| Date Initiated by Firm | August 19, 2020 |
| Create Date | September 23, 2020 |
| Recall Status1 |
Terminated 3 |
| Recall Number | Z-3005-2020 |
| Recall Event ID |
86185 |
| Product | Software: Syngo.via RT Image Suite software version VA30_SP02 with license T_RT_IMAGE_SUITE_PAT_MARK
Intended Use: used as an advanced application on a SOM10 scanner SMN: 10496180 |
Recalling Firm/ Manufacturer |
Siemens Medical Solutions USA, Inc 40 Liberty Blvd Malvern PA 19355
|
| For Additional Information Contact | SAME 610-219-4834 |
Manufacturer Reason for Recall | Calibration offset between the central CT image plane and the external RT lasers has been set automatically to the factory default of 50.00 cm, if the misadjustment of the offset value is not recognized during the subsequent treatment planning process or remains undetected during the position verification at the treatment system (e.g. linear accelerator), this can potentially result in an incorrect dose or dose applied to the wrong location |
| Action | Siemens Medical Solutions USA, Inc. issued Customer Safety Advisory Letter (CSAN) SY051/20/S on August 19, 2020 via email/hand delivery by Siemens Customer Service Engineer,. The letter states reason for recall, health risk and action to take: we highly recommend performing the laser calibration and the quality assurance procedure for the laser offset after every RT Image Suite software update.Please update the calibration offset in the MM RT Image Suite Configuration dialog accordingly . Details of this procedure are documented in Chapter 13.2 (Laser configuration) of your user manual.
If you have any questions or require technical support, please contact our service organization at 1-800-888-7436.
The technical solution distributed via SY040/20/S. Distribution of SY040/20/S (VB40B_HF01) will be distributed to all affected customers, free of charge. Distribution of SY040/20/S began on August 26, 2020. SY040/20/S will be installed with support from your local service organization |
| Quantity in Commerce | 64 systems US |
| Distribution | US Nationwide distribution. |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|