• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Recognized Consensus Standards

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

New Search Back To Search Results
Part B: Supplementary Information Sheet (SIS)
FR Recognition List Number 034 Date of Entry 01/30/2014 
FR Recognition Number 11-238
Standard
ASTM F2033-12
Standard Specification for Total Hip Joint Prosthesis and Hip Endoprosthesis Bearing Surfaces Made of Metallic, Ceramic, and Polymeric Materials
Scope/Abstract
1.1 This specification covers the requirements for the mating bearing surfaces of total hip prostheses and hip endoprostheses. More specifically, this specification covers hip joint replacement of the ball-and-socket configuration.

1.2 This specification covers the sphericity, surface finish requirements, and dimensional tolerances for the spherical articulating metallic or ceramic femoral heads of total hip joint prostheses.

1.3 This specification covers the sphericity, surface finish requirements, and dimensional tolerances for the spherical concave mating surface of metallic and ceramic acetabular components, including the inner polymeric bearing surface of bipolar heads, and the surface finish requirements and dimensional tolerances for the spherical concave mating surface of polymeric acetabular components.

1.4 This specification covers the sphericity and surface finish requirements for the spherical metallic or ceramic femoral heads of hip endoprostheses, and the outer bearing surface of bipolar heads.

1.5 This specification is intended for standard practice regarding the design of total hip joint bearing surfaces. Additionally, the tolerances imposed on the polymeric portion of the bearing surface are intentionally large due to temperature-induced changes and other manufacturing concerns. Some manufacturing methods or designs may intentionally reduce the diameter of the polymeric bearing to more closely mate with the diameter of the head.
Extent of Recognition
Complete standard
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.
Transition Period
FDA recognition of ASTM F2033-12 [Rec# 11-238] will be superseded by recognition of ASTM F2033-20 [Rec# 11-376]. FDA will accept declarations of conformity, in support of premarket submissions, to [Rec# 11-238] until July 10, 2022. After this transition period, declarations of conformity to [Rec# 11-238] will not be accepted.
Public Law, CFR Citation(s) and Procode(s)*
Regulation Number Device Name Device Class Product Code
§888.3353 Hip, Semi-Constrained, Cemented, Metal/Ceramic/Polymer + Additive, Porous Uncemented Class 2 OQI
§888.3353 Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous Cemented, Osteophilic Finish Class 2 MAY
§888.3353 Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented Class 2 LZO
§888.3353 Prosthesis, Hip, Semi-Constrained, Uncemented, Metal / Polymer, Non-Porous, Calcium Phosphate Class 2 MEH
Unclassified Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Ceramic, Cemented Class 3 LPF
Unclassified Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Ceramic/Metal, Cemented Or Uncemented Class 3 MRA
§888.3300 Prosthesis, Hip, Constrained, Metal Class 3 KXD
§888.3330 Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component) Class 3 KWA
§888.3350 Hip, Semi-Constrained, Cemented, Metal/Polymer + Additive, Cemented Class 2 OQH
§888.3350 Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented Class 2 JDI
§888.3358 Hip Prosthesis, Semi-Constrained, Cemented, Metal/Polymer, + Additive, Porous, Uncemented Class 2 OQG
§888.3358 Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Porous Class 2 MBL
§888.3360 Prosthesis, Hip, Femoral Component, Cemented, Metal Class 2 JDG
§888.3360 Prosthesis, Hip, Hemi-, Femoral, Metal Class 2 KWL
§888.3360 Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented Class 2 LWJ
§888.3360 Prosthesis, Upper Femoral Class 2 JDD
§888.3390 Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented Class 2 KWY
§888.3310 Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer Class 2 KWZ
§888.3320 Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component) Class 3 JDL
§888.3340 Prosthesis, Hip, Semi-Constrained, Composite/Metal Class 2 KMC
Relevant FDA Guidance and/or Supportive Publications*
Guidance Document for the Preparation of Premarket Notification for Ceramic Ball Hip Systems, issued January 1995.
FDA Technical Contact
 John Goode
  FDA/OC/CDRH/OPEQ/OHTVI/DHTVIA/
  301-796-6407
  john.goode@fda.hhs.gov
Standards Development Organization
ASTM ASTM International http://www.astm.org/
FDA Specialty Task Group (STG)
Orthopedic
*These are provided as examples and others may be applicable.
-
-