| Part B: Supplementary Information Sheet (SIS) |
|
FR Recognition List Number
|
047
|
Date of Entry 08/21/2017
|
|
FR Recognition Number
|
6-400
|
| Standard | |
ASTM F3212-16(2023) Standard Test Method for Coring Testing of Huber Needles |
|
Scope/Abstract1.1 This test method covers the qualitative measurement of Huber-type needles' potential to remove septum material during implantable port access (1)2. 1.2 This test method does not address other issues that may include, but are not limited to, force measurement during the perforation/withdrawal, septum integrity, and any safety issues. |
|
| Extent of Recognition
|
Rationale for Recognition
| This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies. |
|
Public Law, CFR Citation(s) and Procode(s)*
| Regulation Number |
Device Name |
Device Class |
Product Code |
| §880.5570 |
Needle, Hypodermic, Single Lumen
|
Class 2
|
FMI
|
|
Relevant FDA Guidance and/or Supportive Publications*
| Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018. |
|
| FDA Technical Contact
|
| Standards Development Organization
|
FDA Specialty Task Group (STG)
| General Plastic Surgery/General Hospital |
|
|
|
| *These are provided as examples and others may be applicable. |