• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Heartstart

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Heartstartsee related information
Date Initiated by FirmNovember 02, 2009
Date PostedMarch 23, 2010
Recall Status1 Terminated 3 on May 23, 2013
Recall NumberZ-1194-2010
Recall Event ID 53816
510(K)NumberK020715 
Product Classification Automated External Defibrillators (Non-Wearable) - Product Code MKJ
ProductPhilips Heartstart (HS1) Onsite Automated External Defibrillator (AED) Catalog number M5066A
Code Information DOMESTIC UNIT SERIAL NUMBERS - A09J-01665, A09J-01237, A09J-02203, A09J-01279, A09J-01353, A09J-01284, A09J-01227, A09J-01593, A09J-01236, A09J-01238, A09J-01667, A09J-02140, A09J-02230, A09J-02238, A09J-01351, A09J-02218, A09J-01183, A09J-01232, A09J-01275, A09J-01277, A09J-01280, A09J-01283, A09J-01285, A09J-01286, A09J-01289, A09J-01290, A09J-01291, A09J-01654, A09J-01663, A09J-01669, A09J-01670, A09J-01727, A09J-02002, A09J-02139, A09J-02147, A09J-02148, A09J-02149, A09J-02202, A09J-02208, A09J-02211, A09J-02213, A09J-02219, A09J-02222, A09J-02224, A09J-02225, A09J-02226, A09J-02227, A09J-02228, A09J-02229, A09J-01354, A09J-01656, A09J-01664, A09J-02201, A09J-01229, A09J-01230, A09J-01287, A09J-01651, A09J-01652, A09J-01653, A09J-01657, A09J-01659, A09J-01660, A09J-01662, A09J-01666, A09J-01668, A09J-01225, A09J-01226, A09J-01231, A09J-01235, A09J-01329, A09J-01339, A09J-01356, A09J-01358, A09J-01655, A09J-01658, A09J-01661, A09J-01674, A09J-01675, A09J-01701, A09J-01708, A09J-01710, A09J-02102, A09J-02138, A09J-01153, A09J-01424, A09J-01426, A09J-01732, A09J-01733, A09J-01734, A09J-01736, A09J-02214, A09J-02221, A09J-02239, A09J-01292, A09J-01296, A09J-01297, A09J-01331, A09J-02235, and A09J-02237.   INTERNATIONAL UNIT SERIAL NUMBERS - A09J-02145, A09J-02204, A09J-01298, A09J-02207, A09J-02209, A09J-02223, A09J-02236, A09J-02142, A09J-02210, A09J-01829, A09J-01839, A09J-01847, A09J-01849, A09J-01860, A09J-01873, A09J-01874, A09J-01876, A09J-01897, A09J-01941, A09J-01942, A09J-01943, A09J-01948, A09J-02012, A09J-02272, A09J-02304, A09J-02348, A09J-02501, A09J-02502, A09J-02505, A09J-02510, A09J-02512, A09J-02614, A09J-02629, A09J-02630, A09J-02634, A09J-02668, A09J-02681, A09J-01878, A09J-01880, A09J-01881, A09J-02515, A09J-02545, A09J-02609, A09J-02686, A09J-01879, A09J-01918, A09J-02618, A09J-01894, A09J-02028, A09J-02282, A09J-02504, A09J-02513, A09J-02698, A09J-01883, A09J-02022, A09J-02047, A09J-02174, A09J-02178, A09J-02184, A09J-02186, A09J-02193, A09J-02356, A09J-02396, A09J-02411, A09J-02412, A09J-02414, A09J-02419, A09J-02421, A09J-02422, A09J-02423, A09J-02425, A09J-02427, A09J-02428, A09J-02429, A09J-02432, A09J-02442, A09J-02445, A09J-02446, A09J-02447, A09J-02498, A09J-02506, A09J-02560, A09J-02576, A09J-02577, A