Class 2 Device Recall HillRom 405 Bed |
 |
Date Initiated by Firm |
September 16, 2009 |
Date Posted |
August 03, 2010 |
Recall Status1 |
Terminated 3 on March 16, 2012 |
Recall Number |
Z-2168-2010 |
Recall Event ID |
56018 |
Product Classification |
AC-Powered Adjustable Hospital Bed - Product Code FNL
|
Product |
Hill-Rom 405 Bed, P1440B, (electric) Hill Rom Batesville, IN.
Intended for low to moderate acuity patients in the low end acute care area of the hospital. |
Code Information |
K110HE1421, K110HE1422, K110HE1423, K110HE1424, K110HE1425, K110HE1426, K110HE1427, K110HE1428, K110HE1429, K110HE1430, K110HE1431, K110HE1432, K110HE1433, K110HE1434, K110HE1435, K110HE1436, K110HE1437, K110HE1438, K110HE1439, K110HE1440, K110HE1441, K110HE1442, K110HE1443, K110HE1444, K110HE1445, K110HE1446, K110HE1447, K110HE1448, K110HE1449, K110HE1450, K110HE1451, K110HE1452, K110HE1453, K110HE1454, K110HE1455, K110HE1456, K110HE1457, K110HE1458, K110HE1459, K110HE1460, K110HE1461, K110HE1462, K110HE1463, K110HE1464, K110HE1465, K110HE1466, K110HE1467, K110HE1468, K110HE1469, K110HE1470, K110HE1471, K110HE1472, K110HE1473, K110HE1474, J145HE0493, J144HE0452, J144HE0462, J144HE0469, J144HE0472, J145HE0476, J145HE0479, J145HE0480, J145HE0487, J145HE0488, J145HE0495, J145HE0475, J144HE0454, J144HE0455, J144HE0457, J144HE0470, J144HE0471, J145HE0474, J145HE0477, J145HE0489, J145HE0490, J145HE0494, K040HE0576, K040HE0577, K040HE0579, K040HE0580, K040HE0581, K04HE0578, J194HE0770, J194HE0771, J194HE0772, J194HE0773, J194HE0774, J194HE0775, J194HE0776, J194HE0777, J194HE0778, J194HE0779, J194HE0780, J194HE0781, J194HE0782, J194HE0783, J194HE0784, J194HE0785, J194HE0786, J194HE0787, J194HE0788, J194HE0789, J194HE0790, J194HE0791, J194HE0792, J194HE0793, J194HE0794, J195HE0795, J195HE0796, J195HE0797, J195HE0798, J195HE0799, J195HE0800, J195HE0801, J195HE0802, J195HE0803, J195HE0804, J195HE0805, J195HE0806, J195HE0807, J195HE0808, J195HE0809, J195HE0810, J195HE0811, J195HE0812, J195HE0813, J195HE0814, J195HE0815, J195HE0816, J196HE0817, J196HE0818, J196HE0819, J196HE0820, J196HE0821, J196HE0822, J196HE0823, J196HE0824, J196HE0825, J196HE0826, J196HE0827, J196HE0828, J196HE0829, J196HE0830, J196HE0831, J196HE0832, J196HE0833, J196HE0834, J196HE0835, J196HE0836, J196HE0837, J196HE0838, J196HE0839, J196HE0840, J196HE0841, K078HE0962, K078HE0963, K078HE0964, K078HE0965, K078HE0966, K078HE0967, K078HE0968, K078HE0969, K078HE0970, K078HE0971, K078HE0972, K078HE0973, K078HE0974, K078HE0975, K078HE0976, K078HE0977, K078HE0978, K078HE0979, K078HE0980, K078HE0981, K078HE0982, K078HE0983, K078HE0984, K078HE0985, K078HE0986, K078HE0987, K078HE0988, K078HE0989, K078HE0990, K078HE0991, K078HE0992, K078HE0993, K078HE0994, K079HE0995, K079HE0996, K079HE0997, K071HE0905, K071HE0906, K077HM0958, K077HM0959, K077HM0960, K077HM0961, K180HE2549, K194HE2841, J200HE0856, J200HE0857, K071HE0907, K071HE0908, K064HE089, K101HE1315, K101HE1316, K101HE1317, K101HE1318, K101HE1319, K101HE1320, K101HE1321, K101HE1322, K101HE1323, K101HE1324, K101HE1325, K101HE1326, K101HE1327, K101HE1328, K101HE1329, K101HE1330, K101HE1331, K101HE1332, K101HE1333, K101HE1334, K101HE1335, K101HE1336, K101HE1337, K017HE0231, K017HE0232, K017HE0233, K017HE0234, K017HE0235, K017HE0236, K017HE0237, K017HE0238, K017HE0239, K017HE0240, K017HE0241, K017HE0242, K017HE0243, K017HE0244, K017HE0245, K017HE0246, K017HE0247, K017HE0248, K017HE0249, K017HE0250, K017HE0251, K017HE0252, K018HE0253, K018HE0254, K018HE0255, K018HE0256, K018HE0257, K018HE0258, K018HE0259, K018HE0260, K018HE0261, K018HE0262, K018HE0263, K018HE0264, K018HE0265, K018HE0266, K018HE0267, K018HE0268, K018HE0269, K018HE0270, K018HE0271, K018HE0272, K018HE0273, K018HE0274, K018HE0275, K018HE0276, K018HE0277, K018HE0278, K018HE0279, K018HE0280, K018HE0281, K018HE0282, J108H30142, K187HE2635, K187HE2636, K187HE2637, K187HE2638, K187HE2639, K187HE2640, K187HE2641, K187HE2642, K187HE2643, K187HE2644, K187HE2645, K187HE2646, K187HE2647, K187HE2648, K187HE2649, K187HE2650, K187HE2651, K187HE2652, K188HE2653, K188HE2654, K188HE2655, K188HE2656, K188HE2657, K188HE2658, K188HE2659, K188HE2660, K188HE2661, K188HE2662, K188HE2663, K188HE2664, K188HE2665, K188HE2666, K188HE2667, K188HE2668, K188HE2669, K188HE2670, J063BE0843, J063BE0844, J063BE0845, J063BE0846, J063BE0847, J063BE0848, J063BE0849, J063BE0850, J063BE0851, J063BE0852, J063BE0853, J063BE0854, J063BE0855, J063BE0856, J063BE0857, J063BE0858, J063BE0859, J063BE0860, J063BE0861, J063BE0862, J063BE0863, J063BE0864, J063BE0865, J063BE0866, J063BE0867, J063BE0868, J063BE0869, J063BE0870, J063BE0871, J063BE0872, J064BE0873, J064BE0874, J064BE0875, J064BE0876, J064BE0877, J064BE0878, J064BE0879, J064BE0880, J064BE0881, J064BE0882, J064BE0883, J064BE0884, J064BE0885, J064BE0886, J077HE0008, J077HE0009, J077HE0010, J077HE0011, J077HE0012, J077HE0013, J077HE0014, J077HE0015, J077HE0016, J077HE0017, J077HE0018, J077HE0019, J078HE0020, J078HE0021, J078HE0022, J078HE0023, J078HE0024, J078HE0025, J078HE0026, J078HE0027, J078HE0028, J078HE0029, J078HE0030, J078HE0031, J078HE0032, J078HE0033, J078HE0034, J078HE0035, J078HE0036, J078HE0037, J079HE0038, J079HE0039, J079HE0040, J079HE0041, J079HE0042, J079HE0043, J079HE0044, J079HE0045, J079HE0046, J079HE0047, J079HE0048, J079HE0049, J079HE0050, J079HE0051, J205HE0884, J205HE0885, J205HE0886, J205HE0887, J205HE0888, J205HE0889, J205HE0890, J205HE0891, J205HE0892, J205HE0893, J205HE0894, J205HE0895, J205HE0896, J205HE0897, J205HE0898, J205HE0899, J205HE0900, J205HE0901, J205HE0902, J205HE0903, J205HE0904, J205HE0905, J205HE0906, J205HE0907, J205HE0908, J208HE0909, J208HE0910, J208HE0911, J208HE0912, J208HE0913, J208HE0914, J208HE0915, J208HE0916, J208HE0917, J208HE0918, J208HE0919, J208HE0920, J208HE0921, J208HE0922, J208HE0923, J302HE1896, J302HE1897, J302HE1899, J302HE1900, J302HE1901, J302HE1902, J302HE1903, J302HE1904, J302HE1905, J302HE1905, J302HE1906, J302HE1907, J302HE1908, J302HE1909, J302HE1910, J302HE1911, J302HE1912, J302HE1913, J302HE1914, J302HE1915, J302HE1916, J302HE1917, J302HE1918, J302HE1919, J302HE1920, J302HE1921, J302HE1922, J302HE1923, J302HE1924, J302HE1925, J302HE1926, J302HE1927, J302HE1928, J302HE1929, J302HE1930, J302HE1931, J302HE1932, J302HE1933, J302HE1934, J302HE1935, J337HE2394, J337HE2395, J337HE2396, J337HE2397, J337HE2398, J337HE2399, J337HE2400, J337HE2401, J337HE2402, J337HE2403, J337HE2404, J337HE2405, J337HE2406, J337HE2407, J337HE2408, J337HE2409, J337HE2410, J337HE2411, J337HE2412, J337HE2413, J337HE2414, J337HE2415, J338HE2416, J338HE2417, J338HE2418, J338HE2419, J338HE2420, J338HE2421, J338HE2422, J338HE2423, J338HE2424, J338HE2425, J338HE2426, J338HE2427, J338HE2428, J338HE2429, J338HE2430, J338HE2431, J338HE2432, J338HE2433, J338HE2434, J338HE2435, J338HE2436, J338HE2437, J338HE2438, J338HE2439, J338HE2440, J338HE2441, J338HE2442, J338HE2443, J338HE2444, J338HE2445, J338HE2446, J338HE2447, J338HE2448, J338HE2449, J338HE2450, J338HE2451, J338HE2452, J338HE2453, J338HE2454, J339HE2455, J339HE2456, J339HE2457, J339HE2458, J339HE2459, J339HE2460, J339HE2461, J339HE2462, J339HE2463, J339HE2464, J339HE2465, J339HE2466, J339HE2467, J339HE2468, J339HE2469, J339HE2470, J339HE2471, J339HE2472, J339HE2473, J339HE2474, J339HE2475, J339HE2476, J339HE2477, J339HE2478, J339HE2479, J339HE2480, J339HE2481, J339HE2482, J339HE2483, J339HE2484, J339HE2485, J339HE2486, J339HE2487, J339HE2488, J339HE2489, J339HE2490, J339HE2491, J339HE2492, J339HE2493, J340HE2494, J340HE2495, J340HE2496, J340HE2497, J340HE2498, J340HE2499, J340HE2500, J340HE2501, J340HE2502, J340HE2503, J340HE2504, J340HE2505, J340HE2506, J340HE2507, J340HE2508, J340HE2509, J340HE2510, J340HE2511, J340HE2512, J340HE2513, J340HE2514, J340HE2515, J340HE2516, J340HE2517, J340HE2518, J340HE2519, J340HE2520, J340HE2521, J340HE2522, J340HE2523, J340HE2524, J340HE2525, J340HE2526, J340HE2527, J340HE2528, J340HE2529, J340HE2530, J340HE2531, J340HE2532, J341HE2533, J341HE2534, J341HE2535, J341HE2536, J341HE2537, J353HE2633, J353HE2634, J353HE2636, J35SHE2635, J244HE1256, J244HE1257, J244HE1258, J244HE1259, J244HE1260, J244HE1261, J244HE1262, J244HE1263, J244HE1264, J244HE1265, J244HE1266, J244HE1267, J244HE1268, J244HE1269, J244HE1270, J244HE1271, J244HE1272, J244HE1273, J244HE1274, J244HE1275, J244HE1276, J244HE1277, J244HE1278, J244HE1279, J244HE1280, J244HE1281, J244HE1282, J244HE1283, J244HE1284, J244HE1285, J244HE1286, J244HE1287, J244HE1288, J244HE1289, J244HE1290, J244HE1291, J244HE1292, J244HE1293, J244HE1294, J244HE1295, J244HE1296, J244HE1297, J244HE1298, J244HE1299, J244HE1300, J245HE1301, J245HE1302, J245HE1303, J245HE1304, J245HE1305, J245HE1306, J245HE1307, J245HE1308, J245HE1309, J245HE1310, J245HE1311, J245HE1312, J245HE1313, J245HE1314, J245HE1315, J245HE1316, J245HE1317, J245HE1318, J245HE1319, J245HE1320, J245HE1321, J245HE1322, J245HE1323, J245HE1324, J245HE1325, J245HE1326, J245HE1327, J245HE1328, J245HE1329, J245HE1330, J245HE1331, J245HE1332, J245HE1333, J245HE1334, J245HE1335, J245HE1336, J245HE1337, J245HE1338, J245HE1339, J245HE1340, J245HE1341, J245HE1342, J245HE1343, J245HE1344, J245HE1345, J246HE1346, J246HE1347, J246HE1348, J246HE1349, J246HE1350, J246HE1351, J246HE1352, J246HE1353, J246HE1354, J246HE1355, J246HE1356, J246HE1357, J246HE1358, J246HE1359, J246HE1360, J246HE1361, J246HE1362, J246HE1363, J246HE1364, J246HE1365, J246HE1366, J246HE1367, J246HE1368, J246HE1369, J