| | Class 1 Device Recall Responder 2023440 automated external defibrillator |  |
| Date Initiated by Firm | November 13, 2009 |
| Date Posted | March 04, 2010 |
| Recall Status1 |
Terminated 3 on October 24, 2014 |
| Recall Number | Z-0776-2010 |
| Recall Event ID |
53827 |
| 510(K)Number | K031987 K040438 K052161 |
| Product Classification |
Automated external defibrillators (non-wearable) - Product Code MKJ
|
| Product | GE Responder 2023440 automated external defibrillator. This semi-automatic model requires the user to press a shock button in order for the device to deliver defibrillation energy (when appropriate).
Intended for use in either in-hospital or out-of-hospital settings for emergency treatment of victims exhibiting symptoms of sudden cardiac arrest who are unresponsive and not breathing. |
| Code Information |
SERIAL NUMBERS: 900650, 900651, 900652, 900653, 900654, 900655, 900656, 900658, 900659, 900660, 900661, 900662, 900663, 900664, 900665, 900666, 900667, 900668, 900670, 900671, 900672, 900673, 900675, 900677, 900678, 900679, 900680, 900681, 900682, 900683, 900684, 900685, 900686, 900687, 900688, 900689, 900690, 900691, 900692, 900693, 900694, 900695, 900696, 900697, 900698, 900940, 900941, 900942, 900943, 900944, 900945, 900946, 900947, 900948, 901026, 901027, 901028, 901030, 901031, 901033, 901034, 901035, 901186, 901187, 901188, 901189, 901190, 901191, 901192, 901193, 901194, 901195, 901196, 901197, 901198, 901199, 901200, 901201, 901202, 901203, 901204, 901205, 901206, 901207, 901208, 901209, 901211, 901212, 901213, 901214, 901215, 901216, 901217, 901218, 901219, 901220, 901221, 901222, 901223, 901224, 901225, 901226, 901227, 901228, 901229, 901230, 901231, 901232, 901233, 901234, 901235, 901296, 901297, 901298, 901299, 901300, 901301, 901302, 901303, 901304, 901305, 901306, 901307, 901308, 901309, 901310, 901311, 901312, 901313, 901314, 901315, 901316, 901317, 901318, 901320, 901321, 901323, 901324, 901325, 901326, 901327, 901328, 901329, 901330, 901331, 901332, 901333, 901334, 901336, 901337, 901338, 901768, 901769, 901770, 901772, 901773, 901774, 901775, 901776, 901777, 901778, 901779, 901780, 901781, 901782, 901783, 901784, 901785, 901786, 901787, 901788, 901789, 901790, 901791, 901792, 901793, 901794, 901795, 901796, 901797, 901798, 901799, 901800, 901801, 901802, 901803, 901804, 901805, 901806, 901807, 901988, 901989, 901990, 901991, 901992, 901993, 901994, 901995, 901996, 901997, 901998, 901999, 902000, 902001, 902002, 902003, 902004, 902005, 902006, 902007, 902008, 902009, 902010, 902011, 902012, 902013, 902014, 902015, 902016, 902017, 902018, 902019, 902020, 902021, 902022, 902023, 902024, 902025, 902026, 902027, 902178, 902179, 902180, 902181, 902182, 902183, 902184, 902185, 902186, 902187, 902188, 902189, 902190, 902191, 902192, 902193, 902194, 902195, 902196, 902198, 902199, 902200, 902201, 902202, 902203, 902204, 902206, 902207, 902358, 902359, 902360, 902361, 902362, 902363, 902364, 902365, 902366, 902367, 902368, 902369, 902370, 902371, 902372, 902373, 902374, 902375, 902376, 902377, 902508, 902509, 902510, 902511, 902512, 902513, 902514, 