H272227 H272269 H272311 H272353 H272395 H272437 H272480 H272826 H272869 H272228 H272270 H272312 H272354 H272396 H272438 H272482 H272827 H272870 H272229 H272271 H272313 H272355 H272397 H272439 H272483 H272828 H272871 H272230 H272272 H272314 H272356 H272398 H272440 H272484 H272829 H272872 H272231 H272273 H272315 H272357 H272399 H272441 H272485 H272830 H272873 H272232 H272274 H272316 H272358 H272400 H272442 H272488 H272831 H272875 H272233 H272275 H272317 H272359 H272401 H272443 H272489 H272833 H272876 H272234 H272276 H272318 H272360 H272402 H272444 H272490 H272834 H272877 H272235 H272277 H272319 H272361 H272403 H272445 H272491 H272835 H272878 H272236 H272278 H272320 H272362 H272404 H272446 H272492 H272836 H272879 H272237 H272279 H272321 H272363 H272405 H272447 H272493 H272837 H272880 H272238 H272280 H272322 H272364 H272406 H272448 H272494 H272838 H272881 H272239 H272281 H272323 H272365 H272407 H272449 H272495 H272839 H272882 H272240 H272282 H272324 H272366 H272408 H272450 H272496 H272840 H272883 H272241 H272283 H272325 H272367 H272409 H272451 H272497 H272841 H272884 H272242 H272284 H272326 H272368 H272410 H272452 H272498 H272842 H272885 H272243 H272285 H272327 H272369 H272411 H272453 H272500 H272843 H272886 H272244 H272286 H272328 H272370 H272412 H272454 H272800 H272844 H272887 H272245 H272287 H272329 H272371 H272413 H272455 H272801 H272845 H272888 H272246 H272288 H272330 H272372 H272414 H272456 H272802 H272846 H272889 H272247 H272289 H272331 H272373 H272415 H272457 H272803 H272847 H272890 H272248 H272290 H272332 H272374 H272416 H272458 H272804 H272848 H272891 H272249 H272291 H272333 H272375 H272417 H272459 H272805 H272849 H272892 H272250 H272292 H272334 H272376 H272418 H272460 H272806 H272850 H272893 H272251 H272293 H272335 H272377 H272419 H272461 H272807 H272851 H272894 H272252 H272294 H272336 H272378 H272420 H272462 H272808 H272852 H272895 H272253 H272295 H272337 H272379 H272421 H272463 H272809 H272853 H272896 H272897 H272939 H290142 H290195 H290234 H290274 H290316 H290358 H290399 H272898 H272940 H290145 H290196 H290235 H290275 H290317 H290359 H290400 H272899 H272941 H290147 H290197 H290236 H290276 H290318 H290360 H290401 H272900 H272942 H290148 H290198 H290237 H290277 H290319 H290361 H290402 H272901 H272943 H290153 H290199 H290238 H290278 H290320 H290362 H290403 H272902 H272944 H290154 H290200 H290239 H290279 H290321 H290363 H290404 H272903 H272946 H290155 H290201 H290240 H290280 H290322 H290364 H290405 H272904 H272947 H290156 H290202 H290241 H290281 H290323 H290365 H290406 H272905 H272948 H290157 H290203 H290242 H290282 H290324 H290366 H290407 H272906 H272949 H290161 H290203 H290243 H290283 H290325 H290367 H290408 H272907 H272950 H290163 H290204 H290244 H290284 H290326 H290368 H290409 H272908 H272951 H290164 H290204 H290245 H290285 H290327 H290369 H290410 H272909 H272952 H290165 H290205 H290245 H290286 H290328 H290370 H290411 H272910 H272953 H290165 H290206 H290246 H290287 H290329 H290371 H290412 H272911 H290002 H290166 H290207 H290247 H290288 H290330 H290372 H290413 H272912 H290006 H290167 H290208 H290248 H290289 H290331 H290372 H290414 H272913 H290008 H290168 H290209 H290249 H290290 H290332 H290373 H290415 H272914 H290010 H290169 H290210 H290250 H290291 H290333 H290374 H290416 H272915 H290012 H290170 H290211 H290251 H290292 H290334 H290375 H290417 H272916 H290017 H290171 H290212 H290251 H290293 H290335 H290376 H290422 H272917 H290018 H290172 H290213 H290252 H290294 H290336 H290377 H290423 H272918 H290024 H290173 H290214 H290253 H290295 H290337 H290378 H290424 H272919 H290026 H290175 H290215 H290254 H290296 H290338 H290379 H290425 H272920 H290030 H290176 H290216 H290255 H290297 H290339 H290380 H290426 H272921 H290031 H290177 H290217 H290256 H290298 H290340 H290381 H290427 H272922 H290032 H290178 H290218 H290257 H290299 H290341 H290382 H290500 H272923 H290033 H290179 H290219 H290258 H290300 H290342 