 Class 3 Device Recall Abbott RealTime HIV1 Assaysee related information
Date Initiated by FirmDecember 20, 2010
Date PostedFebruary 23, 2011
Recall Status1 Terminated 3 on November 19, 2012
Recall NumberZ-1398-2011
Recall Event ID 57431
Product Classification Monitor, test, hiv - Product Code MTL
ProductAbbott RealTime HIV-1 Assay, List 6L18, in vitro diagnostic; Abbott Molecular Inc., Des Plaines, IL 60018 USA The Abbott RealTime HIV-1 assay is an in vitro reverse transcription-polymerase chain reaction (RT-PCR) assay for the quantitation of Human Immunodeficiency Virus type 1 (HIV-1) on the automated m2000 System in human plasma from HIV-1 infected individuals over the range of 40 to 10,000,000 copies/mL. The Abbott RealTime HIV-1 assay is intended for use in conjunction with clinical presentation and other laboratory markers for disease prognosis and for use as an aid in assessing viral response to antiretroviral treatment as measured by changes in plasma HIV-1 RNA levels. This assay is not intended to be used as a donor screening test for HIV-1 or as a diagnostic test to confirm the presence of HIV-1 infection.
Code Information All in date lots
Recalling Firm/
Manufacturer
Abbott Molecular
1300 E Touhy Ave
Des Plaines IL 60018-3315
For Additional Information Contact
224-361-7133
Manufacturer Reason
for Recall
Internal testing identified the potential of under filled and over filled enzyme reagent vials which may result in obtaining less than the total number of tests per kit as described on the package insert.
FDA Determined
Cause 2
Other
ActionAbbott Molecular sent a Field Advisory Notice letter to all RealTime HIV-1, CT/NG and HBV assay customers on December 20, 2010, informing them that internal testing identified the potential of under filled and over filled enzyme vials which may result in obtaining less than the total number of tests per kit as described in the package insert. The customers were told that specimens that meet acceptance criteria, per the product labeling are not affected by the situation and are considered valid results, and that the issue has no impact on results for those runs that generate valid results. Any questions were directed to Abbott Technical Support at 1-800-553-7042, Option 2.
DistributionNationwide Distribution