902515, 902516, 902518, 902519, 902520, 902521, 902522, 902523, 902524, 902525, 902526, 902527, 902528, 902529, 902530, 902531, 902532, 902533, 902534, 902535, 902536, 902537, 902538, 902539, 902540, 902541, 902542, 902543, 902544, 902545, 902546, 902547, 902548, 902549, 902550, 902551, 902552, 903053, 903054, 903055, 903056, 903057, 903058, 903059, 903061, 903062, 903063, 903064, 903065, 903066, 903067, 903068, 903069, 903070, 903071, 903072, 903073, 903074, 903075, 903076, 903077, 903078, 903079, 903080, 903081, 903082, 903083, 903084, 903085, 903086, 903087, 903088, 903089, 903090, 903091, 903092, 903093, 903094, 903095, 903096, 903097, 903098, 903099, 903100, 903101, 903102, 903103, 903104, 903105, 903106, 903107, 903108, 903109, 903110, 903111, 903112, 903113, 903114, 903115, 903116, 903117, 903118, 903119, 903120, 903121, 903122, 903123, 903124, 903125, 903126, 903127, 903128, 903129, 903130, 903131, 903132, 903133, 903134, 903135, 903136, 903137, 903138, 903139, 903140, 903141, 903142, 903143, 903144, 903145, 903146, 903147, 903148, 903149, 903150, 903151, 903152, 903153, 903154, 903155, 903156, 903157, 903158, 903159, 903160, 903161, 903162, 903163, 903164, 903165, 903166, 903167, 903168, 903169, 903171, 903172, 903173, 903174, 903175, 903176, 903177, 903178, 903179, 903180, 903181, 903182, 903183, 903184, 903185, 903186, 903187, 903188, 903189, 903190, 903191, 903192, 4008853, 4008854, 4008855, 4008856, 4008857, 4008858, 4008859, 4008860, 4008861, 4008862, 4008863, 4008864, 4008865, 4008866, 4011483, 4011484, 4011485, 4011486, 4011487, 4011488, 4011489, 4011490, 4011491, 4011492, 4011493, 4011494, 4011495, 4011496, 4011497, 4011498, 4011499, 4011500, 4011501, 4011502, 4011503, 4011504, 4011505, 4011506, 4011507, 4011508, 4011509, 4011510, 4011511, 4011512, 4011513, 4011514, 4011515, 4011516, 4011517, 4011518, 4011519, 4011520, 4011521, 4011522, 4011523, 4011524, 4011525, 4011526, 4011527, 4011528, 4011529, 4011530, 4011531, 4011532, 4011533, 4011534, 4011535, 4011536, 4011537, 4011538, 4011539, 4011540, 4011541, 4011542, 4011543, 4011544, 4011545, 4011546, 4011547, 4011548, 4011549, 4011550, 4011551, 4011552, 4011553, 4011554, 4011555, 4011556, 4011557, 4011558, 4011559, 4011560, 4011561, 4011562, 4011563, 4011564, 4011565, 4011566, 4011567, 4011568, 4011569, 4011570, 4011571, 4011572, 4011573, 4011574, 4011575, 4011576, 4011577, 4011578, 4011579, 4011580, 4011581, 4011582, 4011583, 4011584, 4011585, 4011586, 4011587, 4011588, 4011589, 4011590, 4011591, 4011592, 4011593, 4011594, 4011595, 4011596, 4011597, 4011598, 4011599, 4011600, 4011601, 4011602, 4011603, 4011604, 4011605, 4011606, 4011607, 4011608, 4011609, 4011610, 4011611, 4011612, 4011613, 4011614, 4011615, 4011616, 4011617, 4011618, 4011619, 4011620, 4011621, 4011622, 4011623, 4011624, 4011625, 4011626, 4011627, 4011628, 4011629, 4011630, 4011631, 4011632, 4011633, 4011634, 4011635, 4011636, 4011637, 4011638, 4011639, 4011640, 4011641, 4011642, 4011643, 4011644, 4011645, 4011646, 4011647, 4011648, 4011649, 4011650, 4011651, 4011652, 4011653, 4011654, 4011655, 4011656, 4011657, 4011658, 4011659, 4011660, 4011661, 