246HE1370, J246HE1371, J246HE1372, J246HE1373, J246HE1374, J246HE1375, J246HE1376, J246HE1377, J246HE1378, J246HE1379, J246HE1380, J246HE1381, J246HE1382, J246HE1383, J246HE1384, J246HE1385, J246HE1386, J246HE1387, J246HE1388, J246HE1389, J246HE1390, J247HE1391, J247HE1392, J247HE1393, J247HE1394, J247HE1395, J247HE1396, J247HE1397, J247HE1398, J247HE1399, J247HE1400, J247HE1401, J247HE1402, J247HE1403, J247HE1404, J247HE1405, J247HE1406, J247HE1407, J247HE1408, J247HE1409, J247HE1410, J247HE1411, J247HE1412, J247HE1413, J247HE1414, J247HE1415, J247HE1416, J247HE1417, J247HE1418, J247HE1419, J247HE1420, J247HE1421, J247HE1422, J247HE1423, J247HE1424, J247HE1425, J247HE1426, J128HE0295, J128HE0296, J128HE0297, J128HE0298, J128HE0299, J128HE0300, J128HE0301, J128HE0302, J128HE0303, J128HE0304, J128HE0305, J128HE0306, J128HE0307, J128HE0308, J128HE0309, J128HE0310, J128HE0311, J128HE0312, J128HE0313, J128HE0314, J130HE0315, J130HE0316, J130HE0317, J130HE0318, J130HE0319, J130HE0320, J130HE0321, J130HE0322, J130HE0323, J130HE0324, J130HE0325, J130HE0326, J130HE0327, J130HE0328, J130HE0329, J130HE0330, J130HE0331, J130HE0332, J130HE0333, J130HE0334, J130HE0335, J130HE0336, J130HE0337, J130HE0338, J130HE0339, J131HE0340, J131HE0341, J131HE0342, J131HE0343, J131HE0344, J131HE0345, J131HE0346, J131HE0347, J131HE0348, J131HE0349, J131HE0350, J131HE0351, J131HE0352, J131HE0353, J131HE0354, J131HE0355, J131HE0356, J131HE0357, J131HE0358, J131HE0359, J131HE0360, J131HE0361, J131HE0362, J131HE0363, J131HE0364, J138HE0367, J138HE0368, J138HE0369, J138HE0370, J138HE0371, J138HE0372, J138HE0373, J138HE0374, J138HE0375, J138HE0376, J141HE0377, J141HE0378, J141HE0379, J141HE0380, J142HE0381, J142HE0382, J142HE0383, J142HE0384, J142HE0385, J142HE0386, J142HE0387, J142HE0388, J142HE0389, J142HE0390, J142HE0391, J142HE0392, J142HE0393, J142HE0394, J142HE0395, J142HE0396, J142HE0397, J142HE0398, J142HE0399, J142HE0400, J142HE0401, J142HE0402, J142HE0403, J142HE0404, J143HE0405, J143HE0406, J143HE0407, J143HE0408, J155HE0409, J155HE0410, J155HE0411, J155HE0412, J155HE0413, J155HE0414, J155HE0415, J155HE0416, J155HE0417, J155HE0418, J155HE0419, J155HE0420, J155HE0421, J155HE0422, J155HE0423, J155HE0424, J155HE0425, J155HE0426, J155HE0427, J155HE0428, K194HE2699, K194HE2700, K194HE2701, K194HE2702, K194HE2703, K194HE2704, K194HE2705, K194HE2706, K194HE2707, K194HE2708, K194HE2709, K194HE2710, K194HE2711, K194HE2712, K194HE2713, K194HE2714, K194HE2715, K194HE2716, K194HE2717, K194HE2718, K194HE2719, K194HE2720, K194HE2721, K194HE2722, J158HE0584, J158HE0585, J158HE0586, J158HE0587, J158HE0588, J158HE0589, J158HE0590, J158HE0591, J158HE0592, J158HE0593, J158HE0594, J158HE0595, J158HE0596, J158HE0597, J158HE0598, J158HE0599, J158HE0600, J158HE0601, J158HE0602, J158HE0603, J158HE0604, J158HE0605, J159HE0606, J159HE0607, J159HE0608, J159HE0609, J159HE0610, J159HE0611, J159HE0612, J159HE0613, J159HE0614, J159HE0615, J159HE0616, J159HE0617, J159HE0618, J159HE0619, J159HE0620, J159HE0621, J159HE0622, J159HE0623, J159HE0624, J159HE0625, J159HE0626, J159HE0627, J159HE0628, J159HE0629, J159HE0630, J162HE0631, J162HE0632, J162HE0633, J162HE0634, J162HE0635, J162HE0636, J162HE0637, J162HE0638, J162HE0639, J162HE0640, J162HE0641, J162HE0642, J162HE0643, J162HE0644, J162HE0645, J162HE0646, J162HE0647, J162HE0648, J162HE0649, J162HE0650, J162HE0651, J162HE0652, J163HE0653, J163HE0654, J163HE0655, J163HE0656, J163HE0657, J163HE0658, J163HE0659, J163HE0660, J163HE0661, J163HE0662, J163HE0663, J163HE0664, J163HE0665, J163HE0666, J163HE0667, J163HE0668, J163HE0669, J163HE0670, J163HE0671, J172HE0712, J172HE0713, J172HE0714, J172HE0715, J172HE0716, J172HE0717, J172HE0718, J172HE0719, J172HE0720, J172HE0721, J172HE0722, J172HE0723, J342HE2553, J342HE2554, J342HE2555, J342HE2556, J342HE2557, J342HE2558, J342HE2559, J342HE2560, J342HE2561, J342HE2562, J342HE2563, J342HE2564, J342HE2565, J342HE2566, J342HE2567, J342HE2568, J342HE2569, J342HE2570, J342HE2571, J342HE2572, K113HE1515, K113HE1516, K113HE1517, K113HE1518, K113HE1519, K113HE1520, K113HE1521, K113HE1522, K113HE1523, K113HE1524, K113HE1525, K113HE1526, K113HE1527, K113HE1528, K113HE1529, K113HE1530, K113HE1531, K113HE1532, K113HE1533, K113HE1534, J117HE0163, J117HE0164, J117HE0165, J117HE0166, J117HE0167, J117HE0168, J117HE0169, J117HE0170, J117HE0171, J117HE0172, J118HE0173, J118HE0174, J118HE0175, J118HE0176, J118HE0177, J118HE0178, J118HE0179, J118HE0180, J118HE0181, J118HE0182, J118HE0183, J118HE0184, J118HE0185, J118HE0186, J118HE0187, J118HE0188, J118HE0189, J118HE0190, J118HE0191, J118HE0192, J118HE0193, J118HE0194, J118HE0195, J118HE0196, J118HE0197, J118HE0198, J118HE0199, J118HE0200, J118HE0201, J118HE0202, J