H290383 H290501 H272924 H290034 H290180 H290220 H290259 H290301 H290343 H290384 H290502 H272925 H290104 H290181 H290220 H290260 H290302 H290344 H290385 H290503 H272926 H290106 H290182 H290221 H290261 H290303 H290345 H290386 H290504 H272927 H290108 H290183 H290222 H290262 H290304 H290346 H290386 H290505 H272928 H290109 H290185 H290223 H290263 H290305 H290347 H290387 H290506 H272929 H290110 H290186 H290224 H290264 H290306 H290348 H290388 H290507 H272930 H290114 H290187 H290225 H290265 H290307 H290349 H290389 H290508 H272931 H290115 H290188 H290226 H290266 H290308 H290350 H290390 H290509 H272932 H290117 H290189 H290227 H290267 H290309 H290351 H290391 H290510 H272933 H290120 H290190 H290228 H290268 H290310 H290352 H290392 H290511 H272934 H290122 H290191 H290229 H290269 H290311 H290353 H290393 H290512 H272935 H290123 H290192 H290230 H290270 H290312 H290354 H290394 H290513 H272936 H290128 H290193 H290231 H290271 H290313 H290355 H290395 H290514 H272937 H290132 H290194 H290232 H290272 H290314 H290356 H290396 H290515 H272938 H290135 H290194 H290233 H290273 H290315 H290357 H290397 H290516 H290517 H290561 H720425 H720633 H720738 H720784 H720828 H720872 H720916 H290518 H290600 H720429 H720638 H720739 H720785 H720829 H720873 H720918 H290519 H720040 H720432 H720643 H720741 H720786 H720830 H720874 H720919 H290520 H720060 H720434 H720647 H720742 H720787 H720831 H720875 H720920 H290521 H720070 H720435 H720650 H720743 H720788 H720832 H720876 H720921 H290522 H720076 H720445 H720651 H720744 H720789 H720833 H720877 H720922 H290523 H720102 H720453 H720653 H720745 H720790 H720834 H720878 H720923 H290524 H720109 H720455 H720654 H720746 H720791 H720835 H720879 H720924 H290525 H720113 H720461 H720665 H720747 H720792 H720836 H720880 H720925 H290526 H720114 H720463 H720667 H720748 H720793 H720838 H720881 H720926 H290527 H720121 H720473 H720668 H720749 H720794 H720839 H720882 H720927 H290528 H720130 H720474 H720679 H720750 H720795 H720840 H720883 H720928 H290529 H720173 H720475 H720690 H720751 H720796 H720841 H720884 H720929 H290530 H720182 H720481 H720692 H720752 H720797 H720842 H720885 H720930 H290531 H720213 H720492 H720709 H720753 H720798 H720843 H720886 H720931 H290532 H720233 H720499 H720711 H720754 H720799 H720843 H720888 H720932 H290533 H720247 H720500 H720712 H720755 H720801 H720844 H720889 H720933 H290534 H720248 H720502 H720714 H720756 H720802 H720845 H720890 H720934 H290535 H720262 H720513 H720715 H720757 H720803 H720846 H720891 H720935 H290536 H720269 H720514 H720716 H720759 H720804 H720847 H720892 H720936 H290537 H720276 H720515 H720717 H720760 H720805 H720848 H720893 H720937 H290538 H720278 H720521 H720718 H720761 H720806 H720850 H720894 H720938 H290539 H720282 H720535 H720719 H720762 H720807 H720851 H720895 H720939 H290540 H720291 H720536 h720719 H720763 H720808 H720852 H720896 H720940 H290541 H720294 H720541 H720720 H720765 H720809 H720854 H720897 H720941 H290542 H720297 H720542 H720721 H720766 H720810 H720855 H720898 H720942 H290543 H720308 H720548 H720722 H720767 H720811 H720856 H720899 H720943 H290544 H720309 H720551 H720723 H720768 H720812 H720857 H720900 H720944 H290545 H720319 H720556 H720724 H720769 H720813 H720858 H720901 H720945 H290546 H720347 H720559 H720724 H720770 H720814 H720859 H720902 H720946 H290547 H720349 H720562 H720725 H720772 H720815 H720860 H720903 H720947 H290548 H720350 H720566 H720726 H720773 H720816 H720861 H720904 H720948 H290549 H720356 H720568 H720727 H720774 H720817 H720862 H720905 H720949 H290550 H720361 H720571 H720727 H720775 H720818 H720863 H720906 H720950 H290551 H720364 H720578 H720728 H720776 H720819 H720864 H720907 H720951 H290552 H720367 H720581 H720729 H720777 H720820 H720865 H720908 H720952 H290553 H720377 H720596 H720730 H720778 H720822 H720866 H720909 H720953 H290554 H720379 H720602 H720732 H720779 H720823 H720867 H720910 H720954 H290555 H720387 H720613 H720733 H720780 H720824 