Class 2 Device Recall PHILIPS MODEL MX 16slice Computed Tomography Scanners

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall PHILIPS MODEL MX 16slice Computed Tomography Scanners see related information
Date Initiated by Firm March 10, 2011
Date Posted April 26, 2011
Recall Status1 Terminated 3 on March 01, 2013
Recall Number Z-2059-2011
Recall Event ID 58209
510(K)Number K083498  K091195  
Product Classification System, x-ray, tomography, computed - Product Code JAK
Product MX 16-slice Computed Tomography Scanners, equipped with Software Version 1.1.1. Model #E989605657371.

The MX 16-slice CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously. The acquired X-RAY transmission data is reconstructed by computer into cross-sectional images of the body from the same axial plane taken at different angles. The system is suitable for all patients.
Code Information Serial numbers are: P16C080002, P16E080005, P16E080007, P16E080008, P16E080012, P16E080010, P16E090004, P16E090008, P16E090005, P16E080007, P16E080008, P16E080012, P16E080010, P16E090004, P16E090008, P16E090005, P16E090006, P16E090007, P16E090009, P16E090012, P16E090003, P16E090010, P16E090011, P16E090013, P16EM09003, P16EM09004, P16E090002, P16E090014, P16E090015, P16EM08003, P16E090016, P16E090018, P16E090019, P16E090020, P16E090021, P16E090026, EP16E090001, P16E090028, P16E090027, P16E090017, P16EM09002, P16E090024, P16E090025, EP16E090002, EP16E090003, EP16E090004, EP16E090006, EP16E090007, EP16E090008, EP16E090009, P16C090001, EP16E090010, EP16E090011, EP16E090012, EP16E090014, EP16E090015, EP16E100001, EP16E100002, EP16E100004, EP16E100003, EP16E100005, EP16E100006, EP16E100007, EP16E090013, EP16E100008, EP16E100009, EP16E100010, P16C100001, EP16E100011, EP16E100012, EP16E100013, EP16E100014, EP16E100015, EP16E100016, EP16E100017, EP16E100018, EP16E100019, P16C080001, EP16E100020, EP16E100021, EP16E100022, EP16E100023, EP16E100024, EP16E100027, EP16E100026, EP16E100030, EP16E100031, EP16E100029, P16C100003, EP16E100025, EP16E100028, P16C100002, EP16E100034, EP16E100033, EP16E100032, EP16E100035, EP16E100036, EP16E100037, P16E100002, P16C100004, EP16E100039, EP16E100038, EP16E100042, EP16E100043, EP16E100041, EP16E100044, EP16E100046, EP16E100045, EP16E100048, EP16E100047, EP16E100051, EP16E100049, EP16E100050, EP16E100053, EP16E100052, EP16E100054, EP16E100056, EP16E100055, P16C100005, EP16E100057, EP16E100058, EP16E100059, EP16E100060, EP16E100061, EP16E100062, EP16E100063, P16C100007, EP16E100067, EP16ER100003, EP16ER100004, EP16ER100005, EP16E100070, EP16E100069, EP16E100071, EP16E100072, EP16E100073, EP16E100074, EP16E100078, EP16E100075, EP16E100076, EP16E100077, EP16E100079, EP16E100080, EP16E100082, EP16E100081, EP16E100083, EP16E100084, P16C100008, P16C100009, EP16E100085, EP16E100086, EP16E100087, P16C100010, EP16E100088, EP16E100089, EP16E100090, EP16E100091, EP16E100093, EP16E100094, EP16E100092, P16C100014, P16C100015, EP16E100095, P16C100011, EP16E100096, EP16E100097, EP16E100098, P16C100017, EP16E100099, EP16E100100, EP16E100101, EP16E100102, P16C100013, EP16E100103, EP16E100104, EP16E100105, EP16E100106, EP16E100109, EP16E100107, P16C100012, EP16E100108, EP16E100111, EP16E100110, P16C100018, EP16E100112, EP16ER100008, EP16ER100007, EP16E100113, EP16E110003, EP16E110001, EP16E110002, EP16E110004, and EP16E110005.  
Recalling Firm/
Manufacturer
Philips Medical Systems (Cleveland) Inc
595 Miner Road
Cleveland OH 44143-2131
For Additional Information Contact
440-483-7000
Manufacturer Reason
for Recall
The Philips MX 16-slice CT scanners are subject to recall/field correction due to a problem encountered with the units when using the 'Combine Viewing' Option. The firm determined that inaccurate dimension measurements display when using the Combine Viewing Option. Moreover, if the dimension measurements from merged images created using this option are used, there is a potential for applying inac
FDA Determined
Cause 2
Other
Action Philips Healthcare issued an 'URGENT- Field Safety Notice MX 16-slice' (dated 3/10/2011) concerning a software issue to customers. The notification informed the firm's users/customers of the potential for inaccurate dimension measurements produced by the MX 16-slice CT scanners when using the Combine Function Viewing Option. Moreover, the letter notified customers of the potential for patient misdiagnosis in the event that inaccurate dimension measurements are used by a physician for reference. The Field Safety Notice described the following: 1) How to identify the affected product; 2) What actions should to be taken by the customers with regard to avoiding the issue -discontinue use of the Combine Viewing Option and instead use the MPR Batch for diagnostic reviews until a planned software upgrade is installed by the recalling firm; and 3) The firm's plans to carry out the installation of a field change order kits to correct the issue. The planned software upgrade is expected to be completed on or before 6 months after the new software is released. If you need any further information or support concerning this issue, please contact your local Philips representative or local Philips Healthcare office. For North America and Canada contact the Customer Care Solutions Center (1-800-722-9377, option 5: Enter Site ID or follow the prompts).
Quantity in Commerce 25 units were distributed domestically & 168 system units were distributed both domestically & internationally
Distribution Worldwide Distribution -- United States (AL, AR, CA, FL, GA, IL, IN, MS, NJ, NV, NY, OH, PA, SC and TX) including Puerto Rico and countries of Albania, Argentina, Australia, Belgium Brazil, Chile, Columbia, Dominican Republic, Ecuador, Germany, Greece, India, Indonesia, Italy, japan, Kenya, Lebanon, Malawi, The Netherlands, Philippines, Poland, Romania, Saudi Arabia, South Africa, Spain, Switzerland, Turkey, Uganda, United Arab Emirates and Venezuela
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = JAK and Original Applicant = PHILIPS MEDICAL SYSTEMS(CLEVELAND), INC.
-
-