Part B: Supplementary Information Sheet (SIS)
FR Recognition List Number 041 Date of Entry 04/04/2016 
FR Recognition Number 1-115
Standard(Included in ASCA pilot)
ISO 80601-2-70 First Edition 2015-01-15
Medical electrical equipment - Part 2-70: Particular requirements for the basic safety and essential performance of sleep apnoea breathing therapy equipment
Scope/Abstract
ISO 80601-2-70:2015 is applicable to the basic safety and essential performance of sleep apnoea breathing therapy equipment, hereafter referred to as me equipment, intended to alleviate the symptoms of patients who suffer from obstructive sleep apnoea by delivering a therapeutic breathing pressure to the respiratory tract of the patient. Sleep apnoea breathing therapy equipment is intended for use in the home healthcare environment by lay operators as well as in professional healthcare institutions.
It excludes sleep apnoea breathing therapy equipment intended for use with neonates.
ISO 80601-2-70:2015 is applicable to me equipment or an me system intended for those patients who are not dependent on mechanical ventilation.
ISO 80601-2-70:2015 is not applicable to me equipment or an me system intended for those patients who are dependent on mechanical ventilation such as patients with central sleep apnoea.
ISO 80601-2-70:2015 is also applicable to those accessories intended by their manufacturer to be connected to sleep apnoea breathing therapy equipment, where the characteristics of those accessories can affect the basic safety or essential performance of the sleep apnoea breathing therapy equipment.
Extent of Recognition
Complete standard
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.
Transition Period
FDA recognition of ISO 80601-2-70 First Edition 2015-01 [Rec# 1-115] will be superseded by recognition of ISO 80601-2-70 Second edition 2020-11 [Rec# 1-151]. FDA will accept declarations of conformity, in support of premarket submissions, to [Rec# 1-115] until July 7, 2024. After this transition period, declarations of conformity to [Rec# 1-115] will not be accepted.
Public Law, CFR Citation(s) and Procode(s)*
Regulation Number Device Name Device Class Product Code
§868.5905 Ventilator, Non-Continuous (Respirator) Class 2 BZD
Relevant FDA Guidance and/or Supportive Publications*
1. AAMI CR500:2019 Basic Introduction to the IEC 60601 Series.
2. AAMI COVID-19 Emergency Guidance for the design of ventilator, resuscitator, CPAP/BiPAP systems is available at https://www.aami.org/news-resources/covid-19-updates/covid_cr. Check back frequently for the latest version as these are living documents that evolve with experience.
3. Guidance for Industry and Food and Drug Administration Staff: Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling, Issued March 2015.
4. Guidance for Industry and FDA Staff: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, Issued May 2005.
5. Guidance for Industry and Food and Drug Administration Staff: Applying Human Factors and Usability Engineering to Medical Devices, Issued April 2016.
6. Final Guidance for Industry and FDA Staff: Guidance on Medical Device Patient Labeling, Issued April 2001.
7. Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process," Issued September 2020.
8. Guidance for Industry and Food and Drug Administration Staff: Information to Support a Claim of Electromagnetic Compatibility (EMC) of Electrically-Powered Medical Devices, Issued July 2016.
9. ISO 80601-2-74 First edition 2017-05 Medical electrical equipment - Part 2-74: Particular requirements for basic safety and essential performance of respiratory humidifying equipment.
10. ISO 17510 First Edition 2015-08-01 Medical devices - Sleep apnoea breathing therapy - Masks and application accessories.
11. ISO 10993-3 Third edition 2014-10-1 Biological evaluation of medical devices - Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity.
12. ISO /TR 10993-33 First Edition 2015-03-01 Biological evaluation of medical devices - Part 33: Guidance on tests to evaluate genotoxicity - Supplement to ISO 10993-3.
FDA Technical Contacts
 Amy Levelle
  FDA/OC/CDRH/OPEQ/OHTI/DHTIC/
  301-796-6963
  Amy.Levelle@fda.hhs.gov
 Rachana Visaria
  FDA/OC/CDRH/OPEQ/OHTI/DHTIC/
  240-402-5628
  Rachana.Visaria@fda.hhs.gov
Standards Development Organization
ISO International Organization for Standardization https://www.iso.org/
FDA Specialty Task Group (STG)
Anesthesiology
*These are provided as examples and others may be applicable.
-
-