Part B: Supplementary Information Sheet (SIS)
FR Recognition List Number 049 Date of Entry 06/07/2018 
FR Recognition Number 1-129
Standard
ISO 5359 Fourth edition 2014-10-01
Anaesthetic and respiratory equipment - Low-pressure hose assemblies for use with medical gases [Including AMENDMENT 1 (2017)]
Scope/Abstract
ISO 5359:2014 specifies requirements for low-pressure hose assemblies intended for use with the following medical gases: oxygen, nitrous oxide, medical air, helium, carbon dioxide, xenon, specified mixtures of the gases listed above, oxygen-enriched air, air for driving surgical tools, nitrogen for driving surgical tools, and for use with vacuum. It applies to hose assemblies operating at pressures up to 1 400 kPa and for vacuum systems at pressures not greater than 60 kPa absolute. It does not specify the dimensions and allocation of the gas-specific inlet and outlet connectors for the hose assemblies. It does not specify requirements for coaxial hoses used for the supply and removal of air for driving surgical tools. Nor does it specify the intended uses of hose assemblies.
Extent of Recognition
Complete standard
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.
Public Law, CFR Citation(s) and Procode(s)*
Regulation Number Device Name Device Class Product Code
§868.5160 Gas-Machine, Anesthesia Class 2 BSZ
§868.5895 Ventilator, Continuous, Facility Use Class 2 CBK
§868.5895 Continuous, Minimal Ventilatory Support, Facility Use Ventilator Class 2 MNT
Relevant FDA Guidance and/or Supportive Publications*
Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018.
FDA Technical Contact
 Neel Patel
  FDA/OC/CDRH/OPEQ/OHTI/DHTIC/
  301-796-6274
  Neel.Patel@fda.hhs.gov
Standards Development Organization
ISO International Organization for Standardization https://www.iso.org/
FDA Specialty Task Group (STG)
Anesthesiology
*These are provided as examples and others may be applicable.
-
-