4011662, 4011663, 4011664, 4011665, 4011666, 4011667, 4011668, 4011669, 4011670, 4011671, 4011672, 4011673, 4011674, 4011675, 4011676, 4011677, 4011678, 4011679, 4011680, 4011681, 4011682, 4019948, 4019949, 4019950, 4019951, 4019952, 4019953, 4019954, 4019955, 4019956, 4019957, 4019958, 4019959, 4019960, 4019961, 4019962, 4019963, 4019964, 4020565, 4020566, 4020567, 4020568, 4020569, 4020920, 4020921, 4020922, 4020923, 4020924, 4020925, 4020926, 4020927, 4020928, 4020929, 4020930, 4020931, 4020932, 4020933, 4021534, 4021535, 4021536, 4021537, 4021538, 4021539, 4021540, 4021541, 4021542, 4021543, 4021544, 4021545, 4021546, 4021547, 4021548, 4021549, 4021550, 4021551, 4021552, 4021553, 4021554, 4021555, 4021556, 4021557, 4021558, 4021559, 4021560, 4021561, 4021562, 4021563, 4021564, 4021565, 4021566, 4021568, 4021569, 4021570, 4021571, 4021572, 4021573, 4021574, 4021575, 4021576, 4021577, 4021578, 4021579, 4021580, 4021581, 4021582, 4021583, 4028134, 4028135, 4028136, 4028137, 4028138, 4028139, 4028140, 4028141, 4028142, 4028143, 4028144, 4028145, 4028146, 4028147, 4028148, 4028149, 4028150, 4028151, 4028152, 4028153, 4028154, 4028155, 4028156, 4028157, 4028158, 4028159, 4028160, 4028161, 4028162, 4028163, 4028164, 4028165, 4028166, 4028167, 4028168, 4028169, 4028170, 4028171, 4028172, 4028173, 4028174, 4028175, 4028176, 4028177, 4028178, 4028179, 4028180, 4028181, 4028182, 4028183, 4028184, 4028185, 4028186, 4028187, 4028188, 4028189, 4028190, 4028191, 4028192, 4028193, 4028194, 4028195, 4028196, 4028197, 4028198, 4028199, 4028200, 4028201, 4028202, 4028203, 4028204, 4028205, 4028206, 4028207, 4028208, 4028209, 4028210, 4028211, 4028212, 4028213, 4028214, 4028215, 4028216, 4028217, 4028218, 4028219, 4028220, 4028221, 4028222, 4028223, 4028224, 4028225, 4028226, 4028227, 4028228, 4028229, 4028230, 4028231, 4028232, 4028233, 4039203, 4039204, 4039205, 4039206, 4039207, 4039208, 4039209, 4039210, 4039211, 4039212, 4039213, 4039214, 4039215, 4039216, 4039217, 4039218, 4039219, 4039220, 4039221, 4039222, 4039223, 4039224, 4039225, 4039226, 4039227, 4039228, 4039229, 4039230, 4039231, 4039232, 4039233, 4039234, 4039235, 4039236, 4039237, 4039238, 4039239, 4039240, 4039241, 4039242, 4039243, 4039244, 4039245, 4039246, 4039247, 4039248, 4039249, 4039250, 4039251, 4039252, 4039253, 4039254, 4039255, 4039256, 4039257, 4039258, 4039259, 4039260, 4039261, 4039262, 4039263, 4039264, 4039265, 4039266, 4039267, 4039268, 4039269, 4039270, 4039271, 4039272, 4039273, 4039274, 4039275, 4039276, 4039277, 4039278, 4039279, 4039280, 4039281, 4039282, 4042938, 4042939, 4042940, 4042941, 4042942, 4042943, 4042944, 4042945, 4042946, 4042947, 4042948, 4042949, 4042950, 4042951, 4042952, 4042953, 4042954, 4042955, 4042956, 4042957, 4042958, 4042959, 4042960, 4042961, 4042962, 4042963, 4042964, 4042965, 4042966, 4042967, 4042968, 4042969, 4042970, 4042971, 4042972, 4042973, 4042974, 4042975, 4042976, 4042977, 4042978, 4042979, 4042980, 4042981, 4042982, 4042983, 4042984, 4042985, 4042986, 4042987, 4042988, 4042989, 4042990, 4042991, 4042992, 4042993, 4042994, 4042995, 4042996, 4042997, 4042998, 4042999, 4043000, 