• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Stryker VISUM LED Surgical Lights

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Stryker VISUM LED Surgical Lightssee related information
Date Initiated by FirmApril 15, 2010
Date PostedMay 21, 2010
Recall Status1 Terminated 3 on January 24, 2011
Recall NumberZ-1640-2010
Recall Event ID 55477
510(K)NumberK033132 K031068 
Product Classification Ceiling Mounted Surgical Light - Product Code FSY
ProductStryker VISUM LED Surgical Lights (Ceiling Mounted Surgical Light), Catalog numbers: 0682001250, 0682001251. Intended to illuminate the operative site during surgical procedures with high intensity light.
Code Information Product Numbers: Product Installed Between January 2, 2008 and August 31, 2009 Lot Code: Product Installed Between January 1, 2008 and August 31, 2009
FEI Number 3002975342
Recalling Firm/
Manufacturer
Stryker Communications Corp
1410 Lakeside Pkwy Ste 100
Flower Mound TX 75028-4026
For Additional Information Contact
972-410-7310
Manufacturer Reason
for Recall
Installation records are incomplete.
FDA Determined
Cause 2
Employee error
ActionAn Urgent: Medical Device Product Correction customer notification letter, dated 4/15/2010, has been sent out to all affected accounts via regular mail in the form of a certified, return request receipt letter. Accounts will receive a follow-up phone call from a Stryker Representative to schedule a re-inspection of equipment to see if it mets established critical specifications or not. The firm recommends that customers continue using all Stryker Communications products listed in the notice. Questions can be directed to Katie Blackwell at 972-410-7660 or via email at StrykerCommunicationsRA@stryker.com.
Quantity in Commerce608 (all lights and booms)
DistributionNationwide Distribution
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = FSY
510(K)s with Product Code = FSY
-
-