09J-02579, A09J-02580, A09J-02581, A09J-02584, A09J-02587, A09J-02598, A09J-02622, A09J-02646, A09J-02706, A09J-02709, A09J-02710, A09J-02711, A09J-02712, A09J-02713, A09J-02714, A09J-02722, A09J-02724, A09J-02735, A09J-02737, A09J-02738, A09J-02739, A09J-02748, A09J-02749, A09J-02750, A09J-02751, A09J-02752, A09J-02753, A09J-02754, A09J-02755, A09J-02756, A09J-02757, A09J-02758, A09J-02759, A09J-02760, A09J-02761, A09J-02763, A09J-02764, A09J-02766, A09J-02767, A09J-02768, A09J-02769, A09J-02770, A09J-02771, A09J-02772, A09J-02773, A09J-02774, A09J-02780, A09J-02782, A09J-02791, A09J-02795, A09J-02797, A09J-02798, A09J-02799, A09J-02800, A09J-02814, A09J-02815, A09J-02817, A09J-02818, A09J-02819, A09J-02820, A09J-02825, A09J-02848, A09J-02863, A09J-02867, A09J-02868, A09J-02869, A09J-02870, A09J-02871, A09J-02872, A09J-02873, A09J-02874, A09J-02875, A09J-02876, A09J-02878, A09J-02879, A09J-02880, A09J-02881, A09J-02882, A09J-02883, A09J-02884, A09J-02885, A09J-02886, A09J-02887, A09J-02888, A09J-02889, A09J-02890, A09J-02891, A09J-02892, A09J-02893, A09J-02894, A09J-02895, A09J-02896, A09J-02897, A09J-02898, A09J-02899, A09J-02900, A09J-02901, A09J-02902, A09J-02904, A09J-02905, A09J-02906, A09J-02907, A09J-02910, A09J-02911, A09J-02912, A09J-02914, A09J-02930, A09J-02931, A09J-02932, A09J-02934, A09J-02935, A09J-02937, A09J-02950, A09J-02951, A09J-02952, A09J-02956, A09J-02957, A09J-02959, A09J-02961, A09J-02962, A09J-02963, A09J-02964, A09J-02965, A09J-03001, A09J-03002, A09J-03003, A09J-03004, A09J-03005, A09J-03006, A09J-03007, A09J-03018, A09J-03019, A09J-03024, A09J-03026, A09J-03031, A09J-03032, A09J-03033, A09J-03034, A09J-03037, A09J-03038, A09J-03039, A09J-03040, A09J-03041, A09J-03049.A09J-03050, A09J-03051, A09J-03061, A09J-03066, A09J-00901, A09J-00902, A09J-00903, A09J-00904, A09J-00905, A09J-00906, A09J-00908, A09J-00910, A09J-00911, A09J-00912, A09J-00913, A09J-00914, A09J-00915, A09J-00917, A09J-00920, A09J-00921, A09J-00922, A09J-00925, A09J-00927, A09J-00929, A09J-00931, A09J-00932, A09J-00933, A09J-00934, A09J-00935, A09J-00936, A09J-00937, A09J-00938, A09J-00939, A09J-00941, A09J-00942, A09J-00943, A09J-00944, A09J-00945, A09J-00946, A09J-00947, A09J-00949, A09J-00950, A09J-01116, A09J-01117, A09J-01119, A09J-01120, A09J-01125, A09J-01151, A09J-01155, A09J-01156, A09J-01157, A09J-01158, A09J-01160, A09J-01163, A09J-01164, A09J-01168, A09J-01179, A09J-01180, A09J-01184, A09J-01185, A09J-01186, A09J-01187, A09J-01188, A09J-01189, A09J-01190, A09J-01191, A09J-01193, A09J-01194, A09J-01196, A09J-01197, A09J-01199, A09J-01200, A09J-01228, A09J-01233, A09J-01234, A09J-01239, A09J-01240, A09J-01241, A09J-01243, A09J-01248, A09J-01251, A09J-01252, A09J-01253, A09J-01254, A09J-01255, A09J-01257, A09J-01258, A09J-01259, A09J-01260, A09J-01261, A09