  Medical Device Recalls see related information
Date Initiated by Firm January 19, 2010
Create Date May 26, 2010
Recall Status1 Terminated 3
Recall Number Z-1681-2010
Recall Event ID 54473
Product InSound Medical, Inc, Lyric in-canal hearing aid.
Code Information All devices manufactured between 12/10/2009 and 12/24/2009. Expiration dates 12/10/2010 through 12/24/2010.
Recalling Firm/
Manufacturer
InSound Medical, Inc.
37500 Central Ct
Newark CA 94560-3454
For Additional Information Contact
510-792-4000 Ext. 470
Manufacturer Reason
for Recall
Manufacturing error could result in electrolyte leakage from the product's battery.
Action InSound sent to its consignees a "Dear Lyric Provider" letter of explanation on 2/1/2010.
Quantity in Commerce 2019 units
Distribution Distribution only in US ( nationwide). No foreign consignees involved.