4043001, 4043002, 4043003, 4043004, 4043005, 4043006, 4043007, 4043008, 4043009, 4043010, 4043011, 4043012, 4043013, 4043014, 4043015, 4043016, 4043017, 4043018, 4043019, 4043020, 4043021, 4043022, 4043023, 4043024, 4043025, 4043026, 4043027, 4043028, 4043029, 4043030, 4043031, 4043032, 4043033, 4043034, 4043035, 4043036, 4043037, 4047789, 4047790, 4047791, 4047792, 4047793, 4047794, 4047795, 4047796, 4047797, 4047798, 4047799, 4047800, 4047801, 4047802, 4047803, 4047804, 4047805, 4047806, 4047807, 4047808, 4047809, 4047810, 4047811, 4047812, 4047813, 4047814, 4047815, 4047816, 4047817, 4047818, 4047819, 4047820, 4047821, 4047822, 4047823, 4047824, 4047825, 4047826, 4047827, 4047828, 4047829, 4047830, 4047831, 4047832, 4047833, 4047834, 4047835, 4047836, 4047837, 4047838, 4047949, 4047950, 4047951, 4047952, 4047953, 4047954, 4047955, 4047956, 4047957, 4047958, 4047959, 4047960, 4047961, 4047962, 4047963, 4047964, 4047965, 4047966, 4047967, 4047968, 4047969, 4047970, 4047971, 4047972, 4047973, 4047974, 4047975, 4047976, 4047977, 4047978, 4047979, 4047980, 4047981, 4047982, 4047983, 4047984, 4047985, 4047986, 4047987, 4047988, 4047989, 4047990, 4047991, 4047992, 4047993, 4047994, 4047995, 4047996, 4047997, 4047998, 4047999, 4048000, 4048001, 4048002, 4048003, 4048004, 4048005, 4048006, 4048007, 4048008, 4048009, 4048010, 4048011, 4048012, 4048013, 4048014, 4048015, 4048016, 4048017, 4048018, 4048019, 4048020, 4048021, 4048022, 4048023, 4048024, 4048025, 4048026, 4048027, 4048028, 4048029, 4048030, 4048031, 4048032, 4048033, 4048034, 4048035, 4048036, 4048037, 4048038, 4048039, 4048040, 4048041, 4048042, 4048043, 4048044, 4048045, 4048046, 4048047, 4048048, 4048149, 4048150, 4048151, 4048152, 4048153, 4048154, 4048155, 4048156, 4048157, 4048158, 4048164, 4048165, 4048166, 4048167, 4048168, 4048169, 4048170, 4048171, 4048172, 4048173, 4048244, 4048245, 4048246, 4048247, 4048248, 4048249, 4048250, 4048251, 4048252, 4048253, 4048254, 4048255, 4048256, 4048257, 4048258, 4048259, 4048260, 4048261, 4048262, 4048263, 4048773, 4048774, 4048775, 4048776, 4048777, 4048778, 4048779, 4048780, 4048781, 4048782, 4048783, 4048784, 4048785, 4048786, 4048787, 4048788, 4048789, 4048790, 4048791, 4048792, 4048793, 4048794, 4048795, 4048796, 4048797, 4048798, 4048799, 4048800, 4048801, 4048802, 4048803, 4048804, 4048805, 4048806, 4048807, 4048808, 4048809, 4048810, 4048811, 4048812, 4048813, 4048814, 4048815, 4048816, 4048817, 4048818, 4048819, 4048820, 4048821, 4048822, 4048973, 4048974, 4048975, 4048976, 4048977, 4048978, 4048979, 4048980, 4048981, 4048982, 4048983, 4048984, 4048985, 4048986, 4048987, 4048988, 4048989, 4048990, 4048991, 4048992, 4048993, 4048994, 4048995, 4048996, 4048997, 4048998, 4048999, 4049000, 4049001, 4049002, 4049003, 4049004, 4049005, 4049006, 4049007, 4049008, 4049009, 4049010, 4049011, 4049012, 4049013, 4049014, 4049015, 4049016, 4049017, 4049018, 4049019, 4049020, 4049021, 4049022, 4049303, 4049304, 4049305, 4049306, 4049307, 4049308, 4049309, 4049310, 4049311, 4049312, 4049313, 4049314, 4049315, 4049316, 4049317, 4049318, 4049319, 4049320, 4049321, 4049322, 4049323, 