| Part B: Supplementary Information Sheet (SIS) |
|
FR Recognition List Number
|
063
|
Date of Entry 12/23/2024
|
|
FR Recognition Number
|
1-183
|
| Standard | |
ASTM G175-24 Standard Test Method for Evaluating the Ignition Sensitivity and Fault Tolerance of Oxygen Regulators Used for Medical and Emergency Applications |
|
Scope/Abstract1.1 For the purpose of this test method, a pressure regulator, also called a pressure-reducing valve, is a device intended for medical or emergency purposes that is used to convert a medical or emergency gas pressure from a high, variable pressure to a lower, more constant working pressure [21 CFR 868.2700 (a)]. Some of these oxygen pressure regulators are a combination of a pressure regulator and cylinder valve. These devices are often referred to as valve integrated pressure regulators, or VIPRs.
1.2 This test method provides an evaluation tool for determining the ignition sensitivity and fault tolerance of oxygen pressure regulators and VIPRs used for medical and emergency applications. An ignition-sensitive pressure regulator or VIPR is defined as having a high probability of ignition as evaluated by rapid pressurization testing (Phase 1). A fault-tolerant pressure regulator or VIPR is defined as having a low consequence of ignition as evaluated by forced ignition testing (Phase 2).
Note 1: It is essential that a risk assessment be carried out on breathing gas systems, especially concerning toxic product formation due to ignition or decomposition of nonmetallic materials as weighed against the risk of flammability (refer to Guide G63 and ISO 15001.2). See Appendix X1 and Appendix X2 for details. ... |
|
| Extent of Recognition
|
Rationale for Recognition
| This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies. |
|
Transition Period
| FDA recognition of ASTM G175-13 (Reapproved 2021) [Rec# 1-99] will be superseded by recognition of ASTM G175-24 [Rec# 1-183]. FDA will accept declarations of conformity, in support of premarket submissions, to [Rec# 1-99] until December 20, 2026. After this transition period, declarations of conformity to [Rec# 1-99] will not be accepted. |
|
Public Law, CFR Citation(s) and Procode(s)*
| Regulation Number |
Device Name |
Device Class |
Product Code |
| §868.2700 |
Regulator, Pressure, Gas Cylinder
|
Class 1
|
CAN
|
| §868.5905 |
Conserver, Oxygen
|
Class 2
|
NFB
|
|
Relevant FDA Guidance and/or Supportive Publications*
| Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018. |
|