H720868 H720911 H720955 H290556 H720388 H720617 H720734 H720781 H720825 H720869 H720913 H720956 H290558 H720408 H720621 H720736 H720782 H720826 H720870 H720914 H720957 H290559 H720418 H720632 H720737 H720783 H720827 H720871 H720915 H720958 H720959 H721002 H721045 H770042 H770086 H770129 H770171 H720960 H721003 H721046 H770043 H770087 H770130 H770172 H720961 H721004 H721047 H770045 H770088 H770131 H770173 H720962 H721005 H721048 H770046 H770089 H770132 H770174 H720963 H721006 H770002 H770047 H770090 H770133 H770175 H720964 H721007 H770004 H770048 H770091 H770134 H770176 H720965 H721008 H770005 H770049 H770092 H770135 H770177 H720966 H721009 H770006 H770051 H770093 H770136 H770178 H720967 H721010 H770007 H770052 H770094 H770137 H770179 H720968 H721011 H770009 H770053 H770095 H770138 H770180 H720969 H721012 H770010 H770054 H770096 H770139 H770181 H720970 H721013 H770011 H770055 H770097 H770140 H770182 H720971 H721014 H770012 H770056 H770098 H770141 H770183 H720973 H721015 H770013 H770057 H770099 H770142 H770184 H720974 H721016 H770014 H770058 H770100 H770143 H770185 H720975 H721017 H770015 H770059 H770101 H770144 H770186 H720976 H721018 H770016 H770060 H770102 H770145 H770187 H720977 H721019 H770017 H770061 H770103 H770146 H770188 H720978 H721020 H770018 H770062 H770104 H770147 H770189 H720979 H721022 H770019 H770063 H770105 H770148 H770190 H720980 H721023 H770020 H770064 H770106 H770149 H720981 H721024 H770021 H770065 H770107 H770150 H720982 H721025 H770022 H770066 H770108 H770151 H720983 H721026 H770023 H770067 H770109 H770152 H720984 H721027 H770024 H770068 H770110 H770153 H720985 H721028 H770025 H770069 H770112 H770154 H720986 H721029 H770026 H770070 H770113 H770155 H720987 H721030 H770027 H770071 H770114 H770156 H720988 H721031 H770028 H770072 H770115 H770157 H720989 H721032 H770029 H770073 H770116 H770158 H720990 H721033 H770030 H770074 H770117 H770159 H720991 H721034 H770031 H770075 H770118 H770160 H720992 H721035 H770032 H770076 H770119 H770161 H720993 H721036 H770033 H770077 H770120 H770162 H720994 H721037 H770034 H770078 H770121 H770163 H720995 H721037 H770035 H770079 H770122 H770164 H720996 H721038 H770036 H770080 H770123 H770165 H720997 H721040 H770037 H770081 H770124 H770166 H720998 H721041 H770038 H770082 H770125 H770167 H720999 H721042 H770039 H770083 H770126 H770168 H721000 H721043 H770040 H770084 H770127 H770169 H721001 H721044 H770041 H770085 H770128 and H770170 |
| FEI Number |
2916710
|
Recalling Firm/ Manufacturer |
Varian Medical Systems Oncology Systems 911 Hansen Way Palo Alto CA 94304-1028
|
| For Additional Information Contact | 650-424-5731 |
Manufacturer Reason for Recall | During couch movement, a patient's fingers may get pinched at certain points which could result in broken, disengaged or pinched fingers. |
FDA Determined Cause 2 | Device Design |
| Action | Recall initiated on 08/11/2008. Notification will be sent to Varian Sales, Marketing and Service organizations as well as to all affected users. Contact Varian Medical Systems Oncology Systems, at 1-650-424-5731 for assistance. |
| Quantity in Commerce | 3296 units |
| Distribution | Worldwide including to the US and to Austria, Belgium, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, The Netherlands, United Kingdom, Algeria, Australia, Bahrain, Bangladesh, Belorussia, Bolivia, Brazil, Canada, Chile, China, Colombia, Ecuador, Egypt, Guadeloupe, Hong Kong, India, Indonesia, Iran, Israel, Jamaica, Japan, Kazakhstan, Lebanon, Macau, Macedonia, Malaysia, Martinique, Mexico, Morocco, Nepal, New Zealand, Oman, Pakistan, Panama, Peru, Philippines, Russian Fed., Saudi Arabia, Serbia, Montenegro, Singapore, South Africa, South Korea, Sudan, Syria, Taiwan, R.O.C., Thailand, Turkey, U.A.E, Ukraine, Uruguay, Venezuela, and Vietnam. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = IYE 510(K)s with Product Code = IYE
|
|
|
|