Standard
ASTM F2554-22
Standard Practice for Measurement of Positional Accuracy of Computer-Assisted Surgical Systems
Scope/Abstract
1.1 This document provides procedures for measurement and reporting of basic static performance of surgical navigation and/or robotic positioning devices under defined conditions. They can be performed on a subsystem (for example, tracking only) or a full computer-aided surgery system as would be used clinically. Testing a subsystem does not mean that the whole system has been tested. The functionality to be tested based on this practice is limited to the performance (accuracy in terms of bias and precision) of the system regarding point localization in space by means of a pointer. A point in space has no orientation; only multidimensional objects have orientation. Therefore, orientation of objects is not within the scope of this practice. However, in localizing a point the different orientations of the pointer can produce errors. These errors and the pointer orientation are within the scope of this practice. The aim is to provide a standardized measurement of performance variables by which end users can compare within a system (for example, with different reference elements or pointers) and between different systems (for example, from different manufacturers). Parameters to be evaluated include (based upon the features of the system being evaluated):
(1) Accuracy of a single point relative to a coordinate system.
(2) Sensitivity of tracking accuracy due to changes in pointer orientation.
(3) Relative point-to-point accuracy.

1.1.1 This method covers all configurations of the evaluated system as well as extreme placements across the measurement volume.

1.2 This practice defines a standardized reporting format, which includes definition of the coordinate systems to be used for reporting the measurements, and statistical measures (for example, mean, RMS, and maximum error).
Extent of Recognition
Complete standard
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.
Transition Period
FDA recognition of ASTM F2554-18 [Rec# 11-350] will be superseded by recognition of ASTM F2554-22 [Rec# 11-415]. FDA will accept declarations of conformity, in support of premarket submissions, to [Rec# 11-350] until December 20, 2026. After this transition period, declarations of conformity to [Rec# 11-350] will not be accepted.
Public Law, CFR Citation(s) and Procode(s)*
Regulation Number Device Name Device Class Product Code
§882.4560 Orthopedic Stereotaxic Instrument Class 2 OLO
Relevant FDA Guidance and/or Supportive Publications*
Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018.
FDA Technical Contact
 Sheldon Davis
  FDA/OC/CDRH/OPEQ/OHTVI/DHTVIC/
  301-796-4034
  Sheldon.Davis@fda.hhs.gov
Standards Development Organization
ASTM ASTM International http://www.astm.org/
FDA Specialty Task Group (STG)
Orthopedic
*These are provided as examples and others may be applicable.
-
-