Part B: Supplementary Information Sheet (SIS)
FR Recognition List Number 031 Date of Entry 08/06/2013 
FR Recognition Number 14-383
Standard
ISO ASTM  51702 Third edition 2013-04-15
Practice for dosimetry in a gamma facility for radiation processing
Scope/Abstract
ISO/ASTM 51702:2013 outlines the installation qualification program for an irradiator and the dosimetric procedures to be followed during operational qualification, performance qualification, and routine processing in facilities that process products with ionizing radiation from radionuclide gamma sources to ensure that product has been treated within a predetermined range of absorbed dose. Other procedures related to operational qualification, performance qualification, and routine processing that may influence absorbed dose in the product are also discussed.
NOTE 1 — Dosimetry is only one component of a total quality assurance program for adherence to good manufacturing practices used in radiation processing applications.
NOTE 2 — ISO/ASTM Practices 51818 and 51649 describe dosimetric procedures for low and high enery electron beam facilities for radiation processing and ISO/ASTM Practice 51608 describes procedures for X-ray (bremsstrahlung) facilities for radiation processing.
Extent of Recognition
Complete standard
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.
Relevant FDA Guidance and/or Supportive Publications*
Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018.
FDA Technical Contacts
 Johannetsy Avillan
  FDA/OC/CDRH/OPEQ/OHTIV/DHTIVC/
  --
  Johannetsy.Avillan@fda.hhs.gov
 Paulo R. Laranjeira
  FDA/OC/CDRH/OPEQ/OHTIV/DHTIVB/
  301-796-9024
  Paulo.Laranjeira@fda.hhs.gov
Standards Development Organizations
ISO International Organization for Standardization https://www.iso.org/
ASTM ASTM International http://www.astm.org/
FDA Specialty Task Group (STG)
Sterility
*These are provided as examples and others may be applicable.
-
-