J-01262, A09J-01263, A09J-01264, A09J-01265, A09J-01266, A09J-01267, A09J-01272, A09J-01273, A09J-01288, A09J-01293, A09J-01294, A09J-01295, A09J-01299, A09J-01300, A09J-01301, A09J-01302, A09J-01303, A09J-01311, A09J-01313, A09J-01316, A09J-01318, A09J-01319, A09J-01320, A09J-01321, A09J-01322, A09J-01323, A09J-01324, A09J-01325, A09J-01326, A09J-01327, A09J-01328, A09J-01332, A09J-01333, A09J-01334, A09J-01335, A09J-01336, A09J-01337, A09J-01338, A09J-01341, A09J-01342, A09J-01343, A09J-01344, A09J-01345, A09J-01346, A09J-01347, A09J-01348, A09J-01350, A09J-01352, A09J-01357, A09J-01359, A09J-01360, A09J-01361, A09J-01363, A09J-01364, A09J-01365, A09J-01366, A09J-01367, A09J-01368, A09J-01369, A09J-01370, A09J-01371, A09J-01372, A09J-01373, A09J-01374, A09J-01375, A09J-01376, A09J-01377, A09J-01378, A09J-01379, A09J-01380, A09J-01381, A09J-01382, A09J-01383, A09J-01384, A09J-01385, A09J-01386, A09J-01387, A09J-01388, A09J-01389, A09J-01390, A09J-01391, A09J-01392, A09J-01393, A09J-01394, A09J-01395, A09J-01396, A09J-01397, A09J-01398, A09J-01399, A09J-01400, A09J-01422, A09J-01430, A09J-01445, A09J-01446, A09J-01447, A09J-01448, A09J-01450, A09J-01526, A09J-01527, A09J-01528, A09J-01529, A09J-01530, A09J-01531, A09J-01532, A09J-01533, A09J-01534, A09J-01535, A09J-01536, A09J-01537, A09J-01538, A09J-01539, A09J-01540, A09J-01541, A09J-01542, A09J-01543, A09J-01544, A09J-01545, A09J-01546, A09J-01547, A09J-01548, A09J-01549, A09J-01550, A09J-01551, A09J-01552, A09J-01553, A09J-01555, A09J-01556, A09J-01558, A09J-01559, A09J-01583, A09J-01586, A09J-01595, A09J-01596, A09J-01599, A09J-01600, A09J-01602, A09J-01605, A09J-01607, A09J-01608, A09J-01609, A09J-01612, A09J-01620, A09J-01624, A09J-01628, A09J-01629, A09J-01630, A09J-01632, A09J-01633, A09J-01636, A09J-01637, A09J-01638, A09J-01639, A09J-01640, A09J-01641, A09J-01642, A09J-01646, A09J-01648, A09J-01649, A09J-01676, A09J-01677, A09J-01678, A09J-01679, A09J-01680, A09J-01681, A09J-01682, A09J-01683, A09J-01684, A09J-01685, A09J-01686, A09J-01687, A09J-01688, A09J-01690, A09J-01691, A09J-01692, A09J-01693, A09J-01694, A09J-01695, A09J-01696, A09J-01697, A09J-01702, A09J-01703, A09J-01704, A09J-01705, A09J-01706, A09J-01707, A09J-01709, A09J-01714, A09J-01716, A09J-01723, A09J-01724, A09J-01735, A09J-01737, A09J-01738, A09J-01739, A09J-01740, A09J-01741, A09J-01742, A09J-01743, A09J-01744, A09J-01745, A09J-01746, A09J-01747, A09J-01748, A09J-01750, A09J-01751, A09J-01752, A09J-01753, A09J-01756, A09J-01757, A09J-01759, A09J-01760, A09J-01763, A09J-01764, A09J-01765, A09J-01766, A09J-01767, A09J-01768, A09J-01769, A09J-01771, A09J-01774, A09J-01775, A09J-01776, A09J-01777, A09J-01778, A09J-01779, A09J-01780, A09J-01781, A09J-01782, A09J-01783, A09J-01784, A09J-01785, A09J-01786, A09J-01787, A09J-01788, A09