| | Class 2 Device Recall PHILIPS MODEL MX 16slice Computed Tomography Scanners |  |
| Date Initiated by Firm | March 10, 2011 |
| Date Posted | April 26, 2011 |
| Recall Status1 |
Terminated 3 on March 01, 2013 |
| Recall Number | Z-2059-2011 |
| Recall Event ID |
58209 |
| 510(K)Number | K083498 K091195 |
| Product Classification |
System, x-ray, tomography, computed - Product Code JAK
|
| Product | MX 16-slice Computed Tomography Scanners, equipped with Software Version 1.1.1. Model #E989605657371.
The MX 16-slice CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously. The acquired X-RAY transmission data is reconstructed by computer into cross-sectional images of the body from the same axial plane taken at different angles. The system is suitable for all patients. |
| Code Information |
Serial numbers are: P16C080002, P16E080005, P16E080007, P16E080008, P16E080012, P16E080010, P16E090004, P16E090008, P16E090005, P16E080007, P16E080008, P16E080012, P16E080010, P16E090004, P16E090008, P16E090005, P16E090006, P16E090007, P16E090009, P16E090012, P16E090003, P16E090010, P16E090011, P16E090013, P16EM09003, P16EM09004, P16E090002, P16E090014, P16E090015, P16EM08003, P16E090016, P16E090018, P16E090019, P16E090020, P16E090021, P16E090026, EP16E090001, P16E090028, P16E090027, P16E090017, P16EM09002, P16E090024, P16E090025, EP16E090002, EP16E090003, EP16E090004, EP16E090006, EP16E090007, EP16E090008, EP16E090009, P16C090001, EP16E090010, EP16E090011, EP16E090012, EP16E090014, EP16E090015, EP16E100001, EP16E100002, EP16E100004, EP16E100003, EP16E100005, EP16E100006, EP16E100007, EP16E090013, EP16E100008, EP16E100009, EP16E100010, P16C100001, EP16E100011, EP16E100012, EP16E100013, EP16E100014, EP16E100015, EP16E100016, EP16E100017, EP16E100018, EP16E100019, P16C080001, EP16E100020, EP16E100021, EP16E100022, EP16E100023, EP16E100024, EP16E100027, EP16E100026, EP16E100030, EP16E100031, EP16E100029, P16C100003, EP16E100025, EP16E100028, P16C100002, EP16E100034, EP16E100033, EP16E100032, EP16E100035, EP16E100036, EP16E100037, P16E100002, P16C100004, EP16E100039, EP16E100038, EP16E100042, EP16E100043, EP16E100041, EP16E100044, EP16E100046, EP16E100045, EP16E100048, EP16E100047, EP16E100051, EP16E100049, EP16E100050, EP16E100053, EP16E100052, EP16E100054, EP16E100056, EP16E100055, P16C100005, EP16E100057, EP16E100058, EP16E100059, EP16E100060, EP16E100061, EP16E100062, EP16E100063, P16C100007, EP16E100067, EP16ER100003, EP16ER100004, EP16ER100005, EP16E100070, EP16E100069, EP16E100071, EP16E100072, EP16E100073, EP16E100074, EP16E100078, EP16E100075, EP16E100076, EP16E100077, EP16E100079, EP16E100080, EP16E100082, EP16E100081, EP16E100083, EP16E100084, P16C100008, P16C100009, EP16E100085, EP16E100086, EP16E100087, P16C100010, EP16E100088, EP16E100089, EP16E100090, EP16E100091, EP16E100093, EP16E100094, EP16E100092, P16C100014, P16C100015, EP16E100095, P16C100011, EP16E100096, EP16E100097, EP16E100098, P16C100017, EP16E100099, EP16E100100, EP16E100101, EP16E100102, P16C100013, EP16E100103, EP16E100104, EP16E100105, EP16E100106, EP16E100109, EP16E100107, P16C100012, EP16E100108, EP16E100111, EP16E100110, P16C100018, EP16E100112, EP16ER100008, EP16ER100007, EP16E100113, EP16E110003, EP16E110001, EP16E110002, EP16E110004, and EP16E110005. |
| FEI Number |
1525965
|
Recalling Firm/ Manufacturer |
Philips Medical Systems (Cleveland) Inc
595 Miner Road
Cleveland
OH
44143-2131
|
| For Additional Information Contact | 440-483-7000 |
Manufacturer Reason for Recall | The Philips MX 16-slice CT scanners are subject to recall/field correction due to a problem encountered with the units when using the 'Combine Viewing' Option. The firm determined that inaccurate dimension measurements display when using the Combine Viewing Option. Moreover, if the dimension measurements from merged images created using this option are used, there is a potential for applying inac |
FDA Determined Cause 2 | Other |
| Action | Philips Healthcare issued an 'URGENT- Field Safety Notice MX 16-slice' (dated 3/10/2011) concerning a software issue to customers. The notification informed the firm's users/customers of the potential for inaccurate dimension measurements produced by the MX 16-slice CT scanners when using the Combine Function Viewing Option. Moreover, the letter notified customers of the potential for patient misdiagnosis in the event that inaccurate dimension measurements are used by a physician for reference. The Field Safety Notice described the following: 1) How to identify the affected product; 2) What actions should to be taken by the customers with regard to avoiding the issue -discontinue use of the Combine Viewing Option and instead use the MPR Batch for diagnostic reviews until a planned software upgrade is installed by the recalling firm; and 3) The firm's plans to carry out the installation of a field change order kits to correct the issue.
The planned software upgrade is expected to be completed on or before 6 months after the new software is released.
If you need any further information or support concerning this issue, please
contact your local Philips representative or local Philips Healthcare office.
For North America and Canada contact the Customer Care Solutions Center
(1-800-722-9377, option 5: Enter Site ID or follow the prompts). |
| Quantity in Commerce | 25 units were distributed domestically & 168 system units were distributed both domestically & internationally |
| Distribution | Worldwide Distribution -- United States (AL, AR, CA, FL, GA, IL, IN, MS, NJ, NV, NY, OH, PA, SC and TX) including Puerto Rico and countries of Albania, Argentina, Australia, Belgium Brazil, Chile, Columbia, Dominican Republic, Ecuador, Germany, Greece, India, Indonesia, Italy, japan, Kenya, Lebanon, Malawi, The Netherlands, Philippines, Poland, Romania, Saudi Arabia, South Africa, Spain, Switzerland, Turkey, Uganda, United Arab Emirates and Venezuela |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = JAK
|
|
|
|