• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Recognized Consensus Standards

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

New Search Back To Search Results
Part B: Supplementary Information Sheet (SIS)
FR Recognition List Number 034 Date of Entry 01/30/2014 
FR Recognition Number 11-191
Standard
ISO 14879-1 First edition 2000-06-01 Implants for surgery - Total knee-joint prostheses - Part 1: Determination of endurance properties of knee tibial trays
Scope/Abstract
This part of ISO 14879 specifies a test method for determining the endurance properties, under specified laboratory conditions, of tibial trays used in knee-joint prostheses to support and secure the plastics articulating surface. It applies to tibial trays which cover both the medial and lateral plateaux of the tibia.

The test method does not apply to tibial components manufactured solely from plastics materials.

This part of ISO 14879 does not cover methods of examining and reporting the final condition of the test specimen; these may be the subject of agreement between the test laboratory and the parties submitting the specimen for test.
Extent of Recognition
Complete standard
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.
Public Law, CFR Citation(s) and Procode(s)*
Regulation Number Device Name Device Class Product Code
Unclassified Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Polymer/Metal/Polymer Class 3 LXY
Unclassified Prosthesis, Knee, Patello/Femorotibial, Unconstrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer Class 3 MBD
Regulation Number Device Name Device Class Product Code
§888.3480 Metal Cemented Constrained Femorotibial Knee Prosthesis Class 3 KRN
Regulation Number Device Name Device Class Product Code
§888.3510 Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer Class 2 KRO
Regulation Number Device Name Device Class Product Code
§888.3520 Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer Class 2 HSX
Regulation Number Device Name Device Class Product Code
§888.3530 Prosthesis, Knee, Femorotibial, Semi-Constrained, Cemented, Metal/Polymer Class 2 HRY
Regulation Number Device Name Device Class Product Code
§888.3550 Prosthesis, Knee, Patello/Femorotibial, Constrained, Cemented, Polymer/Metal/Metal Class 3 KRP
§888.3550 Prosthesis, Knee, Patello/Femorotibial, Constrained, Cemented, Polymer/Metal/Polymer Class 2 KRQ
Regulation Number Device Name Device Class Product Code
§888.3560 Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uhmwpe, Pegged, Cemented, Polymer/Metal/Polymer Class 2 MBV
§888.3560 Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer Class 2 JWH
Regulation Number Device Name Device Class Product Code
§888.3565 Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer Class 2 MBH
Relevant FDA Guidance and/or Supportive Publications*
Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA, January 16, 2003
Standards Development Organization
ISO International Organization for Standardization https://www.iso.org/
FDA Specialty Task Group (STG)
Orthopedic
*These are provided as examples and others may be applicable.
-
-