• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Recognized Consensus Standards: Medical Devices

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

New Search Back To Search Results
Part B: Supplementary Information Sheet (SIS)
FR Recognition List Number 052 Date of Entry 07/15/2019 
FR Recognition Number 8-516
Standard
ISO 13779-4 Second edition 2018-12
Implants for surgery - Hydroxyapatite - Part 4: Determination of coating adhesion strength
Scope/Abstract
This document specifies a test method for measurement of the adhesion strength of hydroxyapatite coatings intended for use on metallic-substrate components of surgical implants.


NOTE Requirements for the competence of testing laboratories can be found in ISO/IEC 17025.
Extent of Recognition
Complete standard
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.
Public Law, CFR Citation(s) and Procode(s)*
Regulation Number Device Name Device Class Product Code
§888.3045 Filler, Bone Void, Osteoinduction (W/O Human Growth Factor) Class 2 MBP
§888.3045 Filler, Bone Void, Calcium Compound Class 2 MQV
§888.3045 Calcium Salt Bone Void Filler, Drillable, Non-Screw Augmentation Class 2 OIS
§888.3353 Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented Class 2 LZO
§888.3353 Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous Cemented, Osteophilic Finish Class 2 MAY
§888.3353 Prosthesis, Hip, Semi-Constrained, Uncemented, Metal / Polymer, Non-Porous, Calcium Phosphate Class 2 MEH
§888.3353 Hip, Semi-Constrained, Cemented, Metal/Ceramic/Polymer + Additive, Porous Uncemented Class 2 OQI
§888.3358 Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented Class 2 LPH
§888.3358 Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Porous Class 2 MBL
§888.3358 Hip Prosthesis, Semi-Constrained, Cemented, Metal/Polymer, + Additive, Porous, Uncemented Class 2 OQG
§888.3535 Prosthesis, Knee, Femorotibial, Unicompartmental/Unicondylar, Uncemented, Porous-Coated, Metal/Polymer Class 2 NJD
§888.3565 Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer Class 2 MBH
§888.3670 Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer, Uncemented Class 2 MBF
Relevant FDA Guidance and/or Supportive Publications*
Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement, Issued April 1994

510(k) Information Needed for Hydroxyapatite Coated Orthopedic Implants, Issued March 1995 (revised 2/20/97)

Guidance for Industry and FDA Staff: Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments, Issued May 2004

Guidance for Industry and FDA Staff: Class II Special Controls Guidance Document: Resorbable Calcium Salt Bone Void Filler Device, Issued June 2003

Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018.
FDA Technical Contact
 Peter Allen
  FDA/OC/CDRH/OPEQ/OHTVI/DHTVIA/
  301-796-6402
  peter.allen@fda.hhs.gov
Standards Development Organization
ISO International Organization for Standardization https://www.iso.org/
FDA Specialty Task Group (STG)
Materials
*These are provided as examples and others may be applicable.
-
-