4049324, 4049325, 4049326, 4049327, 4145463, 4145464, 4145465, 4145466, 4145467, 4145468, 4145469, 4145470, 4145471, 4145472, 4145473, 4145474, 4145475, 4145476, 4145477, 4145478, 4145479, 4145480, 4145481, 4145482, 4145483, 4145484, 4145485, 4145486, 4145487, 4145488, 4145489, 4145490, 4145491, 4145492, 4145493, 4145494, 4145495, 4145496, 4145497, 4145498, 4145499, 4145500, 4145501, 4145502, 4145503, 4145504, 4145505, 4145506, 4145507, 4145508, 4145509, 4145510, 4145511, 4145512, 4159116, 4159117, 4159118, 4159119, 4159120, 4159121, 4159122, 4159123, 4159124, 4159125, 4159126, 4159127, 4159128, 4159129, 4159130, 4159131, 4159132, 4159133, 4159134, 4159135, 4159136, 4159137, 4159138, 4159139, 4159140, 4159141, 4159142, 4159143, 4159144, 4159145, 4159146, 4159147, 4159148, 4159149, 4159150, 4159151, 4159152, 4159153, 4159154, 4159155, 4159156, 4159157, 4159158, 4159159, 4159160, 4159161, 4159162, 4159163, 4159164, 4159165, 4159166, 4159167, 4159168, 4159169, 4159170, 4159171, 4159172, 4159173, 4159174, 4159175, 4159176, 4159177, 4159178, 4159179, 4159180, 4159181, 4159182, 4159183, 4159184, 4159185, 4159186, 4159187, 4159188, 4159189, 4159190, 4159191, 4159192, 4159193, 4159194, 4159195, 4159196, 4159197, 4159198, 4159199, 4159200, 4159201, 4159202, 4159203, 4159204, 4159205, 4159206, 4159207, 4159208, 4159209, 4159210, 4159211, 4159212, 4159213, 4159214, 4159215, 4167426, 4167427, 4167428, 4167429, 4167430, 4167431, 4167432, 4167433, 4167434, 4167435, 4167436, 4167437, 4167438, 4167439, 4167440, 4167441, 4167442, 4167443, 4167444, 4167445, 4167446, 4167447, 4167448, 4167449, 4167450, 4167451, 4167452, 4167453, 4167454, 4167455, 4167456, 4167457, 4167458, 4167459, 4167460, 4167461, 4167462, 4167463, 4167464, 4167465, 4167466, 4167467, 4167468, 4167469, 4167470, 4167471, 4167472, 4167473, 4167474, 4167475, 4167476, 4167477, 4167478, 4167479, 4167480, 4167481, 4167482, 4167483, 4167484, 4167485, 4167486, 4167487, 4167488, 4167489, 4167490, 4167491, 4167492, 4167493, 4167494, 4167495, 4167496, 4167497, 4167498, 4167499, 4167500, 4167501, 4167502, 4167503, 4167504, 4167505, 4167506, 4167507, 4167508, 4167509, 4167510, 4167511, 4167512, 4167513, 4167514, 4167515, 4167516, 4167517, 4167518, 4167519, 4167520, 4167521, 4167522, 4167523, 4167524, 4167525, 4174901, 4174902, 4174903, 4174904, 4174905, 4174906, 4174907, 4174908, 4174909, 4174910, 4174911, 4174912, 4174913, 4174914, 4174915, 4174916, 4174917, 4174918, 4174919, 4174920, 4174921, 4174922, 4174923, 4174924, 4174925, 4174926, 4174927, 4174928, 4174929, 4174930, 4174931, 4174932, 4174933, 4174934, 4174935, 4174936, 4174937, 4174938, 4174939, 4174940, 4174941, 4174942, 4174943, 4174944, 4174945, 4174946, 4174947, 4174948, 4174949, 4174950, 4174951, 4174952, 4174953, 4174954, 4174955, 4174956, 4174957, 4174958, 4174959, 4174960, 4174961, 4174962, 4174963, 4174964, 4174965, 4174966, 4174967, 4174968, 4174969, 4174970, 4174971, 4174972, 4174973, 4174974, 4174975, 4174976, 4174977, 4174978, 4174979, 4174980, 4174981, 4174982, 4174983, 4174984, 4174985, 4174986, 4174987, 4174988, 4174989, 4174990, 4174991, 4174992, 4174993, 4174994, 