• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Medical Device Recalls

  • Print
  • Share
  • E-mail
-
  Medical Device Recalls see related information
Date Initiated by Firm June 19, 2012
Create Date September 07, 2012
Recall Status1 Terminated 3
Recall Number Z-2337-2012
Recall Event ID 62925
Product STERRAD 100S Cassette, Batch # 12A077.

The STERRAD 100S and STERRAD NX sterilizations systems are low-temperature general purpose sterilizers used to sterilize heat and moisture sensitive reusable medical devices. The sterilant used is hydrogen peroxide supplied in cassette form as a separate accessory. The STERRAD Cassettes are used in conjunction with the STERRAD Sterilizers, and each cassette model is for use only in the matching sterilizer model.
Code Information Product Code: 10113
Recalling Firm/
Manufacturer
Advanced Sterilization Products
33 Technology Dr
Irvine CA 92618-2346
For Additional Information Contact
949-453-6400
Manufacturer Reason
for Recall
Advanced Sterilization Products (ASP) is recalling (removing) the Sterrad 100S and Sterrad NX System cassettes because the camera used to verify printing on the Sterrad System cassettes was inadvertently disabled. The STERRAD 100S and STERRAD NX System cases were shipped without verification of proper printing.
Action Advanced Sterilization Products sent an "URGENT MEDICAL DEVICE RECALL (REMOVAL) letter dated July 16, 2012 to all their customers who purchased the STERRAD 100S and STERRAD NX System cassettes. The letter informed the customers of the problem identified and the actions to be taken. For the STERRAD NX System cassettes, there should be no impact to the functionality of these affected cassettes. Although printing may be smeared, illegible, or missing, the packaging has multiple references to the information, and the cassettes should be usable despite any printing error. For the STERRAD 100S System cassettes, there is a chance that the barcode will be unreadable and the STERRAD 100S Sterilizer will not accept the cassette. Customers were instructed to complete and return the business reply card and packing slip that are included in the letter. Customers were instructed to contact Stericycle at (888) 406-9303 for additional labels. Customers with questions about the recall letter were instructed to contact Stericycle at (888) 406-9303.
Quantity in Commerce 7,495 units
Distribution Nationwide Distribution.

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-