| | Class 2 Device Recall Enterra Gastric Electrical Stimulation System |  |
| Date Initiated by Firm | August 25, 2009 |
| Date Posted | October 29, 2009 |
| Recall Status1 |
Terminated 3 on May 22, 2012 |
| Recall Number | Z-0088-2010 |
| Recall Event ID |
53086 |
| HDE Number | H990014 |
| Product Classification |
Intestinal stimulator - Product Code LNQ
|
| Product | Medtronic Enterra Therapy, Gastric Electrical Stimulation System, consisting of model 3116 Enterra Therapy Neurostimulator and model 4351 Unipolar Intramuscular Lead. Medtronic, Inc., 710 Medtronic Parkway, Minneapolis, MN 55432-5604 USA.
The Neurostimulator (implantable neurostimulator or INS) is a multiprogrammable device designed to deliver therapy through gastric electrical stimulation when connected to a lead system. Humanitarian Device: Authorized by Federal Law for use in treatment of chronic intractable (drug refractory) nausea and vomiting secondary to gastroparesis of diabetic or idiopathic etiology. |
| Code Information |
SERIAL NUMBERS: NAT139233H, NAT139246H, NAT139368H, NAT139518H, NAT139745H, NAT139770H, NAT139794H, NAT139818H, NAT139881H, NAT139896H, NAT139909H, NAT139921H, NAT139927H, NAT139965H, NAT139972H, NAT140176H, NAT140520H, NAT140650H, NAT140734H, NAT141391H, NAT141409H, NAT141492H, NAT141736H, NAT141767H, NAT141784H, NAT141785H, NAT141794H, NAT141815H, NAT141827H, NAT141836H, NAT141873H, NAT141876H, NAT141878H, NAT141946H, NAT141957H, NAT141958H, NAT141968H, NAT141987H, NAT141989H, NAT142006H, NAT142027H, NAT142051H, NAT142053H, NAT142055H, NAT142056H, NAT142058H, NAT142059H, NAT142061H, NAT142062H, NAT142063H, NAT142065H, NAT142066H, NAT142068H, NAT142071H, NAT142074H, NAT142075H, NAT142076H, NAT142078H, NAT142082H, NAT142084H, NAT142085H, NAT142088H, NAT142090H, NAT142091H, NAT142092H, NAT142093H, NAT142096H, NAT142098H, NAT142099H, NAT142102H, NAT142109H, NAT142110H, NAT142112H, NAT142113H, NAT142115H, NAT142121H, NAT142131H, NAT142134H, NAT142139H, NAT142140H, NAT142142H, NAT142145H, NAT142146H, NAT142147H, NAT142149H, NAT142156H, NAT142158H, NAT142161H, NAT142164H, NAT142165H, NAT142171H, NAT142176H, NAT142177H, NAT142193H, NAT142214H, NAT142219H, NAT142228H, NAT142230H, NAT142231H, NAT142248H, NAT142263H, NAT142264H, NAT142274H, NAT142276H, NAT142278H, NAT142279H, NAT142280H, NAT142287H, NAT142290H, NAT142291H, NAT142292H, NAT142293H, NAT142295H, NAT142296H, NAT142297H, NAT142298H, NAT142300H, NAT142303H, NAT142309H, NAT142310H, NAT142311H, NAT142312H, NAT142313H, NAT142314H, NAT142435H, NAT142505H, NAT142513H, NAT142516H, NAT142543H, NAT142550H, NAT142551H, NAT142584H, NAT142595H, NAT142601H, NAT142621H, NAT142631H, NAT142687H, NAT142691H, NAT142700H, NAT142809H, NAT142847H, NAT142851H, NAT142852H, NAT142853H, NAT142877H, NAT142878H, NAT142895H, NAT142899H, NAT142900H, NAT142908H, NAT142910H, NAT142911H, NAT142913H, NAT142920H, NAT142923H, NAT142925H, NAT142928H, NAT142930H, NAT142932H, NAT142933H, NAT142934H, NAT142941H, NAT142943H, NAT142954H, NAT142960H, NAT142964H, NAT142971H, NAT142973H, NAT142982H, NAT142984H, NAT142991H, NAT143802H, NAT143925H, NAT143950H, NAT143989H, NAT144074H, NAT144201H, NAT144210H, NAT144227H, NAT144228H, NAT144231H, NAT144233H, NAT144241H, NAT144245H, NAT144254H, NAT144255H, NAT144258H, NAT144259H, NAT144260H, NAT144268H, NAT144270H, NAT144273H, NAT144283H, NAT144286H, NAT144289H, NAT144292H, NAT144293H, NAT144299H, NAT144336H, NAT144539H, NAT144662H, NAT144676H, NAT144683H, NAT144704H, NAT144736H, NAT144759H, NAT144775H, NAT144776H, NAT144783H, NAT144786H, NAT144787H, NAT144790H, NAT144796H, NAT144804H, NAT144811H, NAT144812H, NAT144813H, NAT144817H, NAT144820H, NAT144826H, NAT144832H, NAT144836H, NAT144841H, NAT144846H, NAT144847H, NAT144849H, NAT144921H, NAT144927H, NAT144931H, NAT144939H, NAT144942H, NAT144969H, NAT144985H, NAT145054H, NAT145055H, NAT145058H, NAT145061H, NAT145062H, NAT145063H, NAT145080H, NAT145084H, NAT145127H, NAT145131H, NAT145138H, NAT145139H, NAT145143H, NAT145146H, NAT145147H, NAT145148H, NAT145149H, NAT145150H, NAT145151H, NAT145152H, NAT145153H, NAT145155H, NAT145157H, NAT145161H, NAT145168H, NAT145169H, NAT145170H, NAT145172H, NAT145174H, NAT145177H, NAT145178H, NAT145182H, NAT145183H, NAT145184H, NAT145186H, NAT145188H, NAT145190H, NAT145191H, NAT145192H, NAT145194H, NAT145195H, NAT145196H, NAT145198H, NAT145200H, NAT145201H, NAT145202H, NAT145203H, NAT145204H, NAT145205H, NAT145211H, NAT145212H, NAT145217H, NAT145220H, NAT145221H, NAT145223H, NAT145225H, NAT145237H, NAT145300H, NAT145301H, NAT145307H, NAT145310H, NAT145313H, NAT145314H, NAT145317H, NAT145319H, NAT145320H, NAT145321H, NAT145324H, NAT145325H, NAT145327H, NAT145328H, NAT145329H, NAT145330H, NAT145334H, NAT145335H, NAT145336H, NAT145338H, NAT145339H, NAT145342H, NAT145344H, NAT145345H, NAT145347H, NAT145348H, NAT145349H, NAT145350H, NAT145353H, NAT145354H, NAT145356H, NAT145363H, NAT145366H, NAT145377H, NAT145383H, NAT145384H, NAT145386H, NAT145387H, NAT145389H, NAT145390H, NAT145391H, NAT145392H, NAT145396H, NAT145397H, NAT145398H, NAT145399H, NAT145650H, NAT145771H, NAT145861H, NAT145927H, NAT145988H, NAT145990H, NAT145991H, NAT145993H, NAT145994H, NAT145997H, NAT145998H, NAT146002H, NAT146003H, NAT146005H, NAT146006H, NAT146010H, NAT146011H, NAT146012H, NAT146014H, NAT146015H, NAT146016H, NAT146017H, NAT146018H, NAT146019H, NAT146023H, NAT146025H, NAT146026H, NAT146030H, NAT146031H, NAT146032H, NAT146033H, NAT146034H, NAT146035H, NAT146036H, NAT146037H, NAT146038H, NAT146039H, NAT146042H, NAT146044H, NAT146045H, NAT146046H, NAT146052H, NAT146054H, NAT146055H, NAT146056H, NAT146057H, NAT146058H, NAT146059H, NHV100017H, NHV100030H, NHV100032H, NHV100038H, NHV100039H, NHV100040H, NHV100041H, NHV100042H, NHV100043H, NHV100044H, NHV100045H, NHV100046H, NHV100047H, NHV100048H, NHV100049H, NHV100050H, NHV100051H, NHV100052H, NHV100053H, NHV100055H, NHV100056H, NHV100057H, NHV100058H, NHV100059H, NHV100060H, NHV100061H, NHV100062H, NHV100063H, NHV100064H, NHV100065H, NHV100066H, NHV100067H, NHV100068H, NHV100069H, NHV100070H, NHV100071H, NHV100072H, NHV100073H, NHV100074H, NHV100075H, NHV100076H, NHV100077H, NHV100078H, NHV100079H, NHV100080H, NHV100081H, NHV100082H, NHV100083H, NHV100084H, NHV100085H, NHV100086H, NHV100087H, NHV100088H, NHV100089H, NHV100090H, NHV100091H, NHV100092H, NHV100093H, NHV100094H, NHV100095H, NHV100096H, NHV100097H, NHV100098H, NHV100099H, NHV100100H, NHV100101H, NHV100102H, NHV100103H, NHV100104H, NHV100105H, NHV100106H, NHV100107H, NHV100109H, NHV100110H, NHV100111H, NHV100112H, NHV100113H, NHV100114H, NHV100115H, NHV100116H, NHV100117H, NHV100118H, NHV100119H, NHV100120H, NHV100121H, NHV100122H, NHV100123H, NHV100124H, NHV100125H, NHV100126H, NHV100127H, NHV100128H, NHV100129H, NHV100130H, NHV100131H, NHV100132H, NHV100133H, NHV100134H, NHV100135H, NHV100136H, NHV100137H, NHV100138H, NHV100139H, NHV100140H, NHV100141H, NHV100142H, NHV100143H, NHV100144H, NHV100145H, NHV100146H, NHV100147H, NHV100148H, NHV100149H, NHV100150H, NHV100152H, NHV100153H, NHV100154H, NHV100155H, NHV100156H, NHV100157H, NHV100158H, NHV100159H, NHV100160H, NHV100161H, NHV100162H, NHV100164H, NHV100165H, NHV100166H, NHV100167H, NHV100168H, NHV100169H, NHV100170H, NHV100171H, NHV100172H, NHV100173H, NHV100174H, NHV100175H, NHV100176H, NHV100177H, NHV100178H, NHV100179H, NHV100180H, NHV100181H, NHV100182H, NHV100184H, NHV100185H, NHV100186H, NHV100187H, NHV100188H, NHV100189H, NHV100190H, NHV100191H, NHV100192H, NHV100193H, NHV100194H, NHV100195H, NHV100196H, NHV100197H, NHV100198H, NHV100199H, NHV100200H, NHV100201H, NHV100202H, NHV100203H, NHV100204H, NHV100205H, NHV100206H, NHV100207H, NHV100208H, NHV100209H, NHV100210H, NHV100211H, NHV100212H, NHV100213H, NHV100214H, NHV100215H, NHV100216H, NHV100217H, NHV100218H, NHV100219H, NHV100220H, NHV100221H, NHV100222H, NHV100223H, NHV100224H, NHV100225H, NHV100226H, NHV100227H, NHV100228H, NHV100229H, NHV100230H, NHV100231H, NHV100232H, NHV100234H, NHV100235H, NHV100236H, NHV100237H, NHV100238H, NHV100239H, NHV100240H, NHV100241H, NHV100242H, NHV100243H, NHV100244H, NHV100245H, NHV100246H, NHV100247H, NHV100248H, NHV100249H, NHV100250H, NHV100251H, NHV100252H, NHV100253H, NHV100254H, NHV100255H, NHV100256H, NHV100257H, NHV100258H, NHV100259H, NHV100260H, NHV100261H, NHV100262H, NHV100263H, NHV100264H, NHV100265H, NHV100266H, NHV100267H, NHV100268H, NHV100269H, NHV100270H, NHV100271H, NHV100272H, NHV100273H, NHV100274H, NHV100275H, NHV100276H, NHV100277H, NHV100278H, NHV100279H, NHV100280H, NHV100281H, NHV100282H, NHV100283H, NHV100284H, NHV100285H, NHV100286H, NHV100287H, NHV100288H, NHV100289H, NHV100290H, NHV100291H, NHV100292H, NHV100293H, NHV100294H, NHV100295H, NHV100296H, NHV100297H, NHV100298H, NHV100299H, NHV100300H, NHV100301H, NHV100302H, NHV100303H, NHV100304H, NHV100305H, NHV100306H, NHV100307H, NHV100308H, NHV100309H, NHV100310H, NHV100311H, NHV100312H, NHV100313H, NHV100314H, NHV100315H, NHV100316H, NHV100317H, NHV100318H, NHV100319H, NHV100320H, NHV100321H, NHV100322H, NHV100323H, NHV100324H, NHV100325H, NHV100326H, NHV100327H, NHV100328H, NHV100329H, NHV100330H, NHV100331H, NHV100332H, NHV100334H, NHV100335H, NHV100336H, NHV100337H, NHV100338H, NHV100339H, NHV100340H, NHV100341H, NHV100342H, NHV100343H, NHV100344H, NHV100345H, NHV100346H, NHV100347H, NHV100348H, NHV100349H, NHV100350H, NHV100351H, NHV100352H, NHV100353H, NHV100354H, NHV100355H, NHV100356H, NHV100357H, NHV100358H, NHV100359H, NHV100360H, NHV100361H, NHV100362H, NHV100363H, NHV100364H, NHV100365H, NHV100366H, NHV100367H, NHV100368H, NHV100369H, NHV100370H, NHV100372H, NHV100373H, NHV100374H, NHV100375H, NHV100376H, NHV100377H, NHV100378H, NHV100379H, NHV100380H, NHV100381H, NHV100382H, NHV100383H, NHV100384H, NHV100385H, NHV100386H, NHV100387H, NHV100388H, NHV100389H, NHV100390H, NHV100391H, NHV100392H, NHV100393H, NHV100394H, NHV100395H, NHV100396H, NHV100397H, NHV100398H, NHV100399H, NHV100400H, NHV100401H, NHV100402H, NHV100403H, NHV100404H, NHV100405H, NHV100406H, NHV100407H, NHV100408H, NHV100409H, NHV100410H, NHV100411H, NHV100413H, NHV100414H, NHV100415H, NHV100416H, NHV100417H, NHV100418H, NHV100419H, NHV100420H, NHV100421H, NHV100422H, NHV100423H, NHV100424H, NHV100425H, NHV100426H, NHV100427H, NHV100428H, NHV100429H, NHV100430H, NHV100431H, NHV100432H, NHV100433H, NHV100434H, NHV100435H, NHV100436H, NHV100437H, NHV100438H, NHV100439H, NHV100440H, NHV100441H, NHV100442H, NHV100443H, NHV100444H, NHV100445H, NHV100446H, NHV100447H, NHV100448H, NHV100449H, NHV100450H, NHV100451H, NHV100452H, NHV100453H, NHV100454H, NHV100455H, NHV100456H, NHV100457H, NHV100458H, NHV100459H, NHV100460H, NHV100461H, NHV100462H, NHV100463H, NHV100464H, NHV100466H, NHV100468H, NHV100469H, NHV100470H, NHV100472H, NHV100474H, NHV100475H, NHV100476H, NHV100477H, NHV100478H, NHV100479H, NHV100480H, NHV100481H, NHV100482H, NHV100483H, NHV100484H, NHV100485H, NHV100486H, NHV100487H, NHV100488H, NHV100489H, NHV100490H, NHV100491H, NHV100492H, NHV100493H, NHV100494H, NHV100495H, NHV100496H, NHV100497H, NHV100498H, NHV100499H, NHV100500H, NHV100501H, NHV100502H, NHV100503H, NHV100504H, NHV100505H, NHV100506H, NHV100507H, NHV100508H, NHV100509H, NHV100510H, NHV100511H, NHV100512H, NHV100513H, NHV100514H, NHV100515H, NHV100516H, NHV100517H, NHV100518H, NHV100519H, NHV100520H, NHV100521H, NHV100522H, NHV100523H, NHV100524H, NHV100525H, NHV100526H, NHV100527H, NHV100528H, NHV100529H, NHV100530H, NHV100531H, NHV100532H, NHV100533H, NHV100534H, NHV100535H, NHV100536H, NHV100537H, NHV100538H, NHV100539H, NHV100540H, NHV100541H, NHV100542H, NHV100543H, NHV100544H, NHV100545H, NHV100546H, NHV100547H, NHV100548H, NHV100549H, NHV100550H, NHV100551H, NHV100552H, NHV100553H, NHV100554H, NHV100555H, NHV100556H, NHV100557H, NHV100558H, NHV100559H, NHV100560H, NHV100561H, NHV100562H, NHV100563H, NHV100564H, NHV100565H, NHV100566H, NHV100567H, NHV100568H, NHV100569H, NHV100570H, NHV100571H, NHV100572H, NHV100573H, NHV100574H, NHV100575H, NHV100576H, NHV100577H, NHV100578H, NHV100579H, NHV100580H, NHV100581H, NHV100582H, NHV100583H, NHV100584H, NHV100585H, NHV100586H, NHV100587H, NHV100588H, NHV100589H, NHV100590H, NHV100591H, NHV100592H, NHV100593H, NHV100594H, NHV100595H, NHV100596H, NHV100597H, NHV100598H, NHV100599H, NHV100600H, NHV100601H, NHV100602H, NHV100603H, NHV100604H, NHV100605H, NHV100606H, NHV100607H, NHV100608H, NHV100609H, NHV100610H, NHV100611H, NHV100612H, NHV100613H, NHV100614H, NHV100615H, NHV100616H, NHV100617H, NHV100618H, NHV100619H, NHV100620H, NHV100621H, NHV100622H, NHV100623H, NHV100624H, NHV100625H, NHV100626H, NHV100627H, NHV100628H, NHV100629H, NHV100630H, NHV100631H, NHV100632H, NHV100633H, NHV100634H, NHV100635H, NHV100636H, NHV100637H, NHV100638H, NHV100639H, NHV100640H, NHV100641H, NHV100642H, NHV100643H, NHV100644H, NHV100645H, NHV100646H, NHV100647H, NHV100648H, NHV100649H, NHV100650H, NHV100651H, NHV100652H, NHV100653H, NHV100654H, NHV100655H, NHV100656H, NHV100657H, NHV100658H, NHV100659H, NHV100660H, NHV100661H, NHV100662H, NHV100663H, NHV100664H, NHV100665H, NHV100666H, NHV100667H, NHV100668H, NHV100669H, NHV100670H, NHV100671H, NHV100672H, NHV100673H, NHV100674H, NHV100675H, NHV100676H, NHV100677H, NHV100678H, NHV100679H, NHV100680H, NHV100681H, NHV100683H, NHV100685H, NHV100686H, NHV100687H, NHV100688H, NHV100689H, NHV100690H, NHV100691H, NHV100692H, NHV100693H, NHV100694H, NHV100695H, NHV100696H, NHV100697H, NHV100698H, NHV100699H, NHV100700H, NHV100701H, NHV100702H, NHV100703H, NHV100704H, NHV100706H, NHV100707H, NHV100708H, NHV100709H, NHV100710H, NHV100711H, NHV100712H, NHV100713H, NHV100714H, NHV100715H, NHV100716H, NHV100717H, NHV100718H, NHV100719H, NHV100720H, NHV100721H, NHV100722H, NHV100723H, NHV100724H, NHV100725H, NHV100726H, NHV100727H, NHV100728H, NHV100729H, NHV100730H, NHV100731H, NHV100732H, NHV100733H, NHV100734H, NHV100735H, NHV100736H, NHV100737H, NHV100738H, NHV100739H, NHV100740H, NHV100741H, NHV100742H, NHV100743H, NHV100744H, NHV100745H, NHV100746H, NHV100747H, NHV100748H, NHV