J-01789, A09J-01790, A09J-01794, A09J-01795, A09J-01800, A09J-01801, A09J-01802, A09J-01804, A09J-01805, A09J-01806, A09J-01807, A09J-01808, A09J-01809, A09J-01810, A09J-01811, A09J-01812, A09J-01813, A09J-01814, A09J-01815, A09J-01816, A09J-01817, A09J-01818, A09J-01819, A09J-01821, A09J-01822, A09J-01823, A09J-01824, A09J-01856, A09J-01867, A09J-01882, A09J-01884, A09J-01885, A09J-01886, A09J-01887, A09J-01888, A09J-01889, A09J-01890, A09J-01892, A09J-01893, A09J-01895, A09J-01900, A09J-01905, A09J-01907, A09J-01913, A09J-01916, A09J-01917, A09J-01919, A09J-01920, A09J-01921, A09J-01922, A09J-01923, A09J-01924, A09J-01925, A09J-01951, A09J-01953, A09J-01954, A09J-01955, A09J-01957, A09J-01958, A09J-01959, A09J-01960, A09J-01961, A09J-01962, A09J-01964, A09J-01965, A09J-01966, A09J-01969, A09J-01970, A09J-01975, A09J-01977, A09J-01980, A09J-01982, A09J-01983, A09J-01984, A09J-01985, A09J-01986, A09J-01987, A09J-01988, A09J-01990, A09J-01991, A09J-01994, A09J-01995, A09J-01996, A09J-02026, A09J-02043, A09J-02044, A09J-02045, A09J-02051, A09J-02052, A09J-02055, A09J-02072, A09J-02074, A09J-02075, A09J-02081, A09J-02083, A09J-02086, A09J-02087, A09J-02088, A09J-02089, A09J-02091, A09J-02093, A09J-02096, A09J-02097, A09J-02104, A09J-02108, A09J-02110, A09J-02111, A09J-02113, A09J-02114, A09J-02116, A09J-02117, A09J-02118, A09J-02119, A09J-02122, A09J-02123, A09J-02124, A09J-02125, A09J-02127, A09J-02129, A09J-02130, A09J-02133, A09J-02179, A09J-02181, A09J-02187, A09J-02189, A09J-02190, A09J-02195, A09J-02196, A09J-02198, A09J-02212, A09J-02240, A09J-02241, A09J-02242, A09J-02243, A09J-02244, A09J-02245, A09J-02246, A09J-02247, A09J-02248, A09J-02249, A09J-02250, A09J-02253, A09J-02257, A09J-02258, A09J-02259, A09J-02260, A09J-02261, A09J-02262, A09J-02263, A09J-02264, A09J-02265, A09J-02266, A09J-02267, A09J-02270, A09J-02271, A09J-02273, A09J-02274, A09J-02275, A09J-02276, A09J-02278, A09J-02279, A09J-02280, A09J-02284, A09J-02285, A09J-02287, A09J-02289, A09J-02291, A09J-02292, A09J-02293, A09J-02294, A09J-02295, A09J-02296, A09J-02297, A09J-02298, A09J-02301, A09J-02303, A09J-02305, A09J-02308, A09J-02310, A09J-02311, A09J-02312, A09J-02313, A09J-02317, A09J-02319, A09J-02320, A09J-02321, A09J-02322, A09J-02323, A09J-02324, A09J-02325, A09J-02326, A09J-02327, A09J-02328, A09J-02329, A09J-02330, A09J-02331, A09J-02332, A09J-02333, A09J-02334, A09J-02335, A09J-02336, A09J-02337, A09J-02338, A09J-02339, A09J-02340, A09J-02341, A09J-02342, A09J-02343, A09J-02344, A09J-02345, A09J-02346, A09J-02347, A09J-02350, A09J-02352, A09J-02353, A09J-02354, A09J-02355, A09J-02358, A09J-02359, A09J-02362, A09J-02363, A09J-02364, A09J-02365, A09J-02366, A09J-02367, A09J-02370, A09J-02371, A09J-02372, A09J-02374, A09J-02376, A09J-02377, A09J-02378, A09J-02379, A09J-02381, A09J