| | Class 2 Device Recall External Fixation Device |  |
| Date Initiated by Firm | August 08, 2008 |
| Date Posted | January 13, 2009 |
| Recall Status1 |
Terminated 3 on April 26, 2011 |
| Recall Number | Z-0556-2009 |
| Recall Event ID |
49510 |
| 510(K)Number | K051957 |
| Product Classification |
fixation appliance - Product Code KTT
|
| Product | External Fixation Device
Model #'s:
ORT99-36501
SYN390.005
SYN390.008
SYN394.84
SYN394.85 and
SYN394.86
Intended for use in the construction of an external fixation frame for treatment of a various fracture types that require external fixation. |
| Code Information |
n/a |
| FEI Number |
3001237663
|
Recalling Firm/ Manufacturer |
Sterilmed Inc 11400 73rd Ave N Ste 100 Osseo MN 55369-5562
|
| For Additional Information Contact | 763-488-3400 |
Manufacturer Reason for Recall | SteriMed, Inc is conducting a product removal of a small group of devices individually packaged and sealed within a pouch. The removal is being conducted because the possibility exists that some of the pouches were not properly sealed on one end.
A breach in packaging seal or a failure in packaging integrity has the possibility of risk to the patient in terms of transmitting organisms capable |
FDA Determined Cause 2 | Packaging process control |
| Action | Consignees were notified of the recall via verbal communication in person by on-site technician and or their local sales representative on 8/8/08. Others were contacted via fax or e-mail followed by on-site visits 8/11/08. SteriMed also provided consignees (Materials Manager/Risk Manager) with a "Urgent Product Removal" recall letter (August 14, 2008) in person informing them of the potential of receiving affected product. Included in the Product Removal Notification was a Customer Reply Form to inventory quantity of product received and quantity returned. Contact SterilMed, Inc. at 1-888-541-0078 for assistance. |
| Quantity in Commerce | 15 |
| Distribution | AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, MA, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, WA, WI, and WV OUS: CANADA |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = KTT
|
|
|
|