100749H, NHV100750H, NHV100751H, NHV100753H, NHV100754H, NHV100755H, NHV100756H, NHV100757H, NHV100758H, NHV100759H, NHV100760H, NHV100761H, NHV100762H, NHV100763H, NHV100764H, NHV100765H, NHV100766H, NHV100767H, NHV100768H, NHV100769H, NHV100770H, NHV100771H, NHV100772H, NHV100773H, NHV100774H, NHV100775H, NHV100776H, NHV100777H, NHV100778H, NHV100779H, NHV100780H, NHV100781H, NHV100782H, NHV100783H, NHV100784H, NHV100785H, NHV100786H, NHV100787H, NHV100788H, NHV100789H, NHV100790H, NHV100791H, NHV100792H, NHV100793H, NHV100794H, NHV100795H, NHV100796H, NHV100797H, NHV100798H, NHV100799H, NHV100800H, NHV100801H, NHV100802H, NHV100803H, NHV100804H, NHV100805H, NHV100806H, NHV100807H, NHV100808H, NHV100809H, NHV100810H, NHV100811H, NHV100812H, NHV100813H, NHV100814H, NHV100815H, NHV100816H, NHV100817H, NHV100818H, NHV100819H, NHV100820H, NHV100821H, NHV100822H, NHV100823H, NHV100824H, NHV100825H, NHV100826H, NHV100827H, NHV100828H, NHV100829H, NHV100830H, NHV100831H, NHV100832H, NHV100833H, NHV100834H, NHV100835H, NHV100836H, NHV100837H, NHV100838H, NHV100839H, NHV100840H, NHV100841H, NHV100842H, NHV100843H, NHV100844H, NHV100845H, NHV100846H, NHV100847H, NHV100848H, NHV100849H, NHV100850H, NHV100851H, NHV100852H, NHV100853H, NHV100854H, NHV100855H, NHV100856H, NHV100857H, NHV100858H, NHV100859H, NHV100860H, NHV100861H, NHV100862H, NHV100863H, NHV100864H, NHV100865H, NHV100866H, NHV100867H, NHV100868H, NHV100869H, NHV100870H, NHV100871H, NHV100872H, NHV100873H, NHV100875H, NHV100876H, NHV100877H, NHV100878H, NHV100879H, NHV100880H, NHV100881H, NHV100882H, NHV100883H, NHV100884H, NHV100885H, NHV100886H, NHV100887H, NHV100888H, NHV100889H, NHV100890H, NHV100891H, NHV100892H, NHV100893H, NHV100894H, NHV100895H, NHV100896H, NHV100897H, NHV100898H, NHV100899H, NHV100900H, NHV100901H, NHV100902H, NHV100903H, NHV100904H, NHV100905H, NHV100906H, NHV100907H, NHV100908H, NHV100909H, NHV100910H, NHV100911H, NHV100912H, NHV100913H, NHV100914H, NHV100915H, NHV100916H, NHV100917H, NHV100918H, NHV100919H, NHV100920H, NHV100921H, NHV100922H, NHV100923H, NHV100924H, NHV100925H, NHV100926H, NHV100927H, NHV100928H, NHV100929H, NHV100930H, NHV100931H, NHV100932H, NHV100933H, NHV100934H, NHV100935H, NHV100936H, NHV100937H, NHV100938H, NHV100939H, NHV100940H, NHV100941H, NHV100942H, NHV100943H, NHV100944H, NHV100945H, NHV100946H, NHV100947H, NHV100948H, NHV100949H, NHV100950H, NHV100951H, NHV100952H, NHV100953H, NHV100954H, NHV100955H, NHV100956H, NHV100957H, NHV100958H, NHV100959H, NHV100960H, NHV100961H, NHV100962H, NHV100963H, NHV100964H, NHV100965H, NHV100966H, NHV100967H, NHV100968H, NHV100969H, NHV100970H, NHV100971H, NHV100972H, NHV100973H, NHV100974H, NHV100975H, NHV100976H, NHV100977H, NHV100978H, NHV100979H, NHV100980H, NHV100981H, NHV100982H, NHV100983H, NHV100984H, NHV100985H, NHV100986H, NHV100987H, NHV100988H, NHV100989H, NHV100990H, NHV100991H, NHV100992H, NHV100993H, NHV100994H, NHV100995H, NHV100996H, NHV100997H, NHV100998H, NHV100999H, NHV101000H, NHV101001H, NHV101002H, NHV101003H, NHV101004H, NHV101005H, NHV101006H, NHV101007H, NHV101008H, NHV101009H, NHV101010H, NHV101011H, NHV101012H, NHV101013H, NHV101014H, NHV101015H, NHV101016H, NHV101017H, NHV101018H, NHV101019H, NHV101020H, NHV101021H, NHV101022H, NHV101023H, NHV101024H, NHV101025H, NHV101026H, NHV101027H, NHV101028H, NHV101029H, NHV101030H, NHV101031H, NHV101032H, NHV101033H, NHV101034H, NHV101035H, NHV101036H, NHV101037H, NHV101038H, NHV101039H, NHV101040H, NHV101041H, NHV101042H, NHV101043H, NHV101044H, NHV101045H, NHV101046H, NHV101047H, NHV101048H, NHV101049H, NHV101050H, NHV101051H, NHV101052H, NHV101053H, NHV101054H, NHV101055H, NHV101056H, NHV101057H, NHV101058H, NHV101059H, NHV101060H, NHV101061H, NHV101062H, NHV101063H, NHV101064H, NHV101065H, NHV101066H, NHV101067H, NHV101068H, NHV101069H, NHV101070H, NHV101071H, NHV101072H, NHV101073H, NHV101074H, NHV101075H, NHV101076H, NHV101077H, NHV101078H, NHV101079H, NHV101080H, NHV101081H, NHV101082H, NHV101083H, NHV101084H, NHV101085H, NHV101086H, NHV101087H, NHV101088H, NHV101089H, NHV101090H, NHV101091H, NHV101092H, NHV101093H, NHV101094H, NHV101095H, NHV101096H, NHV101097H, NHV101098H, NHV101099H, NHV101100H, NHV101101H, NHV101102H, NHV101103H, NHV101104H, NHV101105H, NHV101106H, NHV101107H, NHV101108H, NHV101109H, NHV101110H, NHV101111H, NHV101112H, NHV101113H, NHV101114H, NHV101115H, NHV101116H, NHV101117H, NHV101118H, NHV101119H, NHV101120H, NHV101121H, NHV101122H, NHV101123H, NHV101124H, NHV101125H, NHV101126H, NHV101127H, NHV101128H, NHV101129H, NHV101130H, NHV101131H, NHV101132H, NHV101133H, NHV101134H, NHV101135H, NHV101136H, NHV101137H, NHV101138H, NHV101139H, NHV101140H, NHV101141H, NHV101142H, NHV101143H, NHV101144H, NHV101145H, NHV101146H, NHV101148H, NHV101149H, NHV101150H, NHV101151H, NHV101152H, NHV101153H, NHV101154H, NHV101155H, NHV101156H, NHV101157H, NHV101158H, NHV101159H, NHV101160H, NHV101161H, NHV101162H, NHV101163H, NHV101164H, NHV101165H, NHV101166H, NHV101167H, NHV101168H, NHV101169H, NHV101170H, NHV101171H, NHV101172H, NHV101173H, NHV101174H, NHV101175H, NHV101176H, NHV101177H, NHV101178H, NHV101179H, NHV101180H, NHV101181H, NHV101182H, NHV101183H, NHV101184H, NHV101185H, NHV101186H, NHV101187H, NHV101188H, NHV101189H, NHV101190H, NHV101191H, NHV101192H, NHV101193H, NHV101194H, NHV101195H, NHV101196H, NHV101197H, NHV101198H, NHV101199H, NHV101200H, NHV101201H, NHV101202H, NHV101203H, NHV101204H, NHV101205H, NHV101206H, NHV101207H, NHV101209H, NHV101210H, NHV101211H, NHV101212H, NHV101213H, NHV101214H, NHV101215H, NHV101216H, NHV101217H, NHV101218H, NHV101219H, NHV101220H, NHV101221H, NHV101222H, NHV101223H, NHV101224H, NHV101225H, NHV101226H, NHV101227H, NHV101228H, NHV101229H, NHV101230H, NHV101231H, NHV101232H, NHV101233H, NHV101234H, NHV101235H, NHV101236H, NHV101237H, NHV101238H, NHV101239H, NHV101240H, NHV101241H, NHV101242H, NHV101243H, NHV101244H, NHV101245H, NHV101246H, NHV101247H, NHV101248H, NHV101249H, NHV101250H, NHV101251H, NHV101252H, NHV101253H, NHV101254H, NHV101255H, NHV101256H, NHV101257H, NHV101258H, NHV101259H, NHV101260H, NHV101261H, NHV101262H, NHV101263H, NHV101264H, NHV101265H, NHV101266H, NHV101267H, NHV101269H, NHV101270H, NHV101271H, NHV101272H, NHV101273H, NHV101274H, NHV101275H, NHV101276H, NHV101277H, NHV101278H, NHV101279H, NHV101280H, NHV101281H, NHV101282H, NHV101283H, NHV101285H, NHV101286H, NHV101288H, NHV101289H, NHV101290H, NHV101291H, NHV101292H, NHV101293H, NHV101294H, NHV101295H, NHV101296H, NHV101297H, NHV101298H, NHV101299H, NHV101300H, NHV101301H, NHV101302H, NHV101304H, NHV101305H, NHV101306H, NHV101307H, NHV101308H, NHV101309H, NHV101310H, NHV101311H, NHV101312H, NHV101313H, NHV101314H, NHV101315H, NHV101316H, NHV101317H, NHV101318H, NHV101319H, NHV101320H, NHV101321H, NHV101322H, NHV101324H, NHV101325H, NHV101326H, NHV101327H, NHV101328H, NHV101329H, NHV101330H, NHV101331H, NHV101332H, NHV101333H, NHV101334H, NHV101335H, NHV101336H, NHV101337H, NHV101338H, NHV101339H, NHV101340H, NHV101341H, NHV101