-02385, A09J-02389, A09J-02390, A09J-02391, A09J-02392, A09J-02394, A09J-02395, A09J-02397, A09J-02398, A09J-02399, A09J-02413, A09J-02415, A09J-02416, A09J-02417, A09J-02418, A09J-02424, A09J-02426, A09J-02430, A09J-02431, A09J-02433, A09J-02434, A09J-02435, A09J-02436, A09J-02437, A09J-02438, A09J-02439, A09J-02440, A09J-02441, A09J-02443, A09J-02444, A09J-02448, A09J-02449, A09J-02508, A09J-02509, A09J-02511, A09J-02557, A09J-02573, A09J-02574, A09J-02575, A09J-02578, A09J-02582, A09J-02583, A09J-02585, A09J-02586, A09J-02588, A09J-02589, A09J-02590, A09J-02591, A09J-02592, A09J-02593, A09J-02594, A09J-02595, A09J-02596, A09J-02597, A09J-02599, A09J-02621, A09J-02623, A09J-02648, A09J-02651, A09J-02652, A09J-02653, A09J-02654, A09J-02655, A09J-02656, A09J-02657, A09J-02658, A09J-02659, A09J-02660, A09J-02662, A09J-02663, A09J-02664, A09J-02665, A09J-02666, A09J-02667, A09J-02669, A09J-02670, A09J-02673, A09J-02674, A09J-02676, A09J-02677, A09J-02678, A09J-02679, A09J-02680, A09J-02682, A09J-02684, A09J-02685, A09J-02687, A09J-02688, A09J-02689, A09J-02690, A09J-02692, A09J-02694, A09J-02695, A09J-02696, A09J-02697, A09J-02699, A09J-02701, A09J-02702, A09J-02703, A09J-02704, A09J-02705, A09J-02707, A09J-02708, A09J-02715, A09J-02716, A09J-02717, A09J-02718, A09J-02719, A09J-02720, A09J-02721, A09J-02723, A09J-02725, A09J-02726, A09J-02728, A09J-02729, A09J-02730, A09J-02731, A09J-02732, A09J-02733, A09J-02734, A09J-02736, A09J-02740, A09J-02741, A09J-02742, A09J-02743, A09J-02744, A09J-02745, A09J-02746, A09J-02747, A09J-02762, A09J-02765, A09J-02776, A09J-02777, A09J-02778, A09J-02779, A09J-02781, A09J-02783, A09J-02786, A09J-02794, A09J-02796, A09J-02801, A09J-02802, A09J-02803, A09J-02804, A09J-02805, A09J-02806, A09J-02816, A09J-02839, A09J-02840, A09J-02843, A09J-02844, A09J-02845, A09J-02846, A09J-02847, A09J-02852, A09J-02854, A09J-02858, A09J-02859, A09J-02860, A09J-02861, A09J-02862, A09J-02865, A09J-02866, A09J-02877, A09J-02903, A09J-02909, A09J-02913, A09J-02915, A09J-02916, A09J-02917, A09J-02918, A09J-02919, A09J-02920, A09J-02921, A09J-02922, A09J-02923, A09J-02924, A09J-02925, A09J-02926, A09J-02927, A09J-02928, A09J-02929, A09J-02933, A09J-02936, A09J-02938, A09J-02939, A09J-02940, A09J-02941, A09J-02942, A09J-02943, A09J-02944, A09J-02945, A09J-02946, A09J-02947, A09J-02948, A09J-02949, A09J-02953, A09J-02954, A09J-02955, A09J-02958, A09J-03009, A09J-03010, A09J-03011, A09J-03012, A09J-03013, A09J-03014, A09J-03015, A09J-03016, A09J-03017, A09J-03020, A09J-03021, A09J-03022, A09J-03023, A09J-03025, A09J-03027, A09J-03030, A09J-03035, A09J-03042, A09J-03048, A09J-03055, A09J-03056, A09J-03070, A09J-00940, A09J-01115, A09J-01174, A09J-01175, A09J-01256, A09J-01421, A09J-01434, A09J-01613, A09J-01614, A09J-01615, A09J-01616, A09J-01617, A09J-