Part B: Supplementary Information Sheet (SIS)
FR Recognition List Number 062 Date of Entry 05/29/2024 
FR Recognition Number 6-500
Standard
IEC 63045 Edition 1.0 2020-05
Ultrasonics - Non-focusing short pressure pulse sources including ballistic pressure pulse sources - Characteristics of fields
Scope/Abstract
This document is applicable to:
- therapy equipment using extracorporeally induced non-focused or weakly focused pressure pulses;
- therapy equipment producing extracorporeally induced non-focused or weakly focused mechanical energy,
where the pressure pulses are released as single events of duration up to 25 µs.

This document does not apply to:
- therapy equipment using focusing pressure pulse sources such as extracorporeal lithotripsy equipment;
- therapy equipment using other acoustic waveforms like physiotherapy equipment, low intensity ultrasound equipment and HIFU/HITU equipment.

This document specifies:
- measurable parameters which are used in the declaration of the acoustic output of extracorporeal equipment producing a non-focused or weakly focused pressure pulse field,
- methods of measurement and characterization of non-focused or weakly focused pressure pulse fields.
Extent of Recognition
Complete standard
Public Law, CFR Citation(s) and Procode(s)*
Regulation Number Device Name Device Class Product Code
§890.5660 Massager, Therapeutic, Electric Class 1 ISA
Relevant FDA Guidance and/or Supportive Publications*
Lohrer, H., Nauck T., "Shock Waves in Sports medicine", Level 10 Publisher, 2018.

Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018.
FDA Technical Contact
 Subha Maruvada
  FDA/OC/CDRH/OSEL/DAM/
  301-796-2524
  subha.maruvada@fda.hhs.gov
Standards Development Organization
IEC International Electrotechnical Commission http://www.iec.ch/
FDA Specialty Task Group (STG)
General Plastic Surgery/General Hospital
*These are provided as examples and others may be applicable.
-
-