342H, NHV101343H, NHV101344H, NHV101345H, NHV101346H, NHV101347H, NHV101348H, NHV101350H, NHV101351H, NHV101352H, NHV101353H, NHV101354H, NHV101355H, NHV101356H, NHV101357H, NHV101358H, NHV101359H, NHV101360H, NHV101361H, NHV101362H, NHV101363H, NHV101364H, NHV101365H, NHV101366H, NHV101367H, NHV101368H, NHV101369H, NHV101370H, NHV101371H, NHV101372H, NHV101373H, NHV101374H, NHV101375H, NHV101376H, NHV101377H, NHV101378H, NHV101379H, NHV101380H, NHV101381H, NHV101382H, NHV101383H, NHV101384H, NHV101385H, NHV101386H, NHV101387H, NHV101388H, NHV101389H, NHV101390H, NHV101391H, NHV101392H, NHV101393H, NHV101394H, NHV101395H, NHV101396H, NHV101397H, NHV101398H, NHV101399H, NHV101400H, NHV101401H, NHV101402H, NHV101403H, NHV101405H, NHV101406H, NHV101408H, NHV101409H, NHV101410H, NHV101411H, NHV101412H, NHV101413H, NHV101414H, NHV101415H, NHV101416H, NHV101417H, NHV101418H, NHV101419H, NHV101420H, NHV101421H, NHV101422H, NHV101423H, NHV101424H, NHV101425H, NHV101426H, NHV101427H, NHV101428H, NHV101429H, NHV101430H, NHV101431H, NHV101432H, NHV101433H, NHV101434H, NHV101435H, NHV101436H, NHV101437H, NHV101438H, NHV101439H, NHV101440H, NHV101441H, NHV101442H, NHV101443H, NHV101444H, NHV101445H, NHV101446H, NHV101447H, NHV101448H, NHV101449H, NHV101450H, NHV101451H, NHV101452H, NHV101453H, NHV101454H, NHV101455H, NHV101456H, NHV101457H, NHV101458H, NHV101459H, NHV101460H, NHV101461H, NHV101462H, NHV101463H, NHV101464H, NHV101465H, NHV101466H, NHV101467H, NHV101468H, NHV101469H, NHV101471H, NHV101472H, NHV101474H, NHV101475H, NHV101476H, NHV101477H, NHV101478H, NHV101479H, NHV101480H, NHV101481H, NHV101482H, NHV101483H, NHV101484H, NHV101485H, NHV101486H, NHV101487H, NHV101488H, NHV101489H, NHV101490H, NHV101491H, NHV101493H, NHV101494H, NHV101495H, NHV101496H, NHV101497H, NHV101498H, NHV101499H, NHV101500H, NHV101501H, NHV101502H, NHV101503H, NHV101504H, NHV101505H, NHV101506H, NHV101507H, NHV101508H, NHV101509H, NHV101510H, NHV101511H, NHV101512H, NHV101513H, NHV101514H, NHV101515H, NHV101516H, NHV101517H, NHV101518H, NHV101519H, NHV101520H, NHV101521H, NHV101522H, NHV101523H, NHV101524H, NHV101526H, NHV101527H, NHV101528H, NHV101529H, NHV101530H, NHV101531H, NHV101532H, NHV101533H, NHV101534H, NHV101535H, NHV101536H, NHV101537H, NHV101538H, NHV101540H, NHV101541H, NHV101542H, NHV101543H, NHV101544H, NHV101545H, NHV101546H, NHV101547H, NHV101548H, NHV101549H, NHV101550H, NHV101551H, NHV101552H, NHV101553H, NHV101554H, NHV101555H, NHV101556H, NHV101557H, NHV101558H, NHV101559H, NHV101560H, NHV101561H, NHV101562H, NHV101563H, NHV101564H, NHV101565H, NHV101566H, NHV101567H, NHV101568H, NHV101569H, NHV101570H, NHV101571H, NHV101572H, NHV101573H, NHV101574H, NHV101575H, NHV101576H, NHV101577H, NHV101578H, NHV101579H, NHV101580H, NHV101581H, NHV101582H, NHV101583H, NHV101584H, NHV101585H, NHV101586H, NHV101587H, NHV101588H, NHV101589H, NHV101590H, NHV101592H, NHV101593H, NHV101594H, NHV101595H, NHV101596H, NHV101597H, NHV101598H, NHV101599H, NHV101600H, NHV101601H, NHV101602H, NHV101603H, NHV101604H, NHV101605H, NHV101606H, NHV101607H, NHV101608H, NHV101609H, NHV101610H, NHV101611H, NHV101612H, NHV101613H, NHV101614H, NHV101615H, NHV101616H, NHV101617H, NHV101618H, NHV101619H, NHV101620H, NHV101621H, NHV101622H, NHV101623H, NHV101624H, NHV101625H, NHV101626H, NHV101627H, NHV101628H, NHV101629H, NHV101630H, NHV101631H, NHV101632H, NHV101633H, NHV101634H, NHV101635H, NHV101636H, NHV101637H, NHV101638H, NHV101639H, NHV101640H, NHV101641H, NHV101642H, NHV101643H, NHV101644H, NHV101645H, NHV101646H, NHV101647H, NHV101648H, NHV101649H, NHV101650H, NHV101651H, NHV101652H, NHV101653H, NHV101654H, NHV101656H, NHV101657H, NHV101658H, NHV101659H, NHV101660H, NHV101661H, NHV101662H, NHV101663H, NHV101664H, NHV101665H, NHV101666H, NHV101667H, NHV101668H, NHV101669H, NHV101670H, NHV101671H, NHV101672H, NHV101673H, NHV101675H, NHV101676H, NHV101677H, NHV101678H, NHV101679H, NHV101680H, NHV101681H, NHV101682H, NHV101683H, NHV101684H, NHV101685H, NHV101686H, NHV101687H, NHV101688H, NHV101689H, NHV101690H, NHV101691H, NHV101692H, NHV101693H, NHV101694H, NHV101695H, NHV101696H, NHV101697H, NHV101698H, NHV101699H, NHV101701H, NHV101702H, NHV101703H, NHV101704H, NHV101705H, NHV101706H, NHV101707H, NHV101708H, NHV101709H, NHV101710H, NHV101711H, NHV101712H, NHV101713H, NHV101714H, NHV101715H, NHV101716H, NHV101717H, NHV101718H, NHV101719H, NHV101720H, NHV101722H, NHV101723H, NHV101724H, NHV101725H, NHV101726H, NHV101727H, NHV101728H, NHV101729H, NHV101730H, NHV101731H, NHV101732H, NHV101733H, NHV101734H, NHV101735H, NHV101736H, NHV101737H, NHV101738H, NHV101739H, NHV101740H, NHV101741H, NHV101742H, NHV101743H, NHV101744H, NHV101745H, NHV101746H, NHV101747H, NHV101748H, NHV101749H, NHV101750H, NHV101751H, NHV101752H, NHV101753H, NHV101754H, NHV101755H, NHV101756H, NHV101757H, NHV101758H, NHV101759H, NHV101760H, NHV101761H, NHV101762H, NHV101763H, NHV101764H, NHV101765H, NHV101766H, NHV101767H, NHV101768H, NHV101769H, NHV101770H, NHV101771H, NHV101772H, NHV101773H, NHV101774H, NHV101775H, NHV101776H, NHV101777H, NHV101778H, NHV101779H, NHV101780H, NHV101781H, NHV101782H, NHV101783H, NHV101784H, NHV101785H, NHV101786H, NHV101787H, NHV101788H, NHV101789H, NHV101790H, NHV101791H, NHV101792H, NHV101794H, NHV101795H, NHV101796H, NHV101797H, NHV101798H, NHV101799H, NHV101800H, NHV101801H, NHV101802H, NHV101803H, NHV101804H, NHV101805H, NHV101806H, NHV101807H, NHV101808H, NHV101809H, NHV101810H, NHV101811H, NHV101812H, NHV101813H, NHV101814H, NHV101815H, NHV101816H, NHV101817H, NHV101818H, NHV101819H, NHV101820H, NHV101821H, NHV101822H, NHV101823H, NHV101824H, NHV101825H, NHV101826H, NHV101827H, NHV101828H, NHV101829H, NHV101830H, NHV101831H, NHV101832H, NHV101833H, NHV101834H, NHV101835H, NHV101836H, NHV101837H, NHV101838H, NHV101839H, NHV101840H, NHV101841H, NHV101842H, NHV101843H, NHV101844H, NHV101845H, NHV101846H, NHV101847H, NHV101848H, NHV101849H, NHV101850H, NHV101851H, NHV101852H, NHV101853H, NHV101854H, NHV101855H, NHV101856H, NHV101857H, NHV101858H, NHV101859H, NHV101860H, NHV101861H, NHV101862H, NHV101863H, NHV101864H, NHV101865H, NHV101866H, NHV101867H, NHV101868H, NHV101869H, NHV101870H, NHV101871H, NHV101872H, NHV101873H, NHV101874H, NHV101875H, NHV101876H, NHV101877H, NHV101878H, NHV101879H, NHV101880H, NHV101881H, NHV101882H, NHV101883H, NHV101884H, NHV101885H, NHV101886H, NHV101887H, NHV101888H, NHV101889H, NHV101890H, NHV101891H, NHV101892H, NHV101893H, NHV101894H, NHV101895H, NHV101896H, NHV101897H, NHV101898H, NHV101899H, NHV101900H, NHV101901H, NHV101902H, NHV101903H, NHV101904H, NHV101905H, NHV101906H, NHV101907H, NHV101908H, NHV101909H, NHV101910H, NHV101911H, NHV101912H, NHV101913H, NHV101915H, NHV101916H, NHV101917H, NHV101918H, NHV101919H, NHV101920H, NHV101921H, NHV101922H, NHV101923H, NHV101924H, NHV101925H, NHV101926H, NHV101927H, NHV101928H, NHV101929H, NHV101930H, NHV101931H, NHV101932H, NHV101933H, NHV101934H, NHV101935H, NHV101936H, NHV101937H, NHV101939H, NHV101940H, NHV101941H, NHV101