118HE0203, J118HE0204, J118HE0205, J118HE0206, J331HE2249, J331HE2250, J331HE2251, J331HE2252, J331HE2253, J331HE2254, J331HE2255, J331HE2256, J331HE2257, J331HE2258, J331HE2259, J331HE2260, J331HE2261, J331HE2262, J331HE2263, J331HE2264, J331HE2265, J331HE2266, J331HE2267, J331HE2268, J331HE2269, J331HE2270, J331HE2271, J331HE2272, J331HE2273, J331HE2274, J331HE2275, J331HE2276, J331HE2277, J331HE2278, J331HE2279, J331HE2280, J331HE2281, J331HE2282, J331HE2283, J331HE2284, J331HE2285, J331HE2286, J331HE2287, J331HE2288, J331HE2289, J331HE2290, J331HE2291, J331HE2292, K014HE0088, K014HE0089, K014HE0090, K014HE0091, K014HE0092, K014HE0093, K014HE0094, K014HE0095, K014HE0096, K014HE0097, K014HE0098, K014HE0099, K014HE0100, K014HE0101, K014HE0102, K014HE0103, K014HE0104, K014HE0105, K014HE0106, K014HE0107, K014HE0108, K014HE0109, K014HE0110, K014HE0111, K014HE0112, K014HE0113, K014HE0114, K014HE0115, K014HE0116, K014HE0117, K014HE0118, K014HE0119, K014HE0120, K014HE0121, K014HE0122, K014HE0123, K014HE0124, K014HE0125, K014HE0126, K015HE0127, K015HE0128, K015HE0129, K015HE0130, K015HE0131, K015HE0132, K015HE0133, K015HE0134, K015HE0135, K015HE0136, K015HE0137, K015HE0138, K015HE0139, K015HE0140, K015HE0141, K015HE0142, K015HE0143, K015HE0144, K015HE0145, K015HE0146, K015HE0147, K015HE0148, K015HE0149, K015HE0150, K015HE0151, K015HE0152, K015HE0153, K015HE0154, K015HE0155, K015HE0156, K015HE0157, K015HE0158, K015HE0159, K015HE0160, K015HE0161, K015HE0162, K015HE0163, K015HE0164, K015HE0165, K015HE0166, K015HE0167, K015HE0168, K016HE0169, K016HE0170, K016HE0171, K016HE0172, K016HE0173, K016HE0174, K016HE0175, K079HE0998, K079HE0999, K079HE1000, K079HE1001, K079HE1002, K079HE1003, K079HE1004, K079HE1005, K079HE1006, K079HE1007, K079HE1008, K079HE1009, K079HE1010, K079HE1011, K079HE1012, K079HE1013, K079HE1014, K079HE1015, K079HE1016, K079HE1017, K079HE1018, K079HE1019, K079HE1020, K113HE1535, K113HE1536, K113HE1537, K113HE1538, K113HE1539, K113HE1540, K113HE1541, K113HE1542, K113HE1543, K113HE1544, K113HE1545, K113HE1546, K113HE1547, K113HE1548, K113HE1549, K113HE1550, K113HE1551, K113HE1552, K113HE1553, K113HE1554, K113HE1555, K113HE1556, K113HE1557, K113HE1558, K113HE1559, K113HE1560, K113HE1561, K113HE1562, K113HE1563, K113HE1564, K113HE1565, K113HE1566, K113HE1567, K113HE1568, K117HE1569, K117HE1570, K117HE1571, K117HE1572, K117HE1573, K117HE1574, K117HE1575, K117HE1576, K117HE1577, K117HE1578, K117HE1579, K117HE1580, K117HE1581, K117HE1582, K117HE1583, K117HE1584, K117HE1585, K117HE1586, K117HE1587, K117HE1588, K117HE1589, K117HE1590, K117HE1591, K117HE1592, K117HE1593, K117HE1594, K117HE1595, K117HE1596, K117HE1597, K117HE1598, K117HE1599, K117HE1600, K117HE1601, K117HE1602, K117HE1603, K117HE1604, K117HE1605, K117HE1606, K118HE1607, K118HE1608, K118HE1609, K118HE1610, K118HE1611, K118HE1612, K118HE1613, K118HE1614, K118HE1615, K118HE1616, K118HE1617, K118HE1618, K118HE1619, K118HE1620, K118HE1621, K118HE1622, K118HE1623, K118HE1624, K118HE1625, K118HE1626, K118HE1627, K118HE1628, K118HE1629, K118HE1630, K118HE1631, K118HE1632, K118HE1633, K118HE1634, K118HE1635, K118HE1636, K118HE1637, K118HE1638, K118HE1639, K118HE1640, K118HE1641, K118HE1642, K118HE1643, K118HE1644, K118HE1645, K118HE1646, K118HE1647, K118HE1648, K119HE1649, K119HE1650, K119HE1651, K119HE1652, K119HE1653, K119HE1654, K119HE1655, K119HE1656, K119HE1657, K119HE1658, K119HE1659, K119HE1660, K119HE1661, K119HE1662, K119HE1663, K119HE1664, K119HE1665, K119HE1666, K177HE2419, K177HE2420, K177HE2421, K177HE2422, K177HE2423, K177HE2424, K177HE2425, K177HE2426, K177HE2427, K177HE2428, K177HE2429, K177HE2430, K177HE2431, K177HE2432, K177HE2433, K177HE2434, J133HE0365, J133HE0366, K164HE2208, K164HE2209, K164HE2210, K164HE2211, K164HE2212, K164HE2213, K164HE2214, K164HE2215, K164HE2216, K164HE2217, K164HE2218, J228HE1237, J228HE1238, J228HE1239, J228HE1240, J228HE1241, J228HE1242, J228HE1243, J228HE1244, J228HE1245, J228HE1246, J228HE1247, J228HE1248, J228HE1249, J228HE1250, J228HE1251, J228HE1252, J228HE1253, J228HE1254, J228HE1255, K105HE1383, K105HE1384, K105HE1385, K105HE1386, K105HE1387, K105HE1388, K105HE1389, K105HE1390, K105HE1391, K105HE1392, K105HE1393, K105HE1394, K105HE1395, K105HE1396, K105HE1397, K105HE1398, K105HE1399, K105HE1400, K105HE1401, K105HE1402, K105HE1403, K105HE1404, K105HE1405, K105HE1406, K105HE1407, K105HE1408, K105HE1409, K105HE1410, K105HE1411, K105HE1412, K105HE1413, K105HE1414, K105HE1415, K105HE1416, K105HE1417, K105HE1418, K105HE1419, K105HE1420, K016HE0176, K016HE0177, K016HE0178, K016HE0179, K016HE0180, K016HE0181, K016HE0182, K016HE0183, K016HE0184, K016HE0185, K016HE0186, K016HE0187, K016HE0188, K016HE0189, K016HE0190, K016HE0191, K016