• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Recognized Consensus Standards: Medical Devices

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

New Search Back To Search Results
Part B: Supplementary Information Sheet (SIS)
FR Recognition List Number 062 Date of Entry 05/29/2024 
FR Recognition Number 1-172
Standard
ISO 11712 Second edition 2023-11
Anaesthetic and respiratory equipment - Supralaryngeal airways and connectors
Scope/Abstract
This document provides the essential requirements for the design of supralaryngeal airways and connectors. These devices are intended to provide a distinct respiratory pathway to the top of the larynx to provide an unobstructed airway in patients during spontaneous, assisted or controlled ventilation.

This document specifies the dimensions, basic properties and method of size designation of the available types of supralaryngeal airways. Airways devised for specialized applications are not specifically covered, although most may be classified by the sizing and dimensions (or other characteristics) required by this document.

The following devices are outside the scope of this document: nasal and oropharyngeal airways, anesthetic masks, oro- and naso-tracheal tubes, cricothyrotomy devices, dental appliances, tracheal stents, tracheal tubes, ventilating laryngoscopes, CPAP devices, esophageal obturators, bougies and devices that require surgical placement.

This document specifies dimensional disclosure so the operator will know which auxiliary devices, such as tracheal tubes and bronchoscopes will be size-compatible.

Flammability of airways, for example if used with certain flammable anesthetic gases, electrosurgical units or lasers, is a well-recognized hazard that is outside the scope of this document. (See E.1.7).
Extent of Recognition
Complete standard
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.
Public Law, CFR Citation(s) and Procode(s)*
Regulation Number Device Name Device Class Product Code
§868.5110 Airway, Oropharyngeal, Anesthesiology Class 1 CAE
§868.5110 Oral Airway Assist Class 1 PHS
Relevant FDA Guidance and/or Supportive Publications*
Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018.
FDA Technical Contact
 Annie Abraham
  FDA/OC/CDRH/OPEQ/OHTI/DHTIC/
  240-402-5219
  Annie.Abraham@fda.hhs.gov
Standards Development Organization
ISO International Organization for Standardization https://www.iso.org/
FDA Specialty Task Group (STG)
Anesthesiology
*These are provided as examples and others may be applicable.
-
-