942H, NHV101943H, NHV101946H, NHV101947H, NHV101948H, NHV101949H, NHV101950H, NHV101951H, NHV101952H, NHV101953H, NHV101954H, NHV101955H, NHV101956H, NHV101957H, NHV101958H, NHV101959H, NHV101960H, NHV101961H, NHV101963H, NHV101964H, NHV101965H, NHV101966H, NHV101967H, NHV101968H, NHV101969H, NHV101970H, NHV101971H, NHV101972H, NHV101973H, NHV101974H, NHV101975H, NHV101976H, NHV101977H, NHV101978H, NHV101979H, NHV101980H, NHV101981H, NHV101982H, NHV101983H, NHV101984H, NHV101985H, NHV101986H, NHV101987H, NHV101988H, NHV101989H, NHV101990H, NHV101991H, NHV101992H, NHV101993H, NHV101994H, NHV101995H, NHV101996H, NHV101997H, NHV101998H, NHV101999H, NHV102000H, NHV102001H, NHV102002H, NHV102003H, NHV102004H, NHV102005H, NHV102006H, NHV102007H, NHV102008H, NHV102009H, NHV102010H, NHV102011H, NHV102012H, NHV102013H, NHV102014H, NHV102015H, NHV102017H, NHV102018H, NHV102019H, NHV102020H, NHV102021H, NHV102022H, NHV102023H, NHV102024H, NHV102025H, NHV102026H, NHV102027H, NHV102028H, NHV102029H, NHV102030H, NHV102031H, NHV102032H, NHV102033H, NHV102034H, NHV102035H, NHV102036H, NHV102037H, NHV102038H, NHV102039H, NHV102040H, NHV102041H, NHV102042H, NHV102043H, NHV102044H, NHV102045H, NHV102046H, NHV102047H, NHV102048H, NHV102049H, NHV102051H, NHV102052H, NHV102053H, NHV102054H, NHV102055H, NHV102058H, NHV102059H, NHV102060H, NHV102061H, NHV102062H, NHV102063H, NHV102064H, NHV102065H, NHV102066H, NHV102067H, NHV102068H, NHV102069H, NHV102070H, NHV102071H, NHV102072H, NHV102073H, NHV102074H, NHV102075H, NHV102076H, NHV102077H, NHV102078H, NHV102079H, NHV102080H, NHV102081H, NHV102082H, NHV102083H, NHV102084H, NHV102085H, NHV102086H, NHV102087H, NHV102088H, NHV102089H, NHV102090H, NHV102091H, NHV102092H, NHV102093H, NHV102094H, NHV102095H, NHV102096H, NHV102097H, NHV102098H, NHV102099H, NHV102100H, NHV102101H, NHV102102H, NHV102103H, NHV102104H, NHV102105H, NHV102106H, NHV102107H, NHV102108H, NHV102109H, NHV102110H, NHV102111H, NHV102112H, NHV102113H, NHV102114H, NHV102115H, NHV102116H, NHV102117H, NHV102118H, NHV102119H, NHV102120H, NHV102121H, NHV102122H, NHV102123H, NHV102124H, NHV102125H, NHV102126H, NHV102127H, NHV102128H, NHV102129H, NHV102130H, NHV102131H, NHV102132H, NHV102133H, NHV102134H, NHV102135H, NHV102136H, NHV102137H, NHV102139H, NHV102140H, NHV102141H, NHV102142H, NHV102143H, NHV102144H, NHV102145H, NHV102146H, NHV102147H, NHV102148H, NHV102149H, NHV102150H, NHV102151H, NHV102152H, NHV102153H, NHV102154H, NHV102155H, NHV102156H, NHV102157H, NHV102158H, NHV102160H, NHV102161H, NHV102162H, NHV102163H, NHV102164H, NHV102165H, NHV102166H, NHV102167H, NHV102168H, NHV102169H, NHV102170H, NHV102171H, NHV102172H, NHV102173H, NHV102174H, NHV102175H, NHV102176H, NHV102177H, NHV102178H, NHV102179H, NHV102180H, NHV102181H, NHV102182H, NHV102183H, NHV102184H, NHV102185H, NHV102186H, NHV102187H, NHV102188H, NHV102189H, NHV102190H, NHV102191H, NHV102192H, NHV102193H, NHV102194H, NHV102195H, NHV102197H, NHV102198H, NHV102199H, NHV102200H, NHV102201H, NHV102202H, NHV102203H, NHV102204H, NHV102205H, NHV102206H, NHV102207H, NHV102208H, NHV102209H, NHV102210H, NHV102211H, NHV102212H, NHV102213H, NHV102214H, NHV102215H, NHV102216H, NHV102217H, NHV102219H, NHV102220H, NHV102221H, NHV102222H, NHV102223H, NHV102224H, NHV102225H, NHV102226H, NHV102227H, NHV102228H, NHV102229H, NHV102230H, NHV102231H, NHV102232H, NHV102233H, NHV102234H, NHV102235H, NHV102236H, NHV102237H, NHV102238H, NHV102239H, NHV102240H, NHV102241H, NHV102242H, NHV102243H, NHV102244H, NHV102245H, NHV102246H, NHV102247H, NHV102248H, NHV102249H, NHV102250H, NHV102251H, NHV102252H, NHV102253H, NHV102254H, NHV102255H, NHV102256H, NHV102257H, NHV102258H, NHV102259H, NHV102260H, NHV102261H, NHV102262H, NHV102263H, NHV102264H, NHV102265H, NHV102266H, NHV102267H, NHV102268H, NHV102269H, NHV102270H, NHV102271H, NHV102272H, NHV102273H, NHV102274H, NHV102275H, NHV102276H, NHV102277H, NHV102278H, NHV102279H, NHV102280H, NHV102281H, NHV102282H, NHV102283H, NHV102284H, NHV102285H, NHV102286H, NHV102287H, NHV102288H, NHV102289H, NHV102290H, NHV102291H, NHV102292H, NHV102293H, NHV102294H, NHV102295H, NHV102296H, NHV102297H, NHV102298H, NHV102299H, NHV102300H, NHV102301H, NHV102302H, NHV102303H, NHV102304H, NHV102305H, NHV102306H, NHV102307H, NHV102308H, NHV102309H, NHV102310H, NHV102311H, NHV102312H, NHV102313H, NHV102314H, NHV102315H, NHV102316H, NHV102317H, NHV102318H, NHV102319H, NHV102320H, NHV102321H, NHV102322H, NHV102323H, NHV102324H, NHV102325H, NHV102326H, NHV102327H, NHV102328H, NHV102329H, NHV102330H, NHV102331H, NHV102332H, NHV102333H, NHV102334H, NHV102335H, NHV102336H, NHV102337H, NHV102338H, NHV102339H, NHV102340H, NHV102341H, NHV102342H, NHV102343H, NHV102344H, NHV102345H, NHV102347H, NHV102348H, NHV102349H, NHV102350H, NHV102351H, NHV102352H, NHV102353H, NHV102354H, NHV102355H, NHV102356H, NHV102357H, NHV102358H, NHV102359H, NHV102360H, NHV102362H, NHV102363H, NHV102364H, NHV102365H, NHV102366H, NHV102367H, NHV102368H, NHV102369H, NHV102370H, NHV102371H, NHV102372H, NHV102373H, NHV102374H, NHV102375H, NHV102376H, NHV102377H, NHV102378H, NHV102379H, NHV102380H, NHV102381H, NHV102382H, NHV102383H, NHV102384H, NHV102385H, NHV102386H, NHV102387H, NHV102388H, NHV102389H, NHV102390H, NHV102391H, NHV102392H, NHV102393H, NHV102394H, NHV102395H, NHV102396H, NHV102397H, NHV102398H, NHV102399H, NHV102400H, NHV102401H, NHV102402H, NHV102403H, NHV102404H, NHV102405H, NHV102406H, NHV102407H, NHV102408H, NHV102409H, NHV102410H, NHV102411H, NHV102412H, NHV102413H, NHV102414H, NHV102415H, NHV102416H, NHV102417H, NHV102418H, NHV102419H, NHV102420H, NHV102421H, NHV102422H, NHV102423H, NHV102424H, NHV102425H, NHV102426H, NHV102427H, NHV102428H, NHV102429H, NHV102430H, NHV102431H, NHV102432H, NHV102433H, NHV102434H, NHV102435H, NHV102436H, NHV102437H, NHV102438H, NHV102439H, NHV102440H, NHV102441H, NHV102442H, NHV102443H, NHV102444H, NHV102445H, NHV102446H, NHV102447H, NHV102448H, NHV102449H, NHV102451H, NHV102452H, NHV102453H, NHV102454H, NHV102455H, NHV102456H, NHV102457H, NHV102458H, NHV102459H, NHV102460H, NHV102461H, NHV102462H, NHV102463H, NHV102464H, NHV102465H, NHV102466H, NHV102467H, NHV102468H, NHV102469H, NHV102470H, NHV102471H, NHV102472H, NHV102473H, NHV102474H, NHV102476H, NHV102477H, NHV102478H, NHV102479H, NHV102480H, NHV102481H, NHV102482H, NHV102483H, NHV102484H, NHV102485H, NHV102486H, NHV102487H, NHV102488H, NHV102489H, NHV102490H, NHV102491H, NHV102492H, NHV102493H, NHV102494H, NHV102495H, NHV102496H, NHV102497H, NHV102498H, NHV102499H, NHV102500H, NHV102501H, NHV102502H, NHV102503H, NHV102504H, NHV102505H, NHV102506H, NHV102507H, NHV102508H, NHV102509H, NHV102510H, NHV102511H, NHV102512H, NHV102513H, NHV102514H, NHV102515H, NHV102516H, NHV102517H, NHV102518H, NHV102519H, NHV102520H, NHV102521H, NHV102522H, NHV102523H, NHV102524H, NHV102525H, NHV102526H, NHV102527H, NHV102528H, NHV102529H, NHV102530H, NHV102531H, NHV102532H, NHV102533H, NHV102534H, NHV102535H, NHV102536H, NHV102537H, NHV102538H, NHV102539H, NHV102540H, NHV102541