4174995, 4174996, 4174997, 4174998, 4174999, 4175000, 4199578, 4199579, 4199580, 4199581, 4199582, 4199583, 4199584, 4199585, 4199586, 4199587, 4199588, 4199589, 4199590, 4199591, 4199592, 4199593, 4199594, 4199595, 4199596, 4199597, 4199598, 4199599, 4199600, 4199601, 4199602, 4199603, 4199604, 4199605, 4199606, 4199607, 4199608, 4199609, 4199610, 4199611, 4199612, 4199613, 4199614, 4199615, 4199616, 4199617, 4199618, 4199619, 4199620, 4199621, 4199622, 4199623, 4199624, 4199625, 4199626, 4199627, 4201928, 4201929, 4201930, 4201931, 4201932, 4201933, 4201934, 4201935, 4201936, 4201937, 4201938, 4201939, 4201940, 4201941, 4201942, 4201943, 4201944, 4201945, 4201946, 4201947, 4201948, 4201949, 4201950, 4201951, 4201952, 4201953, 4201954, 4201955, 4201956, 4201957, 4201958, 4201959, 4201960, 4201961, 4201962, 4201963, 4201964, 4201965, 4201966, 4201967, 4201968, 4201969, 4201970, 4201971, 4201972, 4201973, 4201974, 4201975, 4201976, 4201977, 4201978, 4201979, 4201980, 4201981, 4201982, 4201983, 4201984, 4201985, 4201986, 4201987, 4201988, 4201989, 4201990, 4201991, 4201992, 4201993, 4201994, 4201995, 4201996, 4201997, 4201998, 4201999, 4202000.4202001, 4202002, 4202003, 4202004, 4202005, 4202006, 4202007, 4202008, 4202009, 4202010, 4202011, 4202012, 4202013, 4202014, 4202015, 4202016, 4202017, 4202018, 4202019, 4202020, 4202021, 4202022, 4202023, 4202024, 4202025, 4202026, 4202027, 4218155, 4218156, 4218157, 4218158, 4218159, 4218160, 4218161, 4218162, 4218163, 4218164, 4218165, 4218166, 4218167, 4218168, 4218169, 4218170, 4218171, 4218172, 4218173, 4218174, 4218175, 4218176, 4218177, 4218178, 4218179, 4218180, 4218181, 4218182, 4218183, 4218184, 4218185, 4218186, 4218187, 4218188, 4218189, 4218190, 4218191, 4218192, 4218193, 4218194, 4218195, 4218196, 4218197, 4218198, 4218199, 4218200, 4218201, 4218202, 4218203, 4218204, 4218205, 4218206, 4218207, 4218208, 4218209, 4218210, 4218211, 4218212, 4218213, 4218214, 4218215, 4218216, 4218217, 4218218, 4218219, 4218220, 4218221, 4218222, 4218223, 4218224, 4218225, 4218226, 4218227, 4218228, 4218229, 4218230, 4218231, 4218232, 4218233, 4218234, 4218235, 4218236, 4218237, 4218238, 4218239, 4218240, 4218241, 4218242, 4218243, 4218244, 4218245, 4218246, 4218247, 4218248, 4218249, 4218250, 4218251, 4218252, 4218253, 4218254, 4218255, 4218256, 4218257, 4218258, 4218259, 4218260, 4218261, 4218262, 4218263, 4218264, 4218265, 4218266, 4218267, 4218268, 4218269, 4218270, 4218271, 4218272, 4218273, 4218274, 4218275, 4218276, 4218277, 4218278, 4218279, 4218280, 4218281, 4218282, 4218283, 4218284, 4218285, 4218286, 4218287, 4218288, 4218289, 4218290, 4218291, 4218292, 4218293, 4218294, 4218295, 4218296, 4218297, 4218298, 4218299, 4218300, 4218301, 4218302, 4218303, 4218304, 4218305, 4218306, 4218307, 4218308, 4218309, 4218310, 4218311, 4218312, 4218313, 4218314, 4218315, 4218316, 4218317, 4218318, 4218319, 4218320, 4218321, 4218322, 4218323, 4218324, 4218325, 4218326, 4218327, 4218328, 4218329, 4218330, 4218331, 4218332, 4218333, 4218334, 4218335, 4218336, 4218337, 4218338, 4218339, 4218340, 4218341, 4218342, 4218343, 4218344, 4218345, 4218346, 4218347, 