HE0192, K016HE0193, K016HE0194, K016HE0195, K016HE0196, K016HE0197, K016HE0198, K016HE0199, K016HE0200, K016HE0201, K016HE0202, K016HE0203, K016HE0204, K016HE0205, K016HE0206, K016HE0207, K016HE0208, K016HE0209, K016HE0210, K017HE0211, K017HE0212, K017HE0213, K017HE0214, K017HE0215, J214HE0929, J214HE0930, J214HE0931, J214HE0932, J214HE0933, J214HE0934, J214HE0935, J214HE0936, J214HE0937, J214HE0938, J214HE0939, J214HE0940, J214HE0941, J214HE0942, J214HE0943, J214HE0944, J214HE0945, J214HE0946, J214HE0947, J214HE0948, J214HE0949, J214HE0950, J214HE0951, J214HE0952, J214HE0953, J215HE0954, J215HE0955, J215HE0956, J215HE0957, J215HE0958, J215HE0959, J215HE0960, J215HE0961, J215HE0962, J215HE0963, J215HE0964, J215HE0965, J215HE0966, J215HE0967, J215HE0968, J215HE0969, J215HE0970, J215HE0971, J215HE0972, J215HE0973, J215HE0974, J215HE0975, J215HE0976, J215HE0977, J215HE0978, J216HE0979, J216HE0980, J216HE0981, J216HE0982, J216HE0983, J216HE0984, J216HE0985, J216HE0986, J216HE0987, J216HE0988, J216HE0989, J216HE0990, J216HE0991, J216HE0992, J216HE0993, J216HE0994, J216HE0995, J216HE0996, J216HE0997, J216HE0998, J216HE0999, J216HE1000, J216HE1001, J216HE1002, J216HE1003, J217HE1004, J217HE1005, J217HE1006, J217HE1007, J217HE1008, J217HE1009, J217HE1010, J217HE1011, J217HE1012, J217HE1013, J217HE1014, J217HE1015, J217HE1016, J217HE1017, J217HE1018, J217HE1019, J217HE1020, J217HE1021, J217HE1022, J217HE1023, J217HE1024, J217HE1025, J217HE1026, J217HE1027, J217HE1028, J088HE0052, J088HE0053, J088HE0054, J088HE0055, J088HE0056, J088HE0057, J088HE0058, J088HE0059, J088HE0060, J088HE0061, J088HE0062, J088HE0063, J088HE0064, J088HE0065, J088HE0066, J088HE0067, J088HE0068, J088HE0069, J088HE0070, J088HE0071, J089HE0072, J089HE0073, J089HE0074, J089HE0075, J089HE0076, J089HE0077, J089HE0078, J089HE0079, J089HE0080, J089HE0081, J089HE0082, J089HE0083, J089HE0084, J089HE0085, J089HE0086, J089HE0087, J089HE0088, J089HE0089, J089HE0090, J089HE0091, J090HE0092, J090HE0093, J090HE0094, J090HE0095, J090HE0096, J090HE0097, J090HE0098, J090HE0099, J090HE0100, J090HE0101, J090HE0102, J090HE0103, J090HE0104, J090HE0105, J090HE0106, J090HE0107, J090HE0108, J090HE0109, J090HE0110, J091HE0111, J091HE0112, J091HE0113, J091HE0114, J091HE0115, J091HE0116, J091HE0117, J091HE0118, J091HE0119, J091HE0120, J091HE0121, J091HE0122, J091HE0123, J091HE0124, J091HE0125, J092HE0126, J092HE0127, J092HE0128, J092HE0129, J092HE0130, J092HE0131, J092HE0132, J092HE0133, J092HE0134, J092HE0135, J092HE0136, J092HE0137, J092HE0138, J092HE0139, J117HE0143, J117HE0144, J117HE0145, J117HE0146, J117HE0147, J117HE0148, J117HE0149, J117HE0150, J117HE0151, J117HE0152, J117HE0153, J117HE0154, J117HE0155, J117HE0156, J117HE0157, J117HE0158, J117HE0159, J117HE0160, J117HE0161, J117HE0162, K119HE1670, K119HE1671, K119HE1672, K119HE1673, K119HE1674, K119HE1675, K119HE1676, K119HE1677, K119HE1678, K119HE1679, K119HE1680, K119HE1681, K119HE1682, K119HE1683, K119HE1684, K119HE1685, K119HE1686, K120HE1687, K120HE1688, K120HE1689, K120HE1690, K120HE1691, K120HE1692, K120HE1693, K120HE1694, K120HE1695, K120HE1696, K120HE1697, K120HE1698, K120HE1699, K120HE1700, K120HE1701, K120HE1702, K120HE1703, K120HE1704, K120HE1705, K120HE1706, K120HE1707, K120HE1708, K120HE1709, K120HE1710, K120HE1711, K120HE1712, K120HE1713, K120HE1714, K120HE1715, K120HE1716, K120HE1717, K120HE1718, K120HE1719, K120HE1720, K120HE1721, K120HE1722, K120HE1723, K120HE1724, K120HE1725, K124HE1726, K124HE1727, K124HE1728, K124HE1729, K124HE1730, K124HE1731, K124HE1732, K124HE1733, K124HE1734, K124HE1735, K124HE1736, K124HE1737, K124HE1738, K124HE1739, K124HE1740, K124HE1741, K124HE1742, K124HE1743, K124HE1744, K124HE1745, K124HE1746, K124HE1747, K124HE1748, K124HE1749, K124HE1750, K124HE1751, K124HE1752, K124HE1753, K124HE1754, K124HE1755, K124HE1756, K124HE1757, K124HE1758, K124HE1759, K124HE1760, K124HE1761, K124HE1762, K124HE1763, K124HE1764, K125HE1765, K125HE1766, K125HE1767, K125HE1768, K125HE1769, K125HE1770, K125HE1771, K125HE1772, K125HE1773, K125HE1774, K125HE1775, K125HE1776, K125HE1777, K125HE1778, K125HE1779, K125HE1780, K125HE1781, K125HE1782, K125HE1783, K125HE1784, K125HE1785, K125HE1786, K125HE1787, K125HE1788, K125HE1789, K125HE1790, K125HE1791, K125HE1792, K125HE1793, K125HE1794, K125HE1795, K125HE1796, K125HE1797, K125HE1798, K125HE1799, K125HE1800, K125HE1801, K125HE1802, K125HE1803, K126HE1804, K126HE1805, K126HE1806, K126HE1807, K126HE1808, K126HE1809, K126HE1810, K126HE1811, K126HE1812, K126HE1813, K126HE1814, K126HE1815, K126HE1816, K126HE1817, K126HE1818, K126HE1819, K126HE1820, K126HE1821, K126HE1822, K126HE1823, K126HE1824, K126HE1825, K126HE1826, K126HE1827, K126HE1828, K126HE1829, K126HE1830, K126