H, NHV102542H, NHV102543H, NHV102544H, NHV102546H, NHV102547H, NHV102548H, NHV102549H, NHV102550H, NHV102551H, NHV102552H, NHV102553H, NHV102554H, NHV102555H, NHV102556H, NHV102557H, NHV102558H, NHV102559H, NHV102560H, NHV102561H, NHV102562H, NHV102563H, NHV102564H, NHV102565H, NHV102566H, NHV102567H, NHV102568H, NHV102569H, NHV102570H, NHV102571H, NHV102572H, NHV102573H, NHV102574H, NHV102575H, NHV102576H, NHV102577H, NHV102578H, NHV102579H, NHV102580H, NHV102581H, NHV102582H, NHV102583H, NHV102584H, NHV102585H, NHV102586H, NHV102587H, NHV102588H, NHV102589H, NHV102590H, NHV102591H, NHV102592H, NHV102593H, NHV102594H, NHV102595H, NHV102596H, NHV102597H, NHV102598H, NHV102599H, NHV102600H, NHV102601H, NHV102602H, NHV102603H, NHV102604H, NHV102605H, NHV102606H, NHV102607H, NHV102608H, NHV102609H, NHV102610H, NHV102611H, NHV102612H, NHV102613H, NHV102614H, NHV102615H, NHV102616H, NHV102617H, NHV102618H, NHV102619H, NHV102620H, NHV102621H, NHV102622H, NHV102623H, NHV102624H, NHV102625H, NHV102626H, NHV102627H, NHV102628H, NHV102629H, NHV102630H, NHV102631H, NHV102632H, NHV102633H, NHV102634H, NHV102635H, NHV102636H, NHV102637H, NHV102638H, NHV102639H, NHV102640H, NHV102641H, NHV102642H, NHV102643H, NHV102644H, NHV102645H, NHV102646H, NHV102647H, NHV102648H, NHV102649H, NHV102650H, NHV102651H, NHV102652H, NHV102653H, NHV102654H, NHV102655H, NHV102656H, NHV102657H, NHV102658H, NHV102659H, NHV102660H, NHV102661H, NHV102662H, NHV102663H, NHV102664H, NHV102665H, NHV102667H, NHV102668H, NHV102669H, NHV102670H, NHV102671H, NHV102672H, NHV102673H, NHV102674H, NHV102675H, NHV102676H, NHV102677H, NHV102678H, NHV102679H, NHV102680H, NHV102681H, NHV102682H, NHV102683H, NHV102684H, NHV102685H, NHV102686H, NHV102687H, NHV102688H, NHV102689H, NHV102690H, NHV102691H, NHV102692H, NHV102693H, NHV102694H, NHV102695H, NHV102696H, NHV102697H, NHV102698H, NHV102699H, NHV102700H, NHV102701H, NHV102702H, NHV102703H, NHV102704H, NHV102705H, NHV102706H, NHV102707H, NHV102708H, NHV102709H, NHV102710H, NHV102711H, NHV102712H, NHV102713H, NHV102714H, NHV102715H, NHV102716H, NHV102717H, NHV102718H, NHV102719H, NHV102720H, NHV102721H, NHV102722H, NHV102723H, NHV102724H, NHV102725H, NHV102726H, NHV102727H, NHV102728H, NHV102729H, NHV102730H, NHV102731H, NHV102732H, NHV102733H, NHV102734H, NHV102735H, NHV102736H, NHV102737H, NHV102738H, NHV102739H, NHV102740H, NHV102741H, NHV102743H, NHV102744H, NHV102745H, NHV102746H, NHV102747H, NHV102748H, NHV102749H, NHV102750H, NHV102751H, NHV102752H, NHV102753H, NHV102754H, NHV102755H, NHV102756H, NHV102757H, NHV102758H, NHV102759H, NHV102760H, NHV102761H, NHV102762H, NHV102763H, NHV102764H, NHV102765H, NHV102766H, NHV102767H, NHV102768H, NHV102769H, NHV102770H, NHV102771H, NHV102772H, NHV102773H, NHV102774H, NHV102775H, NHV102776H, NHV102777H, NHV102778H, NHV102779H, NHV102780H, NHV102781H, NHV102782H, NHV102783H, NHV102784H, NHV102785H, NHV102786H, NHV102787H, NHV102788H, NHV102789H, NHV102790H, NHV102791H, NHV102792H, NHV102793H, NHV102794H, NHV102795H, NHV102796H, NHV102797H, NHV102798H, NHV102799H, NHV102800H, NHV102801H, NHV102802H, NHV102803H, NHV102804H, NHV102805H, NHV102806H, NHV102807H, NHV102808H, NHV102809H, NHV102810H, NHV102811H, NHV102812H, NHV102813H, NHV102814H, NHV102815H, NHV102816H, NHV102817H, NHV102818H, NHV102819H, NHV102820H, NHV102821H, NHV102822H, NHV102823H, NHV102824H, NHV102825H, NHV102826H, NHV102827H, NHV102828H, NHV102829H, NHV102830H, NHV102831H, NHV102832H, NHV102833H, NHV102834H, NHV102835H, NHV102836H, NHV102837H, NHV102838H, NHV102839H, NHV102840H, NHV102841H, NHV102842H, NHV102843H, NHV102844H, NHV102845H, NHV102846H, NHV102847H, NHV102848H, NHV102849H, NHV102850H, NHV102851H, NHV102852H, NHV102853H, NHV102854H, NHV102855H, NHV102856H, NHV102857H, NHV102858H, NHV102859H, NHV102861H, NHV102862H, NHV102863H, NHV102864H, NHV102865H, NHV102866H, NHV102867H, NHV102868H, NHV102869H, NHV102870H, NHV102871H, NHV102872H, NHV102873H, NHV102874H, NHV102875H, NHV102876H, NHV102877H, NHV102878H, NHV102879H, NHV102880H, NHV102881H, NHV102882H, NHV102883H, NHV102884H, NHV102885H, NHV102886H, NHV102887H, NHV102888H, NHV102889H, NHV102890H, NHV102891H, NHV102892H, NHV102893H, NHV102894H, NHV102895H, NHV102896H, NHV102897H, NHV102898H, NHV102899H, NHV102900H, NHV102901H, NHV102902H, NHV102903H, NHV102904H, NHV102905H, NHV102906H, NHV102907H, NHV102908H, NHV102910H, NHV102911H, NHV102912H, NHV102913H, NHV102914H, NHV102915H, NHV102916H, NHV102917H, NHV102918H, NHV102919H, NHV102920H, NHV102921H, NHV102922H, NHV102923H, NHV102924H, NHV102925H, NHV102926H, NHV102927H, NHV102928H, NHV102929H, NHV102930H, NHV102931H, NHV102932H, NHV102933H, NHV102934H, NHV102935H, NHV102936H, NHV102937H, NHV102938H, NHV102939H, NHV102940H, NHV102941H, NHV102942H, NHV102943H, NHV102944H, NHV102945H, NHV102946H, NHV102947H, NHV102948H, NHV102949H, NHV102950H, NHV102951H, NHV102952H, NHV102953H, NHV102954H, NHV102955H, NHV102956H, NHV102957H, NHV102958H, NHV102959H, NHV102960H, NHV102961H, NHV102962H, NHV102963H, NHV102964H, NHV102965H, NHV102966H, NHV102967H, NHV102968H, NHV102969H, NHV102970H, NHV102971H, NHV102972H, NHV102973H, NHV102974H, NHV102975H, NHV102976H, NHV102977H, NHV102979H, NHV102980H, NHV102981H, NHV102982H, NHV102983H, NHV102984H, NHV102985H, NHV102986H, NHV102987H, NHV102988H, NHV102989H, NHV102990H, NHV102991H, NHV102992H, NHV102993H, NHV102994H, NHV102995H, NHV102996H, NHV102997H, NHV102998H, NHV102999H, NHV103000H, NHV103001H, NHV103002H, NHV103003H, NHV103004H, NHV103005H, NHV103006H, NHV103007H, NHV103008H, NHV103009H, NHV103010H, NHV103011H, NHV103012H, NHV103013H, NHV103014H, NHV103015H, NHV103016H, NHV103017H, NHV103018H, NHV103019H, NHV103020H, NHV103021H, NHV103022H, NHV103023H, NHV103024H, NHV103025H, NHV103026H, NHV103027H, NHV103028H, NHV103029H, NHV103030H, NHV103031H, NHV103032H, NHV103033H, NHV103034H, NHV103035H, NHV103036H, NHV103037H, NHV103038H, NHV103039H, NHV103040H, NHV103041H, NHV103042H, NHV103043H, NHV103044H, NHV103045H, NHV103046H, NHV103047H, NHV103048H, NHV103049H, NHV103050H, NHV103051H, NHV103052H, NHV103053H, NHV103054H, NHV103055H, NHV103056H, NHV103057H, NHV103058H, NHV103059H, NHV103060H, NHV103061H, NHV103062H, NHV103063H, NHV103064H, NHV103065H, NHV103066H, NHV103067H, NHV103068H, NHV103069H, NHV103070H, NHV103071H, NHV103072H, NHV103073H, NHV103074H, NHV103075H, NHV103076H, NHV103077H, NHV103078H, NHV103079H, NHV103080H, NHV103081H, NHV103082H, NHV103083H, NHV103084H, NHV103085H, NHV103086H, NHV103087H, NHV103088H, NHV103089H, NHV103090H, NHV103091H, NHV103092H, NHV103093H, NHV103094H, NHV103095H, NHV103096H, NHV103097H, NHV103098H, NHV103099H, NHV103100H, NHV103101H, NHV103102H, NHV103103H, NHV103104H, NHV103105H, NHV103106H, NHV103107H, NHV103108H, NHV103109H, NHV103110H, NHV103111H, NHV103112H, NHV103113H, NHV103114H, NHV103115H, NHV103116H, NHV103117H, NHV103118H, NHV103119H, NHV103120H, NHV103121H, NHV103122H, NHV103123H, NHV103124H, NHV103125H, NHV103126H, NHV103127H, NHV103128H, NHV103129H, NHV103130H, NHV103131