• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall GE Precision XR/a

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall GE Precision XR/asee related information
Date Initiated by FirmMarch 29, 2007
Date PostedSeptember 30, 2008
Recall Status1 Terminated 3 on February 04, 2012
Recall NumberZ-1713-2008
Recall Event ID 49533
510(K)NumberK993090 
Product Classification X-Ray System - Product Code KPR
ProductGE Healthcare Proteus XR/a Eclipse Collimator model # 2379827.
Code Information 42948HL0, 42949HL8, 43054HL6, 43055HL3, 43056HL1, 43057HL9, 43132HL0, 43133HL8, 43134HL6, 43135HL3, 43201HL3, 43202HL1, 43203HL9, 44605HL4, 44606HL2, 44607HL0, 44608HL8, 44609HL6, 44610HL4, 44681HL5, 45046HL0, 45047HL8, 45048HL6, 45128HL6, 45129HL4, 45130HL2, 45131HL0, 48119HL2, 48120HL0, 48121HL8, 48122HL6, 48123HL4, 48124HL2, 48125HL9, 48126HL7, 48127HL5, 48128HL3, 48164HL8, 48181HL2, 48274HL5, 48276HL0, 48277HL8, 48278HL6, 48680HL3, 48682HL9, 48683HL7, 48684HL5, 48685HL2, 48686HL0, 48687HL8, 48688HL6, 48689HL4, 48876HL7, 48877HL5, 48878HL3, 48995HL3, 48997HL1, 48998HL9, 49000HL3, 49208HL2, 49209HL0, 49210HL8, 49211HL6, 49391HL6, 49392HL4, 49393HL2, 49395HL7, 49494HL8, 49496HL3, 49497HL1, 49498HL9, 49499HL7, 49500HL2, 49501HL0, 49502HL8, 49503HL6, 49681HL0, 49682HL8, 49683HL6, 49684HL4, 49685HL1, 49686HL9, 49687HL7, 49688HL5, 49850HL1, 49851HL9, 49852HL7, 49853HL5, 49855HL0, 49856HL8, 49857HL6, 50027HL2, 50028HL0, 50029HL8, 50030HL6, 50031HL4, 50177HL5, 50178HL3, 50179HL1, 50180HL9, 50321HL9, 50322HL7, 50323HL5, 50324HL3, 50444HL9, 50445HL6, 50446HL4, 50447HL2, 50448HL0, 50449HL8, 50450HL6, 50451HL4, 50452HL2, 50453HL0, 50790HL5, 50791HL3, 50792HL1, 50793HL9, 50794HL9, 50795HL4, 50797HL0, 50798HL8, 50801HL0, 51143HL6, 51144HL4, 51145HL1, 51147HL7, 51486HL9, 51487HL7, 51488HL5, 52014HL8, and 70RD05001
FEI Number 2126677
Recalling Firm/
Manufacturer
GE Healthcare
3000 N Grandview Blvd
Waukesha WI 53188-1615
For Additional Information Contact
608-221-1551
Manufacturer Reason
for Recall
The actual average illuminance for the collimators are approximately 140-lux. This does not meet the 160-lux requirement. GE will be correcting the units by replacing all collimators.
FDA Determined
Cause 2
Radiation Control for Health and Safety Act
ActionGE will be correcting the units by replacing all collimators that don't meet the 160-lux requirement. The field correction will be initiated and is scheduled to be completed by October 2008.
Quantity in Commerce132
DistributionWorldwide Distribution --- USA including states of CA, DE, FL, GA, ID, IL, KS, KY, MD, MI, MO, MT, NC, NJ, NM, NY, OH, PA, SC, TX, UT, VA, and WV, and countries of Australia, Canada, China, France, Germany, Greece, Italy, Korea, Poland, Saudi Arabia, United Kingdom, and Venezuela.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = KPR
-
-