Part B: Supplementary Information Sheet (SIS) |
FR Recognition List Number
|
034
|
Date of Entry 01/30/2014
|
FR Recognition Number
|
9-89
|
Standard | |
ISO 8638 Third edition 2010-07-01 Cadiovascular implants and extracorporeal blood circuit for hemodialyzers, hemodialfilters, and hemofilters |
|
U.S. Identical AdoptionANSI AAMI ISO 8638:2010/(R)2015 Cardiovascular implants and Extracorporeal blood circuit for hemodialyzers, hemodiafilters, and hemofilters |
|
Scope/AbstractISO 8638:2010 specifies requirements for haemodialysers, haemodiafilters, haemofilters and haemoconcentrators (hereafter referred to as "the device") and (integral and non-integral) transducer protectors which are intended for use in haemodialysis, haemodiafiltration and haemofiltration. |
|
Extent of Recognition
Partial recognition. The following part(s) of the standard is (are) not recognized |
Section 4.4.9.1; Section 4.4.9.2; Section 5.5.6.1; Section 5.5.6.2; Section 5.5.9;Section 5.5.10; Section 6.1 NOTE; Section 6.2e NOTE; Section 6.2 NOTE; Section 6.3 NOTE; and Section 6.4 NOTE. |
|
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.
FDA does not recognize these sections listed above because they are intended for quality control in hemodialysis clinics and not intended for a manufacturer to meet premarket requirements for these devices under the FD&C Act. |
|
Public Law, CFR Citation(s) and Procode(s)*
Regulation Number |
Device Name |
Device Class |
Product Code |
§876.5820 |
Accessories, Blood Circuit, Hemodialysis |
Class 2 |
KOC |
§876.5820 |
Adaptor, Fistula |
Class 2 |
FKM |
§876.5820 |
Adaptor, Y |
Class 2 |
FJP |
§876.5820 |
Connector, Blood Tubing, Infusion T |
Class 2 |
FKB |
§876.5820 |
Connector, Tubing, Dialysate |
Class 2 |
FKY |
§876.5820 |
Declotting Tray, Kit (Including Contents) |
Class 2 |
FJZ |
§876.5820 |
Filter, Blood, Dialysis |
Class 2 |
FKJ |
§876.5820 |
Protector, Transducer, Dialysis |
Class 2 |
FIB |
§876.5820 |
Set, Dialysis, Single Needle With Uni-Directional Pump |
Class 2 |
FIF |
§876.5820 |
Set, Dialysis, Single Needle With Uni-Directional Pump, Reprocessed |
Class 2 |
NNG |
§876.5820 |
Set, Tubing, Blood, With And Without Anti-Regurgitation Valve |
Class 2 |
FJK |
§876.5820 |
System Accessories, Extracorporeal |
Class 2 |
KXM |
§876.5820 |
System, Blood, Extracorporeal And Accessories |
Class 2 |
LLB |
§876.5820 |
Tray, Start/Stop (Including Contents), Dialysis |
Class 1 |
FKG |
§876.5820 |
Tubing, Dialysate |
Class 2 |
FID |
§876.5820 |
Tubing, Dialysate (And Connector) |
Class 2 |
KQQ |
|
Relevant FDA Guidance and/or Supportive Publications*
Guidance for Industry and FDA Staff: Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a Risk Management Process, Issued June 2016.
Guidance for Industry and FDA Staff, Hemodialysis Blood Tubing Sets - Premarket Notification [510(k)] Submissions, Issued April 2008. |
|
FDA Technical Contact
|
Standards Development Organization
|
FDA Specialty Task Group (STG)
ObGyn/Gastroenterology/Urology |
|
|
*These are provided as examples and others may be applicable. |
|
|