• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Recognized Consensus Standards

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

New Search Back To Search Results
Part B: Supplementary Information Sheet (SIS)
FR Recognition List Number 051 Date of Entry 01/14/2019 
FR Recognition Number 4-248
Standard
ISO 10477 Third edition 2018-06
Dentistry - Polymer-based crown and veneering materials
Scope/Abstract
This document classifies polymer-based crown and veneering materials used in dentistry and specifies their requirements. It also specifies the test methods to be used to determine conformity to these requirements.

This document is applicable to polymer-based crown and veneering materials for laboratory-fabricated permanent veneers or crowns. It also applies to polymer-based dental crown and veneering materials for which the manufacturer claims adhesion to the substructure without macro-mechanical retention such as beads or wires.
Extent of Recognition
Complete standard
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.
Public Law, CFR Citation(s) and Procode(s)*
Regulation Number Device Name Device Class Product Code
§872.3690 Material, Tooth Shade, Resin Class 2 EBF
§872.3770 Crown And Bridge, Temporary, Resin Class 2 EBG
Relevant FDA Guidance and/or Supportive Publications*
Guidance for Industry and FDA Staff: Dental Composite Resin Devices - Premarket Notification [510(k)] Submissions, Issued October 26, 2005.

Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018.
FDA Technical Contacts
 Michael Adjodha
  FDA/OC/CDRH/OPEQ/OHTI/DHTIB/
  301-796-6276
  michael.adjodha@fda.hhs.gov
 Joel Anderson
  FDA/OC/CDRH/OPEQ/OHTI/DHTIB/
  301-796-6520
  joel.anderson@fda.hhs.gov
Standards Development Organization
ISO International Organization for Standardization https://www.iso.org/
FDA Specialty Task Group (STG)
Dental/ENT
*These are provided as examples and others may be applicable.
-
-