H, NHV103132H, NHV103133H, NHV103134H, NHV103135H, NHV103136H, NHV103137H, NHV103138H, NHV103139H, NHV103140H, NHV103141H, NHV103142H, NHV103143H, NHV103144H, NHV103145H, NHV103146H, NHV103147H, NHV103148H, NHV103149H, NHV103150H, NHV103151H, NHV103152H, NHV103153H, NHV103154H, NHV103155H, NHV103156H, NHV103157H, NHV103158H, NHV103159H, NHV103160H, NHV103161H, NHV103162H, NHV103163H, NHV103164H, NHV103165H, NHV103166H, NHV103167H, NHV103168H, NHV103169H, NHV103170H, NHV103171H, NHV103172H, NHV103173H, NHV103174H, NHV103175H, NHV103176H, NHV103177H, NHV103178H, NHV103179H, NHV103180H, NHV103181H, NHV103182H, NHV103183H, NHV103184H, NHV103185H, NHV103186H, NHV103187H, NHV103188H, NHV103189H, NHV103190H, NHV103191H, NHV103192H, NHV103193H, NHV103194H, NHV103195H, NHV103196H, NHV103197H, NHV103198H, NHV103199H, NHV103200H, NHV103201H, NHV103202H, NHV103203H, NHV103204H, NHV103205H, NHV103206H, NHV103207H, NHV103208H, NHV103209H, NHV103210H, NHV103211H, NHV103212H, NHV103213H, NHV103214H, NHV103215H, NHV103216H, NHV103217H, NHV103218H, NHV103219H, NHV103220H, NHV103221H, NHV103222H, NHV103223H, NHV103224H, NHV103225H, NHV103226H, NHV103227H, NHV103228H, NHV103229H, NHV103230H, NHV103231H, NHV103232H, NHV103233H, NHV103234H, NHV103235H, NHV103236H, NHV103237H, NHV103238H, NHV103239H, NHV103240H, NHV103241H, NHV103242H, NHV103243H, NHV103244H, NHV103245H, NHV103246H, NHV103247H, NHV103248H, NHV103249H, NHV103250H, NHV103251H, NHV103252H, NHV103253H, NHV103254H, NHV103255H, NHV103256H, NHV103257H, NHV103258H, NHV103259H, NHV103260H, NHV103261H, NHV103262H, NHV103263H, NHV103264H, NHV103265H, NHV103266H, NHV103267H, NHV103268H, NHV103269H, NHV103270H, NHV103271H, NHV103272H, NHV103273H, NHV103274H, NHV103275H, NHV103276H, NHV103277H, NHV103278H, NHV103279H, NHV103280H, NHV103282H, NHV103283H, NHV103284H, NHV103285H, NHV103286H, NHV103287H, NHV103288H, NHV103289H, NHV103290H, NHV103291H, NHV103292H, NHV103293H, NHV103294H, NHV103295H, NHV103296H, NHV103297H, NHV103298H, NHV103299H, NHV103300H, NHV103301H, NHV103302H, NHV103303H, NHV103304H, NHV103305H, NHV103306H, NHV103307H, NHV103308H, NHV103309H, NHV103310H, NHV103311H, NHV103312H, NHV103313H, NHV103314H, NHV103315H, NHV103316H, NHV103317H, NHV103318H, NHV103319H, NHV103320H, NHV103321H, NHV103322H, NHV103323H, NHV103324H, NHV103325H, NHV103326H, NHV103327H, NHV103328H, NHV103329H, NHV103330H, NHV103331H, NHV103332H, NHV103333H, NHV103334H, NHV103335H, NHV103336H, NHV103337H, NHV103338H, NHV103339H, NHV103340H, NHV103341H, NHV103342H, NHV103343H, NHV103344H, NHV103345H, NHV103346H, NHV103347H, NHV103348H, NHV103349H, NHV103350H, NHV103351H, NHV103352H, NHV103353H, NHV103354H, NHV103355H, NHV103356H, NHV103357H, NHV103358H, NHV103359H, NHV103360H, NHV103361H, NHV103362H, NHV103363H, NHV103364H, NHV103365H, NHV103366H, NHV103367H, NHV103368H, NHV103369H, NHV103370H, NHV103371H, NHV103372H, NHV103373H, NHV103374H, NHV103375H, NHV103376H, NHV103377H, NHV103378H, NHV103379H, NHV103380H, NHV103381H, NHV103382H, NHV103383H, NHV103384H, NHV103385H, NHV103386H, NHV103387H, NHV103388H, NHV103389H, NHV103390H, NHV103391H, NHV103392H, NHV103393H, NHV103394H, NHV103395H, NHV103396H, NHV103397H, NHV103398H, NHV103399H, NHV103400H, NHV103401H, NHV103402H, NHV103403H, NHV103404H, NHV103405H, NHV103406H, NHV103407H, NHV103408H, NHV103409H, NHV103410H, NHV103411H, NHV103412H, NHV103413H, NHV103414H, NHV103415H, NHV103416H, NHV103417H, NHV103418H, NHV103419H, NHV103420H, NHV103421H, NHV103422H, NHV103423H, NHV103424H, NHV103425H, NHV103426H, NHV103427H, NHV103428H, NHV103429H, NHV103430H, NHV103431H, NHV103432H, NHV103433H, NHV103434H, NHV103435H, NHV103436H, NHV103437H, NHV103438H, NHV103439H, NHV103440H, NHV103441H, NHV103442H, NHV103443H, NHV103444H, NHV103445H, NHV103446H, NHV103447H, NHV103448H, NHV103449H, NHV103450H, NHV103451H, NHV103452H, NHV103453H, NHV103454H, NHV103455H, NHV103456H, NHV103457H, NHV103458H, NHV103459H, NHV103460H, NHV103461H, NHV103462H, NHV103463H, NHV103464H, NHV103465H, NHV103466H, NHV103467H, NHV103468H, NHV103469H, NHV103470H, NHV103471H, NHV103472H, NHV103473H, NHV103475H, NHV103477H, NHV103478H, NHV103479H, NHV103480H, NHV103481H, NHV103482H, NHV103483H, NHV103484H, NHV103485H, NHV103486H, NHV103487H, NHV103489H, NHV103490H, NHV103491H, NHV103492H, NHV103493H, NHV103494H, NHV103495H, NHV103496H, NHV103497H, NHV103498H, NHV103499H, NHV103500H, NHV103501H, NHV103502H, NHV103503H, NHV103504H, NHV103505H, NHV103506H, NHV103507H, NHV103508H, NHV103509H, NHV103510H, NHV103511H, NHV103512H, NHV103513H, NHV103514H, NHV103515H, NHV103516H, NHV103517H, NHV103518H, NHV103519H, NHV103520H, NHV103521H, NHV103522H, NHV103523H, NHV103524H, NHV103525H, NHV103526H, NHV103527H, NHV103528H, NHV103529H, NHV103530H, NHV103531H, NHV103532H, NHV103534H, NHV103535H, NHV103536H, NHV103537H, NHV103538H, NHV103539H, NHV103540H, NHV103541H, NHV103542H, NHV103543H, NHV103544H, NHV103545H, NHV103546H, NHV103547H, NHV103548H, NHV103549H, NHV103550H, NHV103551H, NHV103552H, NHV103553H, NHV103554H, NHV103555H, NHV103556H, NHV103557H, NHV103558H, NHV103559H, NHV103560H, NHV103562H, NHV103563H, NHV103564H, NHV103565H, NHV103566H, NHV103567H, NHV103568H, NHV103569H, NHV103570H, NHV103571H, NHV103572H, NHV103573H, NHV103574H, NHV103575H, NHV103576H, NHV103577H, NHV103578H, NHV103579H, NHV103580H, NHV103581H, NHV103582H, NHV103583H, NHV103584H, NHV103585H, NHV103586H, NHV103587H, NHV103588H, NHV103589H, NHV103590H, NHV103591H, NHV103592H, NHV103593H, NHV103594H, NHV103595H, NHV103596H, NHV103597H, NHV103598H, NHV103599H, NHV103600H, NHV103601H, NHV103602H, NHV103603H, NHV103604H, NHV103605H, NHV103606H, NHV103607H, NHV103608H, NHV103609H, NHV103610H, NHV103611H, NHV103612H, NHV103613H, NHV103614H, NHV103615H, NHV103616H, NHV103617H, NHV103618H, NHV103619H, NHV103620H, NHV103621H, NHV103622H, NHV103623H, NHV103624H, NHV103625H, NHV103626H, NHV103627H, NHV103628H, NHV103629H, NHV103630H, NHV103631H, NHV103632H, NHV103633H, NHV103634H, NHV103635H, NHV103636H, NHV103637H, NHV103638H, NHV103639H, NHV103640H, NHV103641H, NHV103642H, NHV103643H, NHV103644H, NHV103645H, NHV103646H, NHV103647H, NHV103648H, NHV103649H, NHV103650H, NHV103651H, NHV103652H, NHV103653H, NHV103654H, NHV103655H, NHV103657H, NHV103658H, NHV103659H, NHV103660H, NHV103661H, NHV103662H, NHV103663H, NHV103664H, NHV103665H, NHV103667H, NHV103668H, NHV103669H, NHV103670H, NHV103671H, NHV103672H, NHV103673H, NHV103674H, NHV103675H, NHV103676H, NHV103677H, NHV103678H, NHV103679H, NHV103680H, NHV103681H, NHV103682H, NHV103683H, NHV103684H, NHV103685H, NHV103686H, NHV103687H, NHV103688H, NHV103689H, NHV103690H, NHV103691H, NHV103692H, NHV103693H, NHV103694H, NHV103695H, NHV103696H, NHV103697H, NHV103698H, NHV103699H, NHV103700H, NHV103701H, NHV103702H, NHV103703H, NHV103704H, NHV103705H, NHV103706H, NHV103707H, NHV103708H, NHV103710H, NHV103711H, NHV103712H, NHV103713H, NHV103714H, NHV103715H, NHV103716H, NHV103717H, NHV103718H, NHV103719H, NHV103720H, NHV103721H, NHV103722H, NHV103723H, NHV103724