01621, A09J-01622, A09J-01623, A09J-01625, A09J-01626, A09J-01627, A09J-01643, A09J-01644, A09J-01645, A09J-01647, A09J-01650, A09J-01711, A09J-01712, A09J-01713, A09J-01715, A09J-01718, A09J-01719, A09J-01726, A09J-01731, A09J-01758, A09J-01761, A09J-01762, A09J-01770, A09J-01772, A09J-01773, A09J-01791, A09J-01792, A09J-01799, A09J-01852, A09J-01853, A09J-01854, A09J-01855, A09J-01857, A09J-01858, A09J-01861, A09J-01863, A09J-01864, A09J-01866, A09J-01968, A09J-01971, A09J-01972, A09J-01978, A09J-01979, A09J-01981, A09J-01997, A09J-01998, A09J-01999, A09J-02000, A09J-02007, A09J-02014, A09J-02015, A09J-02016, A09J-02024, A09J-02031, A09J-02032, A09J-02048, A09J-02053, A09J-02054, A09J-02090, A09J-02092, A09J-02095, A09J-02101, A09J-02103, A09J-02105, A09J-02106, A09J-02107, A09J-02109, A09J-02112, A09J-02115, A09J-02120, A09J-02121, A09J-02126, A09J-02128, A09J-02131, A09J-02132, A09J-02134, A09J-02135, A09J-02136, A09J-02141, A09J-02143, A09J-02144, A09J-02215, A09J-02216, A09J-02220, A09J-02231, A09J-02232, A09J-02233, A09J-02234, A09J-02277, A09J-02302, A09J-02306, A09J-02307, A09J-02315, A09J-02316, A09J-02318, A09J-02357, A09J-02368, A09J-02369, A09J-02700, A09J-01304, A09J-01340, A09J-01896, A09J-01928, A09J-02168, A09J-02518, A09J-02527, A09J-02541, A09J-01926, A09J-01803, A09J-01832, A09J-01833, A09J-01841, A09J-01929, A09J-01932, A09J-02037, A09J-02157, A09J-02281, A09J-02300, A09J-02519, A09J-02522, A09J-02523, A09J-02525, A09J-02528, A09J-02612, A09J-02349, A09J-01869, A09J-01875, A09J-02029, A09J-02050, A09J-02649, A09J-01898, A09J-02507, A09J-02524, A09J-02526, A09J-01820, A09J-01834, A09J-01843, A09J-01927, A09J-02154, A09J-02155, A09J-02158, A09J-02160, A09J-02166, A09J-02516, A09J-02517, A09J-02520, A09J-02529, A09J-02548, A09J-02620, A09J-02671, A09J-02672, A09J-02683, A09J-01842, A09J-01899, A09J-01967, A09J-02167, A09J-02283, and A09J-02530.
FEI Number 1000524572
Recalling Firm/
Manufacturer
Philips Medical Systems
2301 5th Ave Ste 200
Seattle WA 98121
For Additional Information Contact
206-664-5290
Manufacturer Reason
for Recall
Potential for failure of a capacitor during use which would prevent the AED from delivering effective defibrillation therapy when indicated.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionThe firm notified their US distributors of the recall and instructed them to contact customers and provide information how to return the product to Philips for a free replacement. Firm contacted distributors via phone and email on 11/2/09 and by letter on 11/9/09. A web site providing information related to the recall was activated on 11/3/09.
Quantity in Commerce99 units to domestic consignees, 1178 units to international consignees
DistributionWorldwide Distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = MKJ
-
-