4218348, 4218349, 4218350, 4218351, 4218352, 4218353, 4218354, 4218355, 4218356, 4218357, 4218358, 4218359, 4218360, 4218361, 4218362, 4218363, 4218364, 4218365, 4218366, 4218367, 4218368, 4218369, 4218370, 4218371, 4218372, 4218373, 4218374, 4218375, 4218376, 4218377, 4218378, 4218379, 4218380, 4218381, 4218382, 4218383, 4218384, 4218385, 4218386, 4218387, 4218388, 4218389, 4218390, 4218391, 4218392, 4218393, 4218394, 4218395, 4218396, 4218397, 4218398, 4218399, 4218400, 4218401, 4218402, 4218403, 4218404, 4252004, 4252005, 4252006, 4252007, 4252008, 4252009, 4252010, 4252011, 4252012, 4252013, 4252014, 4252015, 4252016, 4252017, 4252018, 4252019, 4252020, 4252021, 4252022, 4252023, 4252024, 4252025, 4252026, 4252027, 4252028, 4252029, 4252030, 4252031, 4252032, 4252033, 4252034, 4252035, 4252036, 4252037, 4252038, 4252039, 4252040, 4252041, 4252042, 4252043, 4252044, 4252045, 4252046, 4252047, 4252048, 4252049, 4252050, 4252051, 4252052, 4252053, 4258659, 4258660, 4258661, 4258662, 4258663, 4258664, 4258665, 4258666, 4258667, 4258668, 4258669, 4258670, 4258671, 4258672, 4258673, 4258675, 4258676, 4258680, 4258681, 4258682, 4258683, 4258684, 4258685, 4258686, 4258687, 4258688, 4258689, 4258690, 4258691, 4258693, 4258694, 4258695, 4258696, 4258697, 4258698, 4258699, 4258700, 4258702, 4258703, 4258705, 4258706, 4258707 AND 4258708. |
Recalling Firm/ Manufacturer |
Cardiac Science Corporation 3303 Monte Villa Pkwy Bothell WA 98021-8969
|
| For Additional Information Contact | Michael Matysik 425-402-2206 |
Manufacturer Reason for Recall | Potential for devices not to deliver therapy |
FDA Determined Cause 2 | Process control |
| Action | Cardiac Science issued an "Urgent - Voluntary Medical Device Correction AED Resistor Issue" letter dated November 13, 2009. Consignees were informed of the affected product and instructed on preventive measures.
For further information, contact Cardiac Science at 1-877-901-1788 in the US and +44-161-926-0011 for outside the US. |
| Quantity in Commerce | 2215 |
| Distribution | Worldwide Distribution -- UNITED STATES, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHAMAS, BAHRAIN, BELARUS, BELGIUM, BERMUDA, BOLIVIA, BRAZIL, BULGARIA, CANADA, CAYMAN ISLANDS, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CYPRUS, CZECH REPUBLIC, CZECHOSLOVAKIA, DENMARK, DOMINICAN REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, FINLAND, FRANCE, GERMANY, GRECE, GUAM, GUATEMALA, HOLLAND, HONDURAS, HONG KONG, HUNGARY, ICELAND, INDIA, INDONESIA, IRAQ, IRELAND, ISRAEL, ITALY, JAMAICA, JAPAN, JORDAN, KENYA, KOREA, KUWAIT, LEBANON, LIBYA, LITHUANIA, MALAYSIA, MALTA, MEXICO, MONACO, MORROCCO, NAMIBIA, NETHERLANDS, NEW CALEDONIA, NEW ZEALAND, NIGERIA, NORWAY, OMAN, PAKISTAN, PANAMA, PARAGUAY, PERU, PHILIPPINES, POLAND, PORTUGAL, PUERTO RICO, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SCOTLAND, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TRINIDAD & TOBAGO, TURKEY, UAE, UKRAINE, UNITED KINGDOM, URUGUAY, VENEZUELA, VIETNAM, VIRGIN ISLANDS AND YEMEN. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = MKJ 510(K)s with Product Code = MKJ
|
|
|
|