HE1831, K126HE1832, K126HE1833, K126HE1834, K126HE1835, K126HE1836, K126HE1837, K126HE1838, K126HE1839, K126HE1840, K126HE1841, K126HE1842, K127HE1843, K127HE1844, K127HE1845, K127HE1846, K127HE1847, K127HE1848, K127HE1849, K127HE1850, K127HE1851, K127HE1852, K127HE1853, K127HE1854, K127HE1855, K127HE1856, K127HE1857, K127HE1858, K127HE1859, K127HE1860, K127HE1861, K127HE1862, K127HE1863, K127HE1864, K127HE1865, K127HE1866, K127HE1867, K127HE1868, K127HE1869, K127HE1870, K127HE1871, K127HE1872, K127HE1873, K127HE1874, K127HE1875, K127HE1876, K127HE1877, K127HE1878, K127HE1879, K073HE0949, K073HE0950, K073HE0951, K073HE0952, K073HE0953, K073HE0954, K073HE0955, K073HE0956, K073HE0957, J200HE0858, J149HM0516, J149HM0517, J149HM0518, J149HM0519, J149HM0520, J149HM0521, J149HM0522, J149HM0523, J175HE0735, J175HE0736, J175HE0737, J175HE0738, J245HE1430, J245HE1431, J245HE1432, J245HE1433, J245HE1434, J245HE1435, J245HE1436, J245HE1437, J245HE1438, J245HE1439, J245HE1440, J245HE1441, J245HE1442, J245HE1443, J245HE1444, J245HE1445, J245HE1446, J245HE1447, J245HE1448, J245HE1449, J245HE1450, K072HE0909, K072HE0910, K072HE0911, K072HE0912, K072HE0913, K072HE0914, K072HE0915, K072HE0916, K072HE0917, K072HE0918, K072HE0919, K072HE0920, K072HE0921, K072HE0922, K072HE0923, K072HE0924, K072HE0925, K072HE0926, K072HE0927, K072HE0928, K072HE0929, K072HE0930, K072HE0931, K073HE0932, K073HE0933, K073HE0934, K073HE0935, K073HE0936, K073HE0937, K073HE0938, K073HE0939, K073HE0940, K073HE0941, K073HE0942, K073HE0943, K073HE0944, K073HE0945, K073HE0946, K073HE0947, K073HE0948, K079HE1021, K079HE1022, K079HE1023, K079HE1024, K079HE1025, K079HE1026, K111HE1475, K111HE1476, K111HE1477, K111HE1478, K111HE1479, K111HE1480, K111HE1481, K111HE1482, K111HE1483, K111HE1484, K154HE2201, K154HE2202, K154HE2203, K154HE2204, K154HE2205, K154HE2206, K154HE2207, K045HE0582, K045HE0583, K045HE0584, K045HE0585, K045HE0586, K045HE0587, K045HE0588, K045HE0589, K045HE0590, K045HE0591, K045HE0592, K045HE0593, K045HE0594, K045HE0595, K045HE0596, K045HE0597, K045HE0598, K045HE0599, K045HE0600, K045HE0601, K045HE0602, K045HE0603, K045HE0604, K045HE0605, K045HE0606, K045HE0607, K045HE0608, K045HE0609, K045HE0610, K045HE0611, K045HE0612, K045HE0613, K045HE0614, K045HE0615, K045HE0616, K045HE0617, K045HE0618, K047HE0619, K047HE0620, K047HE0621, K047HE0622, K047HE0623, K047HE0624, K047HE0625, K177HE2435, K177HE2436, K177HE2437, K177HE2438, K177HE2439, K177HE2440, K177HE2441, K177HE2442, K177HE2443, K177HE2444, K177HE2445, K177HE2446, K177HE2447, K177HE2448, K177HE2449, K177HE2450, K177HE2451, K177HE2452, K177HE2453, K177HE2454, K177HE2455, K177HE2456, K177HE2457, K177HE2458, K178HE2459, K178HE2460, K178HE2461, K178HE2462, K178HE2463, K178HE2464, K178HE2465, K178HE2466, K178HE2467, K178HE2468, K178HE2469, K178HE2470, K178HE2471, K178HE2472, K178HE2473, K178HE2474, K178HE2475, K178HE2476, K178HE2477, K178HE2478, K178HE2479, K178HE2480, K178HE2481, K178HE2482, K178HE2483, K178HE2484, K178HE2485, K178HE2486, K178HE2487, K178HE2488, K178HE2489, K178HE2490, K178HE2491, K178HE2492, K178HE2493, K178HE2494, K178HE2495, K178HE2496, K178HE2497, K178HE2498, K178HE2499, K178HE2500, K178HE2501, K178HE2502, K178HE2503, K178HE2504, K178HE2505, K178HE2506, K178HE2507, K178HE2508, J208HE0924, J208HE0925, J208HE0926, J208HE0927, J208HE0928, K104HE1339, K104HE1340, K104HE1341, K104HE1342, K104HE1343, K104HE1344, K104HE1345, K104HE1346, K104HE1347, K104HE1348, K104HE1349, K104HE1350, K104HE1351, K104HE1352, K104HE1353, K104HE1354, K104HE1355, K104HE1356, K104HE1357, K104HE1358, K104HE1359, K104HE1360, K104HE1361, K104HE1362, K104HE1363, K104HE1364, K104HE1365, K104HE1366, K104HE1367, K104HE1368, K104HE1369, K104HE1370, K104HE1371, K104HE1372, K104HE1373, K104HE1374, K104HE1375, K104HE1376, K104HE1377, K104HE1378, K104HE1379, K104HE1380, K104HE1381, K104HE1382, J175HE0739, J308HE1936, J308HE1937, J308HE1938, J308HE1939, J308HE1940, J308HE1941, K117HE1668, K117HE1669, J099HE0140, K104HE1338, J143HE0429, J143HE0430, J143HE0431, J143HE0432, J143HE0433, J143HE0434, J143HE0435, J143HE0436, J143HE0437, J143HE0438, J143HE0439, J143HE0440, J143HE0441, J143HE0442, J143HE0443, J143HE0444, J143HE0445, J143HE0446, J143HE0447, J143HE0448, J143HE0449, J143HE0450, J143HE0451, J155HM0582, J257HE1451, J257HE1452, J257HE1453, J257HE1454, J257HE1455, J257HE1456, J257HE1457, J257HE1458, J257HE1459, J257HE1460, J257HE1461, J257HE1462, J257HE1463, J257HE1464, J257HE1465, J257HE1466, J257HE1467, J257HE1468, J257HE1469, J257HE1470, J257HE1471, J257HE1472, J257HE1473, J257HE1474, J257HE1475, J257HE1476, J257HE1477, J257HE1478, J257HE1479, J257HE1480, J257HE1481, J257HE1482, J257HE1483, J257HE1484, J257HE1485, J146HE0504, J146