• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Recognized Consensus Standards: Medical Devices

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

New Search Back To Search Results
Part B: Supplementary Information Sheet (SIS)
FR Recognition List Number 049 Date of Entry 06/07/2018 
FR Recognition Number 1-130
Standard
ISO 18082 First edition 2014-06-15
Anaesthetic and respiratory equipment - Dimensions of noninterchangeable screw-threaded (NIST) low-pressure connectors for medical gases [Including AMENDMENT 1 (2017)]
Scope/Abstract
ISO 18082:2014 specifies the dimensions, the allocation and marking of non-interchangeable screw-threaded (NIST) connectors intended to be used at operating pressures up to 1 400 kPa, and for vacuum systems at pressures not greater than 60 kPa absolute.
ISO 18082:2014 specifies NIST connectors intended for use with the following medical gases:
-oxygen;
-nitrous oxide;
-medical air;
-helium;
-carbon dioxide;
-xenon;
-specified mixtures of the gases listed above;
-oxygen-enriched air;
-air for driving surgical tools;
-nitrogen for driving surgical tools;
-use with vacuum.
Extent of Recognition
Complete standard
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.
Public Law, CFR Citation(s) and Procode(s)*
Regulation Number Device Name Device Class Product Code
§868.2700 Regulator, Pressure, Gas Cylinder Class 1 CAN
§868.5750 Cuff, Tracheal Tube, Inflatable Class 2 BSK
§868.5895 Ventilator, Continuous, Facility Use Class 2 CBK
§868.5895 Continuous, Minimal Ventilatory Support, Facility Use Ventilator Class 2 MNT
Relevant FDA Guidance and/or Supportive Publications*
Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018.
FDA Technical Contact
 Neel Patel
  FDA/OC/CDRH/OPEQ/OHTI/DHTIC/
  301-796-6274
  Neel.Patel@fda.hhs.gov
Standards Development Organization
ISO International Organization for Standardization https://www.iso.org/
FDA Specialty Task Group (STG)
Anesthesiology