New Search Back To Search Results
Part B: Supplementary Information Sheet (SIS)
FR Recognition List Number 057 Date of Entry 12/20/2021 
FR Recognition Number 8-567
Standard
ASTM F86-21
Standard Practice for Surface Preparation and Marking of Metallic Surgical Implants
Scope/Abstract
1.1 This practice provides descriptions of surface characteristics, surface preparation, and marking for metallic surgical implants, with the purpose of improving the corrosion resistance of the implant surfaces and markings.

1.2 Marking nomenclature and neutralization of endotoxin are not specified in this practice (see X1.4).

1.3 Surface requirements and marking methods included in the implant specification shall take precedence over requirements listed in this practice, where appropriate.
Extent of Recognition
Complete standard
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.
Public Law, CFR Citation(s) and Procode(s)*
IMPLANTABLE METALLIC COMPONENTS
Relevant FDA Guidance and/or Supportive Publications*
Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018.
FDA Technical Contacts
 David M. Saylor
  FDA/OC/CDRH/OSEL/DBCMS/
  301-796-2626
  david.saylor@fda.hhs.gov
 Matthew Di Prima
  FDA/OC/CDRH/OSEL/DAM/
  301-796-2507
  matthew.diprima@fda.hhs.gov