HE0505, J146HE0506, J146HE0507, J146HE0508, J146HE0509, J146HE0510, J146HE0511, J146HE0512, J146HE0513, J146HE0514, J146HE0515, J151HE0524, J151HE0525, J151HE0526, J151HE0527, J151HE0528, J151HE0529, J151HE0530, J151HE0531, J151HE0532, J151HE0533, J151HE0534, J151HE0535, J151HE0536, J151HE0537, J151HE0538, J151HE0539, J151HE0540, J151HE0541, J151HE0542, J151HE0543, J152HM0557, J152HM0558, J152HM0559, J152HM0560, J152HM0561, J152HM0562, J152HM0563, J152HM0564, J152HM0565, J152HM0566, J152HM0567, J152HM0568, J152HM0569, J171HE0692, J171HE0693, J171HE0694, J171HE0695, J171HE0696, J171HE0697, J171HE0698, J171HE0699, J171HE0700, J171HE0701, J171HE0702, J171HE0703, J171HE0704, J171HE0705, J171HE0706, J171HE0707, J171HE0708, J171HE0709, J171HE0710, J171HE0711, J308HE1943, J308HE1944, J308HE1945, J308HE1946, J308HE1947, J308HE1948, J308HE1949, J308HE1950, J308HE1951, J308HE1952, J308HE1953, J308HE1954, J308HE1955, J308HE1956, J308HE1957, K052HE0626, K052HE0627, K052HE0628, K052HE0629, K052HE0630, K052HE0631, K052HE0632, K052HE0633, K052HE0634, K052HE0635, K052HE0636, K052HE0637, K052HE0638, K052HE0639, K052HE0640, K052HE0641, K052HE0642, K052HE0643, K052HE0644, K052HE0645, K052HE0646, K052HE0647, K180HE2509, K180HE2510, K180HE2511, K180HE2512, K180HE2513, K180HE2514, K180HE2515, K180HE2516, K180HE2517, K180HE2518, K180HE2519, K180HE2520, K180HE2521, K180HE2522, K180HE2523, K180HE2524, K180HE2525, K180HE2526, K180HE2527, K180HE2528, J249H31428, J249HE1429, J341HE2538, J341HE2539, J341HE2540, J341HE2541, J341HE2542, J341HE2543, J341HE2544, J341HE2545, J341HE2546, J341HE2547, J341HE2548, J341HE2549, J341HE2550, J341HE2551, J341HE2552, K112HE1485, K112HE1486, K112HE1487, K112HE1488, K112HE1489, K112HE1490, K112HE1491, K112HE1492, K112HE1493, K112HE1494, K112HE1495, K112HE1496, K112HE1497, K112HE1498, K112HE1499, K112HE1500, K112HE1501, K112HE1502, K112HE1503, K112HE1504, K112HE1505, K112HE1506, K112HE1507, K112HE1508, K112HE1509, K112HE1510, K112HE1511, K112HE1512, K112HE1513, K112HE1514, J332HE2294, J332HE2295, J332HE2296, J332HE2297, J332HE2298, J332HE2299, J332HE2300, J332HE2301, J332HE2302, J332HE2303, J332HE2304, J332HE2305, J332HE2306, J332HE2307, J332HE2308, J332HE2309, J332HE2310, J332HE2311, J332HE2312, J332HE2313, J332HE2314, J332HE2315, J332HE2316, J332HE2317, J332HE2318, J332HE2319, J332HE2320, J332HE2321, J332HE2322, J332HE2323, J332HE2324, J332HE2325, J332HE2326, J332HE2327, J332HE2328, J332HE2329, J332HE2330, J332HE2331, J332HE2332, J332HE2333, J332HE2334, J332HE2335, J333HE2336, J333HE2337, J333HE2338, J333HE2339, J333HE2340, J333HE2341, J333HE2342, J333HE2343, J333HE2344, J333HE2345, J333HE2346, J333HE2347, J333HE2348, J333HE2349, J333HE2350, J333HE2351, J333HE2352, J333HE2353, J333HE2354, J333HE2355, J333HE2356, J333HE2357, J333HE2358, J333HE2359, J333HE2360, J333HE2361, J333HE2362, J333HE2363, J333HE2364, J333HE2365, J333HE2366, J333HE2367, J333HE2368, J333HE2369, J333HE2370, J333HE2371, J333HE2372, J333HE2373, J333HE2374, J333HE2375, J333HE2376, J333HE2377, J334HE2378, J334HE2379, J334HE2380, J334HE2381, J334HE2382, J334HE2383, J334HE2384, J334HE2385, J334HE2386, J334HE2387, J334HE2388, J334HE2389, J334HE2390, J334HE2391, J334HE2392, J334HE2393, J282HE1687, J249H31427, J269HE1486, J269HE1487, J269HE1488, J269HE1489, J269HE1490, J269HE1491, J269HE1492, J269HE1493, J269HE1494, J269HE1495, J269HE1496, J269HE1497, J269HE1498, J269HE1499, J269HE1500, J269HE1501, J269HE1502, J269HE1503, J269HE1504, J269HE1505, J269HE1506, J269HE1507, J269HE1508, J269HE1509, J269HE1510, J269HE1511, J269HE1512, J269HE1513, J269HE1514, J269HE1515, J269HE1516, J269HE1517, J269HE1518, J269HE1519, J269HE1520, J269HE1521, J270HE1522, J270HE1523, J270HE1524, J270HE1525, J270HE1526, J270HE1527, J270HE1528, J270HE1529, J270HE1530, J270HE1531, J270HE1532, J270HE1533, J270HE1534, J270HE1535, J270HE1536, J270HE1537, J270HE1538, J270HE1539, J270HE1540, J270HE1541, J270HE1542, J270HE1543, J270HE1544, J270HE1545, J270HE1546, J270HE1547, J270HE1548, J270HE1549, J270HE1550, J270HE1551, J270HE1552, J270HE1553, J270HE1554, J270HE1555, J270HE1556, J270HE1557, J270HE1558, J270HE1559, J270HE1560, J270HE1561, J270HE1562, J270HE1563, J270HE1564, J270HE1565, J270HE1566, J270HE1567, J270HE1568, J270HE1569, J270HE1570, J270HE1571, J270HE1572, J270HE1573, J270HE1574, J270HE1575, J270HE1576, J270HE1577, J270HE1578, J270HE1579, J270HE1580, J270HE1581, J271HE1582, J271HE1583, J271HE1584, J271HE1585, J271HE1586, J271HE1587, J271HE1588, J271HE1589, J271HE1590, J271HE1591, J271HE1592, J271HE1593, J271HE1594, J271HE1595, J271HE1596, J271HE1597, J271HE1598, J271HE1599, J271HE1600, J271HE1601, J271HE1602, J271HE1603, J271HE1604, J271HE1605, J271HE1606, J271HE1607, J271HE1608, J271