H, NHV103725H, NHV103726H, NHV103727H, NHV103728H, NHV103729H, NHV103730H, NHV103731H, NHV103732H, NHV103733H, NHV103734H, NHV103735H, NHV103736H, NHV103737H, NHV103738H, NHV103739H, NHV103740H, NHV103741H, NHV103742H, NHV103743H, NHV103744H, NHV103745H, NHV103746H, NHV103747H, NHV103749H, NHV103750H, NHV103751H, NHV103752H, NHV103753H, NHV103754H, NHV103755H, NHV103756H, NHV103757H, NHV103758H, NHV103759H, NHV103760H, NHV103761H, NHV103762H, NHV103763H, NHV103764H, NHV103765H, NHV103766H, NHV103767H, NHV103768H, NHV103769H, NHV103770H, NHV103771H, NHV103772H, NHV103773H, NHV103774H, NHV103775H, NHV103776H, NHV103777H, NHV103778H, NHV103780H, NHV103781H, NHV103782H, NHV103783H, NHV103784H, NHV103785H, NHV103786H, NHV103787H, NHV103788H, NHV103789H, NHV103790H, NHV103791H, NHV103792H, NHV103793H, NHV103794H, NHV103795H, NHV103796H, NHV103797H, NHV103798H, NHV103799H, NHV103800H, NHV103801H, NHV103802H, NHV103803H, NHV103804H, NHV103805H, NHV103806H, NHV103807H, NHV103808H, NHV103809H, NHV103810H, NHV103811H, NHV103812H, NHV103813H, NHV103814H, NHV103815H, NHV103816H, NHV103817H, NHV103818H, NHV103819H, NHV103820H, NHV103821H, NHV103822H, NHV103823H, NHV103824H, NHV103825H, NHV103826H, NHV103827H, NHV103828H, NHV103829H, NHV103830H, NHV103831H, NHV103832H, NHV103833H, NHV103834H, NHV103835H, NHV103836H, NHV103837H, NHV103838H, NHV103839H, NHV103841H, NHV103842H, NHV103843H, NHV103844H, NHV103845H, NHV103846H, NHV103847H, NHV103848H, NHV103849H, NHV103850H, NHV103851H, NHV103852H, NHV103853H, NHV103854H, NHV103855H, NHV103856H, NHV103857H, NHV103858H, NHV103859H, NHV103860H, NHV103861H, NHV103862H, NHV103863H, NHV103864H, NHV103865H, NHV103866H, NHV103867H, NHV103868H, NHV103869H, NHV103870H, NHV103871H, NHV103872H, NHV103873H, NHV103874H, NHV103875H, NHV103876H, NHV103877H, NHV103878H, NHV103879H, NHV103880H, NHV103881H, NHV103882H, NHV103883H, NHV103884H, NHV103885H, NHV103886H, NHV103887H, NHV103888H, NHV103889H, NHV103890H, NHV103891H, NHV103892H, NHV103893H, NHV103894H, NHV103895H, NHV103896H, NHV103897H, NHV103898H, NHV103899H, NHV103900H, NHV103901H, NHV103902H, NHV103903H, NHV103904H, NHV103905H, NHV103906H, NHV103907H, NHV103908H, NHV103909H, NHV103910H, NHV103911H, NHV103912H, NHV103913H, NHV103914H, NHV103915H, NHV103916H, NHV103917H, NHV103918H, NHV103919H, NHV103920H, NHV103921H, NHV103922H, NHV103923H, NHV103924H, NHV103925H, NHV103926H, NHV103927H, NHV103928H, NHV103929H, NHV103930H, NHV103931H, NHV103932H, NHV103933H, NHV103934H, NHV103935H, NHV103936H, NHV103937H, NHV103938H, NHV103939H, NHV103940H, NHV103941H, NHV103942H, NHV103943H, NHV103944H, NHV103945H, NHV103946H, NHV103947H, NHV103948H, NHV103949H, NHV103950H, NHV103951H, NHV103952H, NHV103953H, NHV103955H, NHV103956H, NHV103957H, NHV103958H, NHV103959H, NHV103960H, NHV103961H, NHV103962H, NHV103963H, NHV103964H, NHV103965H, NHV103966H, NHV103967H, NHV103968H, NHV103969H, NHV103970H, NHV103971H, NHV103972H, NHV103973H, NHV103974H, NHV103975H, NHV103976H, NHV103977H, NHV103978H, NHV103979H, NHV103980H, NHV103981H, NHV103982H, NHV103983H, NHV103984H, NHV103985H, NHV103986H, NHV103987H, NHV103988H, NHV103989H, NHV103990H, NHV103991H, NHV103992H, NHV103993H, NHV103994H, NHV103995H, NHV103996H, NHV103997H, NHV103998H, NHV103999H, NHV104000H, NHV104001H, NHV104002H, NHV104003H, NHV104004H, NHV104005H, NHV104006H, NHV104007H, NHV104008H, NHV104009H, NHV104010H, NHV104011H, NHV104012H, NHV104013H, NHV104014H, NHV104015H, NHV104016H, NHV104017H, NHV104018H, NHV104019H, NHV104020H, NHV104021H, NHV104022H, NHV104023H, NHV104024H, NHV104025H, NHV104026H, NHV104027H, NHV104028H, NHV104029H, NHV104030H, NHV104032H, NHV104033H, NHV104034H, NHV104035H, NHV104036H, NHV104037H, NHV104038H, NHV104039H, NHV104040H, NHV104041H, NHV104042H, NHV104043H, NHV104044H, NHV104045H, NHV104046H, NHV104047H, NHV104048H, NHV104049H, NHV104050H, NHV104051H, NHV104052H, NHV104053H, NHV104054H, NHV104055H, NHV104056H, NHV104057H, NHV104058H, NHV104059H, NHV104060H, NHV104061H, NHV104062H, NHV104063H, NHV104064H, NHV104065H, NHV104066H, NHV104067H, NHV104068H, NHV104069H, NHV104070H, NHV104071H, NHV104072H, NHV104073H, NHV104074H, NHV104075H, NHV104076H, NHV104077H, NHV104078H, NHV104079H, NHV104080H, NHV104081H, NHV104082H, NHV104083H, NHV104084H, NHV104085H, NHV104086H, NHV104087H, NHV104088H, NHV104089H, NHV104090H, NHV104091H, NHV104092H, NHV104093H, NHV104094H, NHV104095H, NHV104096H, NHV104097H, NHV104099H, NHV104100H, NHV104101H, NHV104102H, NHV104103H, NHV104104H, NHV104105H, NHV104106H, NHV104107H, NHV104108H, NHV104109H, NHV104110H, NHV104111H, NHV104112H, NHV104113H, NHV104114H, NHV104115H, NHV104116H, NHV104117H, NHV104118H, NHV104120H, NHV104121H, NHV104122H, NHV104123H, NHV104124H, NHV104125H, NHV104126H, NHV104127H, NHV104128H, NHV104129H, NHV104130H, NHV104131H, NHV104132H, NHV104133H, NHV104134H, NHV104135H, NHV104136H, NHV104137H, NHV104138H, NHV104140H, NHV104141H, NHV104142H, NHV104143H, NHV104144H, NHV104145H, NHV104146H, NHV104147H, NHV104148H, NHV104149H, NHV104150H, NHV104151H, NHV104152H, NHV104153H, NHV104154H, NHV104155H, NHV104156H, NHV104157H, NHV104158H, NHV104159H, NHV104160H, NHV104161H, NHV104162H, NHV104163H, NHV104164H, NHV104165H, NHV104166H, NHV104167H, NHV104168H, NHV104169H, NHV104170H, NHV104171H, NHV104172H, NHV104173H, NHV104174H, NHV104175H, NHV104176H, NHV104178H, NHV104179H, NHV104180H, NHV104181H, NHV104182H, NHV104183H, NHV104184H, NHV104185H, NHV104186H, NHV104187H, NHV104188H, NHV104189H, NHV104190H, NHV104191H, NHV104192H, NHV104193H, NHV104194H, NHV104195H, NHV104196H, NHV104197H, NHV104198H, NHV104199H, NHV104200H, NHV104201H, NHV104202H, NHV104203H, NHV104204H, NHV104205H, NHV104206H, NHV104207H, NHV104208H and NHV104209H. |
| FEI Number |
2182207
|
Recalling Firm/ Manufacturer |
Medtronic Neuromodulation 7000 Central Ave NE Minneapolis MN 55432-3568
|
| For Additional Information Contact | 763-526-9684 |
Manufacturer Reason for Recall | Medtronic is in the process of updating the Enterra Therapy labeling to include information regarding the risks of bowel obstruction and/or perforation. Medtronic has received 15 reports since November 2002 of bowel obstruction and/or perforation of the bowel associated with Enterra Therapy System leads. Of the 15 reported events, 11 required surgical intervention (all patients recovered withou |
FDA Determined Cause 2 | Device Design |
| Action | Medtronic Neuromodulation issued a "Medical Device Correction" letter beginning August 25, 2009 addressed to Healthcare Professionals. The letter described the Nature of the Issue, Scope, Potential Severity of the Issue, Enterra Therapy System labeling updating information warnings, Recommendations and other Additional Information.
For further information, contact Medronic Neuromodulation Technical Services at 1-800-707-0933. |
| Quantity in Commerce | 1909 (1095 US, 814 OUS) |
| Distribution | Worldwide Distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| HDE Database | HDEs with Product Code = LNQ
|
|
|
|