HE1609, J271HE1610, J271HE1611, J271HE1612, J271HE1613, J271HE1614, J271HE1615, J271HE1616, J271HE1617, J271HE1618, J271HE1619, J271HE1620, J271HE1621, J271HE1622, J271HE1623, J271HE1624, J271HE1625, J271HE1626, J271HE1627, J271HE1628, J271HE1629, J271HE1630, J271HE1631, J271HE1632, J271HE1633, J271HE1634, J271HE1635, J271HE1636, J271HE1637, J271HE1638, J271HE1639, J271HE1640, J271HE1641, J272HE1642, J272HE1643, J272HE1644, J272HE1645, J272HE1646, J272HE1647, J272HE1648, J272HE1649, J272HE1650, J272HE1651, J272HE1652, J272HE1653, J272HE1654, J272HE1655, J272HE1656, J272HE1657, J272HE1658, J272HE1659, J272HE1660, J272HE1661, J272HE1662, J272HE1663, J272HE1664, J272HE1665, J272HE1666, J272HE1667, J272HE1668, J272HE1669, J272HE1670, J272HE1671, J272HE1672, J272HE1673, J272HE1674, J272HE1675, J272HE1676, J272HE1677, J272HE1678, J272HE1679, J272HE1680, J272HE1681, J272HE1682, J272HE1683, J272HE1684, J272HE1685, K007H30043, K007HE0041, K007HE0042, K119HE1667, K189HE2671, K189HE2672, K189HE2673, K189HE2674, K189HE2675, K189HE2676, K189HE2677, K189HE2678, K189HE2679, K189HE2680, K189HE2681, K189HE2682, K189HE2683, K190HE2684, K190HE2685, K190HE2686, K194HE2687, K194HE2688, K194HE2689, K194HE2690, K194HE2691, K195HE2692, K195HE2693, K195HE2694, K195HE2695, K195HE2696, K195HE2697, K195HE2698, J282HE1686, J187HE0740, J187HE0741, J187HE0742, J187HE0743, J187HE0744, J187HE0745, J187HE0746, J187HE0747, J187HE0748, J187HE0749, J187HE0750, J187HE0751, J187HE0752, J187HE0753, J187HE0754, J187HE0755, J187HE0756, J187HE0757, J187HE0758, J187HE0759, J187HE0760, J187HE0761, J187HE0762, J187HE0763, J187HE0764, J187HE0765, J187HE0766, J187HE0767, J187HE0768, J187HE0769, K164HE2219, K164HE2220, J321HE2157, J321HE2158, J321HE2159, J321HE2160, J321HE2161, J321HE2162, J321HE2163, J321HE2164, J321HE2165, J321HE2166, J321HE2167, J321HE2168, J321HE2169, J321HE2170, J321HE2171, J321HE2172, J321HE2173, J321HE2174, J321HE2175, J321HE2176, J321HE2177, J321HE2178, J321HE2179, J321HE2180, J321HE2181, J321HE2182, J321HE2183, J321HE2184, J321HE2185, J321HE2186, J321HE2187, J321HE2188, UNKNOWN, J291HE1792, J291HE1793, J291HE1794, J291HE1795, J291HE1796, J291HE1797, J291HE1798, J291HE1799, J291HE1800, J291HE1801, K008HE0044, K008HE0045, K008HE0046, K008HE0047, K008HE0048, K008HE0049, K008HE0050, K008HE0051, K008HE0052, K008HE0053, K008HE0054, K008HE0055, K008HE0056, K008HE0057, K008HE0058, K008HE0059, K008HE0060, K008HE0061, K008HE0062, K008HE0063, K008HE0064, K008HE0065, K008HE0066, K008HE0067, K008HE0068, K008HE0069, K008HE0070, K009HE0071, K009HE0072, K009HE0073, K009HE0074, K009HE0075, K009HE0076, K009HE0077, K139HE1880, K139HE1881, K139HE1882, K139HE1883, K139HE1884, K139HE1885, K139HE1886, K139HE1887, K139HE1888, K139HE1889, K139HE1890, K139HE1891, K139HE1892, K139HE1893, K139HE1894, K139HE1895, K139HE1896, K139HE1897, K139HE1898, K139HE1899, K139HE1900, K139HE1901, K139HE1902, K139HE1903, K139HE1904, K139HE1905, K139HE1906, K139HE1907, K139HE1908, K139HE1909, K139HE1910, K139HE1911, K139HE1912, K139HE1913, K139HE1914, K139HE1915, K139HE1916, K139HE1917, K139HE1918, K139HE1919, K139HE1920, K139HE1921, K139HE1922, K139HE1923. |
Recalling Firm/ Manufacturer |
Hill-Rom, Inc. 125 E Pearl St 1069 State Route 46 East Batesville IN 47006
|
For Additional Information Contact |
812-934-7777
|
Manufacturer Reason for Recall |
The flexing of the junction of the sleep deck's head and mid sections can create excess pressure on the pivot pin E-ring, which could cause the E-ring to come off.
|
FDA Determined Cause 2 |
Device Design |
Action |
The firm sent an Urgent Field Safety Notice to its customers dated 9/16/2009. The letter identified the affected product, explained the reason for the recall, actions to be taken, and a contact reference. Customers are to locate the affected product, examine the pivot points and ensure the E-ring is present. If the E-ring is missing, then the affected bed is to be taken out of service. Customers' distributors or agents will contact them with a permanent fix. Questions should be directed towards Hill-Rom Technical Support. |
Quantity in Commerce |
2,973 beds |
Distribution |
Worldwide Distribution -- Argentina, Bahrain, Bhutan, Brazil, Bulgaria, Chili, China, Columbia, Costa Rica, Ecuador, Egypt, Eritrea, France, Hong Kong, India, Indonesia, Iran, Iraq, Jordan, Kenya, Kuwait, Lebanon, Libya, Malaysia, Mexico, Morocco, Oman, Pakistan, Palestine, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Puerto Rico, Qatar, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Syria, Taiwan, Thailand, Trinidad & Tobago, Turkey, United Arab Emirates, United States, Uruguay, Venezuela, and Vietnam. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|